manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Nortek
  6. •
  7. Measuring Instrument
  8. •
  9. Nortek Vector User manual

Nortek Vector User manual

© 2022 Nortek AS
This Quick Guide provides you with the basic information you need to get the instrument up and running as
quickly as possible. If you are familiar with the Nortek instruments, this guide may be all you need. If you are
new to our instruments and software, you should check out the Nortek Comprehensive Manual, which includes
principles of operation, mounting guidelines, troubleshooting, service manual and simple system integration.
You will also find a lot of useful information on the Nortek Support site.
Preparation
·
Connect the battery/power cable.
·
Install the software located on the USB memorystick. Once the installation is
complete, launch the application related to your instrument.
·
Run a Functional Test, described in the Service Manual chapter in the
Comprehensive Manual.
·
Set the Pressure Offset: On-line > Set Pressure Offset.
·
If there are magnetic materials in proximityto the instrument when mounted, a
calibration of the compass is necessary: On-line > Compass Calibration.
·
Erase the recorder in advance of everydeployment.
Deployment Planning
·
Select Deployment Planning.
·
The activated dialog contains all parameters required to specifythe operation of the
instrument. Adescription of each parameter can be found in the application help,
accessible from the Deployment Planning dialog box. One parameter that requires
extra attention is the Nominal velocity range, which is a simple wayof describing
the maximum unambiguous velocitythe instrument is be able to measure. The
velocityrange should be set as low as possible, to minimize measurement
uncertaintyand noise. If the actual values measured exceed the range, your data
will be impaired and maybecome unusable. See the Comprehensive Manual
(Velocimeters) for more information.
·
The Deployment Planning frame on the right-hand side of the dialog displays
performance parameters that are automatically updated as you change the
settings. Use the Standard configuration to plan your deployment, unless you have
a very good reason to move to Advanced. The latter increases the flexibilitybut so
does the risk of making mistakes. Note that the Use Advanced Settings box
(Standard tab) must be checked for the advanced settings to be effective.
·
When finished, press Update to accept the changes. By using the Open/Save
commands in the File menu (or the corresponding toolbar buttons) the deployment
parameters can be saved to file at any given time and re-loaded when it is time to
deploythe instrument.
Start Recorder Deployment
·
Establish communication with the instrument.
·
Select Start Recorder Deployment, either from the pull-down menu or from the
toolbar.
·
Specifythe date and time to start the data recording.
·
In the Name field, enter up to sixcharacters of text to identifythe data file stored in
the recorder.
·
In the File Comment box, enter up to 180 characters of text that will be included in
the data file for documenting the data set.
·
After pressing OK, the Set Clock dialog box is displayed. We recommend setting
the clock to PC time. The time reference is especially important to ensure proper
synchronization between the instrument and other datasets collected
independently. Time spent on setting the time properlybefore a deployment can
save a lot of time in post processing and analysis.
Please check that all
expected items are included
in the shipment:
·
The instrument
·
External pow er/signal cable
·
NORTEK equipment storage
box/toolkit
·
USB to serial converter
·
AC/DC voltage transformer
·
Pow er cable (European or
American)
·
Softw are on USB memory
stick
·
Final test check list
·
Packing list
Install the Nortek software
The instrument’s softw are is
located on the attached memory
stick.
Please check the web regularly
for updates on both softw are
and firmw are.
·
Press Yes to continue. A confirmation dialog will display the deployment
parameters and allow you to verifythat the set-up is what you intended.
·
Press Confirm to send the configuration to the instrument. The instrument will verify
that the deployment has started. The configuration information from the deployment
sequence is saved in a log file that has the same name as the deployment..
Stop Recorder Deployment
·
Select Stop Recorder Deployment. When data acquisition has stopped a dialog
will display the current instrument and PC clock time.
Data Retrieval
·
Select Recorder Data Retrieval.
·
The recorder overview dialog will show a list of all files currently stored in the
recorder, including the time of the first measurement and the size for each file.
·
Select a file in the list byclicking on the filename. Press Retrieve to specifythe
name and disk location of the file.
Tip! Check the retrieved data before deploying the instrument for the second time. If
something is wrong with the instrument, the configuration or the deployment, this will
be detected and the possibilityof making the same mistake twice is avoided.
Data Conversion
·
To convert binary data files retrieved from the recorder to a readable ASCII format,
