PASCO ES-9070 User manual

012-03760E
05/99
©1989 PASCO scientific $5.00
COULOMB BALANCE
Instruction Manual and
Experiment Guide for
the PASCO scientific
Model ES-9070
Includes
Teacher's Notes
and
Typical
Experiment Results


012-03760E Coulomb Balance
i
Table of Contents
Section Page
Copyright, Warranty and Equipment Return...................................................ii
Introduction .....................................................................................................1
Theory .............................................................................................................2
Equipment........................................................................................................3
Additional Equipment Recommended: .....................................................3
Tips for Accurate Results ................................................................................4
Setup .............................................................................................................5
Experiments:
Part A Force Versus Distance...................................................................7
Part B Force Versus Charge ......................................................................8
Part C The Coulomb Constant...................................................................9
Replacing the Torsion Wire........................................................................... 13
Teacher’s Guide............................................................................................. 15
Technical Support.................................................................Inside Back Cover

Coulomb Balance 012-03760E
ii
Credits
Author: Bruce Lee
Editor: Dave Griffith
Equipment Return
Should the product have to be returned to PASCO
scientific for any reason, notify PASCO scientific
by letter, phone, or fax BEFORE returning the
product. Upon notification, the return authorization
and shipping instructions will be promptly issued.
When returning equipment for repair, the units
must be packed properly. Carriers will not accept
responsibility for damage caused by improper
packing. To be certain the unit will not be damaged
in shipment, observe the following rules:
➀The packing carton must be strong enough for the
item shipped.
➁Make certain there are at least two inches of
packing material between any point on the
apparatus and the inside walls of the carton.
➂Make certain that the packing material cannot shift
in the box or become compressed, allowing the
instrument come in contact with the packing
carton.
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone: (916) 786-3800
FAX: (916) 786-3292
email: [email protected]
web: www.pasco.com
ä
NOTE: NO EQUIPMENT WILL BE
ACCEPTED FOR RETURN WITHOUT AN
AUTHORIZATION FROM PASCO.
Copyright Notice
The PASCO scientific 012-03760E Coulomb
Balance manual is copyrighted and all rights
reserved. However, permission is granted to non-
profit educational institutions for reproduction of
any part of the manual providing the reproductions
are used only for their laboratories and are not sold
for profit. Reproduction under any other circum-
stances, without the written consent of PASCO
scientific, is prohibited.
LimitedWarranty
PASCO scientific warrants the product to be free
from defects in materials and workmanship for a
period of one year from the date of shipment to the
customer. PASCO will repair or replace at its option
any part of the product which is deemed to be
defective in material or workmanship. The warranty
does not cover damage to the product caused by
abuse or improper use. Determination of whether a
product failure is the result of a manufacturing
defect or improper use by the customer shall be
made solely by PASCO scientific. Responsibility
for the return of equipment for warranty repair
belongs to the customer. Equipment must be
properly packed to prevent damage and shipped
postage or freight prepaid. (Damage caused by
improper packing of the equipment for return
shipment will not be covered by the warranty.)
Shipping costs for returning the equipment after
repair will be paid by PASCO scientific.
Copyright, Warranty, and Equipment Return
Please—Feel free to duplicate this manual
subject to the copyright restrictions below.

012-03760E Coulomb Balance
1
The PASCO Model ES-9070 Coulomb Balance
(Figure 1) is a delicate torsion balance that can be used to
investigate the force between charged objects. A
conductive sphere is mounted on a rod, counterbalanced,
and suspended from a thin torsion wire. An identical
sphere is mounted on a slide assembly so it can be
positioned at various distances from the suspended sphere.
To perform the experiment, both spheres are charged, and
the sphere on the slide assembly is placed at fixed
Figure 1. Experimenting with the Coulomb Balance
slideassembly
torsionwire
chargedspheres
distances from the equilibrium position of the suspended
sphere. The electrostatic force between the spheres causes
the torsion wire to twist. The experimenter then twists the
torsion wire to bring the balance back to its equilibrium
position. The angle through which the torsion wire must be
twisted to reestablish equilibrium is directly proportional
to the electrostatic force between the spheres.
All the variables of the Coulomb relationship
(F = kq1q2/R2) can be varied and measured using the
Coulomb Balance. You can verify the inverse square
relationship and the charge dependence using the balance
and any electrostatic charging source.
However, for best results, we recommend you charge the
spheres with a stable kilovolt power supply to ensure a
reproducible charge throughout the experiment. To
determine the Coulomb constant with reasonable
accuracy, we recommend you use an electrometer and a
Faraday ice pail to accurately measure the charge on the
spheres. For more information about accuracy, read the
section Tips for Accurate Results.
Introduction

