Always make sure the weigh cell
is disconnected from AC power
before performing any installa-
tion, cleaning, maintenance or repair work.
If the equipment housing is opened by
anyone other than persons authorized by
Sartorius, all claims under the manufactur-
er’s warranty are forfeited. Use only origi-
nal Sartorius spare parts.
Disconnecting equipotential
bonding conductors is not per-
mitted.
If you see any indication that
the weigh cell cannot be oper-
ated safely (for example, due to
damage), turn off the unit and lock it in a
secure place so that it cannot be used for
the time being. Observe the relevant safety
precautions and inform personnel as
required.
The casing on all connecting
cables, as well as the casing on
wires inside the equipment
housing, is made of PVC. Chemicals that
corrode these materials must be kept away
from these cables.
Make sure the weighing instru-
ment is not exposed to sub-
stances that release chlorine ions
at the place of use. If such exposure
cannot be ruled out, the operator is
responsible for establishing and observing
appropriate safety precautions, to
be checked at regular intervals for contin-
ued effectiveness.
The weigh cell may be used and
operated by qualified personnel
only. The permissible use of the
equipment is defined by the equipment
specifications and the relevant safety regu-
lations. Operating the weigh cell beyond
the specifications listed in the type-
approval certificate is not permitted, and
is considered use of the equipment for
other than its intended purpose. All restric-
tions listed in the type-examination certifi-
cate must be strictly observed.
t Once the weigh cell has been installed,
it must be checked to ensure:
−Observance of the guidelines and stand-
ards for using electrical equipment
−Electromagnetic compatibility of the
entire device
−Compliance with mandatory safety reg-
ulations.
−Installation in a Zone 1, 2, 21 or 22
hazardous area or Class 1, Division 1 or
2 hazardous location must be per-
formed by a trained technician who is
familiar with the assembly and opera-
tion of the equipment, as well as with
the procedure for putting the system
into operation. Furthermore, the trained
technician must have the required qual-
ifications and must be familiar with the
relevant guidelines and regulations.
If you need assistance, contact the
system installer, your Sartorius dealer or
the Sartorius Service Center.
Any installation work that does not
conform to the instructions in this
manual will result in forfeiture of
all claims under the manufacturer’s
warranty. Be sure to observe all restric-
tions listed in the type-examination
certificate. Operating the weigh cell
beyond the limits imposed by these
restrictions is not permitted and is con-
sidered use of the equipment for other
than its intended purpose.
Always make sure the weigh cell
is disconnected from AC power
before performing any installa-
tion, cleaning, maintenance or repair work.
The electronics module and the
optional YAC01NX display and
control unit may not be opened.
The weigh cell may be opened only by
authorized service technicians who have
been trained by Sartorius and who follow
Sartorius’ standard operating procedures
for maintenance and repair work.
Avoid interchanging weigh cells
and electronics modules.
Connect the parts that fit
together, ensuring that the serial numbers
match.
Do not expose the weigh cell
unnecessarily to extreme tem-
peratures, aggressive chemical
vapors, moisture, temperature fluctuations
or vibrations. The permissible operating
temperature range during operation is
+5°C to +40°C.
Be sure to unplug the system
from AC power before connect-
ing or disconnecting any cables.
If you are using the optional
YAC01NX display and control
unit, it must be installed a way
that limits the risk of mechanical damage.
Never open the display and
control unit.
If there is visible damage to the
components: disconnect the
equipment from AC power and
replace the weigh cell along with the elec-
tronics module.
If you use cables purchased from
another manufacturer in the
non-hazardous area/location,
check the pin assignments. Before con-
necting the cable to Sartorius equipment,
check the pin assignments in the cable
against those specified by Sartorius and
disconnect any wires that are assigned
differently. The operator shall be solely
responsible for any damage or injuries that
occur when using cables not supplied by
Sartorius. On request, Sartorius will provide
information on the minimum operating
specifications (in accordance with the
Standards for defined immunity to inter-
ference).
When using the weigh cell in
hazardous areas/locations, make
sure there is no current or volt-
age in the equipment before connecting or
disconnecting current-carrying cables to or
from the weigh cell. Disconnect the weigh
cell from AC power before connecting or
disconnecting cables.
Avoid exposing the weigh cell to static
electricity.
All components of the system
must be grounded, including
any draft shields used. Be sure
to connect an equipotential bonding
conductor.
−If any problems arise with the weigh
cell that require service, please contact
your local Sartorius dealer.
Hotline
§If you have technical questions about
the construction, specifications or
installation of this weigh cell, just call
or fax your local Sartorius office.
For the address, please visit our Internet
website at: www.sartorius. com