Lintec MC-3100 Series User manual

LN0324E2103F0
Mass Flow Controller
Instruction Manual
MC-3100/3200L Series
Safety Precautions
Incorrect handling may cause death or injury
(1) Before connecting the fittings, check that no damage or defects are found on
the fittings. Make connections properly and make sure that a leak test is
conducted before actual operation to prevent fluid from leaking into the
atmosphere (Hereafter, the measured fluid is called “gas” or “fluid”).
(2) DO NOTapply any fluidscorrosive tomaterials exposedto gas. Corrosion may
cause fluid to leak into the atmosphere. Check the gas type to be used in advance.
(3) This device is not designed as an explosion proof structure. DO NOT use this
device in a place where explosion-proof structures are required. Doing so may
cause fire or explosion.
Incorrecthandlingmaylead tomedium orslightinjury ormay
causedamageto,orlossof,facilitiesorequipment
(1) Observe the precautions listed in the WARNING (above)
(2) Strictly observe the electrical specifications. Not doing so may cause fire,
damage to sensors or malfunction.
(3) This device is not designed to be waterproof. DO NOT locate this device
outdoors orin a place whereit may besplashed with water. Doing so may cause
fire, trouble, or malfunction of the device.
(4) DO NOT modify this device. It may cause fire or other problems.
(5) This device is not designed to handle hot swap. Please avoid attaching and
removing the power supply connector and interface connector with the power
switched on. Attachment and/or removal with the power on may result in failure
of the device.
(6) While a power supply is applied to MFC, ±15VDC must be applied
simultaneously. If only +15VDC or -15VDC is applied, electronic circuits will
become unstable and it may cause a malfunction of MFC.
(7) This device is a precious device, please handle it carefully. Dropping down or
handling it carelessly will cause damage. Please use assist instrument while
moving or setting the device.
(8) Regular maintenance is recommended for steady use of this device
(Recommended proofreading frequency is once a year).
1. Introduction
This manual explains basic operation of the MC-3100/3200L series (Hereafter, it is
called "MFC"). Please read through this manual and other separate volumes (Digital
Interface Manual, Special Function Manual, Command Chart) carefully to
familiarize yourself with the features of this device.
2. Summary
The MFC is high performance mass flow controller for gas using thermal flow
sensor and high response piezoelectric valve. In the MFC model, MO type and
MC type are available depending on the kind of valve operation mode.
3. Features
The MFC is the following features.
(1)
LINTEC’s proprietary ambient temperature compensation type flow sensor is carried.
· Low sensor temperature, long-time steady monitor, and few problems such
as degradation of gas to be monitored.
·The effect of ambient temperature is less because the sensor temperature is
kept at the ambient temperature.
·Steady temperature distribution of sensor and high-speed response.
(2)Digital computing system with microprocessor and high-resolution A/D, D/A
converter is carried
· High functionality
·By setting of device number (address), multiple devices can be controlled
through a single interface
· Many additional are provided as standard, e.g., totalizer function, zero adjust,
ramping function
(3) Universal digital Interface (RS-485)
(4) Small dead space structure using a diaphragm valve
(5) Superior corrosion resistant sealing materials make maintenance easy. Metal
seals (Au)
(6) Particle-free structure
(7) By using a metal case and various types of filters, steady operation can achieve
even in an environment of high-frequency noise and stationary magnetic field
(8) Based on RoHS
4. Structure
The MFC consists of sensors, bypass, valves, and a microcomputer for signal
processing. A digital PID feedback control system controls the valve action so that
flow rate output from the sensor agrees with flow rate setting value.
5. Specification / Dimensions
(1) Specification
Name
Mass Flow Controller
Model
MC-3102L
MC-3202L
MC-3252L
Standard flow rate in
nitrogen
5SCCMto5SLM
~20SLM
~50SLM
Valve operation mode
Normally open valve / Normally closed valve
Internal surface treatment
No treatment / Precision polishing (LEP)
Minimumcontrollableflowrate
2%F.S.
Analog flow rate output signal
Proportional to flow rate 0 to 5VDC
Analog flow rate setting signal
Accuracy
100%F.S. to 25%F.S. ±1.0%S.P.
25%F.S. to 2%F.S. ±0.25%F.S.
Repeatability
±0.2%F.S.
Response time
1 second (Typical)
Operating differential
pressure
50 to 300kPa
Normally open valve
50 to 300kPa (~
10SLM)
100 to 300kPa (10 to 20SLM)
Normally closed valve
100 to 300kPa (~
10SLM)
200 to 300kPa (10 to 20SLM)
100 to 300kPa
(~ 30SLM)
150 to 300kPa
(~ 50SLM)
Maximum operating
pressure (Note1)
300kPa(G)
Withstanding pressure
1MPa(G)
Operating
temperature
&
Humidity
5 to 50ºC·0 to 80%RH (Condensation should be avoided)
Temperature
coefficient
(Standard
Temp.25 ºC)
Zero
±0.02%F.S./ºC
Span
±0.02%F.S./ºC
±0.04%F.S./ºC
±0.08%F.S./ºC
Storage temperature
&
Humidity
0 to 60ºC·0 to 80%RH (Condensation should be avoided)
Leak integrity
