Maico MA 33 User manual

Operation Manual
MA 33

Operation Manual MA 33
8100092 Rev. 9
i
19/08/2019

Operation Manual MA 33
8100092 Rev.
1
08/19/2019
Table of Contents
1Introduction...........................................................................................................................4
1.1 General...........................................................................................................................4
1.2 Intended Use and Indications for Use...........................................................................4
1.3 Contraindications of Use................................................................................................4
1.4 Features and Benefits....................................................................................................4
1.5 Description......................................................................................................................5
1.6 PC-System Requirements.............................................................................................6
2For your Safety.....................................................................................................................7
2.1 How to Readthis Operation Manual.............................................................................7
2.2 Customer Responsibility................................................................................................8
2.3 Manufacturer’s Liability ..................................................................................................8
2.4 Regulatory Symbols.......................................................................................................9
2.5 Device Control ...............................................................................................................9
2.6 General Precautions....................................................................................................10
2.7 Electrical Safety and Measuring Security...................................................................10
2.8 Electromagnetic Compatibility (EMC).........................................................................11
3Warranty, Maintenance and After-Sales Service..........................................................13
3.1 Warranty.......................................................................................................................13
3.2 Maintenance.................................................................................................................15
3.3 Cleaning and Disinfection Recommendations...........................................................15
3.4 Components and Replacement Parts ........................................................................16
3.5 Recycling and Disposal................................................................................................16
4Unpacking and Installation...............................................................................................17
4.1 Unpacking the System.................................................................................................17
4.2 Hardware and Components........................................................................................19
4.3 MA 33 Software Installation.........................................................................................20
5Operating the MA 33..........................................................................................................24
5.1 Getting Started with the MA 33...................................................................................24
5.2 Starting the MA 33 Software........................................................................................24
5.3 Turning Off the MA 33..................................................................................................31
5.4 Using the MA 33 Software...........................................................................................31
5.5 Tone Audiometry..........................................................................................................33
5.6 Speech Audiometry (MA 33 Speech).........................................................................54
5.7 Pilot Test.......................................................................................................................64

Operation Manual MA 33
8100092 Rev. 9
2
19/08/2019
6Technical Specifications...................................................................................................74
6.1 MA 33 Hardware and Software...................................................................................74
6.2 Calibration Values and Maximum Levels...................................................................77
6.3 Connections..................................................................................................................80
6.4 Steckerbelegung..........................................................................................................80
6.5 Electromagnetic Compatibility (EMC).........................................................................81
6.6 Electrical Safety, EMC and Associated Standards....................................................84
6.7 Checklist for subjective Audiometer Testing ..............................................................85

Operation Manual MA 33
8100092 Rev. 9
3
19/08/2019
Title: MA 33 –Operation Manual
Date of issue/last revision: 19/08/2019
MAICO Diagnostics GmbH
Sickingenstr. 70-71
10553 Berlin
Germany
Tel.: + 49 30 / 70 71 46-50
Fax: + 49 30 / 70 71 46-99
E-mail: sales@maico.biz
Internet: www.maico.biz
Copyright © 2019 MAICO Diagnostics
All rights reserved. No part of this publication may be reproduced or transmitted in any
form or by any means without the prior written permission of MAICO. The information in
this publication is proprietary to MAICO.
Compliance
MAICO Diagnostics is an ISO 13485 certified corporation.
Caution for USA
Federal Law restricts this device to sale by or on the order of a licensed
medical professional.
Trademark Notice
Windows is a registered trademark of Microsoft Corp.

