This Quick Guide provides you with the basic information you need to get the instrument up and running as
quickly as possible. If you are familiar with the Nortek instruments, this guide may be all you need. If you are
new to our instruments and software, you should check out the Nortek Comprehensive Manual, which includes
principles of operation, mounting guidelines, troubleshooting, service manual and simple system integration.
You will also find a lot of useful information on the Nortek Support site.
Preparation
·
Connect the battery/power cable.
·
Install the software located on the USB memorystick. Once the installation is
complete, launch the application related to your instrument.
·
Run a Functional Test, described in the Service Manual chapter in the
Comprehensive Manual.
·
Set the Pressure Offset: On-line > Set Pressure Offset.
·
If there are magnetic materials in proximityto the instrument when mounted, a
calibration of the compass is necessary: On-line > Compass Calibration.
·
Erase the recorder in advance of everydeployment.
Deployment Planning
·
Select Deployment Planning.
·
The activated dialog contains all parameters required to specifythe operation of the
instrument. The Deployment Planning frame on the right-hand side of the dialog
displays performance parameters that are automaticallyupdated as you change
the settings. Use the Standard configuration to plan your deployment unless you
have a very good reason to move to Advanced. The latter increases the flexibilitybut
also the risk of making mistakes. Note that the Use Advanced Settings box (in the
Standard tab) must be checked for the advanced settings to be effective. A
description of each parameter can be found in the application help, accessible
from the dialog box.
·
When finished, press Update to accept the changes. By using the Open/Save
commands in the File menu (or the corresponding toolbar buttons) the deployment
parameters can be saved to file at any given time and re-loaded when it is time to
deploythe instrument.
Start Recorder Deployment
·
Establish communication with the instrument.
·
Select Start Recorder Deployment, either from the pull-down menu or from the
toolbar.
·
Specifythe date and time to start the data recording.
·
In the Name field, enter up to sixcharacters of text to identifythe data file stored in
the recorder.
·
In the File Comment box, enter up to 180 characters of text that will be included in
the data file for documenting the data set.
·
After pressing OK, the Set Clock dialog box is displayed. We recommend setting
the clock to PC time. The time reference is especially important to ensure proper
synchronization between the instrument and other datasets collected
independently. Time spent on setting the time properlybefore a deployment can
save a lot of time in post processing and analysis.
·
Press Yes to continue. A confirmation dialog will display the deployment
parameters and allow you to verifythat the set-up is what you intended.
·
Press Confirm to send the configuration to the instrument. The instrument will verify
that the deployment has started. The configuration information from the deployment
sequence is saved in a log file that has the same name as the deployment..
Please check that all
expected items are included
in the shipment:
·
The instrument
·
External pow er/signal cable
·
NORTEK equipment storage
box/toolkit
·
USB to serial converter
·
AC/DC voltage transformer
·
Pow er cable (European or
American)
·
Softw are on USB memory
stick
·
Final test check list
·
Packing list
Install the Nortek software
The instrument’s softw are is
located on the attached memory
stick.
Please check the web regularly
for updates on both softw are
and firmw are.
Stop Recorder Deployment
·
Select Stop Recorder Deployment. When data acquisition has stopped a dialog
will display the current instrument and PC clock time.
Data Retrieval
·
Select Recorder Data Retrieval.
·
The recorder overview dialog will show a list of all files currently stored in the
recorder, including the time of the first measurement and the size for each file.
·
Select a file in the list byclicking on the filename. Press Retrieve to specifythe
name and disk location of the file.
Tip! Check the retrieved data before deploying the instrument for the second time. If
something is wrong with the instrument, the configuration or the deployment, this will
be detected and the possibilityof making the same mistake twice is avoided.
Data Conversion
·
To convert binary data files retrieved from the recorder to a readable ASCII format,
select Data Conversion.
·
Add files to the Files to convert list byusing the Add file... file selection dialog.
Select the file to convert in the list and press the blue arrow. Specifythe location for
