3/3
Fig.4-2
4-1. Check the leg plate of operating table is in the horizontal position.
4-2. Attach the clampto the side rail of the leg plate.
4-3. Insert the insertion rodto the clamp, and rotate the fixing
handleto secure the insertion rod.
4-4. Attach the cassette board to the board.
4-5. Attach the padto the cassette board.
Warning/Caution
1. Warning
Ensure all fixing handles are tight. Loose condition can cause the
product to come off or move, which may result in a patient injury.
2. Important Caution
2-1. When working with an operating table, take care not to allow this
product to make contact or interfere with the table top or with other
tools and appliances used in combination with this product.
2-2. Use a pair of Knee rests (right and left) to support the patient.
2-3. When using this product to support patient's leg (Fig.2-1 and
Fig.4-1), read instructions for use of operating table carefully and
pay close attention to the patient and devices. Patient could fall
down by the condition or position of the devices.
2-4. For hygiene, be sure to use sterilized drapes on the areas on this
product where the patient comes into contact with it.
3. Interaction
3-1. Combination contraindicated
Do not use the Knee Rests (08-072-01) which are incompatible
with cassettes with the cassette board attached to the operating
table. When the board is flexed upwards, the board will bump into
the cassette board, and it may get damaged.
Storage/Life
1. Do not store the device in high temperatures or in areas with high
humidity where the temperature has drastic variations.
2. Service life of this product: 7 years
(Subject to the specified maintenance and inspection and is stored
properly.)
Maintenance/Inspection
<By the user>
1. Check before and after use
Check that this product is not damaged or broken before and after each
use.
2. Cleaning and disinfection
Wipe off blood, chemicals, contaminants, and other stains with water,
and clean the device with gauze or other materials soaked in
disinfectants. In addition, use the same disinfectants asthe authorized
disinfectants on Mizuho's operating table described in the operating
table operator's manual.
3. In case of a malfunction
When this product is broken, clearly label the device as “Broken”, “Do
not use”, “Need repair”, etc., and contact your local dealer or Mizuho.
<By agents>
Maintenance and inspection can only be carried out by certified agents of
Mizuho.
Packing
1 set per pack
Warranty
MIZUHO Corporation will repair defective parts of this product without
charge for one year from the date of delivery/installment except forcases
of damage caused by a third party’s repair, act ofnature, improperuse or
intentional damage. All other warranty terms and conditions are subject
to regulations of MIZUHO Corporation
Disposal
This device must be disposed of in accordance with local regulations.
Please contact your local distributor for proper disposal.
Notice
Any serious incident that has occurred in relation to the device should be
reported to the manufacturer and the competent authority in which the
user and/or patient is established.