Keypad Functions
Short press = <2 seconds
Long press = >2 seconds
Calibration
1. Rinse the probe in distilled water and gently dry with a
clean tissue. Short press /MEAS to turn on the tester.
2. Long press CAL/to enter calibration mode;
short press /MEAS to exit.
3. Dip the probe into 1413 µS/cm calibration
solution. Stir gently, leave it to stand. Wait
for the measurement stability icon () to
appear and stay on the display (see Fig 1);
then short press CAL/to complete the 1st
calibration. Tester returns to measurement
mode, and calibration icon “M” appears
on bottom left side of display.
4. Rinse probe in distilled water and dry it. Repeat steps
#2 and #3 to complete 2nd calibration in 12.88 mS/cm
calibration solution. Tester returns to measurement
mode, and calibration icons “M” and “H” will appear
on bottom left side of display.
Operating Manual
EcoTestr™EC1 Pocket Tester
Setting the Parameters
When tester is off, long press /MEAS to enter setup mode.
Short press /MEAS to switch from P1 to P2 to P3. Short
press CAL/and parameter will flash, then short press
/MEAS to choose desired parameter. Short press
CAL/to confirm parameter selection. Long press /MEAS
to return to measurement mode.
Notes:
• Once the readings are greater than 2000 µS/cm, the unit
will automatically become mS/cm.
• Select “Yes” for P3 to restore the calibration to the
theoretical values and parameter setting to original
values. When tester’s calibration or measurement
performs abnormally, this function can be adopted so
the tester returns to factory default setting and then
users can conduct calibration or take measurements
again.
Measurement
1. Short press /MEAS to turn on the tester. Rinse probe
in distilled water and dry it gently with a clean tissue to
remove excess water.
2. Stir the probe in the sample solution gently, leave it to
stand. Wait for the stability icon () to remain on, then
take a reading.
3. Rinse off the tester thoroughly in distilled water after
each test.
Notes:
• If this is the rst-time use or the tester hasn’t been used
for a long time, we recommend soaking the probe in
12.88 mS/cm solution for 15 to 30 minutes to restore its
sensitivity and speed up the response time.
• The tester adopts 1413 µS/cm and 12.88 mS/cm standard
calibration solutions. User can use 1 and 2 point calibra-
tions as needed. For most circumstances, calibrating in
1413 µS/cm to complete the 1st point calibration will
meet testing requirements.
• The tester has already been calibrated after manufac-
ture. User can use the tester immediately or can test it in
the calibration solutions to test its accuracy. If error is
large, calibrate the tester before using.
• We recommend replacing the calibration solution after
5 to 10 calibrations to maintain accuracy.
1433
1413
µS
Cond


Figure 1
Symbol Menu setting Selection Factory
default
P1 Select conductivity
range* Aut* – µS –
mS Aut
P2 Select temperature
unit °F – °C °C
P3 Restore to
factory default No – Yes No
*Aut = all ranges (0 to 199.9 µS/cm, 20 to 1999 µS/cm, 2.0 to 20.0 mS/cm)
1. Short press to turn on; long press to
turn off.
2. When turned off, long press to enter
setup mode.
3. In mode setting, short press to change
parameter.
1. When turned on, long press to enter
calibration mode.
2. In calibration mode, short press to
confirm calibration.
3. In mode setting, short press to confirm
parameter selection.
CAL
<
MEAS
1065O101_MAN_EC1_35634-06.indd 1 03/22/2018 3:49:52 PM
Specifications
Power: four AAA batteries (included); >400 hours of
continuous operation
Low-voltage warning: battery status icon flashes
Auto power-off: tester automatically turns off after 8 minutes
of nonuse
IP rating: IP67 (waterproof), oats on water when sensor cap
is on
Dimensions (L x W x H): 7" x 1.5" x 1.25" (17.8 x 4 x 3.1 cm)
Weight: 3.8 oz (107 g)
1065O101_MAN_EC1_35634-06 March 2018
www.4oakton.com
Conductivity
Range 0 to 200.0 µS/cm,
0 to 2000 µS/cm,
0 to 20.00 mS/cm
Resolution 0.1/1 µS/cm, 0.01 mS/cm
Accuracy ±1% full-scale ± 1 digit
Calibration
points 1 or 2 points; auto
standard recognition
Automatic
temperature
compensation 32 to 122°F (0 to 50°C)
Temperature
Range 32 to 122°F (0 to 50°C)
Resolution 0.1°F/°C
Accuracy ±0.9°F (0.5°C)
Symbol Self-diagnosis
information How to fix
Er 1 Wrong calibration
solution, or measured
value is not within the
range of the tester.
• Check if calibration
solution is correct.
• Check if probe is
damaged.
Er 2 CAL/is pressed
before measurement
is stable (appears).
Wait for the
measurement
stability icon () to
appear and stay, then
press CAL/.
Ordering Information
Battery Installation
The tester uses four AAA batteries. Please install batteries
according to the following steps. Note the correct direction
of battery installation: the positive side (+) of every single
battery must face up. Incorrect installation of batteries will
cause damage to the tester and create a potential hazard.
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of
arrow.
7. Fit the tester’s cap while making sure to push all the
way down. The tester’s waterproof design may be
compromised if the cap is not fitted correctly.
Self-Diagnostic Messages
Model Product description Catalog number
EC1 EcoTestr pocket
conductivity tester 35634-06
4 x AAA Batteries
+
+
+
+
__
__
1
2
3
4
5
6
7
Warranty
We warrant this instrument to be free from defects in
material and workmanship and agrees to repair or replace
free of charge, at option of Oakton Instruments, any
malfunctioned or damaged product attributable to
responsibility of Oakton Instruments, for a period of
two years from the delivery (a six-month limited warranty
applies to probes). This warranty does not apply to defects
resulting from actions such as misuse (violation of the
instructions in this manual or operations in the manner
not specified in this manual), improper maintenance, and
unauthorized repairs. Warranty period is the time limit to
provide free service for the products purchased by
customers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the
information in this manual without giving notice in
advance.
1065O101_MAN_EC1_35634-06.indd 2 03/22/2018 3:49:57 PM

