Safety precaution
■Other
ㅡProhibited
• Do not disassemble and/or modify the operating table. Otherwise, malfunction may
occur.
ㅡPatient position
• When using the tabletop or accessories to secure a patient’s body position, always
observe the patient’s condition. Being in the same position for long periods of time
causes neuroparalysis or bedsores.
ㅡOther medical electrical equipment to be used together with the operating table
• Whenusingahigh-frequencysurgicalequipmentand/oracardiacdebrillatoretc.with
the operating table, refer to their operator’s manual provided by the manufacturers and
observe the precautions and usage. Improper precautions and usage may cause the
operator or the patient to get burned and/or devices to malfunction.
• When using medical electrical equipment etc. with the operating table, check that the
operating table does not malfunction before using it. Electromagnetic interference may
result in malfunctioning of the operating table.
ㅡWeight capacity
• Do not apply loads which exceed the weight capacity (total of the patient and
accessories)*. The operating table may not function, which may result in failures.
* Elevation: 250 kg (550 lbs) / Operations other than elevation: 250 kg (550 lbs)
ㅡPreventive maintenance and inspections
• Make sure to inspect and maintain the operating table before and after use. The
operatingtablemayrequirereplacementofthepartsduetosignicantwear,
deterioration, and/or breakage depending on the usage period and frequency of use.
• For preventive maintenance and inspections, contact your distributor or Mizuho
directly.
ㅡAntistatic measure
• Donotusetheoperatingtableonoorsand/orwithaccessoriesthatdonotpossess
static electricity countermeasures. This may impede surgical operations.
ㅡDevices and accessories used together with this product
• Before using other devices or accessories, thoroughly read the instruction manual of
thedevicesandmakesurethattheoperatingtableisnotaectedadversely.Before
ttingonaccessoriesfromthirdpartycompanies,contactyourdistributororMizuho.
Someaccessoriescannotbettedon.
• While operating the operating table, check the position of other devices or the
accessories used with them. They may come in contact with each other during the
operation, the operating table, devices and/or accessories may get damaged.
• For hygiene, be sure to use sterilized drapes on the areas on this product where the
patient comes into contact with it.