select Data Conversion.
·
Add files to the Files to convert list byusing the Add file... file selection dialog.
Select the file to convert in the list and press the blue arrow. Specifythe location for
the converted files in the Save in folder field. An optional prefixor suffix that will be
added to the recorder filename may be specified. Select from the View files drop-
down list to open the converted (ASCII) files in e.g. Notepad.
On-line Data Collection
·
Select Start Data Collection from the On-line menu (or press the toolbar button) to
start data acquisition. The real time displayshows velocitydata and sensor data as
it is collected.
·
To capture the data that comes in over the serial port to disk select Disk
Recording… from the On-line menu and define the name of a file (without
extension). Once this disk file is defined, the Start Disk Recording and Stop Disk
Recording menu items and toolbar buttons will be enabled.
Mounting
Proper mounting of the instrument is crucial for successful collection of data. Please
mount the instrument in accordance with the Mounting Guidelines in the
Comprehensive Manual.
Cables
Our instruments are mostlyshipped with RS232 communication. RS422 is used for
cables longer than 50 m. The wiring is shown in the tables below. The output can
either be in ASCII, in fixed NMEA format, or binaryformat. Communication baud rate:
300-115200 (user setting), recorder download baud rate: 600/1200 kBaud. Ensure
that the connectors and dummyplugs are lubricated with dielectric silicone. We
recommend 3Msilicone spray. Note that the silicone grease included in the shipment
should be used on the O-rings only.
Analog Input (default for most instruments)
The instrument can read two analog inputs at the same time. The input range is 0-5V,
corresponding to 0-65535 counts in the data file.
Basic principles
Pulse-coherent instruments
utilize a pair of acoustic pulses
w ith a know n time lag to
determine a Doppler induced
phase shift. This measured
phase shift is converted to
velocity by scaling w ith the
speed of sound in w ater. The
Vector use separate
transducers to transmit and
receive the acoustic beams
(bistatic system).
Measuring waves
The Vector can measure w aves
using the PUV method. The
pressure measurement provides
a means of estimating all the
nondirectional w ave
parameters, w hile the combined
P, U, and V measurements allow
for estimating the directional
w ave parameters. These data
need to be post processed.
Data analysis
The softw are supplied has been
designed to provide you w ith the
ability to set the instrument up
for deployment, upgrade the
firmw are, dow nload acquired
data and do a quality assurance
of them. The data can be
converted to ASCIIformat that
may be imported into various
post processing programs.
© 2022 Nortek AS
Power
·
DC Input: 9-15 VDC
·
Battery DC-input, nominal voltage: 13.5 V
System Integrator Manual
For more information on how to control a Nortek product
with a non-PC controller, please take a look at the
System Integrator Manual.
Troubleshooting
As most problems are caused by simple mistakes,
please initially check if
·
you have forgotten to power the system
·
the connector has fallen out of the computer
·
you are using the wrong serial port
You mayfind that there is a lot of noise in the velocity
data. This maybe caused by
·
not enough particles in the water
·
the velocityrange set too high
·
the sampling volume located beyond the bottom
·
reflection and reverberation conditions at the bottom
·
probe malfunction
Test the probe with the Probe Check feature to identify
issues.
Nortek online
At our website you will find technical support, user
manuals, and the latest software and firmware. General
information, technical notes, and user experience can
also be found here.
Email: inquiry@nortekgroup.com for general inquiries or
support@nortekgroup.com for technical support.
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS232 Tx
Twisted
Pair
2
4
White
RS232 Rx
3
5
Black
RS232 Gnd
Twisted
Pair
5
6
White/purple
Power Output
7
Black
Analogue Ch. 2
Twisted
Pair
8
White/orange
Analogue Ch. 1
1
Black
Power Gnd
Twisted
Pair
2
White
Power +
Screen terminated at pin 1 in underwater connector, unterminated at PC
side.
Ref: N2100-002A
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS422 Tx +
Twisted
Pair
7
4
White
RS422 Tx -
2
7
Black
RS422 Rx -
Twisted
Pair
8
8
White/orange
RS422 Rx +
3
5
Black
Sync. Out
Twisted
Pair
6
White/purple
Sync. In.
1
Black
Power Gnd
Twisted
Pair
2
White
Power +
Screen terminated at pin 1 in underwater connector, unterminated at PC
side.
Ref: N2100-108
Wiring of RS232 (upper) and RS422 (lower) cables.
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS232 Tx
Twisted Pair
2
4
White
RS232 Rx
3
5
Black
RS232 Gnd
Twisted Pair
5
6
White/purple
Analog Z*
Red wire
7
Black
Analog X
Twisted Pair
Green wire
8
White/orange
Analog Y
Yellow wire
1
Black
Power Gnd
Twisted Pair
Black
wire
2
White
Power +
White
wire
Screen
Bare
Ground
3 bare wires f or grounds, connected internally
to power ground
Ref: N2100-004
Wiring of RS232 cable with option for analog inputs. *Optional version outputs pressure instead of Z velocity.

