Oakton CTSTestr 5 User manual

CTSTestr™5 Pocket Tester
Features
• Measuring parameters:
conductivity, TDS, salinity,
temperature
• Large easy-to-read LCD with
2-color backlight
• Replaceable single-junction
sensor saves you money
• Auto power-off function
conserves battery life
• IP67 waterproof rating
Operating Manual
Thank you for purchasing the Oakton® CTSTestr™ 5 Multiparameter
Pocket Tester. Please carefully read this operating manual before
using the product to obtain an accurate and reliable test result
and avoid unnecessary damage to the tester or probe.
Conductivity Measurement
1. Press /MEAS key to turn on the tester. Rinse the probe in
distilled water and dry it.
2. Dip the probe in sample solution, stir gently, and allow it to
stand still in the solution until a stable reading is reached and
the stability indicator () remains on, then take a conductivity
reading.
3. Press MODE/ key to switch from Conductivity to TDS to Salinity
parameters.
Notes
• The TDS and salinity measurements are converted from the
conductivity measurements via a certain conversion factor.
• The tester can calibrate 84 μS/cm, 1413 μS/cm, and 12.88 mS/cm
conductivity calibration solution. User can conduct 1 to 3 point
calibrations (see table below). Usually calibrating the tester with
1413 μS/cm solution alone meet testing requirements.
Calibration
indication icon
Calibration
standard
Measuring
range
L84 μS/cm 0 to 200 μS/cm
M1413 μS/cm 200 to 2000 μS/cm
H12.88 mS/cm 2 to 20 mS/cm
• The tester has been calibrated before leaving the factory. Generally,
users can use the tester directly or users can test conductivity buffer
solutions rst. If the error is large, then calibration is needed.
• For conductivity calibration solutions, we recommend that users
replace with new solution after each calibration to keep the
standard solution’s accuracy. Do NOT pour the used calibration
solution back into the solution bottle in case of contamination.
• For temperature compensation factor, the default setting of the
temperature compensation factor is 2.0%/°C. User can adjust the
factor based on test solution and experimental data in parameter
setting P3.
Solution Temperature compensation factor
NaCl 2.12%/°C
5% NaOH 1.72%/°C
Dilute ammonia 1.88%/°C
10% Hydrochloric acid 1.32%/°C
5% Sulfuric acid 0.96%/°C
• TDS and conductivity are linear related, and the conversion factor
is 0.40 to 1.00. Adjust the factor in parameter setting P4 based
on the requirements in different industries. The factory default
setting is 0.71. Salinity and conductivity are linear related, and the
conversion factor is 0.5. The tester only needs to be calibrated in
Conductivity mode, then after calibration of conductivity, the tester
can switch from conductivity to TDS or salinity.
• 1000 μS/cm = 1 mS/cm; 1000 ppm = 1 ppt
• Conversion Example: If conductivity measurement is 1000 μS/cm2,
then the default TDS measurement will be 710 ppm (under the
default 0.71 conversion factor), and the salinity be 0.5 ppt.
Setting the Parameters
When tester is off, long press /MEAS key to enter parameter set-
ting mode. Short press MODE/ key to switch from P1 to P2…P7.
Short press CAL/ key and parameter will ash, then short press
MODE/ key to choose desired parameter. Short press CAL/ key to
conrm selection. Long press /MEAS key to exit parameter setting
mode.
Symbol Menu setting Selection Factory default
P1 Auto lock (HOLD) Off – On Off
P2 Backlight Off – 1 – On 1 (1-min auto-off)
P3 Temperature
compensation factor 0.00 to 4.00% 2.00%
P4 TDS factor 0.40 to 1.00 0.71
P5 Salinity unit ppt – mg/L ppt
P6 Temperature unit °C – °F °C
P7 Restore to factory
default No – Yes No
Notes
• For Automatic Lock (P1), select “On” to activate auto lock function.
When reading is stable for more than 10 seconds, the tester will
lock the value automatically, and “HOLD” icon will appear on the
bottom left of the screen. Press CAL/ key to cancel HOLD on
reading.
• For Backlight (P2), select “Off” to turn off backlight function,
“On” to turn on backlight function, or “1” to have backlight last
for 1 minute.
• For Factory Default (P7), select “Yes” to restore the calibration to
theoretical values and parameter settings to initial values. When
tester’s calibration or measurement performs abnormally, this
function can be adopted so the tester returns to factory default
setting and then users can conduct calibration or take measure-
ments again.
Keypad Functions
Short press = <2 seconds Long press = >2 seconds
MEAS
1. Short press to turn on the tester and long press to turn off the
tester.
2. When turned off, long press to enter parameter setting mode.
3. In measurement mode, short press to turn on backlight.
MODE