Other Oakton Test Equipment manuals

Oakton WD-35634-06 User manual

Oakton

Oakton WD-35634-06 User manual

Oakton TempTestr 35628-00 User manual

Oakton

Oakton TempTestr 35628-00 User manual

Oakton TDS 1 User manual

Oakton

Oakton TDS 1 User manual

Oakton ORPTestr 30 User manual

Oakton

Oakton ORPTestr 30 User manual

Oakton CTSTestr 5 User manual

Oakton

Oakton CTSTestr 5 User manual

Oakton PCTSTestr 50 User manual

Oakton

Oakton PCTSTestr 50 User manual

Oakton EcoTestr pH 2 User manual

Oakton

Oakton EcoTestr pH 2 User manual

Oakton PCTS5 User manual

Oakton

Oakton PCTS5 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton CTSTestr 50 User manual

Oakton

Oakton CTSTestr 50 User manual

Oakton WD-35662-35 User manual

Oakton

Oakton WD-35662-35 User manual

Oakton ORP5 User manual

Oakton

Oakton ORP5 User manual

Oakton pH 5+ User manual

Oakton

Oakton pH 5+ User manual

Oakton pHTestr 1 User manual

Oakton

Oakton pHTestr 1 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton ORPTestr 50 User manual

Oakton

Oakton ORPTestr 50 User manual

Oakton pHTestr 1 Technical manual

Oakton

Oakton pHTestr 1 Technical manual

Oakton pHTestr 20 User manual

Oakton

Oakton pHTestr 20 User manual

Oakton Eutech Instruments TDSTestr11+ User manual

Oakton

Oakton Eutech Instruments TDSTestr11+ User manual

Oakton ORPTestr10 User manual

Oakton

Oakton ORPTestr10 User manual

Oakton EcoTestr CTS1 User manual

Oakton

Oakton EcoTestr CTS1 User manual

Oakton EC1 User manual

Oakton

Oakton EC1 User manual

Oakton pHTestr 30 User manual

Oakton

Oakton pHTestr 30 User manual

Oakton pH BNC 10 User manual

Oakton

Oakton pH BNC 10 User manual

Popular Test Equipment manuals by other brands

SKF TKRS 11 Instructions for use

SKF

SKF TKRS 11 Instructions for use

Tektronix 515 instruction manual

Tektronix

Tektronix 515 instruction manual

Sealey BT91/3 instructions

Sealey

Sealey BT91/3 instructions

ADTRAN DS3 NIU MTC Job aid

ADTRAN

ADTRAN DS3 NIU MTC Job aid

WILLTEK 4403 Getting started manual

WILLTEK

WILLTEK 4403 Getting started manual

insize ISH-MRD200 Operation manual

insize

insize ISH-MRD200 Operation manual

BW Technologies MicroDock II user manual

BW Technologies

BW Technologies MicroDock II user manual

Keysight Technologies N4877A Getting started

Keysight Technologies

Keysight Technologies N4877A Getting started

Extech Instruments 422123 user manual

Extech Instruments

Extech Instruments 422123 user manual

Rohde & Schwarz RTP-K110 user manual

Rohde & Schwarz

Rohde & Schwarz RTP-K110 user manual

ED&D LT-952HC user manual

ED&D

ED&D LT-952HC user manual

VeEX VePAL TX320s user manual

VeEX

VeEX VePAL TX320s user manual

Tektronix MSO4104B Service manual

Tektronix

Tektronix MSO4104B Service manual

Pundit PI8000 quick reference

Pundit

Pundit PI8000 quick reference

Millipore Integritest Exacta Operating and maintenance manual

Millipore

Millipore Integritest Exacta Operating and maintenance manual

WHALETEQ MECG 2.0 user manual

WHALETEQ

WHALETEQ MECG 2.0 user manual

Triplett CamView5 user manual

Triplett

Triplett CamView5 user manual

Nieaf-Smitt EazyVolt Basic instruction manual

Nieaf-Smitt

Nieaf-Smitt EazyVolt Basic instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.