Other manuals for Vector

1

Other Nortek Measuring Instrument manuals

Nortek Aquadopp User manual

Nortek

Nortek Aquadopp User manual

Nortek Signature VM User manual

Nortek

Nortek Signature VM User manual

Nortek DVL 1000 300m User manual

Nortek

Nortek DVL 1000 300m User manual

Nortek Vectrino User manual

Nortek

Nortek Vectrino User manual

Nortek Vectrino Profiler User manual

Nortek

Nortek Vectrino Profiler User manual

Nortek Signature Series User manual

Nortek

Nortek Signature Series User manual

Nortek Signature55 User manual

Nortek

Nortek Signature55 User manual

Nortek AWAC User manual

Nortek

Nortek AWAC User manual

Nortek Aquadopp Profiler User manual

Nortek

Nortek Aquadopp Profiler User manual

Nortek Vectrino User manual

Nortek

Nortek Vectrino User manual

Nortek Vector User manual

Nortek

Nortek Vector User manual

Nortek Scour Monitor User manual

Nortek

Nortek Scour Monitor User manual

Popular Measuring Instrument manuals by other brands

Bosch GIM 120 Professional Original instructions

Bosch

Bosch GIM 120 Professional Original instructions

Hanna Instruments HI 718 quick start guide

Hanna Instruments

Hanna Instruments HI 718 quick start guide

Tektronix VM700T instructions

Tektronix

Tektronix VM700T instructions

Wavelength References ClarityPlus Operator's manual

Wavelength References

Wavelength References ClarityPlus Operator's manual

DATREND Systems Vision Pad quick start guide

DATREND Systems

DATREND Systems Vision Pad quick start guide

TESTO 0555 6451 operating instructions

TESTO

TESTO 0555 6451 operating instructions

KJ Instrument KJ6013 Operation manual

KJ Instrument

KJ Instrument KJ6013 Operation manual

Delta F NANOTRACE DF-745 instruction manual

Delta F

Delta F NANOTRACE DF-745 instruction manual

CHOWIS dp viso+ user manual

CHOWIS

CHOWIS dp viso+ user manual

Kirchner und Tochter RA 65 Installation and operating instructions

Kirchner und Tochter

Kirchner und Tochter RA 65 Installation and operating instructions

Graco 26C266 manual

Graco

Graco 26C266 manual

Omntec FuelCheck Installation & operating manual

Omntec

Omntec FuelCheck Installation & operating manual

Emerson Rosemount 2051CF series quick start guide

Emerson

Emerson Rosemount 2051CF series quick start guide

DATREND Systems vPad-Rugged 2 quick start guide

DATREND Systems

DATREND Systems vPad-Rugged 2 quick start guide

Philips VoiceTracer DVT4110 user manual

Philips

Philips VoiceTracer DVT4110 user manual

BRONKHORST EL-FLOW Prestige series instruction manual

BRONKHORST

BRONKHORST EL-FLOW Prestige series instruction manual

Refco Refmate quick guide

Refco

Refco Refmate quick guide

Apogee Instruments SP-110 owner's manual

Apogee Instruments

Apogee Instruments SP-110 owner's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.