Operation Manual MA 33
8100092 Rev. 9
4
19/08/2019
1 Introduction
This section offers you important information about:
▪the intended use of the device and indications for use
▪contraindications
▪features and benefits
▪a description of the device functions
1.1 General
Thank you for selecting one of our quality products from the MAICO family range. The
MA 33 is designed and manufactured to meet all quality and safety requirements, and
has been certified with the CE-symbol according to Medical Device Directive
(93/42/EEC).
Particular attention has been taken during the designing phase of the MA 33 to ensure its
user-friendliness, meaning that its operation is simple, easy to learn and to understand.
As all the functions are software-controlled, upgrading the software and/or adding
additional functions at a later date, will be simple and cost-effective. By purchasing the
MAICO MA 33, you have made a decision towards long-term investment.
This operation manual aims to make learning and understanding the different MAICO MA 33
functionsasquickandaseasyaspossible.Shouldyouencounteranyproblemsorhaveideasfor
anyfurther improvements,weareonlya phone callaway.Pleasedo nothesitate tocontactus.
Your MAICO-Team
NOTE: Operation of the MA 33 requires knowledge of the Windows operating system.
1.2 Intended Use and Indications for Use
Screening audiometers are designed for determine hearing thresholds levels. The
instrument is intended for all patient populations over 5 years age and able to response
to test signal in a rational way.
Audiometers are intended to be used by an audiologist, hearing healthcare professional,
or trained technician.
1.3 Contraindications of Use
The patient is too young, sick oruncooperative toperform the tasks.
1.4 Features and Benefits
1.4.1 General
The MA 33 is available in 3 versions:
•MA 33 AC
•MA 33 BC
•MA 33 Speech

Operation Manual MA 33
8100092 Rev. 9
5
19/08/2019
The MA 33 gives the benefit of:
•PC-controlledaudiometerforAirConduction
•BoneConduction(MA33BConly)
•Pulse and Warble Tone
•Masking
•Individual test configurations
•Compatibility with MAICO Database, Noah 4, EssiConnect, or your existing
practice management software via BDT/GDT
•Power supply via USB
•Keyboard and/or mouse operation
•SISI (MA 33 BC and MA 33 Speech only)
•Speech Audiometry (MA 33 Speech only)
•German, English, French, Italian, Spanish, Netherlands, Polish
Additionally available tests for each version are:
•Module Pilot Test (Select Picture Audiometry for children)
•Module Hughson-Westlake (automatic test according to Hsughson-Westlake)
1.4.2 Language Pack for Module Pilot Test
The MA 33 with the Module Pilot Test comes with the following languages
•International configuration: Afrikaans, Arabic, Basque, Catalan, Croatian, Czech,
Danish, Dutch, English, Finnish, French, Galician, German, Greek, Hungarian,
Italian, Japanese, Korean, Norwegian, Polish/Polish (Audifon), Portuguese,
Romanian, Russian, Serbian, Slovakian, South Sotho, Spanish, Swedish, Swiss
German, Turkish, Vietnamese, Xhosa, Zulu
•US configuration: English (US) and Spanish (US)
1.5 Description
1.5.1 General
1.5.2 Air Conduction Testing
Hearing threshold levels can be determined by presenting test signals to the test subject
with the included headphones (Air Conduction –AC). The purpose of AC audiometry is
to establish the hearing sensitivity at various frequencies. The test can specify the AC
loss but cannot distinguish between a conductive versus a sensorineural abnormality.

Operation Manual MA 33
8100092 Rev. 9
6
19/08/2019
1.5.3 Bone Conduction Testing
Hearing threshold levels can be determined by presenting test signals to the test subject
with the included bone conductor (Bone Conduction –BC). The purpose of BC
audiometry is to establish the hearing sensitivity at various frequencies. The test can
specify the BC loss in combination with AC loss it can distinguish between conductive
versus a sensorineural abnormality.
1.5.4 Speech Audiometry
Speech recognition screening incorporates the child listening to a series of words at
different decibel levels and point to the picture. The purpose of speech audiometry is to
establish an ear specific hearing level when standard pure tone testing cannot be
performed. The level at which a patient can understand spoken language can be a valuable
screening tool, especially with young children. This speech recognition level can be
determined easily by the MA 33.
1.5.5 Masking
Masking is required if there is a notable threshold difference between the left and right
ears. It is possible for sound to be transmitted to both ears via bone conduction while
testing the poorer ear. This is called “crossover”.
Crossover occurs often while testing bone conduction, but it can also occur during air
conduction testing. Relevant to crossover is the sound level received by the opposite ear.
The difference between the original test signal in the test ear and the received signal in
the opposite ear is called “interaural attenuation”.
For Bone Conduction measurements the interaural attenuation is 0 dB to 15 dB. Bone
Conduction crossover is therefore possible even with a slight difference in hearing loss
between ears.
1.6 PC-System Requirements
PC connection: USB port
Operating system: Windows®10 SP1 (x86 and x64)
Windows®8 / 8.1 (x86 and x64)
Windows®7 SP1 (x86 and x64)
.NET Framework 3.5
Processor: 2 GHz Intel Core 2 Duo CPU
Memory: 2 GB RAM
Graphic display: 1280 x 1024 (optimal), min. 1024 x 768
Silent PC for use in audiometric room
Optional use of a touchscreen for certain functions.

