Oakton ORPTestr 30 User manual

Operating Manual
ORPTestr®30 Pocket Tester
Keypad Functions
Short press = <2 seconds Long press = >2 seconds
1. Short press to turn on the tester and long press to turn
off the tester.
2. When turned off, long press to enter parameter setting
mode.
3. In measurement mode, short press toturn on backlight.
1. In calibration mode, short press to change calibration
value (unidirectional).
2. In mode setting, short press to change parameter
(unidirectional).
1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3. When measured value is locked (HOLD icon), short
press to unlock.
CAL
<
MEAS
Thank you for purchasing the Oakton®ORPTestr®30 Pocket Tester.
Please carefully read this operating manual before using the
product to obtain an accurate and reliable test result and avoid
unnecessary damage to the tester or probe.
Features
• Large easy-to-read LCD with 3-color backlight
• Replaceable double-junction sensor saves you money
• Stability and battery status icons
• Auto power-off function conserves battery life
• IP67 waterproof rating
Preparation Before Use
1. Pour certain amount of storage solution (KCl) into the probe cap,
soak the probe for 3 to 5 minutes to activate the sensor (if frequently
used, there is no need to soak).
2. The soaking solution is 3M KCl solution; preferably Oakton
Electrode Storage Solution (SKU 00653-04). Users can also
make it themselves by dissolving 25 g KCl in 100 ml of distilled
water.
ORP Calibration
1. Clean electrode in distilled water and dry it.
2. Pour a small amount of 222 mV ORP
standard solution (SKU 80055-90, 500-mL
bottle) into the calibration bottle.
3. Short press /MEAS key to turn the meter
on. Long press CAL/ key to enter calibration
mode. If you need to exit calibration mode,
short press /MEAS key.
4. Dip electrode in the ORP standard solution
(Zobell’s solution), stir gently, and allow it to
stand still in the calibration solution until a
stable reading is reached. When measurement stability icon ( )
appears and remains on the screen (see Fig. 1), press MODE/△ key to
adjust the calibration value according to the table below (also on the
buffer bottle’s label).
Temperature (ºC)
10
15
20
25
Temperature (ºC)
30
35
38
40
mV
215
209
205
201
mV
242
235
227
222
For example, if the temperature is around 25ºC, then adjust the
calibration value to 222 mV. If the temperature is around 20ºC, then
adjust the calibration value to 227 mV. Short press CAL/ key to
complete the calibration.
ORP Measurement
Press /MEAS key to power on, clean the probe in distilled water,
and gently dry it. Place the probe in the tested solution and stir
gently and let it stand still. Read the measurement after the stability
icon ( ) appears and remains on.
Notes
Usually there’s no need to calibrate for ORP testing. When not used
for a long time, users can test the unit in 222 mV ORP standard
buffer. If the error is large, recalibrate according to the steps above.
Cleaning and Activating the ORP Probes
After long-time use, the surface of the platinum sensor can be
contaminated and cause inaccuracy and slow response time. If that
happens, the followed methods can be adopted to clean and
activate:
A. For contamination caused by inorganic substances, dip the
probe into 0.1 mol/L dilute hydrochloric acid for 30 minutes.
Clean it with distilled water, soak in KCL storage solution (SKU
00653-04) for 6 hours, and then it will be ready for use.
B. For contamination caused by organic substances and oiliness,
clean the platinum surface with dishwashing liquid and wash
with distilled water. Soak for 6 hours in storage solution (SKU
00653-04) and then it will be ready for use.
C. If the platinum surface is so severely contaminated that an
oxide film forms, use toothpaste to clean the platinum surface,
and then wash it with distilled water. The probe will be ready for
use again after 6 hours of soaking in the storage solution (SKU
00653-04).
Setting the Parameters
When tester is off, long press /MEAS key to enter parameter
settingmode. Short press MODE/△key to switch from P1 to
P2…P6. Short press CAL/ key and parameter will flash, then short
press MODE/△key to choose desired parameter. Short press
CAL/ key to confirm selection. Long press /MEAS key to exit
parameter setting mode.
ORP Standard Buffer Solution Calibration Reference Table
Symbol Selection
°C
Menu setting
Restore to
factory default
Set low
measurement alarm
Set high
measurement alarm
Select Automatic
Lock (HOLD)
Select backlight
Select temperature
unit
No – Yes
–1000 to +1000 mV
–1000 to +1000 mV
Off – On
Off – 1 – On
°C – °F
P1
P2
P3
P4
P5
P6
Factory default
–1000
1000
Off
1 (1-min auto-off)
No
A. High and Low Measurement Heads-Up Examples (P1 and P2)
Preset low value =
mV (stability icon ( ) displays on LCD) the red backlight appears
on display.
Preset high value =
mV (stability icon ( ) displays on LCD) the red backlight appears