1. In measurement mode, short press to switch from COND
to TDS to SAL.
2. In parameter setting mode, short press to change parameter
(unidirectional).
CAL
<
1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3. When measured value is locked (HOLD icon), short press to
unlock.
Conductivity Calibration
1. Press MODE/ key to switch to conductivity measurement mode.
Rinse the probe in distilled water and dry it.
2. Pour a small amount of 1413 μS/cm and 12.88 mS/cm conductivity
calibration solution into calibration bottles.
3. Long press CAL/ key to enter calibration mode, short press
/MEAS key to exit.
4. Dip the probe in 1413 μS/cm conductivity calibration solution,
stir gently and allow it to stand still in the solution until a stable
reading is reached. When stability icon () appears and remains
on the screen, short press CAL/ key to complete the one-point
calibration. The tester returns to measurement mode and
indication icon “M” will appear at the bottom left of the screen.
5. For a 2-point calibration. After 1st point calibration, dip the probe
in 12.88 mS/cm conductivity calibration solution. If the value is
accurate, it is not necessary to conduct a 2nd point calibration. If
it is inaccurate, repeat steps #3 and #4 above to complete the
2nd point of calibration using 12.88 mS/cm buffer solution.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
1065O101_MAN_CTS5_35634-42 March 2018
Sensor Replacement
Screw off the sensor ring, unplug the sensor, plug in the new
replacement sensor (pay attention to the probe’s position), and
rescrew on the sensor ring.
Battery Installation
The tester uses four AAA batteries. Please install batteries according to
the following steps. Note the correct direction of battery installation: the
positive side (+) of every single battery must face up. Incorrect installa-
tion of batteries will cause damage to the tester and create a potential
hazard.
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3.③Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7. Fit the tester’s cap while making sure to push all the way
down. The tester’s waterproof design may be compromised if
the cap is not tted correctly.
4 x AAA Batteries
+
+
+
+
__
__
1
2
3
4
5
6
7
Warranty
We warrant this instrument to be free from defects in material and
workmanship and agrees to repair or replace free of charge, at
option of Oakton Instruments, any malfunctioned or damaged prod-
uct attributable to responsibility of Oakton Instruments, for a period
of two years from the delivery (a six-month limited warranty applies
to sensors). This warranty does not apply to defects resulting from
actions such as misuse (violation of the instructions in this manual
or operations in the manner not specied in this manual), improper
maintenance, and unauthorized repairs. Warranty period is the time
limit to provide free service for the products purchased by custom-
ers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the information in
this manual without giving notice in advance.
Self-Diagnostic Messages
Symbol Self-diagnostic
information How to fix
Er 1
Wrong conductivity
standard solution, which
exceeds the recognizable
range of the meter.
1. Check if standard solution is
correct.
2. Check if electrode is damaged.
Er 2 CAL/ is pressed before
measurement is stable
(smiley face ()appears).
Wait for the measurement stability
icon ()to appear and stay, then
press CAL/ key.
Specifications
Conductivity
Range 0 to 200.0 μS/cm, 0 to 2000 μS/cm,
0 to 20.00 mS/cm
Resolution 0.1 μS/cm, 1 μS/cm, 0.01 mS/cm
Accuracy ±1% full-scale
Calibration 1, 2, or 3 points; automatic recognition of
standards
TDS Range 0 to 100.0 ppm, 0 to 1000 ppm,
0 to 10.00 ppt
Resolution 0.1 ppm, 1 ppm, 0.01 ppt
Salinity Range 0 to 10.00 ppt
Resolution 0.01 ppt
Temperature
Range 32 to 122°F (0 to 50°C)
Resolution 0.1°F/°C
Accuracy ±0.9°F (0.5°C)
Display: LCD with two-color backlight. Blue = measurement mode;
Green = calibration mode
Reading lock: HOLD icon
Power: four AAA batteries (included); >400 hours of continuous
operation
Low-voltage warning: battery status icon ashes
Auto power-off: tester automatically turns off after 8 minutes
of nonuse
IP rating: IP67 (waterproof), oats on water when sensor cap is on
Dimensions (L x W x H): 7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm)
Weight: 4.7 oz (133 g)
Ordering Information
Model Product description Catalog
number
CTS5 CTSTestr 5 pocket multiparameter tester 35634-42
— Replacement sensor for CTSTestr 5 35634-44
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