Less than 1×10-11Pa · m3/sec (He)
Mounting direction
Free
Wetted materials
Stainless steel 316L, PCTFE, Au
Seal Materials
Au
Actuator
Piezoelectric actuator
Fittings
6.35VCR (106, 124) 6.35UJR (106, 124)
1.125W-Seal (79.8mm, 92mm) 1.5W-Seal (79.8mm, 92mm)
Power supply
+15V DC±3%: 100mA -15VDC±3%: 50mA
Digital interface
RS-485 (2-wire)
Weight
Approx. 1kg
Note1)There is a possibility that the gas does not flow in case of the gas pressure of more than
490kPa was used on the MC-3252L- MO /NO.
(2) Dimensions
6. Ordering information
MC-3102L –MO –4JR1 AA0A0 –dd –ee –ff
[1] [2] [3] [4] [5] [6] [7]
[1] Series model
MC: Mass Flow Controller Series
MC-3102L, MC-3202L, MC-3252L
[2] Valve mode
NO: Normally open (No treatment), MO: Normally open (LEP)
NC: Normally close (No treatment), MC: Normally close (LEP)
[3] Fitting
4JR1:6.35UJR106,
4JR2:6.35UJR124, 4VR1:6.35VCR106, 4VR2:6.35VCR124, 4SWL:6.35SWL
SPW1: Surface mount 1.5W-Seal 79.8mm,
SPW2: 1.5W-Seal 92mm
SPW3: Surface mount 1.1.25W-Seal 79.8mm,
SPW4: Surface mount 1.125W-Seal 92mm
※ Please consult for more information.
[4] Optional
B: Applied to liquefied gas,
D: Medium temperature specifications (45 to 55 ºC)
E: Small Range Control (1 to 4SCCM F.S. in nitrogen)
G:
Wide Range specifications (Control range 0.5 to 100 %F.S.)
J:
Valve is closed at below 2% setting
※ Default setting is labeled
“AA0A0”.
Please consult for more information.
[5] Gas type
[6] Full scale flow rate
[7] Flow rate unit SCCM (0 ºC standard), SLM (0 ºC standard)
7. Connection
(1) Analog interface connector
Mounted connector : D-Sub 9 pin(male)
Pair connector : D-Sub 9 pin(female) (Fixed size of screw : M3)
Pin No.
Single name
Function
1
Valve on-off input (Note2)
+15VDC: OPEN,-15VDC: CLOSE
2
Flow rate output signal 0 to 5VDC
Flow rate output voltage plus side 0 to 5VDC
3
Power supply input +15VDC±3%
Power supply (plus) 100mA
4
Power supply COMMON (Note3)
COMMON ±15VDC
5
Power supply input -15VDC±3%
Power supply (minus) 50mA
6
Flow rate setting signal 0 to 5VDC
(Note2)
Flow rate setting voltage plus side 0 to 5VDC
7
Flow rate output COMMON (Note3)
Flow rate output voltage COMMON
8
Flow rate setting COMMON (Note3)
Flow rate setting voltage COMMON
9
N.C.
N.C.
Note2) Inputimpedance of flow rate settingsignal is 100kΩ. Valve override impedance is 100kΩ.
Note3) Pin No.4,7,8 are connected internally.
Wirings should be done as shown below in order to remove the effect of potential difference
among the COMMON.
Do not connect. FlowratesettingsignalCOMMON[8],Flow rateoutputsignal COMMON[7],
andPowersupplyCOMMON[4]inthe powersupply unit
(2) Digital interface connector
Mounted connector : RJ-45 Modular jack
Pair connector : RJ-45 Modular plug
Pin No.
Signal name
Function
1
Signal COMMON
RS-485 Signal COMMON
2
Signal COMMON
RS-485 Signal COMMON
3
N.C.
N.C.
4
Signal [- Txd / Rxd ]
RS-485 2-wire system transmitter. receiver minus signal
5
Signal [+Txd / Rxd ]
RS-485 2-wire system transmitter. receiver plus signal
6
N.C.
N.C.
7
N.C.
N.C.
8
N.C.
N.C.
8. Alarm functions
This MFC featurestwo alarm modes: alarm A and alarm B. Using a digital interface,
alarm output of the digital connector or LED indicator located on top of the MFC
housing can monitor the alarm status. As alarm settings can only be changed using
digital communication, please carry out necessary changes using this method. For
details, please refer to the Interface Instruction Manual.
Alarm
Set off
LED indicator
A
Setting value≠ flow rate output
Continuous Red
Flow totalizer Alarm level 2
Power supply voltage (+15VDC) drop
Off
B
Voltage of valve control change
Red (every 0.5sec flashing)
Abnormal zero offset (Note 4)
Flow totalizer Alarm level 1
Other
Command error
Continuous Red (every 0.5sec flashing)
Normal
-
Green (every 1sec flashing)
9. Software switch (factory shipped value)
This MFC is provided with a software switch for operation mode setting. Before
operation, input the necessary data for various functions by using the digital
interface. Details of the functions are described in the special function manual.
Software switch
Available values
Function
Default
Device number
00 to 99
MFC No. Registration
00
Alarm A range
0 to 99 [%]
Setting Alarm A range
5%
Alarm B range
0 to 99 [%]
Setting of Alarm B range
20%
Alarm timer
0 to 99 [seconds]
Setting of Alarm timer
5 seconds
Alarm A output
Enable / Disable
Setting of Alarm A output
Enable
Alarm B output
Enable / Disable
Setting of Alarm B output
Disable
(Note 4)
Operation mode
Analog / Digital
Switching of Analog/digital
Analog
Power on mode
Analog / Preset
Switching of Operation mode
in power turning on
Analog
Valve control
C / O / H / S
Close / Open / Hold / Servo
Servo
Control speed
FAST / SLOW
Switching of Response speed
FAST
Control mode
2%Close/2%Hold/ Normal
Control mode switching
(at flow rate setting <2%)
2%Close: Valve Close
2%Hold: 2% Control
Normal: Normal Control
Normal
Conversion factor
0.6666 to 1.500
Conversion factor setting
1.0000
Communication
protocol
9600 to 38400bps
8bit / 7bit
Odd/Even/ None
1bit / 2bit
Baud rate
Character length
Parity
Stop bit
9600bps
7bit
None
2bit
Note4)Zero adjustment error alarms regardless of the “Enable”/ “Disable”of the alarm.
·