the converted files in the Save in folder field. An optional prefixor suffix that will be
added to the recorder filename may be specified. Select from the View files drop-
down list to open the converted (ASCII) files in e.g. Notepad.
On-line Data Collection
·
Select Start Data Collection from the On-line menu (or press the toolbar button) to
start data acquisition. The real time displayshows velocitydata and sensor data as
it is collected.
·
To capture the data that comes in over the serial port to disk select Disk
Recording… from the On-line menu and define the name of a file (without
extension). Once this disk file is defined, the Start Disk Recording and Stop Disk
Recording menu items and toolbar buttons will be enabled.
Mounting
Proper mounting of the instrument is crucial for successful collection of data. Please
mount the instrument in accordance with the Mounting Guidelines in the
Comprehensive Manual.
Cables
Our instruments are mostlyshipped with RS232 communication. RS422 is used for
cables longer than 50 m. The wiring is shown in the tables below. The output can
either be in ASCII, in fixed NMEA format, or binaryformat.
Communication baud rate: 300-115200 (user setting).
Recorder download baud rate: 600/1200 kBaud.
Ensure that the connectors and dummyplugs are lubricated with dielectric silicone.
We recommend 3Msilicone spray. Note that the silicone grease included in the
shipment should be used on the O-rings only.
Basic principles
The instrument measures
current speed and direction by
transmitting sound waves that
are reflected by particles drifting
w ith the currents in the w ater.
The Doppler shift of the return
signal is proportional to the
velocity component along the
beams.
Measuring waves
Both the Aquadopp Current
Meter and Aquadopp Current
Profiler can measure waves
using the PUV method. The
pressure measurement provides
a means of estimating all the
nondirectional w ave
parameters, w hile the combined
P, U, and V measurements allow
for estimating the directional
w ave parameters. These data
need to be post processed.
Data analysis
The softw are supplied has been
designed to provide you w ith the
ability to set the instrument up
for deployment, upgrade the
firmw are, dow nload acquired
data and do a quality assurance
of them. The data can be
converted to ASCIIformat that
may be imported into various
post processing programs. In
addition, Nortek has developed a
series of real time and post
processing softw are packages
that simplify your analysis w ork.
These can be dow nloaded from
our w ebsite; contact Nortek for
a valid license key.
© 2022 Nortek AS
Analog Input (default for most instruments)
The instrument can read two analog inputs at the same
time. The input range is 0-5V, corresponding to 0-65535
counts in the data file.
Power
·
DC Input: 9-15 VDC
·
Battery DC-input, nominal voltage: 13.5 V
System Integrator Manual
For more information on how to control a Nortek product
with a non-PC controller, please take a look at the
System Integrator Manual.
Troubleshooting
As most problems are caused by simple mistakes,
please initially check if
·
you have forgotten to power the system
·
the connector has fallen out of the computer
·
you are using the wrong serial port
Nortek online
At our website you will find technical support, user
manuals, and the latest software and firmware. General
information, technical notes, and user experience can
also be found here.
Email: inquiry@nortekgroup.com for general inquiries or
support@nortekgroup.com for technical support.
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS232 Tx
Twisted
Pair
2
4
White
RS232 Rx
3
5
Black
RS232 Gnd
Twisted
Pair
5
6
White/purple
Power Output
7
Black
Analogue Ch. 2
Twisted
Pair
8
White/orange
Analogue Ch. 1
1
Black
Power Gnd
Twisted
Pair
2
White
Power +
Screen terminated at pin 1 in underwater connector, unterminated at PC
side.
Ref: N2100-002A
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS422 Tx +
Twisted
Pair
7
4
White
RS422 Tx -
2
7
Black
RS422 Rx -
Twisted
Pair
8
8
White/orange
RS422 Rx +
3
5
Black
Sync. Out
Twisted
Pair
6
White/purple
Sync. In.
1
Black
Power Gnd
Twisted
Pair
2
White
Power +
Screen terminated at pin 1 in underwater connector, unterminated at PC
side.
Ref: N2100-108
Wiring of RS232 (upper) and RS422 (lower) cables.
Underwater Connector
Termination
Pin #
Wire Color
Purpose
Pins
3
Black
RS232 Tx
Twisted Pair
2
4
White
RS232 Rx
3
5
Black
RS232 Gnd
Twisted Pair
5
6
White/purple
Analog Z*
Red wire
7
Black
Analog X
Twisted Pair
Green wire
8
White/orange
Analog Y
Yellow wire
1
Black
Power Gnd
Twisted Pair
Black
wire
2
White
Power +
White
wire
Screen
Bare
Ground
3 bare wires f or grounds, connected internally
to power ground
Ref: N2100-004
Wiring of RS232 cable with option for analog inputs. *Optional version outputs pressure instead of Z velocity.

