BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
Signal conditioner (body + 2 CJC sensors) ........................(1)
Mounting screw (M3.5 × 10)...............................................(2)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
24V DC rating: 24V ±10%, approx. 50mA
■GENERAL PRECAUTIONS
• Before you remove the unit or mount it, turn off the power
supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
• Be sure to put the terminal cover on while the power is
supplied.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
Output ch.1
Output ch.2 Zero Adj.
Body
Specifications
Connection Diagram
(side)
Output ch.1
Span Adj. Output ch.2
Terminal Cover
INSTALLATION
Use Standard Rack (model: 15BX).
THERMOCOUPLE TRANSMITTER MODEL 15TS
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0612 Rev.6 P. 1 / 3
INSTRUCTION MANUAL
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
130 (5.12)
3 (.12)
51 (2.01)
82 (3.23)
2–4.3 (.17) dia.
MTG HOLE
TERMINAL COVER
7–M3.5 SCREW
8 (.31)
91 (3.58)
99 (3.90 )
25 (.98)
12
3
56
78
■CONNECTION DIAGRAM
CONNECTOR
POWER
Ch.1 OUTPUT
Ch.2 OUTPUT
+
–
Ch.1 T/C
CJC
SENSOR
metal leg
comp. leadwire
1
2
+
–
model: CJM
+
–
Ch.2 T/C
CJC
SENSOR
metal leg
comp. leadwire
5
6
+
–
■TERMINAL ASSIGNMENT
12
56
78
CJC SENSOR
(model: CJM)
15TS
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0612 Rev.6 P. 2 / 3
WIRING INSTRUCTIONS
■SCREW TERMINAL
Torque: 0.8 N·m
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage supplied to the rack
(model: 15BX). For the DC power source, be sure that the
ripple level is within 10% p-p.
3) Input: Check that the input voltage is within 0 – 100% of
full-scale.
If the thermocouple or its extension wires are broken, the
output goes over 100% (below 0% with downscale) due to
burnout function. Check leadwires in such a case.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
Use extender card (model: 10EC) for SPAN adjustment.
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
5) Go through the same procedure for the Ch. 2 Output.
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.
15TS
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-0612 Rev.6 P. 3 / 3

Other M-system Transmitter manuals

M-system M5VS Series User manual

M-system

M-system M5VS Series User manual

M-system CT TRANSMITTER User manual

M-system

M-system CT TRANSMITTER User manual

M-system M5XCRS User manual

M-system

M-system M5XCRS User manual

M-system 10VS User manual

M-system

M-system 10VS User manual

M-system Mini-M M2AMS User manual

M-system

M-system Mini-M M2AMS User manual

M-system 27RS User manual

M-system

M-system 27RS User manual

M-system M3LU2 User manual

M-system

M-system M3LU2 User manual

M-system 10MS User manual

M-system

M-system 10MS User manual

M-system F2TS User manual

M-system

M-system F2TS User manual

M-system 6M User manual

M-system

M-system 6M User manual

M-system MFS2 User manual

M-system

M-system MFS2 User manual

M-system B3VS/2 User manual

M-system

M-system B3VS/2 User manual

M-system KCNA User manual

M-system

M-system KCNA User manual

M-system M3LM User manual

M-system

M-system M3LM User manual

M-system W2VS User manual

M-system

M-system W2VS User manual

M-system 10TS User manual

M-system

M-system 10TS User manual

M-system M5VF User manual

M-system

M-system M5VF User manual

M-system F2VS User manual

M-system

M-system F2VS User manual

M-system W2TS User manual

M-system

M-system W2TS User manual

M-system M5PT User manual

M-system

M-system M5PT User manual

M-system Mini-M M2VF2 User manual

M-system

M-system Mini-M M2VF2 User manual

M-system M3LM User manual

M-system

M-system M3LM User manual

M-system VOS2T-R User manual

M-system

M-system VOS2T-R User manual

M-system M2VF User manual

M-system

M-system M2VF User manual

Popular Transmitter manuals by other brands

Greyline Instruments LIT25 user guide

Greyline Instruments

Greyline Instruments LIT25 user guide

Sage BT77D user manual

Sage

Sage BT77D user manual

sauermann Si-CPE320 user manual

sauermann

sauermann Si-CPE320 user manual

Comet T0213D instruction manual

Comet

Comet T0213D instruction manual

WIKA IPT-1 series operating instructions

WIKA

WIKA IPT-1 series operating instructions

DVDO DVDO-RS-3 user manual

DVDO

DVDO DVDO-RS-3 user manual

Transystem PL-12-1-28ND Operation manual

Transystem

Transystem PL-12-1-28ND Operation manual

Dwyer Instruments 605 Installation and operating instructions

Dwyer Instruments

Dwyer Instruments 605 Installation and operating instructions

ESI Cases 4FM915 user manual

ESI Cases

ESI Cases 4FM915 user manual

Telkom Telmor MTX-1 Series user manual

Telkom Telmor

Telkom Telmor MTX-1 Series user manual

Pretorian SimplyWorks SEND instructions

Pretorian

Pretorian SimplyWorks SEND instructions

Emerson PT5N FLR Series Operating	 instruction

Emerson

Emerson PT5N FLR Series Operating instruction

JUMO 404304 Series operating instructions

JUMO

JUMO 404304 Series operating instructions

HumanTechnik lisa operating instructions

HumanTechnik

HumanTechnik lisa operating instructions

Shure T1 Service manual

Shure

Shure T1 Service manual

Sonus Faber TRANSMITTER manual

Sonus Faber

Sonus Faber TRANSMITTER manual

Sutron XLink 100/500 quick start guide

Sutron

Sutron XLink 100/500 quick start guide

Community Controls 390LMR2V manual

Community Controls

Community Controls 390LMR2V manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.