Operation Manual MA 33
8100092 Rev. 9
7
19/08/2019
2 For your Safety
This section offers you important information about:
▪how to read the operation manual
▪where to spend special attention
▪the customer responsibility
▪the explanation of all regulatory symbols used
▪important cautions and warnings that have to be considered during
the whole time handling and operating your device
2.1 How to Read this Operation Manual
This operation manual contains information pertinent to the use of the MAICO MA 33
system including safety information as well as maintenance and cleaning
recommendations.
READ THIS ENTIRE OPERATION MANUAL BEFORE ATTEMPTING
TO USE THIS SYSTEM!
Use this device only as described in this manual.
All images and screenshots are only examples and may differ in
appearance from the actual device settings.
In this operation manual, the following two labels identify potentially dangerous or
destructive conditions and procedures:
The WARNING label identifies conditions or practices that may
present danger to the patient and/or user.
The CAUTION label identifies conditions or practices that could
result in damage to the equipment.
NOTE: Notes help you identify areas of possible confusion and avoid potential
problems during system operation.
WARNING
CAUTION

Operation Manual MA 33
8100092 Rev. 9
8
19/08/2019
2.2 Customer Responsibility
All safety precautions given in this operation manual must be observed at all times.
Failure to observe these precautions could result in damage to the equipment and injury
to the operator or subject.
The employer should instruct each employee in the recognition and avoidance of unsafe
conditions and the regulations applicable to his or her work environment to control or
eliminate any hazards or other exposure to illness or injury.
It is understood that safety rules within individual organizations vary. If a conflict exists
between the material contained in this operation manual and the rules of the organization
using this device, the more stringent rules should take precedence.
This product and its components will perform reliably only when
operated and maintained in accordance with the instructions
contained in this operation manual, accompanying labels,
and/or inserts. A defective product should not be used. Make
sure all connections to external accessories are snug and
secured properly. Parts which may be broken or missing or are
visibly worn, distorted, or contaminated should be replaced
immediately with clean, genuine replacement parts
manufactured by or available from MAICO.
NOTE: Customer responsibility includes proper maintenance and cleaning of the device
(see sections 3.2 and 3.3). Breach of customer responsibility can lead to limitations of
Manufacturer’s Liability and Warranty (see sections 2.3 and 3.1).
NOTE: In the unlikely case of a serious incident, inform MAICO as well as your local
distributor.
2.3 Manufacturer’s Liability
Usage of the device in a way deviant from the intended use will lead to a limitation or
termination of the manufacturer’s liability in case of damage. Improper use includes
disregarding the operation manual, the operation of the device by underqualified
personnel as well as making unauthorized alterations on the device.
WARNING

Operation Manual MA 33
8100092 Rev. 9
9
19/08/2019
2.4 Regulatory Symbols
The following Table 1 gives an explanation of the symbols used on the device itself, on the
packaging and the accompanying documents including the Operation Manual.
Table 1 Regulatory Symbols
REGULATORY SYMBOLS
SYMBOL
DESCRIPTION
Serial number
Date of manufacture
Manufacturer
Caution, consult accompanying documents
Warning, consult accompanying documents
Returntoauthorizedrepresentative,specialdisposalrequired
Reference number
Applied part type B according to IEC 60601-1
Refer to operation manual (mandatory)
Keep away from rain
Transport and storage temperature range
Transport and storage humidity limitations
Transport and storage atmosphericpressure limitations
Voltage transformer
Conforms to European Medical Device Directive 93/42/EEC
ETL listed mark
Logo
2.5 Device Control
The user of the device should perform a subjective device check once a week
(ISO 8253-1). See Section 6.5 for a checklist.
For annual calibration see Sections 2.6 and 3.2.

