20-25mm
CR2032
Always use CR2032
battery.
PLEASE NOTE
Notch
CR2032
+
DIGITAL S-LINE 3-POINT MICROMETER
Thank you for purchasing the DIGITAL S-LINE 3-POINT MICROMETER. Please read this manual thoroughly before use
for proper operation.
Ring gauge calibration standard (not included) is required for zero setting. Please prepare ring gauge appropriate for the
instrument’s measurement range.
INSTRUCTION MANUAL
Easy to read digital display.
High accuracy 3-point ID measurements.
Capable of storing up to three preset values.
With tolerance setting capability.
Auto display off function (approx. 4 hrs.)
●
●
●
●
●
FEATURES
SPECIFICATIONS
Resolution:0.001mm
Accuracy:4μm
Operating Temperature Range:15〜40℃
Storage Temperature Range:0〜50℃
Electronic Unit Waterproof Grade:IP65 Spec.
Power:CR2032 (Lithium Battery)
   ※1x included for testing
●
●
●
●
●
●
Gauge Head
Rod
[SET] Button
[MODE]Button
LCD Display (side)
Probe Tips
Battery Cover Ratchet Stop
(Constant Force Mechanism)
PART IDENTIFICATION
REPLACING THE BATTERY
【
ACCESSORIES
】
・CR2032 (Lithium Battery)
・
○
Driver for battery replacement
※Provided for testing.
 Please replace with new
 battery before use.
4 1
Press and hold the [MODE]button and
[SET]button simultaneously to change
the number of digits displayed to the
right of the decimal point, and the
direction of increasing display value.
The options are represented as:
0.001/ 0.01 /-0.001 / -0.01 on the
display for the available choices.
Always allow sufficient time for instrument and object to be measured to reach the same temperature before taking
measurement.
Always remove dirt from measuring surfaces with a lens cloth before use.
Keep instrument away from direct sunlight or high temperatures such as in a car, or near a stove or heat source.
Do not turn ratchet stop above or below measurement range.
When low battery indicator is displayed, please replace battery as soon as possible. Continued use may cause reading
errors.
This is a precision instrument, handle with care. Do not drop or subject to excessive forces.
Remove dust and cutting chips after use and apply rust preventative oil to metallic surfaces. Please keep out of reach
of children.
Please do not use organic solvents to clean the instrument body.
Please do not disassemble or modify.
Only use as directed. Improper use may cause accident or injury.
●
●
●
●
●
●
●
●
●
●
 CAUTION
(Ring gauge not included.)
①Using accessory  driver, remove the two battery
cover screws. A flat-head driver can be used in the
notch to gently pry open the cover. Pull out the battery
cover.
②Insert the new battery following the marks on the
battery cover.
③Replace the cover and tighten the screws.
※When battery level is low,
indicator will display on the LCD. With
continued use, the display will become
unstable and readout may not be
correct.
※The battery cover screws
 are very small, care must be
 taken to prevent loss.
※Use care not to bend the terminals
 when inserting the battery.
CONTROLLING READOUT DIGITS AND POLARITY
B
B
①Remove any grease and dust from the measuring surfaces with a clean
cloth, such as a lens cleaning cloth. Make sure no foreign objects are
adhered to the surfaces.
②Turn on power and make sure measurement screen is displayed.
③Insert the measurement head into object to be measured and turn spindle
until the measuring surfaces make contact with the inside diameter.
Continue turning until ratchet stop『clicks』1
〜
3 times and read the
measurement off the LCD display.
OPERATION
【SETTING SCREEN】 【
MEASUREMENT SCREEN
】
Press and
Hold Both
Buttons at
Same Time
[SET]
Button
[MODE]
Button
Measurement Range
(mm)
Measuring Part
Material
−
−
−
−
−
−
−
−
Carbide Tip
Alloy Steel Tip
Battery
Notch
MCD-0608IPS
MCD-0810IPS
MCD-10125IPS
MCD-12516IPS
MCD-1620IPS
MCD-2025IPS
MCD-2535IPS
MCD-3550IPS
6
8
10
12.5
16
20
25
35
8
10
12.5
16
20
25
35
50
Model No. : MCD-****IPS <combined>
Niigata seiki Co., Ltd.
6-15-22, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-31-5660 Fax. : +81-256-39-7730
MAIL:[email protected]
URL
http://www.niigataseiki.co.jp
I221-K 1411
○
『』
Value displayed for
spreading gauge head
Increase
Decrease
Increase
Decrease
Object To Be Measured
20-25mm
“Click, Click”
Inser t
[MODE]Button
Long
Press
[SET]
Button
[SET]
Button
[SET]
Button
[SET]
Button
[MODE]
Button
[SET]
Button
[SET]
Button
Setting Upper Limit
Setting Upper Limit
Setting Lower Limit
Setting Lower Limit
[MODE]
Button
[MODE]
Button
[MODE]
Button
[MODE]
Button
[SET]
Button
[MODE]
Button
[SET]
Button
[SET]
Button
[MODE]
Button
[MODE]
Button
[MODE]
Button
[MODE]
Button
[MODE]Button
Long
Press
20-25mm
