
Recommended Fluid Speeds & Times
G-Force and Spin Time are the most important considerations for proper fluid separations. G-force is a function of radius and speed,
and varies with different centrifuges and rotor configurations. The following are commonly recommended separation settings.
Please refer to your tube manufacturer and/or medical procedures manuals for the correct G-force and spin times for each fluid,
tube type, and procedure.
Blood 4,000 rpm (2,682 g) 8-10 min.
Fecals 1,300 rpm (280 g) 6 min.
Urine 1,600 rpm (400 g) 5 min.
Semen 1,600 rpm (400 g) 10 min.
Care & Maintenance
Use only high quality test tubes. Lower quality or inexpensive glass or plastic tubes may fracture and release their contents into the
tube chamber. Make sure to know the maximum force allowed for the tubes.
Never force a tube into the tube shield. The tube shields and cushions were designed to accommodate most common sizes of tubes.
Keep the rotor inserts clean. If a tube breaks inside an insert, clean all the debris from the insert, bucket, and bowl and disinfect.
Motor and electrical maintenance: The ComboXL uses a maintenance-free brushless motor and its bearings are permanently
lubricated. It should not need servicing for the life of the unit. Likewise, the electrical components were designed for high reliability
and should not need regular service. However, if repairs are needed, please contact LW Scientific.
Because of the safety issues with high g-forces in a centrifuge, it is recommended that rotors, buckets, and inserts be inspected every
6 months for corrosion and fatigue. If there is any indication of wear, the rotor, buckets, and inserts should be removed from service.
Contact LW Scientific for return instructions, so the rotor can be evaluated by an LW Scientific technician for repair or replacement.
It is also recommended that after 2 years of service rotors, buckets, and tube shields be returned to LW Scientific for inspection.
Following these procedures will ensure safety of lab personnel as well as extend the life of the centrifuge.
Troubleshooting
No Power:
• Ensure power switch on right side is on.
• Ensure outlet has power or move to another outlet on a different circuit.
• Disconnect power cord from outlet and check fuse located on rear of unit. Visually inspect fuse filament, if broken or glass is
discolored, replace fuse.
Wobbles or shakes:
• Ensure unit is properly balanced.
• To test unit is balance, ensure rotor, tube shields and inserts are clean, and run unit without samples.
• If unit is still imbalanced, remove components (e.g., inserts, then tube shields) and run to determine where the
imbalance is occurring.
• If you still experience imbalance, contact LW Scientific.
Makes excessive noise:
• Ensure rotor and rotor retention screw is secure.
Breaking tubes:
• Ensure your tubes are adequate for the speed/G-force selected.
• Discard tubes and try new ones.
Lid won’t open:
• Ensure unit is powered on.
• If unit does not have power, use included release tool or small screwdriver. The unit has a manual release port on the right side.
Gently push the release tool into the port while applying light pressure down on the lid to release.
Memory Function
There are 9 memory presets available for each rotor selected above. Press MEMO. button once and the memory icon will flash. Use
the UP arrows to select which memory preset number (1-9). While flashing, you can set the appropriate speed and time. Press
MEMO. again to save the settings. Note: The memory presets are tied to the rotor selection, and will change as you change rotors.
100 Lauman Lane, Suite A, Hicksville, NY 11801
Tel: (877) 877-7274 | Fax: (516) 801-2046
www.microscopeinternational.com