Operation Manual MA 33
8100092 Rev. 9
10
19/08/2019
2.6 General Precautions
Before starting a measurement make sure, that the device works
properly.
Use and store the device indoors only. For operation, storage and
transport conditions see table in section 6.
For operation in certain places, a recalibration maybe necessary.
No modification of this equipment is allowed.
Equipment is not user repairable. Repairs must be performed by a
qualified service representative only. No modifications of the
equipment are allowed by anyone other than a qualified MAICO
representative. Modification of the equipment could be hazardous.
No part of the equipment can be serviced or maintained while in
use with the patient.
Do not drop or otherwise cause undue impact to this device. If the
device is dropped or otherwise damaged, return it to the
manufacturer for repair and/or calibration. Do not use the device if
anydamage is suspected.
Calibration of the device: The device and the transducers
complement each other and share the same serial number (i.e.
MA7663252). Therefore, the device shall not be used with any
other transducer prior to recalibration. Recalibration also needs to
be conducted, when a defected headphone is replaced.
Uncalibrated devices may lead to faulty measurement results and
could even damage the hearing of the examinee.
Do not immerse the device in any fluids. Should the user suspect
fluids have contacted the system components or accessories, the
unit should not be used until deemed safe by a MAICO certified
service technician.
2.7 Electrical Safety and Measuring Security
The device complies with international standard EN 60601-1
and conforms with Type B applied part.
The protection class according EN 60601-1 depends on the
used computer (USB connection).
In Case of Emergency
In case of emergency, disconnect the device from the
computer.
Position the device in such a way that it can be easily
disconnected from the USB cable at any time. Do not use the
device if the connection cable is damaged.
WARNING
WARNING
WARNING
WARNING
WARNING

Operation Manual MA 33
8100092 Rev. 9
11
19/08/2019
To operate the device, establishing a PC-connection via USB is
required.
This equipment is intended to be connected to other equipment
thus forming a Medical Electrical System. External equipment
intended for connection to signal input, signal output or other
connectors shall comply with the relevant product standard e.g.
IEC 60950-1 for IT equipment and the IEC 60601-series for
medical electrical equipment. In addition, all such combinations
–Medical Electrical Systems –shall comply with the safety
requirements stated the general standard IEC 60601-1, edition
3.1, clause 16.
Any person who connects external equipment to signal input,
signal output or other connectors has formed a Medical
Electrical System and is therefore responsible for the system
to comply with the requirements. If in doubt, contact qualified
medical technician or your local representative. If the device is
connected to a PC (IT equipment forming a system) ensure not
to touch t
he patient whileoperating the PC or PC equipment.
The device is not intended for operation in areas with an
explosion hazard. Do NOT use the device in a highly oxygen-
enriched environment, such as a hyperbaric chamber, oxygen
tent, etc. If the device is not used switch it off and disconnect it
from the power supply.
Never short-circuit the terminals.
In order to maintain a high level of safety and to ensure the
device works properly, it is necessary to have the device
checked according to the medical electrical safety standard
IEC 60601-1 by a qualified service technician at least every
twelve months. For more information see Section 3.2.
The use of non-calibrated devices can lead to incorrect test
results and is not advisable.
Prevent cable breakage: cables must not be bend or buckled.
2.8 Electromagnetic Compatibility (EMC)
This device is suitable in hospital environments except for
near active HF surgical equipment and RF shielded rooms of
systems for magnetic resonance imaging, where the intensity
of electromagnetic disturbance is high.
WARNING
WARNING
CAUTION
WARNING

