ROHO MID PROFILE User manual

Supplier: This manual must be given to the user of this product.
Operator (Individual or Caregiver): Before using this product, read the instructions and save for future reference.
ROHO®Single Compartment Cushion
with Sensor Ready™Technology
Operation Manual
shape fitting technology
®

Table of Contents
Intended Use
Important Safety Information 2-3
Parts Detail
Product Specifications 4-5
Cushion Adjustment 6-7
Cover Removal and Replacement 8
Cleaning and Disinfecting
Troubleshooting
Disposal
Limited Warranty
9-10
Español 11
Français 21
Deutsch 31
Italiano 41
Nederlands 51
Dansk 61
Svenska 71
Norsk 81
Suomi 91
Português 101

2
ROHO Sensor Ready Cushion Operation Manual
Intended Use
The ROHO®Single Compartment Cushion with Sensor Ready™Technology (Sensor Ready
Cushion) is an adjustable, air-filled, cellular-design, wheelchair support surface that
utilizes DRY FLOATATION®technology and is intended to conform to an individual’s
seated shape to provide skin/soft tissue protection and an environment to facilitate
wound healing. There is no weight limit, yet the cushion must be properly sized to the
individual. Cushion models include:
• ROHO®MID PROFILE™Single Compartment Cushion with Sensor Ready™Technology
(MID PROFILE Sensor Ready Cushion)
• ROHO®HIGH PROFILE®Single Compartment Cushion with Sensor Ready™Technology
(HIGH PROFILE Sensor Ready Cushion)
ROHO, Inc. recommends evaluation by a clinician who is experienced in seating,
positioning and mobility: 1) to determine whether the cushion is appropriate for the
individual; and 2) to determine whether a solid seat platform is recommended if using
the cushion on a sling-seat wheelchair.
The Sensor Ready Cushion is compatible with Smart Check®by ROHO®.
ROHO medical products are intended to be part of an overall care regimen. ROHO, Inc.
recommends that a clinician make product recommendations based upon an evaluation
of the individual’s medical and therapeutic needs and overall condition.
Important Safety Information
Warnings
• Skin/soft tissue breakdown can occur due to a number of factors, which vary by
individual. Check skin frequently, at least once a day. Redness, bruising, or darker
areas (when compared to normal skin) may indicate the beginning of soft tissue
breakdown and may need to be addressed. If there is any discoloration to skin/soft
tissue, STOP USE immediately. If the discoloration does not disappear within
30 minutes after disuse, immediately consult a healthcare professional.
• The cushion and the cover MUST be compatible sizes and MUST be used as directed
in this manual. IF THEY ARE NOT: 1) The benefits provided by the cushion may be
reduced or eliminated, increasing the risk to skin or to other soft tissue, and 2) the
individual may become unstable and vulnerable to falling.
• If you are unable to perform any tasks described in this manual, seek assistance.

ROHO Sensor Ready Cushion Operation Manual
3
Important Safety Information, Continued
Warnings
• DO NOT use your cushion on top of, or in conjunction with, any other products or
materials, except as indicated in this manual. Doing so may cause the individual to
become unstable and vulnerable to falling.
• DO NOT use this cushion as a water flotation device (e.g. a Life Preserver). It will
NOT support you in water.
Cautions:
• DO NOT use a cover or a repair kit other than those provided by ROHO, Inc. Doing so
may void your product’s warranty.
• The ROHO Sensor Ready Cushion is intended to be inflated only with the ROHO Hand
Pump.
• DO NOT modify the cushion or any components. Doing so may result in damage to the
product and will void the warranty.
• DO NOT allow your cushion to come in contact with oil-based lotions or lanolin. They
may degrade the material.
• Prolonged exposure to ozone may degrade materials used in the cushion, affect the
performance of the cushion, and void the product warranty.
• Changes in altitude may require adjustment to your cushion.
• Keep the cushion away from sharp objects.
• DO NOT expose the product to high heat, open flames, or hot ashes.
• If the cushion has been in temperatures lower than 32º F (0º C) and exhibits unusual
stiffness, allow the cushion to warm to room temperature.
• DO NOT use the inflation valve, the quick disconnect, or hoses as a handle for
carrying or pulling the cushion. Carry the cushion by the cushion base or by the
carrying handle on the cover.