Other Nortek Measuring Instrument manuals

Nortek Signature Series User manual

Nortek

Nortek Signature Series User manual

Nortek Vectrino User manual

Nortek

Nortek Vectrino User manual

Nortek Signature55 User manual

Nortek

Nortek Signature55 User manual

Nortek Vector User manual

Nortek

Nortek Vector User manual

Nortek Signature VM User manual

Nortek

Nortek Signature VM User manual

Nortek Vectrino User manual

Nortek

Nortek Vectrino User manual

Nortek Aquadopp Profiler User manual

Nortek

Nortek Aquadopp Profiler User manual

Nortek Vectrino Profiler User manual

Nortek

Nortek Vectrino Profiler User manual

Nortek Vector User manual

Nortek

Nortek Vector User manual

Nortek Scour Monitor User manual

Nortek

Nortek Scour Monitor User manual

Nortek AWAC User manual

Nortek

Nortek AWAC User manual

Nortek DVL 1000 300m User manual

Nortek

Nortek DVL 1000 300m User manual

Popular Measuring Instrument manuals by other brands

Rechner Sensors TRUE LEVEL instruction manual

Rechner Sensors

Rechner Sensors TRUE LEVEL instruction manual

OmniMetrix TrueGuard-Pro installation guide

OmniMetrix

OmniMetrix TrueGuard-Pro installation guide

Gossen MetraWatt SECULIFE IM Quick setup guide

Gossen MetraWatt

Gossen MetraWatt SECULIFE IM Quick setup guide

Microtrac S3500 Operation and maintenance manual

Microtrac

Microtrac S3500 Operation and maintenance manual

Horizon Fitness Autometers Systems HT-100 manual

Horizon Fitness

Horizon Fitness Autometers Systems HT-100 manual

LATNEX HM-400 manual

LATNEX

LATNEX HM-400 manual

Vega VEGACAL 63 Specification sheet

Vega

Vega VEGACAL 63 Specification sheet

Ponpe 8602 Operation manual

Ponpe

Ponpe 8602 Operation manual

Wohler KM 410 Operation manual

Wohler

Wohler KM 410 Operation manual

Keysight N432A installation guide

Keysight

Keysight N432A installation guide

Frankford Arsenal INTELLIDROPPER 1082250 user manual

Frankford Arsenal

Frankford Arsenal INTELLIDROPPER 1082250 user manual

SCHUNK IFT operating instructions

SCHUNK

SCHUNK IFT operating instructions

Greenlee AirScout Gigacheck instruction manual

Greenlee

Greenlee AirScout Gigacheck instruction manual

Seca mVSA 535 Instructions for use

Seca

Seca mVSA 535 Instructions for use

Dakota Digital Odyssey ODY-08-1 quick start guide

Dakota Digital

Dakota Digital Odyssey ODY-08-1 quick start guide

METREL Energy Master MI 2883 instruction manual

METREL

METREL Energy Master MI 2883 instruction manual

sauermann Si-HH3 quick start guide

sauermann

sauermann Si-HH3 quick start guide

Keysight Technologies 85059B Operating and service manual

Keysight Technologies

Keysight Technologies 85059B Operating and service manual