Coulomb Balance 012-03760E
2
Theory
Take one gram of protons and place them one meter away
from one gram of electrons. The resulting force is equal to
1.5 x 1023 newtonsroughly the force it would take to
lift an object from the surface of the Earth that had a
mass about 1/5 that of the moonnot a small force.
So, if such small amounts of charge produce such
enormous forces, why does it take a very delicate torsion
balance to measure the force between charged objects in
the laboratory? In a way, the very magnitude of the forces
is half the problem. The other half is that the carriers of the
electrical forcethe tiny proton and the even tinier
electronare so small, and the electrons are so mobile.
Once you separate them, how do you keep them
separated? The negatively charged electrons are not only
drawn toward the positively charged protons; they also
repel each other. Moreover, if there are any free electrons
or ions between the separated charges, these free charges
will move very quickly to reduce the field caused by the
chargeseparation.
So, since electrons and protons stick together with such
tenacity,onlyrelativelysmallchargedifferentialscanbe
sustained in the laboratory. This is so much the case that,
even though the electrostatic force is more than a billion-
billion-billion-billiontimesasstrongasthegravitational
force, it takes a very delicate torsion balance to measure
the electrical force, whereas we can measure the
gravitationalforce by weighinganobject with aspring
balance.
ä
NOTE: The torsion balance gives a direct and
reasonably accurate measurement of the Coulomb
force. The most accurate determinations of
Coulomb's law, however, are indirect. It can be
shown mathematically that if the inverse square law
holds for the electrostatic force, the electric field
inside a uniformly charged sphere must be
everywhere zero. Measurements of the field inside a
charged sphere have shown this to be true with
remarkable accuracy. The Coulomb force can be
expressedby the formula:
F = kq1q2/R2+n.
Using this indirect method, it has been demonstrated
experimentally that n 2 x 1016.

012-03760E Coulomb Balance
3
The Coulomb Balance and the included accessories are
shown in Figure 2.
Equipment
(The Coulomb Balance and the slide assembly should be
shipped with one of the conductive spheres unattached.
See the Setup section of this manual.)
Additional Equipment Recommended:
A stable kilovolt power supply for charging the
spheresAny electrostatic charger can be used to
charge the spheres, but a power supply lets you
replenish the charge to a fixed value throughout an
experiment. Ideally the supply would have a
momentary power on button so that you can
conveniently turn it off whenever you are not
charging the spheres.
Figure 2. The Coulomb Balance
chargingprobe
slideassembly
allenwrench for theslide
assembly
conductivesphereoninsulating
thread
torsionwire
clamp
COULOMB
BALANCE magnetic
dampingarm
indexarm
calibration
supporttube
toolboxwith:
one 50 mg mass
two 20 mg masses
one hex key
torsionwire
retainer
(pendulum)
coupling
Plate
sparetorsion wire
(3meters)
ä
NOTE: The balance may
be shipped with the copper
rings unattached.
An electrometer and Faraday ice pail (such as
PASCO Models ES-9054A and ES-9058) for
accurately measuring the charge on the spheres.
A spring balance capable of measuring a force of
approximately 4 newtons (400 gram mass). This is
not necessary for the experiment itself, but is helpful
in setting the tension of the torsion wire.

Coulomb Balance 012-03760E
4
IMPORTANT: If you live in an area where
humidity is always high, and if you have no facilities
for controlling humidity, the experiment will be
difficult, if not impossible, to perform. Static
charges are very hard to maintain in a humid
atmosphere because of surface conductivity.
Experiments with the Coulomb Balance are
straightforward and quite accurate, yet, as with any
quantitative electrostatic experiment, frustration lurks just
around the corner. A charged shirt sleeve, an open
window, an excessively humid dayany of these and
more can affect your experiment. However, if you
carefully follow the tips listed below, youve got a good
start toward a successful experiment.
Perform the experiment during the time of year when
humidity is lowest.
Perform the experiment in a draft-free room.
The table on which you set up the experiment should
be made of an insulating materialwood, masonite,
plastic, etc. If a metal table is used, image charges
will arise in the table that will significantly affect the
results. (This is also true for insulating materials, but
the effect is significantly reduced.)
Position the torsion balance at least two feet away
from walls or other objects which could be charged
or have a charge induced on them.
When performing experiments, stand directly behind
the balance and at a maximum comfortable distance
from it. This will minimize the effects of static
charges that may collect on clothing.
Avoid wearing synthetic fabrics, because they tend to
acquire large static charges. Short sleeve cotton
clothes are best, and a grounding wire connected to
the experimenter is helpful.
Use a stable, regulated kilovolt power supply to
charge the spheres. This will help ensure a constant
charge throughout an experiment.
When charging the spheres, turn the power supply
on, charge the spheres, then immediately turn the
supply off. The high voltage at the terminals of the
supply can cause leakage currents which will affect
the torsion balance. A supply with a momentary
power on button is ideal.
When charging the spheres, hold the charging probe
near the end of the handle, so your hand is as far
from the sphere as possible. If your hand is too close
to the sphere, it will have a capacitive effect,
increasing the charge on the sphere for a given
voltage. This effect should be minimized so the
charge on the spheres can be accurately reproduced
when recharging during the experiment.
If you are using a PASCO Electrometer (Model
ES-9035 or ES9054A) to measure the charge on the
spheres, connect the voltage output to a digital
multimeter so that values can be measured more
accurately. It is also useful to calibrate the
electrometer. This is done by applying a calibrating
voltage to the input and measuring the electrometer
output on the digital multimeter. Your measured
values can then be adjusted as necessary.
Surface contamination on the rods that support the
charged spheres can cause charge leakage. To
prevent this, avoid handling these parts as much as
possible and occasionally wipe them with alcohol to
removecontamination.
There will always be some charge leakage. Perform
measurements as quickly as possible after charging,
to minimize the leakage effects.
Recharge the spheres before each measurement.
Tips for Accurate Results

012-03760E Coulomb Balance
5
Setup
ä
Note Threading the Torsion Wire: The
Torsion Balance is shipped to you with the wire
already threaded. However, if it ever breaks,
you can thread it using the spare wire that is sup-
plied. See the procedure at the end of this manual.
TorsionBalanceSetup
➀One of the conductive spheres is not attached when the
Coulomb Balance is shipped. To attach it, just slip the
stem of the sphere over the fiber glass rod of the
pendulumassembly.
➁Slide the copper rings onto the counterweight vane, as
shown in the bottom of Figure 3. Then release the
packing clamp that holds the counterweight vane, as
shown in the top of Figure 3. Adjust the position of the
copper rings so the pendulum assembly is level.
➂Reposition the index arm so it is parallel with the base
of the torsion balance and at the same height as the
vane.
ä
Important: When storing the torsion balance,
always clamp the counterweight vane to protect the
torsion wire. When you do this, be sure to adjust the
height and angle of the index arm so that you can
clamp the vane without pulling on the torsion wire.
➃Adjust the height of the magnetic damping arm so the
counterweight vane is midway between the magnets.
➄Turn the torsion knob until the index line for the degree
scale is aligned with the zero degree mark.
counterweight
vane 1. Loosen the top
thumbscrew
2. Loosen the side
thumbscrew and
rotate the index
arm so it is parallel
with the base of the
Coulomb Balance.
Then tighten the
thumbscrew.
packing
clamp
Releasing the packing clamp
degreescale
torsionknob
lateralsupport
bar
Figure 3. Setting Up the Coulomb Balance
magnetic
dampingarm
Index arm
supporttube
torsionwireretainer
pendulum
assembly counterweight
vane
copperrings
copperrings
support
tube
lateralsupportbar
Figure 4. Zeroing the Torsion Balance

Coulomb Balance 012-03760E
6
1
2
3
4
5
side view top view
➅Rotate the bottom torsion wire retainer (do not loosen
or tighten the thumbscrew) until the index line on the
counterweight vane aligns with the index line on the
index arm.
➆Carefully turn the torsion balance on its side,
supporting it with the lateral support bar, as shown in
Figure 4. Place the support tube under the sphere, as
shown.
➇Adjust the positions of the copper rings on the
counterweight vane to realign the index line on the
counterweight with the index line on the index arm.
➈Place the torsion balance upright.
Slide Assembly Setup
(Refer to Figure 5)
➀Connect the slide assembly to the torsion balance as
shown in Figure 5, using the coupling plate and
thumbscrews to secure it in position.
➁Align the spheres vertically by adjusting the height of
the pendulum assembly so the spheres are aligned: Use
the supplied allen wrench to loosen the screw that
anchors the pendulum assembly to the torsion wire.
Adjust the height of the pendulum assembly as needed.
Readjust the height of the index arm and the magnetic
damping arm as needed to reestablish a horizontal
relationship.
➂Align the spheres laterally by loosening the screw in
the bottom of the slide assembly that anchors the
vertical support rod for the sphere, using the supplied
allen wrench (the vertical support rod must be moved
to the end of the slide assembly, touching the white
plastic knob to access the screw). Move the sphere on
the vertical rod until it is laterally aligned with the
suspended sphere and tighten the anchoring screw.
➃Position the slide arm so that the centimeter scale
reads 3.8 cm (this distance is equal to the diameter of
the spheres).
➄Position the spheres by loosening the thumbscrew on
top of the rod that supports the sliding sphere and
sliding the horizontal support rod through the hole in
the vertical support rod until the two spheres just
touch. Tighten the thumbscrew.
You're now ready to experiment. The degree scale should
read zero, the torsion balance should be zeroed (the index
lines should be aligned), the spheres should be just
touching, and the centimeter scale on the slide assembly
should read 3.8 cm. (This means that the reading of the
centimeter scale accurately reflects the distance between
the centers of the two spheres.)
Figure 5. Slide Assembly Setup

012-03760E Coulomb Balance
7
Experiment: (PartA) Force Versus Distance
suspended
sphere sliding
sphere
slide assembly
pendulum
assembly
Procedure
➀Set up the Coulomb Balance as described in the
previoussection.
➁Be sure the spheres are fully discharged (touch
them with a grounded probe) and move the sliding
sphere as far as possible from the suspended sphere.
Set the torsion dial to 0×C. Zero the torsion balance
by appropriately rotating the bottom torsion wire
retainer until the pendulum assembly is at its zero
displacement position as indicated by the index
marks.
➂With the spheres still at maximum separation, charge both the spheres to a potential of
6-7 kV, using the charging probe. (One terminal of the power supply should be grounded.)
Immediately after charging the spheres, turn the power supply off to avoid high voltage leakage
effects.
ä
IMPORTANT: Read the section Tips for Accurate Results. It has some helpful hints about
charging the spheres.
➃Position the sliding sphere at a position of 20 cm. Adjust the torsion knob as necessary to balance
the forces and bring the pendulum back to the zero position. Record the distance (R) and the angle
(q) in Table 1.
➄Separate the spheres to their maximum separation, recharge them to the same voltage, then
reposition the sliding sphere at a separation of 20 cm. Measure the torsion angle and record your
results again. Repeat this measurement several times, until your result is repeatable to within ± 1
degree. Record all your results.
➅Repeat steps 3-5 for 14, 10, 9, 8, 7, 6 and 5 cm.
Analysis
ä
NOTE: In this part of the experiment, we are assuming that force is proportional to the
torsion angle. If you perform Part C of the experiment, you will test this assumption when you
calibrate the torsion balance.
Determine the functional relationship between the force, which is proportional to the torsion angle
(q); and the distance (R). This can be done in the following ways:
➀Plot log q versus log R.
Explanation: If q= bRn, where b and n are unknown constants, then log q= n log R + log b.
The slope of the graph of log qversus log R will therefore be a straight line. Its slope will be equal
to n and its y intercept will be equal to log b. Therefore, if the graph is a straight line, the
functionisdetermined.
Figure 6. Experimental Setup

Coulomb Balance 012-03760E
8
➁Plot qversus R2
Either of these methods will demonstrate that, for relatively large values of R, the force is propor-
tional to 1/R2. For small values of R, however, this relationship does not hold.
Corrections to the data
The reason for the deviation from the inverse square relationship at short distances is that the charged
spheres are not simply point charges. A charged conductive sphere, if it is isolated from other
electrostatic influences, acts as a point charge. The charges distribute themselves evenly on the
surface of the sphere, so that the center of the charge distribution is just at the center of the sphere.
However, when two charged spheres are separated by a distance that is not large compared to the size
of the spheres, the charges will redistribute themselves on the spheres so as to minimize the electro-
static energy. The force between the spheres will therefore be less than it would be if the charged
spheres were actual point charges.
A correction factor B, can be used to correct for his deviation. Simply multiply each value of qby 1/
B, where
B
=1–4a
3
R
3
;
where a equals the radius of the spheres and R is the separation between spheres.
To correct your data:
➀Calculate the correction factor B for each of the separations R that you used. Record your results in
Table 1.
➁Multiply each of your collected values of qby 1/B and record your results as qcorrected.
➂Reconstruct your graphs relating force and separation, but this time use qcorrected instead of q. Make
your new plot on the same graph as your original plot. How does the correction factor affect your
results?
(Part B) Force Versus Charge
With the sphere separation (R) held at a constant value (choose a value between 7 and 10 cm), charge
the spheres to different values and measure the resulting force. Keep the charge on one sphere
constant, and vary the charge on the other. Then graph angle versus charge to determine the relation-
ship.
The charge can be varied using either of two methods:
Method I:
If your power supply is adjustable, simply charge the spheres to different potentials, such as 7, 6, 5,
4, and 3 kV. (When charging the spheres, they should always be at their maximum separation.) The
charge on the sphere is proportional to the charging potential.
Method II:
If your power supply voltage is not adjustable, the charge can be changed by touching one or both of
the spheres with an identical sphere that is discharged. The charge will be shared equally between the
charged and discharged sphere. Therefore, touch the charged sphere once to reduce the charge by
half, twice to reduce the charge by 1/4, etc.

012-03760E Coulomb Balance
9
(Part C) The Coulomb Constant
In parts A and B of this lab, you determined (if all went well) that the electrostatic force between two point
charges is inversely proportional to the square of the distance between the charges and directly proportional
to the charge on each sphere. This relationship is stated mathematically in Coulombs Law:
F=kq1q2
R2
;
where F is the electrostatic force, q1and q2are the charges, and R is the distance between the charges. In
order to complete the equation, you need to determine the value of the Coulomb constant, k. To accom-
plish this, you must measure three additional variables: the torsion constant of the torsion wire (Ktor), so you
can convert your torsion angles into measurements of force, and the charges, q1and q2. Then, knowing F,
q1, q2, and R, you can plug these values into the
Coulomb equation to determine k.
Measuring the Torsion constant, K
➀Carefully turn the Torsion Balance on its side,
supporting it with the lateral support bar, as shown
in Figure 7. Place the support tube under the
sphere, as shown.
➁Zero the torsion balance by rotating the torsion
dial until the index lines are aligned. Record the
angle of the degree plate in Table 2.
➂Carefully place the 20 mg mass on the center line
of the conductive sphere.
➃Turn the degree knob as required to bring the
indexlines back intoalignment.Read the torsion
angle on the degree scale. Record the angle in
Table 2.
➄Repeat steps 3 and 4, using the two 20 mg masses and the 50 mg mass to apply each of the masses shown
in the table. Each time record the mass and the torsion angle.
➅Complete the table as follows to determine the torsion constant for the wire:
a. Calculate the weight for each set of masses that you used.
b. Divide the weight by the torsion angle to determine the torsion constant at each weight.
c. Average your measured torsion constants to determine the torsion constant for the wire. Use the
variance in your measured values as an indication of the accuracy of your measurement.
ä
NOTE: A torsion constant for a wire usually expresses the torque required to twist the wire a unit
angle, and is normally expressed in newton meters per degree. However, when using the torsion balance,
the torque arm is always the same (the distance from the center of the conductive sphere to the torsion
wire), so the torsion constant for the balance is more conveniently expressed in newtons per degree.
copperrings
support
tube
lateralsupportbar
Figure 7. Calibrating the Torsion Balance
center
line
mass

