9
2
Before use
When used together with other medical equipment or accessories
- Before using other devices or accessories, read carefully the instruction manual of the device
and make sure that the operating table will not be affected adversely. Before fitting an accessory
of third party company, make sure to contact your distributor or us. Some accessories cannot be
fitted.
- For hygiene, be sure to use sterile drapes on the areas on this product where the patient comes
into contact with it.
- When using the operating table, make sure to check the positions of other equipment. They may
come in contact with each other during the operation, and may get damaged.
Cleaning and disinfection
- After the table has been used, make sure to perform cleaning and disinfection for the table.
Do not use hydrogen peroxide solution to clean the hand console. Using it may cause
discoloration.
Procedure of cleaning and disinfection
1. Turn off power and remove the power cord from the medical grade wall outlet.
2. Remove the mattress from the table.
3. Wipe the surface, the sides and the back of the mattress with a clean cloth dampened with
disinfectant.
4. In the same way as step 3, wipe the table plate and the side rails as well.
5. Wipe the parts again with a dry cloth within 15 minutes.
Usable disinfectant solutions which we have already conrmed as compatible.
Medicine Name
1Sodium hypochlorite 0.1% (halogen containing compound)
2Sodium thiosulfate
3Chlorhexidine (chlorhexidine gluconate 0.5%)
4Benzalkonium chloride (invert soap 10%)
5Povidone iodine
6Ethanol 80%
7Hydrogen peroxide
8Saline 0.9%
9Isopropyl alcohol (IPA) 99.5%
Moving and transportation
- Do not move the table while a patient is on it.
- Follow the procedure below for moving the table.
1. Power off and remove the power cord from medical grade wall outlet.
2. Check if handles and levers are in the xed position, and each section is xed rmly.
3. Unlock a brake to move the table.
4. Before moving the operating table, disinfect the whole operating table in order to prevent
infection.
- Transportation should be carried out with the following conditions met.
1. Disinfect the whole operating table before transportation.
2. Take measures against falling down, like moving down the table to the lowest position.
3. Actuate the brake.
4. Use cushioning materials to prevent damage during transportation.
5. Protect against dust, wind and rain.
Disposal
- Always follow the local country regulation in disposing the operating table.
CAUTION
Others