on display.
Preset low value = 120 mV, highest value = 150 mV, when
the measured value is <120 mV or >150 mV (stability icon ( )
displays on LCD) the red backlight appears on display.
B. Automatic Lock (P3)
Select “On” to activate auto lock function. When reading is stable for
more than 10 seconds, the tester will lock the value automatically,
and “HOLD” icon will appear on the bottom left of the screen.Press
CAL/ key to cancel HOLD on reading.
C. Backlight (P4)
Select “Off” to turn off backlight function, “On” to turn on backlight
function, or “1” to have backlight last for 1 minute.
D. Temperature Unit (P5)
Select °C or °F; the factory default is °C.
E. Factory Default Setting (P6)
Select “Yes” to recover tester calibration to theoretical value (pH
value in zero potential is 7.00, slope is 100%), parameter setting
return to initial value. This function can be used when instrument
does not work well in calibration or measurement. Calibrate and
measure again after resetting the unit to factory default status.
Self-Diagnostic Messages
Symbol Self-diagnosis
information How to fix
Wait for the measurement stability
icon ( ) to appear and stay, then
press CAL/ .
CAL/ is pressed
before measurement
is stable ( appears)
Specifications
ORP (mV)
Temperature
Range
Resolution
Accuracy
Range
Resolution
Accuracy
±1000 mV
1 mV
±0.2% full-scale
32 to 122°F (0 to 50°C)
0.1°F/°C
±0.9°F (0.5°C)
Display: LCD with three-color backlight. Blue = measurement;
Green = calibration; Red = alarm
Reading lock: HOLD icon
Power: four AAA batteries (included); >400 hours of continuous
operation
Low-voltage warning: battery status icon flashes
Auto power-off: tester automatically turns off after 8 minutes of
nonuse
IP rating: IP67 (waterproof), floats on water when sensor cap is on
Dimensions (L x W x H): 7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm)
Weight: 4.7 oz (133 g)
Sensor Replacement
Screw off the sensor ring, unplug the sensor, plug in the new
replacement sensor (pay attention to the probe’s position), and
rescrew on the sensor ring
Cap Lock Clip Battery Lid
Lift the clip to unlock
4×AAA
Batteries
Push the clip to open the battery lid
①
②
③
④
⑤
⑥
⑦
1. Open battery lid.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7. Fit the tester’s cap while making sure to push all the way down.
The tester’s waterproof design may be compromised if the cap is
not fitted correctly.
Warranty
We warrant this instrument to be free from defects in material and
workmanship and agrees to repair or replace free of charge, at
option of Oakton Instruments, any malfunctioned or damaged
product attributable to responsibility of Oakton Instruments, for a
period of two years from the delivery (a six-month limited warranty
applies to sensors). This warranty does not apply to defects
resulting from actions such as misuse (violation of the instructions in
this manual or operations in the manner not specified in this
manual), improper maintenance, and unauthorized repairs.
Warranty period is the time limit to provide free service for the
products purchased by customers, not the service life of the tester
or probe.
Oakton Instruments reserves the right to update the information in
this manual without giving notice in advance.
V: Preset high value =
150mV, lowest value = –1000 mV
mV (stability icon ( ) displays on LCD) the red backlight appears
on display.
mV: Preset low value = 120 mV, highest value = 150 mV, when
the measured value is <120 mV or >150 mV (stability icon ( )
displays on LCD) the red backlight appears on display.
B. Automatic Lock (P3)
Select “On” to activate auto lock function. When reading is stable for
more than 10 seconds, the tester will lock the value automatically,
and “HOLD” icon will appear on the bottom left of the screen.Press
CAL/ key to cancel HOLD on reading.
C. Backlight (P4)
Select “Off” to turn off backlight function, “On” to turn on backlight
function, or “1” to have backlight last for 1 minute.
D. Temperature Unit (P5)
Select °C or °F; the factory default is °C.
E. Factory Default Setting (P6)
Select “Yes” to recover tester calibration to theoretical value (pH
value in zero potential is 7.00, slope is 100%), parameter setting
return to initial value. This function can be used when instrument
does not work well in calibration or measurement. Calibrate and
measure again after resetting the unit to factory default status.
Ordering Information
Product description
ORPTestr 30 pocket tester
Replacement sensor for ORPTestr 30
Model
ORP30
—
Catalog number
35634-32
35634-68
Battery Installation
The tester uses four AAA batteries. Please install batteries
according tothe following steps. Note the correct direction of battery
installation: thepositive side (+) of every single battery must face up.
Incorrect installa-tion of batteries will cause damage to the tester
and create a potential hazard.
www.coleparmer.com

