CAUTION
2
Safety precaution
*
6
Devices and accessories used together with this product
Before using other devices or accessories, thoroughly read the instruction manual
While operating the operating table, check the position of other devices or the
accessories used with them. They may come in contact with each other during the
operation, operating table, devices, and/or accessories may get damaged.
For hygiene, be sure to use sterilized drapes on the areas on this product where
the patient comes into contact with it.
Cleaning and disinfection
After using the operating table, make sure to follow the steps below to clean up
and disinfect the operating table.
upper, side, and back side of the mattresses.
Make sure to use Mizuho authorized disinfectants. The disinfectants are as shown below.
Moving and transporting
Do not move the operating table with a patient on it.
Follow the procedures below to move the operating table.
Before moving the operating table, disinfect the entire operating table in order to
prevent infection.
2. Unlock the brakes, and move the operating table.
* Do not move the operating table by pulling on the head plate or leg plate.
* Do not move the operating table by holding onto the operating handles (right and left).
The operating table should be transported with the following conditions met.
Disinfect the entire operating table before transporting it.
Take measures to prevent it from tipping over, such as lowering the tabletop to
the bottom position.
Actuate the brake.
Suitably position cushioning on the product to prevent it from getting damaged
during transport.
Store the product in a container so that it does not get exposed to dust, and the weather.
Disposal
All components must be disposed of in accordance with local regulations. Please contact
your local distributor for proper disposal.
1.
1.
2.
3.
4.
5.
a)
b)
c)
d)
e)
f)
g)
h)
i)
Sodium hypochlorite 0.1% (halogen containing compound)
Hypo Alcohol (iodine color removing agent)
Chlorhexidine (chlorhexidine gluconate 0.5%)
Benzalkonium chloride (invert soap 10%)
Povidone iodine
Ethanol 80%
Hydrogen peroxide
Saline (0.9%)
Isopropyl alcohol (IPA)
of the devices and make sure that the operating table is not affected adversely.
Before fitting on accessories from third party companies, contact your distributor or
Mizuho. Some accessories cannot be fitted on.
Use a lint-free cloth soaked with proper volume of disinfectant to wipe off the
Wipe off the operating table with a clean dry cloth 15 minutes after disinfecting it.
Check if the handles and levers are in fixed positions, and each section is fixed firmly.
As with step 2, disinfect the surfaces of the tables and side rails.
Detach all the mattresses from the operating table.1.
2.
3.
4.