※If problem does not resolve,or if you have any questions, please
contact distributor or place of purchase.
※Please note, manufacturer is unable to respond to inquires or provide
service directly. Please contact distributor or place of purchase.
CORRECTIVE ACTIONERROR CONDITION
Measured value is
wrong
Check if origin point is set properly.
Re-set.
Check if measuring surfaces are
dirty. Clean with lens cloth.
LCD displays: “ERR 0” Sensor reading error. Please re-set
origin point.
Nothing is
displayed on LCD
Display is not stable
Display is not clear
Remove and reinsert battery to
cycle power.
Replace the new battery.
Make sure the battery cover
terminals are not damaged.
Gauge head does not
retract
When stored at low temperature, or
not used for a while, the internal
grease may become stiff. Allow
sufficient time for grease to warm up.
Wearing dust-free gloves, use hand
to push in.
【POWER ON】
Press either [MODE]or [SET]button to turn on instrument --
screen will display “CAL”. Press the [SET]button again to
recall the Preset Value and display the measurement screen.
【AUTO DISPLAY OFF】
After approx. 4 hours of inactivity, the display will turn off.
Press [MODE]or [SET]button to turn it on.
【POWER OFF】
Press the [SET]button and hold for 1 sec. to turn off power.
※When inputting preset values, pressing the [SET]button for
 more than one sec. does not turn off power.
※Use ring gauge appropriate for instrument measuring range.
The Origin point will be reinitialized after power off or battery
replacement. Please reset origin using ring gauge before
measurement.
1.Prepare instrument and ring gauge by cleaning     
 measurement head and ring gauge inner diameter with  
 appropriate cleaner.
2.Turn on instrument power and enter the ring gauge size as
 a preset.
 ( Refer to        -③Preset Function)
3.Insert the measurement head into the ring gauge and turn
 the ratchet stop until the measuring surfaces make contact
 with the inside of the ring gauge. Continue turning until  
 ratchet stop “clicks” 1~3 times.
4.Press the [SET]button to recall the preset value. The origin
 point is now set and instrument is ready for use.
23
ORIGIN-POINT SETTING SETTINGS and MODES
TROUBLESHOOTING
POWER
Ring Gauge Standard
(Example:) 20.005mm
Ring Gauge Size stored
as Preset Value and used
for setting the origin point.
Ratchet Stop
A long press of [MODE]button selects the display mode. [MODE], [SET]buttons change the settings.
①
【Measurement Screen】【Measurement Screen】
【Display OFF】
②
③
※Press the [MODE]button to move
cursor, press the [SET]button change
the value.
④
⑤
【Display OFF】
●
LCD displays: “ERR 6”
Oscillation circuit error.Please
remove and reinsert battery to
cycle power.
●
LCD displays: “ERR 8” Memory error.Please remove and
reinsert battery to cycle power.
●
LCD displays: “ERR 9” Please contact distributor or place
of purchase for service.
●
●
●
●
●
●
●
●
●
●
●
Example)
Above Tolerance Limit
『 』Mark displays
Below Tolerance Limit
『 』Mark displays
POWER ON
Press either [MODE]or [SET]button to turn on power and display “CAL” on
the screen. Press [SET]button again for measurement screen.
MODE SWITCHING
From the measurement screen, press the [MODE]button and hold for more
than one second to switch to: Reference, Preset, MIN MAX, and Tolerance
screens. Release [MODE]button at desired screen to change the settings or
use that mode.
①Preset Recall Feature
When [SET]button is pressed upon power up,
or in the measurement screen, the Preset Value
set in ③below is used.
②Reference Function
Different Preset Values (origin) for REFI, REFII,
and REFIII can be selected using the [MODE]
button. This feature is useful when multiple
measurement criteria are required. While in this
mode, the [SET]button will function to HOLD
the display.
③Preset Function
The Preset Value is displayed in measurement
mode by pressing the [SET]button. Use the
[MODE] button to move the underscore cursor,
and the [SET]button to change the value. When
complete, a long press on the [MODE]button
will fix the value and return to the measurement
screen.
④MIN MAX Mode
This screen displays the Minimum Reading,
Maximum Reading, and Maximum Displacement
during measurement.
A short press of the [MODE]button switches
from “MIN” (minimum reading,) “MAX”
(maximum reading,) and “MIN MAX” (maximum
displacement) on the display and their
corresponding value.
Press the [SET]button to reset the stored
values.
⑤Tolerance Mode
Allows easy verification that a measurement is
within preset tolerance range.
With the measurement range set to
20.005mm ~ 20.000mm
1
2
SETTINGS IN EACH MODE
3
3
SETTINGS and MODES
“Click, Click”

