
11
Safety information
2.3.5 Risks regarding burns
WARNING
Burns to the patient and/or the user
‒The maximum output voltage characteristics of the electrosurgical generator are shown
in the diagrams in section “17.5 Output characteristic diagrams” on page 124. When
setting the power level, rst set it to a low level and increase it gradually. If the output is
initially set to a high level, the electrode’s insulation may be damaged and cause user
and/or patient burns.
However, certain modes may present an unacceptable risk at low output power
settings. For example, with the PulseCut fast mode or PulseCut slow mode, the risk of
an excessive thermal effect rises if the output power setting is too low. Therefore, it is
recommended that you perform basic testing before using the electrosurgical generator.
If the instruction manual of the HF instrument to be used stipulate a rated voltage, the
output should be set so that it does not exceed that voltage.
‒Contact with the tip of the electrodes may cause burns when the electrosurgical
generator is active.
‒During operation, temporarily unused electrodes should be stored in an electrically
insulated container. Unused electrodes or HF instruments should never be placed on the
patient. Otherwise, it may cause patient and/or user burns.
‒To prevent patient burns, the electrosurgical generator and ancillary cords should not
come in contact with the patient or metal parts of the operating table. Furthermore,
the patient should also be kept away from metallic parts of the operating table or other
devices. Remove any metallic items from the patient (wristwatches, jewelry, etc.) before
starting the procedure.
‒If the electrosurgical generator is used in conjunction with another electrosurgical
generator, never use both generators simultaneously. Keep the HF instrument connected
to the not-used electrosurgical generator away from the target area while the other
generator is in operation. Do not activate output of both generators simultaneously.
Patient or user injury may occur due to the concentration of electric current.
‒To prevent patient burns, the patient’s skin surfaces should not touch each other (e.g.
bare arm and side of chest) or any metal items in the procedure room.
‒To prevent patient burns during high frequency treatment, the patient’s clothes must be
dry.
‒During endoscopic treatment be sure that the distal end of the endoscope and/or
HF instruments do/does not contact bridging uids surrounding the target tissue. Electric
current may ow to the surrounding tissue via the uids and cause burns. This does not
apply to instruments intended for use in conjunction with conductive uids.
‒The endoscopic treatment performed should not include an operation in which part of
the treated tissue (polyp head, etc.) or part of the endoscope distal end or endotherapy
instrument is in contact with or close to surrounding tissue during high frequency
exposure. Otherwise, current ows to the tissue through the part of the treated tissue,
the metallic parts at the endoscope’s distal end or endotherapy instrument and may
cause burns.
‒When using an electrocardiograph or other physiological monitoring equipment
simultaneously with the electrosurgical generator on a patient, any monitoring
electrodes should be placed as far away as possible from the electrodes used with
the electrosurgical generator. If placed too close, high frequency signals or spark
discharge noise from the electrosurgical generator may interfere with the operation of an
electrocardiograph or other physiological monitoring equipment.
‒Needle monitoring electrodes should not be used, as they may cause patient burns.
Physiological monitoring equipment incorporating high frequency current limiting
measures is recommended.
‒To prevent burns, the user and assistant should wear surgical gloves during the
procedure.