Coulomb Balance 012-03760E
10
Measuring the Charges, q1and q2
Method I:
The capacitance of an isolated conductive sphere is given by the equation:
C = 4 pe0a;
where C is the capacitance, e0= 8.85 x 1012 F/m, and a = the radius of the sphere.
For a capacitor, charge (q) and charging potential (V) are related by the equation: q = CV. You can
use this equation to determine the charge on the spheres from your applied charging potential.
This is the simplest method for determining the charge on the spheres. Unfortunately, the conducting
spheres of the Coulomb Balance are not isolated in this application, so the measured values of q will
beonlyapproximate.
ä
NOTE: A capacitor normally consists of two conductors. The charge on one conductor is +q and
the charge on the other is q. V is the potential difference between the two conductors. For an
isolated sphere with a charge +q, the second conductor is a hypothetical plane at ground potential
and with charge q, located at a distance infinitely far from the sphere.
Method II:
The charge on the spheres can be measured more
accurately using an electrometer with a Faraday ice
pail. The setup for the measurement is shown in
Figure 8. The electrometer and ice pail can be
modeled as an infinite impedance voltmeter in
parallel with a capacitor. A sphere with a charge q is
touched against the ice pail. Since the capacitance of
the ice pail and electrometer is much greater than
that of the sphere, virtually all of the charge q is
transferred onto the ice pail. The relationship
between the voltage reading of the electrometer and
the charge deposited into the system is given by the
equation q = CV, where C is the combined capaci-
tance of the electrometer, the ice pail, and the
connecting leads. Therefore, in order to determine
the charge, you must know the capacitance of the
system.
The simplest way to measure the capacitance of the
electrometer and ice pail is to use a good capaci-
tance meter connected between the inside and
outside conductors of the ice pail (the electrometer
must be connected to the ice pail during the measurement). A second method is to charge a precision
capacitor with capacitance equal to Ctest (³ 250 pF) to a known voltage Vtest (10 - 30 V). The charge
on the capacitor is then equal to qtest = CtestVtest. Place the leads of the charged capacitor between the
inside and outside conductors of the ice pail. The charge qtest is now distributed across two parallel
capacitors, the precision capacitor and the capacitance of the ice pail and electrometer system.
Therefore: CtestVtest = (C + Ctest)V; where C is the capacitance of the electrometer and Faraday ice pail
and V is the voltage reading of the electrometer.
Figure 8. Measuring the Charge with an
Electrometer and a Faraday Ice Pail
Faraday
ice pail
Conductive sphere on an
insulating thread
Electrometer

012-03760E Coulomb Balance
11
Therefore C = Ctest (Vtest - V)/V. Once youve measured the capacitance C, measure the charge of
thecharged sphere is follows:
➀Discharge the conducting sphere on the insulating thread, by touching it to a grounded probe.
➁Holding the sphere by the insulating thread, touch it to the charged sphere, then to the inner
conductor of the ice pail.
➂The charge on the original charged sphere, q, can now be determined using the equation:
q = 2CV;
where C is the capacitance of the electrometer and ice pail and V is the reading on the electrom-
eter. (The factor of two arises because, in using the test sphere to sample the charge on the
originalsphere, only halftheoriginal charge was transferred.)
Calculations for the Coulomb Constant
The Coulomb constant can now be determined by using any data pair from your force versus
distancedata.
➀Convert your torsion angle measurement (qcorrected) to a force measurement, using your measured
torsion constant for the torsion wire: F = Ktorqcorrected.
➁Determine the charge that was on the sphere using Method I or Method II above. If you are using
Method II, you will need to recharge the sphere to the voltage previously used while taking data,
so that you can determine the charge using the electrometer and the Faraday ice pail.
➁Plug your collected data into the Coulomb equation, F = k q1q2/R2, to determine the value of k.
Do this for several sets of data. Average your results to determine a value for k.

Coulomb Balance 012-03760E
12
mg mg/q
mq
0 mg
20 mg
40 mg
50 mg
70 mg
Table 2. Force Calibration
Data and Calculations
Torsion constant = Ktor
B
1 - 4a3/R3
R
qq
avg qcorrected 1/R2
Table 1. Force versus Distance
Data and Calculations
C (Capacitance of Electrometer System) =
V (Electrometer Voltage) =
q (Charge on sphere) = 2CV =
Table 3. The Charge on the Sphere

012-03760E Coulomb Balance
13
To replace the torsion wire, follow the numbered steps
in Figure 9. When you're done, follow the instructions
in the Setup section of this manual to balance the pen-
dulum and to zero the torsion balance.
ä
IMPORTANTWhen Replacing the
Wire:
➀Begin with a length of wire at least 50 cm
long (if you've done this before, you may not
need such a long piece).
➁Be careful not to kink the wire.
➂As you thread the wire, the end may become
bent or kinked. It will help to clip the end off
so it remains straight.
➃Tighten the screws that hold the torsion wire
gently. Overtightening will break the wire.
➄The tension on the torsion wire is not a
critical variable, as long as the wire is
reasonably taut. The advantage of the spring
balance is that it helps you adjust the tension,
without pulling too hard and breaking the
wire. If you don't have a spring balance, you
can adjust the tension by feel. Just take care
not to break the wire.
Replacing the Torsion Wire