Operation Manual MA 33
8100092 Rev. 9
12
19/08/2019
The device fulfills the relevant EMC requirements.
Avoid unnecessary exposure toelectromagnetic fields, e.g. from
mobile phones etc.
The MA 33 has been verified by an independent laboratory to
conform to international standards for EMC (electromagnetic
emissionsandimmunity).Theuserisadvisedtoavoidinstallation
and use of this device in proximity with other devices or
equipment that may emit or be susceptible to electromagnetic
interference, including mobile phones. If the device is used
adjacent to other devices or equipment, the user is instructed to
verify that no disturbanceisfound in the operation of thisorother
equipment in proximity.
Please also refer to EMC consideration in Section 6.5.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and
result in improper operation.
The list of accessories, transducers and cables can be found
in Section 6.5 of this instruction.
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used
no closer than 30 cm (12 inches) to any part of the MA 33,
including cablesspecified by the manufacturer.
Otherwise, degradation of the performance of this equipment
could result in improper operation.
WARNING
WARNING
WARNING

Operation Manual MA 33
8100092 Rev. 9
13
19/08/2019
3 Warranty, Maintenance and After-Sales Service
This Section offers you important information about:
▪warranty conditions
▪maintenance
▪cleaning and disinfection recommendations
▪component and replacement parts
▪recycling and disposal of the device
3.1 Warranty
3.1.1 General
The MAICO device is guaranteed for at least one year. Ask your authorized local
distributor for more information.
This warranty is extended to the original purchaser of the device by MAICO through the
distributor from whom it was purchased and covers defects in material and workmanship for a
period of at least one year from date of delivery to the originalpurchaser.
The device shall only be repaired and serviced by your distributor or by an authorized service
center. Opening the device case will void the warranty.
Do not modify this equipment without authorization of the
manufacturer.
In the event of repair during the guarantee period, please enclose evidence of purchase with
the device.
3.1.2 Ownership, Warranty and Disclaimer (Software)
Ownership
The MA 33 Software (hereinafter the “SOFTWARE”) is solely owned by MAICO
Diagnostics GmbH, Sickingenstr. 70-71, D-10553 Berlin, Germany. By purchasing the
SOFTWARE the buyer is entitled the right of usage, but not ownership of the
SOFTWARE. The SOFTWARE is to be used in accordance to the agreed terms of usage
provisioned by MAICO.
Copyrights
MAICO’s ownership of the SOFTWARE covers worldwide and is therefore, protected against
any unauthorized copying of the SOFTWARE. Non conformity of use of the SOFTWARE is
strictly prohibited.
Restrictions
You may not:
Reverse engineer or attempt in any mannerto discover the source code of the SOFTWARE.
Attempt to defeat any mechanisms in the SOFTWARE, including those mechanisms
responsible for password protection of data and limiting the number of concurrent users.
WARNING