This manual suits for next models

7

Other Niigata seiki Measuring Instrument manuals

Niigata seiki TVC-60 User manual

Niigata seiki

Niigata seiki TVC-60 User manual

Niigata seiki PROFESSIONAL DI-10KD User manual

Niigata seiki

Niigata seiki PROFESSIONAL DI-10KD User manual

Niigata seiki WCDI-50D User manual

Niigata seiki

Niigata seiki WCDI-50D User manual

Niigata seiki VH-30D User manual

Niigata seiki

Niigata seiki VH-30D User manual

Niigata seiki DTW-DG02 User manual

Niigata seiki

Niigata seiki DTW-DG02 User manual

Niigata seiki DP-90G EX User manual

Niigata seiki

Niigata seiki DP-90G EX User manual

Niigata seiki GDCS-150DS User manual

Niigata seiki

Niigata seiki GDCS-150DS User manual

Niigata seiki LEVELNIC User manual

Niigata seiki

Niigata seiki LEVELNIC User manual

Niigata seiki DT-150 User manual

Niigata seiki

Niigata seiki DT-150 User manual

Niigata seiki MC122-25C User manual

Niigata seiki

Niigata seiki MC122-25C User manual

Niigata seiki QEM394 Series User manual

Niigata seiki

Niigata seiki QEM394 Series User manual

Niigata seiki JLVC-30 User manual

Niigata seiki

Niigata seiki JLVC-30 User manual

Niigata seiki GDCS-150IK2 User manual

Niigata seiki

Niigata seiki GDCS-150IK2 User manual

Niigata seiki GDCS-150ST User manual

Niigata seiki

Niigata seiki GDCS-150ST User manual

Niigata seiki SCDI S Series User manual

Niigata seiki

Niigata seiki SCDI S Series User manual

Niigata seiki FTM-52FN User manual

Niigata seiki

Niigata seiki FTM-52FN User manual

Niigata seiki TVC-15S User manual

Niigata seiki

Niigata seiki TVC-15S User manual

Niigata seiki GDCS-150DCD User manual

Niigata seiki

Niigata seiki GDCS-150DCD User manual

Niigata seiki KMC-1001 User manual

Niigata seiki

Niigata seiki KMC-1001 User manual

Niigata seiki EOM133 Series User manual

Niigata seiki

Niigata seiki EOM133 Series User manual

Niigata seiki TVC-15S User manual

Niigata seiki

Niigata seiki TVC-15S User manual

Niigata seiki DT Series User manual

Niigata seiki

Niigata seiki DT Series User manual

Niigata seiki WCDI-35 User manual

Niigata seiki

Niigata seiki WCDI-35 User manual

Niigata seiki GDCS-150NP User manual

Niigata seiki

Niigata seiki GDCS-150NP User manual

Popular Measuring Instrument manuals by other brands

PICO TA189 user guide

PICO

PICO TA189 user guide

MEI KC-321C quick start guide

MEI

MEI KC-321C quick start guide

Motrona AX320 operating instructions

Motrona

Motrona AX320 operating instructions

Advanced Instruments OsmoTECH Installation and Training Guide

Advanced Instruments

Advanced Instruments OsmoTECH Installation and Training Guide

Fluke 189II/AN Service manual

Fluke

Fluke 189II/AN Service manual

FloCat C-TJ60 Smart Series Operator's manual

FloCat

FloCat C-TJ60 Smart Series Operator's manual

levr LH-20S user manual

levr

levr LH-20S user manual

Hobo RX3000 Series manual

Hobo

Hobo RX3000 Series manual

Tsuruga Electric Corporation 3566 instruction manual

Tsuruga Electric Corporation

Tsuruga Electric Corporation 3566 instruction manual

REED DW-6060 instruction manual

REED

REED DW-6060 instruction manual

OpenEye OE-MM Hardware manual

OpenEye

OpenEye OE-MM Hardware manual

Biosan DEN-1B User instructions

Biosan

Biosan DEN-1B User instructions

Tintometer Lovibond MD100 instruction manual

Tintometer

Tintometer Lovibond MD100 instruction manual

Legacy LCG90111 Owner's instructions

Legacy

Legacy LCG90111 Owner's instructions

Thermo Scientific Orion Star A211 reference guide

Thermo Scientific

Thermo Scientific Orion Star A211 reference guide

VOLTCRAFT DSA-115 Safety instructions

VOLTCRAFT

VOLTCRAFT DSA-115 Safety instructions

HT HT9014 user manual

HT

HT HT9014 user manual

Harvia sentiotec 260-TH Instructions for use

Harvia

Harvia sentiotec 260-TH Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.