4
ROHO Sensor Ready Cushion Operation Manual
Parts Detail
Package contents: Sensor Ready Cushion, cover, operation manual, product registration
card, ROHO Hand Pump, repair kit
To order replacement parts, contact your product provider or ROHO, Inc., or visit therohostore.com.
Product Specifications
Cushion: neoprene rubber; made without latex
Inflation valve: nickel-plated brass; polyurethane hose
Yellow Rope: nylon
Quick Disconnect: acetal coupling for connecting the cushion to Smart Check by ROHO.
Standard Cover: two-way stretch top, spacer fabric sides, with a zipper located in the
back of the cover; and a non-skid bottom with hook and loop fasteners to help secure the
cushion in the chair. Refer to cover care label for material content.
Heavy Duty Cover: two-way stretch, fluid-resistant top and sides, with a zipper located in
the front of the cover; and a non-skid bottom with hook and loop fasteners to help secure
the cushion in the chair. Refer to cover care label for material content.
Heavy Duty Cover
Zipper Flap
Zipper
Inflation Valve
Back
Front
Quick Disconnect
ROHO Hand Pump Repair Kit
Operation Manual
Product Registration
Card
Cushion
Back
Front
Air Cells
Yellow Rope®
Inflation Valve
Quick Disconnect
Standard Cover
Quick Disconnect
Inflation Valve
Back
Front
Zipper
Carrying Handle

ROHO Sensor Ready Cushion Operation Manual
5
Product Specifications, Continued
Standard Cushion Dimensions and Weights
Cushion Height Average Weight*
MID PROFILE (MP) 3.25 in. (8.5 cm) 3.0 lb. (1.4 kg)
HIGH PROFILE (HP) 4.25 in. (10.5 cm) 3.25 lb. (1.5 kg)
*based on 9 x 9 cell cushions: 16.5 in. x 16.5 in. (42 cm x 42 cm)
Min. Width Min. Depth Max. Width Max. Depth
#Cells in.
(cm) #Cells in.
(cm) #Cells in.
(cm) #Cells in.
(cm)
MP 7 13.25
(33.5) 713.25
(33.5) 13 23.5
(59.5) 12 21.75
(55)
HP 6 11.5
(29) 611.5
(29) 15 26.75
(68) 12 21.75
(55)
NOTE: Contact Customer Care for specific cushion sizes.

6
ROHO Sensor Ready Cushion Operation Manual
Cushion Adjustment
1. Place the cushion in the
wheelchair with the air
cells facing up and the
non-skid surface of the
cover facing down.
2. Slide the rubber nozzle of
the hand pump over the
inflation valve.
3. Pinch the nozzle and open the inflation valve.
4. Inflate the cushion
with the hand pump
until the cushion
bulges.
Open
This cushion is equipped with Sensor Ready Technology and can be used with
Smart Check by ROHO, which is a cushion feedback system. Smart Check can
be used to find the individual’s inflation level and help check their cushion. For
more information, contact ROHO Customer Care (see ContaCt InformatIon.)
NOTE: If you own Smart Check by ROHO, STOP HERE and refer to the Smart Check®
by ROHO®Operation Manual.
Cushion Cover
Quick Disconnect Inflation Valve
Important Safety Information
• DO NOT use an under-inflated cushion or an over-inflated cushion because the
product benefits will be reduced or eliminated, resulting in an increased risk
to skin and other soft tissue. If the cushion does not appear to be holding air,
or if you are not able to inflate or deflate the cushion, see TroubleshooTing.
Immediately contact your clinical caregiver, distributor, or ROHO, Inc. if the
problem persists.
• When possible, use the hook or loop fasteners to further secure the cushion to
the seating surface.
• Check inflation frequently, at least once a day!

