Ensure that test tubes are supported from the bottom and not hanging by their caps. Use inserts as follows:
Set the speed and time as recommended on page 4.
Press the START/STOP button to run the cycle. The unit will stop on its own, beep, and open the lid upon completion.
To program P1 and P2 buttons, select desired speed and time. Hold programmable button for 3 seconds. Once the
programmable button has been set, the unit will beep as a confirmation.
WARNING: Failure to secure rotor could lead to personal injury or damage to the centrifuge.
CAUTION: Spinning unbalanced loads could damage unit and destroy samples.
WARNING: Use 6-place test tube rotor ONLY with this centrifuge. The unit is not designed to accept additional rotors.
Use of other rotors could lead to personal injury or damage to the centrifuge.
Loading and Operation Continued
Care, Maintenance, and Troubleshooting
3
2
The Zip-IQ motor is designed to be maintenance-free. With proper care, this centrifuge will provide years of service.
However, if repairs should be needed, please contact LW Scientific.
Use only quality test tubes that are rated for the g-forces utilized. Lower-quality tubes may fracture and allow contents
to leak out of tube.
Never force tubes into the rotor. The rotor was designed to hold the most common sized test tubes.
Clean with common laboratory disinfectants regularly. Do not allow moisture to seep into the centrifuge and do not
immerse the electrical components in any liquid during the cleaning process.
Because of safety issues with high g-forces in a centrifuge, it is recommended that rotors be inspected monthly for
wear and fatigue. If there is any indication of wear, the rotor should be removed from service. Contact LW Scientific
for return instructions so the rotor can be evaluated by a technician for repair or replacement. After 2 years of service,
it is recommended that rotors be returned to LW Scientific for inspection or replacement.
Following these procedures will ensure safety of lab personnel as well as extend the life of the centrifuge.
CAUTION: If corrosion, scratches, or other abnormalities are found on rotor or tube shields, discontinue use and
contact our service center.
1
3
4
5
Black Inserts
5ml-6ml tubes (13mm x 100mm)
These support tall, skinny tubes
No Inserts
7ml-15ml tubes(16mm x100mm - 120mm)
Use the openings in the rotor
White Inserts
3ml-4ml tubes (13mm x 75mm)
These support short, skinny tubes
Note:
Not hanging
by cap
Note:
Not hanging
by cap
Green Inserts
0.5ml - 2.0ml microtubes
Microtubes ARE designed to hang
by the rim of the tube.
Note:
Hanging
by cap
4
6