Operation Manual MA 33
8100092 Rev. 9
14
19/08/2019
Rent, lease, sublicense or in any manner, copy or transfer (except as permitted above) the
SOFTWARE.
Obscure or obliterate any MAICO copyright or trademark notices which appear on the
SOFTWARE, the documentation, the screen-display, or otherwise in connection with the
SOFTWARE.
MAICO specifically calls your attention to the fact that, any violation or infringement of above
restrictions will result in legal action.
The SOFTWARE can be used by any number of users, on any number of computers, and in
anyplace, provided but not on more than one display screen at the same time.
Limited Warranty
MAICO warrantsthat anyphysicalmedia and physicaldocumentation providedbyMAICO are
free of defects in materials and workmanship. This limited warranty is effective for a period of
ninety (90) days from the original purchase date.
IfMAICOreceivesnotificationwithinthewarrantyperiodofdefectsinmaterialsorworkmanship
and determines that such notifications are correct, MAICO will replace defective media or
documentation.
Do not return any product until you have obtained authorization to do so from your supplier.
The entire and exclusive liability and remedy for breach of thislimited warranty shall be limited
to replacementofdefective media or documentation supplied by MAICO, and shallnot include
or extend to any claim for or right to recover any other damages, including but not limited to,
lossofprofit,data,oruseoftheSOFTWARE,orspecial,incidentalorconsequentialdamages,
or other similar claims, even if MAICO has been specifically advised of possibility of such
damages. In no event will MAICO’s liability for any damages to you or any other person ever
exceed the lowest list price or the actual price paid for the license to use the SOFTWARE,
regardless of the formof the claim.
Disclaimer
MAICO covers; including but not limited to; all warranties, representations and terms and
conditions, either expressed or implied; under specified terms of use ans application of the
SOFTWARE for its specific purpose. All otherterms and conditions shall not apply.
Furthermore, MAICO does not guarantee that the SOFTWARE or Documentation is free of
bugs, or fulfill the relevant standards, requirement or needs of a user. In this case, all the
warranties, guarantees and terms and conditions on all MAICO delivered physical disk and
documentation shall be limited to the 90 days warranty period.
MAICO is not liable for any third party’s product, disks, SOFTWARE or documentation that is
used in conjunction with MAICO’s SOFTWARE or programs, but is not directly manufactured
or supplied by MAICO.
General Terms and Conditions
Any change made to this Agreement shall be notified in writing, agreed and signed between
both parties, namely the purchaser of the SOFTWARE and a representative of MAICO.
In the event that the essential purpose of the above remedy (limited warranty) is not fulfilled,
all other limited liability including the liability limits and exclusions of damage claims shall
continue to apply.

Operation Manual MA 33
8100092 Rev. 9
15
19/08/2019
This SOFTWARE License Agreement shall be interpreted and construed according to, and
governed by, the laws of Jurisdiction of Federal Republic of Germany.
In the event that any legal or commercial dispute or controversy arising out of, or relating to
this agreement; provided MAICO is in all case violated of the rights, to the SOFTWARE or
otherintellectualpropertyprotectionrightrelatedtotheSOFTWARE;shallbepresentedunder
the Jurisdiction of Federal Republic of Germany in the court of Berlin.
The SOFTWARE is protected under both Copyright Law and the International Copyright
Treaties. Copying of the SOFTWARE is strictly prohibited except for copies made of the
SOFTWARE for backup purposesto protect data loss.
3.2 Maintenance
In order to ensure that the device works properly, it has to be checked and calibrated at
least every twelve months.
The service and calibration must be performed by your dealer or a service center
authorized by MAICO.
When returning the device for repairs or calibration it is essential to send the acoustic
transducers with the device. Please include a detailed description of faults. In order to
prevent damage in transit, please use the original packing when returning the device.
3.3 Cleaning and Disinfection Recommendations
3.3.1 General
It is recommended that parts (device and components like headphones, ear cushions)
which come in direct contact with the patient be subjected to standard cleaning and
disinfecting procedure between patients.
Recommendations for cleaning and disinfection of MAICO device presented in this
document are not intended to replace or contradict policies in effect or procedures
required for infection control at the facility.
If there is nota high infection potential, MAICO recommends:
•Before cleaning always switch off and disconnect the device from power supply
(USB).
•For cleaning use a lightly dampened cloth with soap water solution.
•Disinfect the plastic housing of the MA 33 and its accessories by wiping the
surfaces with wet disinfection wipes. Follow the instructions on the specific
disinfection product.
oWipe before and after each patient
oAfter contamination
•Disinfect computer, keyboard, etc. with wet disinfection wipes:
oonce a week
oafter contamination
owhen polluted

