-73-
PROBLEM SOLVER CHART
OCC OUT
(repeated)
NO SET
AIR
PUSH STRT
Condition
END OF DOSE
PROG
DOOR
BATT
the pump
remains working
Batter, “E” and “F”
flashing
FILL SET
OCC IN
OCC OUT
• Calibration not et completed
on the current feeding set
• The set is not fitted or wrongl fitted
in the pump.
• The pressure sensor area is dirt
• The air sensor detected an empt
pump set.
• The air sensor area is dirt.
• The feeding set is not inserted cor-
rectl.
• The pump has been untouched in
hold mode for 3 minutes or more.
Cause
• The pump administered the installed
dose = volume.
• No flow rate is installed.
Flow rate = 0 ml/h
• The door is not correctl closed.
• The batter capacit is too low.
The pump is not able to deliver
highl viscous fluids at high flow
rates with the present charge level
of the batter.
• Batter failure.
• The pump is priming the set.
• The pump detected an upstream
occlusion between the pump and
the feeding bag.
• The pressure sensor area is dirt.
• The pump detected a
downstream occlusion between the
pump and the patient.
• The pressure sensor area is dirt.
• Stop the alarm b pressing the “START/STOP” ke.
• Start the pump b pressing the “START/STOP” ke and
let it run onl briefl.
• Stop the pump b pressing the “START/STOP” ke insu-
ring that there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert the
feeding set in the pump.
• Restart the pump b pressing the “START/STOP” ke.
• Stop the alarm b pressing “START/STOP”
• Insert the Flocare®Infinit™cassette into the pump as
indicated on the blister packaging and close the door.
• Restart the pump.
• Clean the sensors, reinsert the cassette in the pump and
restart the pump.
• Replace the empt feeding reservoir and continue feeding.
If necessar prime the set.
• Make sure the air sensor is clean.
• Make sure the feeding set is properl inserted
in the pump.
• Stop the alarm and prolong the hold mode with another 3
minutes b pressing “START/STOP” ke.
• Program the pump and start it b pressing the “START/
STOP”.
Correction
• Turn the pump off b pressing the “ON/OFF” ke
and hold it during 2 seconds or
• Clear the memor of the total volume administered (see
paragraph: clearing the memor), reprogram a new fee-
ding schedule and start the pump b pressing “START/
STOP”.
• Make sure the correct flow rate is programmed.
• Make sure the door is properl closed prior to
starting a feeding program.
• Connect the adaptor to the pump and mains and charge
the pump for approx. 6 hours. During charging the pump
can be used.
• Turn the pump “off”, contact our healthcare provider or
refer the pump to service.
• Press the “FILL SET” ke another time to stop the pump
and bring it back in the hold mode.
• Stop the alarm b pressing the “START/STOP” ke.
• Remove the feeding set out of the pump and check the
permeabilit b flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the cassette in the pump and
restart the pump.
• Stop the alarm b pressing the “START/STOP” ke.
• Check the permeabilit of the feeding tube b
aspirating liquid via the medication port.
• Clean the sensors, reinsert the cassette in the pump and
restart the pump.
UK-IE
Alwas follow the instructions below in case a problem occurs.
Using methods other than those described ma cause the pump to function incorrectl.
ZMNL400336_AW.indd 75 3/18/10 3:47:43 PM