5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-2465 Rev.5 P. 1 / 2
INSTRUCTION MANUAL
POTENTIOMETER TRANSMITTER MODEL M5MS
BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner ...............................................................(1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■CONFORMITY WITH EU DIRECTIVES
•The equipment must be mounted inside a panel.
•The actual installation environments such as panel con-
figurations, connected devices, connected wires, may af-
fect the protection level of this unit when it is integrated
in a panel system. The user may have to review the CE
requirements in regard to the whole system and employ
additional protective measures to ensure the CE conform-
ity.
■POWER INPUT RATING & OPERATIONAL RANGE
•Locate the power input rating marked on the product and
confirm its operational range as indicated below:
85 – 264V AC rating: 85 – 264V, 47 – 66 Hz, approx. 2 – 3VA
24V DC rating: 24V ±10%, approx. 2W
■GENERAL PRECAUTIONS
•Before you remove the unit or mount it, turn off the power
supply and input signal for safety.
■ENVIRONMENT
•Indoor use.
•When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
•Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
•Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 0 to 90% RH
in order to ensure adequate life span and operation.
■WIRING
•Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
•Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
•Install lightning surge protectors for those wires connect-
ed to remote locations. For 24V DC power supply line,
choose a surge protector with its maximum surge volt-
age 40V or less between lines. Recommended M-System
model: MDP-D24.
■AND ....
•The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
Specifications Power LED
Terminal Cover
Zero Adj.
Span Adj.
Body
INSTALLATION
Set the unit so that its DIN rail adapter is at the bottom.
■MOUNTING THE UNIT ON A DIN RAIL
A)Hang the upper hook at the rear side of unit on the DIN
rail.
B)Push in the lower in keeping pressing the unit to the DIN
rail.
A
B
DIN Rail
■REMOVING THE UNIT
A)Push down the DIN rail adaptor using a minus screw-
driver.
B)Pull out the lower part of the unit.
C)Remove the upper part from the DIN rail.
A
C
B
Spring Loaded
DIN Rail Adaptor
M5MS
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-2465 Rev.5 P. 2 / 2
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the con-
nection diagram on the front of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
94 (3.70)3 (.12)
25 (.98) 41 (1.61) [3.3 (.13)]
• When mounting, no extra space is needed between units.
DIN RAIL
35mm wide
8–M3.5 SCREW
7.3 (.29)
12
34
56
78
■CONNECTION DIAGRAM
3
2
1
max.
min.
+
−
POTENTIOMETER
+
–
OUTPUT
5
6
4
3
2
8
7U(+)
V(–)
POWER
WIRING INSTRUCTIONS
■SCREW TERMINAL
Torque: 0.8 N·m
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 8 with a multimeter.
3) Input: Check voltage across the terminal 3(+) – 2(–) with
a voltmeter to show 0 V at 0% potentiometer input and
the same voltage as that across 4(+) – 2(–) at 100% input.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory with the total resist-
ance input, therefore you do not need any calibration if you
use the potentiometer’s total resistance.
When you do not use the total resistance or in case of regu-
lar calibration, adjust the output as explained in the fol-
lowing.
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.

Other M-system Transmitter manuals

M-system Mini-M M2XR2 User manual

M-system

M-system Mini-M M2XR2 User manual

M-system M2EXV User manual

M-system

M-system M2EXV User manual

M-system M6SXU User manual

M-system

M-system M6SXU User manual

M-system Mini-MW Series User manual

M-system

M-system Mini-MW Series User manual

M-system W5FV User manual

M-system

M-system W5FV User manual

M-system M5CT User manual

M-system

M-system M5CT User manual

M-system M6SCTC User manual

M-system

M-system M6SCTC User manual

M-system Mini-M M2VT User manual

M-system

M-system Mini-M M2VT User manual

M-system M5XU-1-R User manual

M-system

M-system M5XU-1-R User manual

M-system 6R User manual

M-system

M-system 6R User manual

M-system B3VS/2 User manual

M-system

M-system B3VS/2 User manual

M-system W2VABS User manual

M-system

M-system W2VABS User manual

M-system Mini-M M2XM2 User manual

M-system

M-system Mini-M M2XM2 User manual

M-system W2VF User manual

M-system

M-system W2VF User manual

M-system KP User manual

M-system

M-system KP User manual

M-system M5CT-5AM User manual

M-system

M-system M5CT-5AM User manual

M-system M6SVS User manual

M-system

M-system M6SVS User manual

M-system 10SP User manual

M-system

M-system 10SP User manual

M-system MXS User manual

M-system

M-system MXS User manual

M-system 6PA User manual

M-system

M-system 6PA User manual

M-system MXT User manual

M-system

M-system MXT User manual

M-system W2TS User manual

M-system

M-system W2TS User manual

M-system 10VS User manual

M-system

M-system 10VS User manual

M-system 10TG User manual

M-system

M-system 10TG User manual

Popular Transmitter manuals by other brands

Emerson Rosemount 644 Reference manual

Emerson

Emerson Rosemount 644 Reference manual

FUTABA 9CAP SUPER instructions

FUTABA

FUTABA 9CAP SUPER instructions

Iflight SucceX-Force 5.8GHz manual

Iflight

Iflight SucceX-Force 5.8GHz manual

Evikon PluraSens E2608-N2O user manual

Evikon

Evikon PluraSens E2608-N2O user manual

Thies CLIMA 5.4041 1.00 Series Instructions for use

Thies CLIMA

Thies CLIMA 5.4041 1.00 Series Instructions for use

Bell 28E ADJUSTMENTS

Bell

Bell 28E ADJUSTMENTS

GF Signet 8850-3 manual

GF

GF Signet 8850-3 manual

Ecreso Helios FM 20/100 W user manual

Ecreso

Ecreso Helios FM 20/100 W user manual

Rexing FMT2 user manual

Rexing

Rexing FMT2 user manual

Supersonic Gate and Door Automation MUC300 installation instructions

Supersonic Gate and Door Automation

Supersonic Gate and Door Automation MUC300 installation instructions

Endress+Hauser mycom COM 121 operating instructions

Endress+Hauser

Endress+Hauser mycom COM 121 operating instructions

Fuji Electric FCX-AII-VG Series instruction manual

Fuji Electric

Fuji Electric FCX-AII-VG Series instruction manual

Niko 05-301-4X manual

Niko

Niko 05-301-4X manual

Optelecom-nkf 9000 Series Installation and operation manual

Optelecom-nkf

Optelecom-nkf 9000 Series Installation and operation manual

BD Sensors EP 500 Mounting instructions

BD Sensors

BD Sensors EP 500 Mounting instructions

Sony WRT-808A operating instructions

Sony

Sony WRT-808A operating instructions

FONESTAR BT-CONVERTER-MULTI quick start guide

FONESTAR

FONESTAR BT-CONVERTER-MULTI quick start guide

Dwyer Instruments 629 Series Installation and operating instructions

Dwyer Instruments

Dwyer Instruments 629 Series Installation and operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.