Operation Manual MA 33
8100092 Rev. 9
16
19/08/2019
To avoid damage of the device and its accessories, please
mind the following:
•Do not autoclave or sterilize.
•Do notusethedeviceinthepresence offluid thatcancome
into contactwithanyof theelectroniccomponentsorwiring.
Should the user suspect fluids have entered the system
components or accessories, the unit should not be used
until deemed safe by a MAICO certified service technician.
Do not use hard or pointed objects on the device or its
accessories.
For more detailed cleaning recommendations see the following Section 3.3.2.
3.3.2 Cleaning the Case and Cables
Use caution while cleaning.
Use adamp cloth to clean the plastic parts of the MA 33.
If disinfection is required, use a disinfectant wipe rather than a
sprayproduct. Makesurethatexcessliquid fromthe wipedoes
not seep into any sensitive areas such as connectors and
seams where plastic pieces connect.
Follow the instructions on the disinfection product.
3.4 Components and Replacement Parts
Some reusable components are subject to wearwith use over time. MAICO recommends that
you keep these replacement parts available (as appropriate for your MA 33 device
configuration). Ask your authorized local distributor when accessories need to be replaced.
3.5 Recycling and Disposal
Within the European Union it is illegalto dispose of electric and
electronicwasteasunsortedmunicipalwaste.Accordingtothis,
allMAICOproductssoldafterAugust13,2005,aremarkedwith
a crossed-out wheeled bin. Within the limits of Article (9) of
DIRECTIVE 2002/96/EC on waste electrical and electronic
equipment (WEEE), MAICO has changed their sales policy.To
avoid additional distribution costs we assign the responsibility
for the proper collection and treatment according to legal
regulations to our customers.
Non-European countries
Outside the European Union, local regulations should be
followed when disposing of the product after its useful life.
CAUTION
CAUTION
CAUTION

Operation Manual MA 33
8100092 Rev. 9
17
19/08/2019
4 Unpacking and Installation
This Section provides information on:
▪unpacking the system
▪components
▪becoming familiar with the hardware inclusive connections
▪MA 33 Software Installation
▪how to establish a PC connection
▪how to store the device
4.1 Unpacking the System
Check Box and Contents for Damage
•It is recommended that you unpack your MA 33 carefully making sure that all
components are removed from the packing materials.
•Verify that all components are included as shown on the packing slip included with
your shipment.
•If any component is missing, contact your distributor immediately to report the
shortage.
•If any component appears to be damaged in shipment, contact your distributor
immediately to report it. Do not attempt to use any component or device that
appears to be damaged.
Reporting Imperfections
Notify the carrierimmediately if anymechanical damage isnoted.This will insure that aproper
claim is made. Save all packaging material so the claim adjuster can inspect it as well.
Report Immediately any Faults
Any missing part or malfunction should be reported immediately to the supplier of the
device together with the invoice, serial number, and a detailed report of the problem.
Keep Packaging for Future Shipment
Save all the original packing material and the shipping container so the device can be
properly packed if it needs to be returned for service or calibration (see Section 3.2).

Operation Manual MA 33
8100092 Rev. 9
18
19/08/2019
The MA 33 comes with different components (see Table 2 and Table 3). The availability of
configurationswiththefollowingcomponentsiscountryandversionspecific.Contactyourlocal
distributor for more information.
Table 2 MA 33 Available Components
Available components
Base Unit
MAICO USB Flash Drive Kit with
MAICO Database and MA 33 Software
AC Headphones DD65v2*
AC Headphones DD45*
AC Headphones DD45 with HB7 headband*
BC Headphones B71W*
Patient Response Switch APS3*
USB Cable
Carrying Bag
Living Speech Microphone
Operation Manual
Quick Guide
Software Modules
Module Pilot Test
Module Hughson Westlake
*Applied parts according to IEC60601-1
Table 3 Accessories for Software Pilot Test
Accessories for Software Pilot Test
Picture board
Roll of stickers
This manual suits for next models
3
Table of contents
Other Maico Medical Equipment manuals

Maico
Maico PILOT TEST User manual

Maico
Maico MA 41 User manual

Maico
Maico easyTymp User manual

Maico
Maico MA 41 User manual

Maico
Maico MB 11 Classic User manual

Maico
Maico ERO SCAN User manual

Maico
Maico easyTymp User manual

Maico
Maico MA 41 User manual

Maico
Maico ERO SCAN User manual

Maico
Maico Ero Scan Pro User manual
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual