SZX–RFL2
10
4. Screw in the burner focusing knob ƒ fully.
5. Using the burner centering knobs …, adjust so that two arc images of the
specimen will be visible as shown in Fig. 16-A .
# During centering with the minimum zoom magnification, the left side
of the illuminated field may be vignetted. However, the illuminated
field in actual use is not affected.
6. Using the collector lens focusing knob † bring either one of the two arc
images into focus. (Fig. 16-B)
7. Using the burner focusing knob ƒ, adjust so that the sizes of two arc
images will be almost identical. (Fig. 16-C)
# At this point, the two arc images may shift. In this case, adjust the
positions of the arc images using the burner centering knobs ….
When it is difficult to determine the sizes of the arc images, adjust by
using the collector lens focusing knob † to bring the arc images into
focus.
8. While turning the collector lens focusing knob † repeatedly, check that
the degree of blurring of one arc image is equal to that of the other arc
image. (That is, when one of two arc image is focused, check that the
other arc image is in focus also.)
If the degree of blurring of one arc image is different from that of the other
arc image (the focusing positions of two arc images do not match each
other), repeat the above steps 6 – 8.
9. After making sure that the degree of blurring of one arc image is equal to
that of the other arc image (the focusing positions of two arc images are
identical), overlay the two arc images using the burner centering knobs …
under in-focus condition.
(Fit section ‡ on the collector lens focusing knob and adjust by turning
knob Š.)
Fig. 15 illuminated field
Fig. 16-A
Fig. 16-B
Fig. 16-C
Fig. 16-D
ƒ
…
†
‡
Š
10. After the two arc images are overlaid, they appear as shown in Fig. 16-D.
This completes the burner centering adjustment.
Note that even if the arc images deviate from the center position to some
extent, it will not pose any trouble in observation practice.
11. Before proceeding to observation, adjust the collector lens focusing knob
† so that illumination on the visual field will be uniform.
}When you have difficulties in operating the collector lens focusing knob,
use the optional U-CLA collector lens adjustment handle which is shown
below.
# To avoid serious injury, never open the lamp housing while the burner
is turned on or immediately after it is turned off.
}Recenter the burner each time it is replaced.
Collector lens adjustment handle
U–CLA