ROHO Sensor Ready Cushion Operation Manual
7
Close
5. Pinch the rubber nozzle of the hand pump and close the
inflation valve completely. Remove the hand pump.
6. The individual should sit in the chair in their usual
position. The cushion should be centered beneath
the individual.
7. Slide your hand between the cushion and the individual. Lift the leg slightly
and feel for the lowest bony prominence, and then lower the leg to a sitting
position.
8. Open the valve to remove air, keeping
your hand beneath the individual’s
lowest bony prominence. Release air
until you are barely able to move your
fingertips. Close the valve.
Cushion Adjustment, Continued
A
B
Proper Cushion Inflation Setting: A) sink down into
the cushion and B) “float” above the cushion base.
½" - 1" (1.5 cm - 2.5 cm)
recommended distance between the
individual and the seating surface.
=

8
ROHO Sensor Ready Cushion Operation Manual
Cover Removal and Placement
Inspect the cover for damage and replace if necessary.
Refer to the label on the bottom of the cushion and to the image on the cover
to correctly orient your cushion and cover.
Keeping a cover on your cushion will help to
protect the cushion components. A cover will
also help to keep the air cells together, which
may aid in slide board transfers.
To remove the cover:
1. Remove Smart Check by ROHO if it is
attached to the cushion.
2. Unzip and remove the cover from the
cushion.
To cover the cushion:
1. Turn the cover and the cushion upside down:
- The non-skid material should face up and the zipped opening should face you.
- The air cells should face down.
2. Insert the cushion into the cover, following the instructions below.
Standard Cover Instructions Heavy Duty Cover Instructions
- The back of the cushion should face the
zipper.
- Make sure all four corners of the
cushion extend through the corner
openings in the cover. Carefully pull the
inflation valve, the quick disconnect,
and the Yellow Rope through the
openings.
- Close the zipper completely. Turn the
cushion over so the air cells face up.
- The front of the cushion (the inflation
valve and the quick disconnect) should
face the zipper.
- Tuck the inflation valve, the quick
disconnect, and the Yellow Rope inside
the cover.
- Close the zipper completely. Turn the
cushion over so the air cells face up.
- When Smart Check by ROHO is attached
to the cushion, close the zipper as
completely as possible.
The cushion should lie flat inside the cover with the non-skid material on the bottom
and all cells covered.
The seated
figure shows
product
orientation.

ROHO Sensor Ready Cushion Operation Manual
9
Cleaning and Disinfecting
Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
Before use by a different individual: clean, disinfect, and check product for proper
functioning.
Cover Care: Remove hook fastener pieces from bottom of cover prior to cleaning and
disinfecting. Wash with like colors.
To Clean the Cover: Remove the cover. Machine wash in cold water (30°C)
with mild detergent, gentle cycle, or wipe clean with neutral detergent and
warm water. Tumble dry low.
To Disinfect the Cover: Hand wash in cold water
(30°C) using 1 part household liquid bleach per
9 parts water. Keep the product wet with bleach
solution for 10 minutes, and then rinse thoroughly.
Or machine wash in hot water (60°C). Tumble dry low.
Follow safety guidelines on bleach container.
Cautions:
Valve(s) must be closed.
DO NOT allow water or cleaning solution to enter the cushion.
DO NOT machine wash or sanitize the cushion or machine dry the cushion.
Use of the following may cause damage to the cushion:
• abrasives (steel wool, scouring pads)
• caustic, automatic dishwasher detergents
• cleaning products that contain petroleum or organic solvents, including acetone,
toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover
• oil-based lotions, lanolin, or phenolic-based disinfectants
• ozone gas; ultraviolet light cleaning methods
To Clean the Cushion: Remove the cover and close the inflation valve. Hand wash, using
mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a multipurpose
detergent (Follow product label instructions.) Gently scrub all cushion surfaces, using
a soft plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT
expose to direct sunlight.
To Disinfect the Cushion: Repeat cleaning instructions, using 1 part household liquid
bleach per 9 parts water. Keep the product wet with the bleach solution for 10 minutes.
Rinse thoroughly with clean water. Air dry.
Follow safety guidelines on bleach container.
Note: Most germicidal disinfectants are safe if used following the disinfectant
manufacturer’s dilution directions.
30C
30C
1:9
+or