Coulomb Balance 012-03760E
14
1. a. Use the allen wrench to
loosen both set screws
one full turn.
b. Thread the wire in the
direction shown.
c. Do not tighten the screws.
8.a. Using the torsion wire
clamp and a spring
balance, pull the wire
with a force of
approximately 4
newtons (400 gram
mass).
b. Tighten the thumb-
screw.
5. a. Carefully lay the torsion
balance on its side.
b. Thread the wire through
the hole in the torsion
knob. It is helpful to hold
the thread against the
small allen wrench,
pushing them through
together, as shown.
allen wrench
4. Thread the wire
through the washer.
3. a. Remove the wire retainer by
unscrewing it.
b. Loosen the set screw and
thumbscrew one full turn.
c. Thread the wire in the
direction shown.
d. Retighten the set screw, but
do not tighten the thumb-
screw. The wire must be
able to turn.
Torsion wire
9. Adjust the height of
the pendulum assem-
bly, then tighten the
set screws with the
allen wrench.
2. a. Using the allen wrench,
loosen the set screw one
full turn.
b. Loosen the thumbscrew
one full turn.
c. Thread the wire in the
direction shown.
d. Tighten both screws.
6. Put the lower wire retainer
back in position, sliding the
wire through the slot in the
bottom bracket.
7. Screw the wire
retainer back in
place.
Figure 9. Replacing the Torsion Wire

012-03760E Coulomb Balance
15
Teacher’s Guide
Experiments: Parts A-C
Torsion Wire Calibration
J
J
J
J
J
J
0
0.0001
0.0002
0.0003
0.0004
0.0005
0.0006
0.0007
0.0008
0.0009
0 100 200 300 400 500 600 700
Force (N)
Angle (degrees)
f(x) = 1.448574E-6*x + 4.496489E-6
R^2 = 9.998832E-1
➤NOTE: The slope of this curve is dependent
on the tension on the wire; thus, it will be slightly
different for each unit.
DistanceDependence
5
5
5
5
555555555
0
50
100
150
200
250
300
350
400
0246810121416
Angle
Distance (cm)
f(x) = 3.994733E+3 * (x^-1.749366E+0 )
R^2 = 9.959050E-1
A power regression of this data shows that there is
an inverse-square dependence, as predicted by
theory.
ChargeDependence
ä
NOTE: There are two ways of verifying the
dependence of force on charge. You may hold
one of the spheres at a constant charge and show
that force is linear with the other charge, or you
may charge both spheres equally and show that
the force is proportional to the square of the
charge. The latter method is easier to control with
a single voltage supply, and was used for this
write-up.
JJ
J
J
J
J
J
J
J
0
20
40
60
80
100
120
0 1000 2000 3000 4000 5000 6000
Angle (degrees)
Potential (V)
f(x) = 1.335407E-5 * (x^1.836206E+0 )
R^2 = 9.954969E-1
y= 3.071653E-6*x^2+4.956344E+0;
R^2 = 9.967336E-1
The first equation given here (a power regression)
shows that the force is dependent on the square of
the charges, as predicted by the equations.
The second curve fit (a programmed least-squares
fit), when converted to SI units, gives us a value of
1.05x1010 for k. This value is 17% higher than the
accepted value of 9 x 109. We do not know the
reason for this error at the time this is being written.
If you have any explanations for this error, or
suggestions about how to improve it, please let us
know. Call PASCO Technical support at (800) 772-
8700.

Notes
16
Coulomb Balance 012-03760D
Table of contents
Other PASCO Science Education Product manuals

PASCO
PASCO //control.Node User manual

PASCO
PASCO EM-3536 User manual

PASCO
PASCO ES-9080B User manual

PASCO
PASCO Node Cart User manual

PASCO
PASCO EM-3535 User manual

PASCO
PASCO ME-7018 User manual

PASCO
PASCO ME-7001 User manual

PASCO
PASCO Basic Electrostatics System User manual

PASCO
PASCO PS-3553 User manual

PASCO
PASCO Spark SLS User manual
Popular Science Education Product manuals by other brands

KOKEN
KOKEN LM-068 instruction manual

Reality Works
Reality Works Cow Udder Model quick start guide

3D Molecular Designs
3D Molecular Designs ENZYMES IN ACTION KIT 6-Group Set manual

Opitec
Opitec 111.563 instructions

3B SCIENTIFIC PHYSICS
3B SCIENTIFIC PHYSICS U21910 instruction sheet

Limbs & Things
Limbs & Things 61141 user guide