This manual suits for next models

1

Other Oakton Test Equipment manuals

Oakton pHTestr 1 Technical manual

Oakton

Oakton pHTestr 1 Technical manual

Oakton EcoTestr EC1 User manual

Oakton

Oakton EcoTestr EC1 User manual

Oakton EC1 User manual

Oakton

Oakton EC1 User manual

Oakton WD-35662-35 User manual

Oakton

Oakton WD-35662-35 User manual

Oakton ORPTestr 30 User manual

Oakton

Oakton ORPTestr 30 User manual

Oakton CTSTestr 50 User manual

Oakton

Oakton CTSTestr 50 User manual

Oakton Eutech Instruments TDSTestr11+ User manual

Oakton

Oakton Eutech Instruments TDSTestr11+ User manual

Oakton WD-35634-06 User manual

Oakton

Oakton WD-35634-06 User manual

Oakton pHTestr 1 User manual

Oakton

Oakton pHTestr 1 User manual

Oakton Salt 1 User manual

Oakton

Oakton Salt 1 User manual

Oakton pHTestr 20 User manual

Oakton

Oakton pHTestr 20 User manual

Oakton ORP5 User manual

Oakton

Oakton ORP5 User manual

Oakton CTS5 User manual

Oakton

Oakton CTS5 User manual

Oakton EcoTestr CTS1 User manual

Oakton

Oakton EcoTestr CTS1 User manual

Oakton EcoTestr pH 2 User manual

Oakton

Oakton EcoTestr pH 2 User manual

Oakton pH 5+ User manual

Oakton

Oakton pH 5+ User manual

Oakton pH BNC 10 User manual

Oakton

Oakton pH BNC 10 User manual

Oakton TDS 1 User manual

Oakton

Oakton TDS 1 User manual

Oakton ORPTestr 50 User manual

Oakton

Oakton ORPTestr 50 User manual

Oakton PCTS5 User manual

Oakton

Oakton PCTS5 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton ORPTestr10 User manual

Oakton

Oakton ORPTestr10 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton PCTSTestr 50 User manual

Oakton

Oakton PCTSTestr 50 User manual

Popular Test Equipment manuals by other brands

Mercury TSL01 user manual

Mercury

Mercury TSL01 user manual

Watts TK-DL manual

Watts

Watts TK-DL manual

Extech Instruments 407744 user manual

Extech Instruments

Extech Instruments 407744 user manual

KNOVA KN 8065-E manual

KNOVA

KNOVA KN 8065-E manual

Dräger Interlock 5 00 Series Instructions for use

Dräger

Dräger Interlock 5 00 Series Instructions for use

Sifam Tinsley Insu 10 user manual

Sifam Tinsley

Sifam Tinsley Insu 10 user manual

HT COMBI521 user manual

HT

HT COMBI521 user manual

ETI PI-800 instruction manual

ETI

ETI PI-800 instruction manual

YOKOGAWA CL250 user manual

YOKOGAWA

YOKOGAWA CL250 user manual

EKO Instruments MP-01 quick start guide

EKO Instruments

EKO Instruments MP-01 quick start guide

Anyload LCT100 user guide

Anyload

Anyload LCT100 user guide

Kenwood CS-2110 instruction manual

Kenwood

Kenwood CS-2110 instruction manual

Tektronix 200 Series instruction manual

Tektronix

Tektronix 200 Series instruction manual

In-situ Aqua TROLL 500 quick start guide

In-situ

In-situ Aqua TROLL 500 quick start guide

Fieldpiece MR45CAN Operator's manual

Fieldpiece

Fieldpiece MR45CAN Operator's manual

Agilent Technologies 2000 X manual&nbsp;

Agilent Technologies

Agilent Technologies 2000 X manual&nbsp;

EMIT SmartLog V5 Installation, operation and maintenance

EMIT

EMIT SmartLog V5 Installation, operation and maintenance

kincrome K8085 quick start guide

kincrome

kincrome K8085 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.