Instructions for use
1. V Attachment Pin Type
1-1. Detach the leg section from the operating table.
1-2. Attach this product to the leg section insertion shafts of the
operating table.
1-3. Tighten the fixing handle to fix this product.
2. V Attachment Side Rail Type
2-1. Detach the leg section from the operating table.
2-2. Attach this product to the side rails of the operating table.
2-3. Tighten the fixing handle to fix this product.
3. Applicable operating tables
Refer to the brochure "OPERATING TABLE ACCESSORIES" of Mizuho.
Warning/Caution
1. Warning
Ensure all fixing handles are tight. Loose condition can cause the
product to come off or move, which may result in a patient injury.
2. Important caution
2-1. When working with an operating table, take care not to allow this
product to make contact or interfere with the tabletop or with other
tools and appliances used in combination with this product.
Otherwise the product could break.
2-2. When attaching this product to the operating table, make sure to
fully insert this product to the leg section insertion shafts or the side
rails of the operating table. Otherwise the product may come off or
move, which may result in a patient injury.
2-3. Do not fix this product to the operating table with the Traction
Device attached. Be sure to fix this product to the operating table by
itself. If this product is not fully fixed, it may fall.
2-4. For hygiene, be sure to use sterilized drapes on the areas on this
product where the patient comes into contact with it.
Storage/Life
1. Do not store the device in high temperatures or in areas with high
humidity where the temperature has drastic variations.
2. Service life of this product: 7 years
(Subject to the specified maintenance and inspection and is stored
properly.)
Maintenance/Inspection
<By the user>
1. Check before and after use
Check that this product is not damaged or broken before and after
each use.
2. Cleaning and disinfection
Wipe off blood, chemicals, contaminants, and other stains with water,
and clean the device with gauze or other materials soaked in
disinfectants. In addition, use the same disinfectants as the
authorized disinfectants on Mizuho's operating table described in the
operating table operator's manual.
3. In case of a malfunction
When this product is broken, clearly label the device as “Broken”, “Do
not use”, “Need repair”, etc., and contact your local distributor or
Mizuho.
<By agents>
Maintenance and inspection can only be carried out by certified
agents of Mizuho.
Packing
1 piece per pack
Warranty
MIZUHO Corporation will repair defective parts of this product without
charge for one year from the date of delivery/installment except for
cases of damage caused by a third party’s repair, act of nature, improper
use or intentional damage. All other warranty terms and conditions are
subject to regulations of MIZUHO Corporation.
Disposal
This device must be disposed of in accordance with local regulations.
Please contact your local distributor for proper disposal.
Notice
Any serious incident that has occurred in relation to the device should be
reported to the manufacturer and the competent authority in which the
user and/or patient is established.