If a zero set input of longer than 5seconds is input the Baud rate will be set to 9600bps and other
communication protocols will be returned to initial value (factory shipped value).
10. Operation
(1) Procedure
1) This product is packed in a clean room before shipment. Please break the
seals in a clean room after taking it out of its box.
2) Check the gas type and flow rate, and check the direction of the gas flow
and the MFC before installation.
3) Check for gas leaks from the tubing with a helium (He) leak detector.
4) Connect the interface connectors according to the Connector table.
5) Power requirements are +15VDC:100mA and -15VDC: 50mA. Check the
voltage, polarity, and capacitance of the power supply voltage.
6) Power of ±15VDC must be applied to the product at the same time, or it
may cause malfunction.
7) Turn on power supply and let the equipment warm up for at least 5 minutes
(Recommended time: 30min).
8) Adjust the zero point by pressing the zero-adjustment switch located on the
top of the MFC. Before zero-point adjustment, check that gas is not being
supplied and the device was warmed up for 30 minutes or more in order to
ensure sensor stability.
9) Input the flow rate setting signal and supply gas with required differential
pressure to the MFC. The MFC will begin to control the gas flow in
proportion to the preset voltage. Full-scale voltage is 5VDC. Maximum
input voltage is ±15.5VDC.
10) When the output flow rate signal is used, the tolerance voltage of the
external device should be more than ±15.5VDC. When it’s connected the
output valve may be within the range of the maximum voltage ±15.5VDC.
11) Complete shut-off cannot be achieved with the gas mass flow controller.
If complete shut-off is desired, a shut-off valve should be installed.
12) When ahighly reactive gas is used, thoroughly purge all foreign matter from
the tubing and the MFC before operation.
13) When contaminated gas is used, install a filter at the equipment inlet.
14) Use the MFC within the range of the operating temperature (5 to 50°C), and
keep it at the same temperature with the gas. If used in any environment that
does not meet the above-mentioned requirements, the flow rate cannot be
measured accurately and the device may fail.
15) Do not switch the power supply on and off within one second. It may cause
failure.
(2) Valve control signal
The MC features a forced valve open/close input function.
The connector pin No.1 is used to input the internal valve open/close signal.
By inputting this signal, a forced opening/closing of the internal valve can be
performed without depending on the value of the flow rate preset signal.
When +15VDC is input: fully open
When –15VDC is input: fully closed
(3) Conversion factor
The MFC ispreset based on the kind of gas and the flow rate. If a gasthat is not
specified is supplied, the controlled flow rate may be different fromthe actual flow rate.
This difference is called a conversion factor, and is normally represented by a ratio
with N2. When gas B is fed into the MFC, which was calibrated with gas A, the
controlled flow rate is shown below.
The accuracy ofthis value may be decreased if the characterofthe gas isgreatly
different. Also, the valueof theMFC has been adjusted basedonthe viscosity and
density ofthe gas,therefore ifa gasother than the calibrationgas isapplied,the
operating pressure range may be altered. The conversion factor can beset at any desired
value, with an operatingrange of0.6666 to 1.5000. Ifthe conversion factor value
exceeds thisrange,theoperation ofthe MFC become unstable.The accuracy isdifferent
from thestandard specifications ofthe MFC, when the conversion factoris changed.
(4) Digital interface
The MFC features the RS-485 serial digital interfaces. Many special functions
can be employed using the digital interfaces. Please refer to other manuals
(Digital Interface Manual, Special Function Manual, Command Chart).
11. Product warranty
(1) Period
This product is guaranteed for a period of 1 year from date of shipment. Defects
are repaired according to the following regulations.
(2) Scope
Warranty coverage is restricted to this product only. Any other damage caused
by this product is not covered.
(3) The following repairs are not covered by the warranty:
1) Failure caused by by-product of fluid used
2) Failure caused by misuse (including careless operation) or incorrect repair or
modification
3) Failure caused by dropping after purchasing
4) Failure caused by a natural disasters
Even if thewarranty period is still in effect, the following items may not be repaired.
1) When the kind of liquid or gas used in the product is unclear
2) The product is returned with liquid or gas remaining inside and
safety cannot be confirmed.
The MFC is a precision instrument. Control may become unstable if electric
noise, temperature change of fluid, pulsation of fluid pressure etc. occurs. Please
be forewarned.
This instruction manual is subject to revision without notice.
http://www.lintec-mfc.co.jp
CorporateHeadquarters
4-1-23Sekinotsu,OtsuCity,ShigaPref.520-2277,Japan
TEL. +81-(0)77-536-2210FAX.+81-(0)77-536-2215
TokyoBranchOffice
3FHattoriBuild.,4-30-14Yotsuya Shinjyuku-kuTokyo160-0004,Japan
TEL. +81-(0)3-5366-2801FAX.+81-(0)3-3341-3513

This manual suits for next models

4

Other Lintec Controllers manuals

Lintec MC-3500L Series User manual

Lintec

Lintec MC-3500L Series User manual

Lintec MC-3000S Series User manual

Lintec

Lintec MC-3000S Series User manual

Lintec MC-700 Series User manual

Lintec

Lintec MC-700 Series User manual

Lintec MC-7000D Series User manual

Lintec

Lintec MC-7000D Series User manual

Lintec MC-3300 Series User manual

Lintec

Lintec MC-3300 Series User manual

Lintec MC-5000L Series User manual

Lintec

Lintec MC-5000L Series User manual

Lintec MC-3000L-TC Series User manual

Lintec

Lintec MC-3000L-TC Series User manual

Popular Controllers manuals by other brands

Schweitzer Engineering SEL-351P quick start

Schweitzer Engineering

Schweitzer Engineering SEL-351P quick start

Dungs FRI 7/6 Series installation instructions

Dungs

Dungs FRI 7/6 Series installation instructions

IAI ACON-C Operation manual

IAI

IAI ACON-C Operation manual

Gefen EXT-CU-LAN user manual

Gefen

Gefen EXT-CU-LAN user manual

SMC Networks ITV1000-CC Series Operation manual

SMC Networks

SMC Networks ITV1000-CC Series Operation manual

EnOcean LEDD installation guide

EnOcean

EnOcean LEDD installation guide

Comet H5421 Quick start manual

Comet

Comet H5421 Quick start manual

EuroLite Color Chief user manual

EuroLite

EuroLite Color Chief user manual

iDTRONIC BLEUBOX R-IN-UHF-5239U manual

iDTRONIC

iDTRONIC BLEUBOX R-IN-UHF-5239U manual

FLIR PTU-D100 E Series user manual

FLIR

FLIR PTU-D100 E Series user manual

ICS Tunnel Master Jr manual

ICS

ICS Tunnel Master Jr manual

Carel AT-th Tune Series quick start guide

Carel

Carel AT-th Tune Series quick start guide

Full Gauge TIC-17 Series quick start guide

Full Gauge

Full Gauge TIC-17 Series quick start guide

Extech Instruments 48VTR user manual

Extech Instruments

Extech Instruments 48VTR user manual

Royal Vendors Merlin IV Installation

Royal Vendors

Royal Vendors Merlin IV Installation

Linak HOMELINE TWINDRIVE TD2 user manual

Linak

Linak HOMELINE TWINDRIVE TD2 user manual

Cleco DGD mPro400GC Quick setup guide

Cleco

Cleco DGD mPro400GC Quick setup guide

Soundoff Signal 400 Series manual

Soundoff Signal

Soundoff Signal 400 Series manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.