Other Oakton Test Equipment manuals

Oakton pHTestr 20 User manual

Oakton

Oakton pHTestr 20 User manual

Oakton EcoTestr EC1 User manual

Oakton

Oakton EcoTestr EC1 User manual

Oakton CTS5 User manual

Oakton

Oakton CTS5 User manual

Oakton TempTestr 35628-00 User manual

Oakton

Oakton TempTestr 35628-00 User manual

Oakton Salt 1 User manual

Oakton

Oakton Salt 1 User manual

Oakton pHTestr 1 Technical manual

Oakton

Oakton pHTestr 1 Technical manual

Oakton CTSTestr 5 User manual

Oakton

Oakton CTSTestr 5 User manual

Oakton WD-35662-35 User manual

Oakton

Oakton WD-35662-35 User manual

Oakton pHTestr 1 User manual

Oakton

Oakton pHTestr 1 User manual

Oakton EcoTestr CTS1 User manual

Oakton

Oakton EcoTestr CTS1 User manual

Oakton CTSTestr 50 User manual

Oakton

Oakton CTSTestr 50 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton ORPTestr10 User manual

Oakton

Oakton ORPTestr10 User manual

Oakton PCTSTestr 50 User manual

Oakton

Oakton PCTSTestr 50 User manual

Oakton PCTS5 User manual

Oakton

Oakton PCTS5 User manual

Oakton TDS 1 User manual

Oakton

Oakton TDS 1 User manual

Oakton ORP5 User manual

Oakton

Oakton ORP5 User manual

Oakton ORPTestr 50 User manual

Oakton

Oakton ORPTestr 50 User manual

Oakton pH 5+ User manual

Oakton

Oakton pH 5+ User manual

Oakton WD-35634-06 User manual

Oakton

Oakton WD-35634-06 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton EcoTestr pH 2 User manual

Oakton

Oakton EcoTestr pH 2 User manual

Oakton pH BNC 10 User manual

Oakton

Oakton pH BNC 10 User manual

Oakton Eutech Instruments TDSTestr11+ User manual

Oakton

Oakton Eutech Instruments TDSTestr11+ User manual

Popular Test Equipment manuals by other brands

Svantek SV258 PRO user manual

Svantek

Svantek SV258 PRO user manual

Rio Grande 705-020/00 manual

Rio Grande

Rio Grande 705-020/00 manual

Time Electronics 5070 user manual

Time Electronics

Time Electronics 5070 user manual

Stoneridge DSRC Tester manual

Stoneridge

Stoneridge DSRC Tester manual

Probewell MT-1/WT3 user guide

Probewell

Probewell MT-1/WT3 user guide

MD Diagnostics RP Check operating manual

MD Diagnostics

MD Diagnostics RP Check operating manual

Tektronix DPO2012 Technical reference

Tektronix

Tektronix DPO2012 Technical reference

YOKOGAWA DLM4000 Series user manual

YOKOGAWA

YOKOGAWA DLM4000 Series user manual

Kikusui TOS3200 Interface manual

Kikusui

Kikusui TOS3200 Interface manual

doble PHENIX 6CB user manual

doble

doble PHENIX 6CB user manual

Baur KSG 200 A user manual

Baur

Baur KSG 200 A user manual

Alcovisor Jupiter operating manual

Alcovisor

Alcovisor Jupiter operating manual

Pro-Set cps TRS21 Series owner's manual

Pro-Set

Pro-Set cps TRS21 Series owner's manual

Omega PCL1200 user guide

Omega

Omega PCL1200 user guide

Dukane Seacom TS500 Technical manual

Dukane Seacom

Dukane Seacom TS500 Technical manual

McGan MM513 Quick operation guide

McGan

McGan MM513 Quick operation guide

biozoom QuickTest Portable quick start guide

biozoom

biozoom QuickTest Portable quick start guide

Katranji FDY10 user manual

Katranji

Katranji FDY10 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.