10
ROHO Sensor Ready Cushion Operation Manual
Cleaning and Disinfecting, Continued
Note About Sterilization: High temperatures accelerate aging and will damage
the product assembly. ROHO products are not packaged in a sterile condition, nor
are they intended or required to be sterilized prior to use. If institution protocol demands
sterilization: First, follow the cleaning and disinfecting instructions; then, open the
inflation valve and use the lowest possible sterilizing temperature, but not to exceed 60°C
(158°F), for the shortest time possible. DO NOT autoclave.
Troubleshooting
For additional assistance, contact your clinical caregiver, distributor, or ROHO, Inc.
Not Holding Air Inspect the quick disconnect, the inflation valve, and hoses for
damage. Confirm that the inflation valve is completely closed
(turn clockwise). Look for holes in the cushion. If very small holes
or no holes are visible, follow the instructions in the Repair Kit
provided with your product. For inflation valve, quick disconnect,
or hose damage, or for large holes or leaks in the cushion, see
the ROHO Limited Warranty supplement, or contact Customer
Care.
Not Able To Inflate
or Deflate Cushion
Confirm that the inflation valve is open. Inspect the quick
disconnect, the inflation valve, and hoses for damage.
Uncomfortable /
Unstable
Confirm that:
- the cushion is not over-inflated (See adjustment.)
- the cushion cells and the top of the cover are facing up.
- the cushion is centered beneath the individual.
- the cushion is properly sized to the individual and the
wheelchair.
Allow the individual at least one hour to adjust to the cushion.
The Cushion Slides
on the Wheelchair
Make sure non-skid bottom of the cushion cover is facing down.
Use directional labels on the cushion and on the cover to orient
the cushion. Use hook and loop fasteners to further secure the
cushion to the wheelchair.
Disposal
The components of the products in this manual are associated with no known
environmental hazards when properly used and when disposed of in accordance with all
local/regional regulations. Incineration must be performed by a qualified, licensed waste
management facility.
Limited Warranty
Limited Warranty Term: 24 months from the date the product was originally purchased.
The warranty does not apply to punctures, tears or burns, nor does it apply to the
removable cover. See also the ROHO Limited Warranty supplement provided with your
product, or contact Customer Care.
This manual suits for next models
3
Table of contents
Other ROHO Wheelchair manuals

ROHO
ROHO AG3-1410A-1 User manual

ROHO
ROHO AirLITE User manual

ROHO
ROHO AGILITY CARBON User manual

ROHO
ROHO AirLITE User manual

ROHO
ROHO Hybrid Elite User manual

ROHO
ROHO Cushion Guide

ROHO
ROHO MOSAIC CUSHION User manual

ROHO
ROHO JetStream Pro User manual

ROHO
ROHO HARMONY User manual

ROHO
ROHO AirLITE User manual
Popular Wheelchair manuals by other brands

Sunrise Medical
Sunrise Medical Guardian Escort User instruction manual & warranty

Küschall
Küschall K-SERIES Service manual

Pride Mobility
Pride Mobility Quantum Jazzy series owner's manual

Nova
Nova 9036-R user guide

Invacare
Invacare TDX SP2 series user manual

Motion Concepts
Motion Concepts M270 user manual