Scale House ATM Series User manual

MULTIFUNCTION SCALE
MULTIFUNKTIONSWAAGE
BALANCE MULTIFONCTION
BILANCIA MULTIFUNZIONE
ATM-ATMI
USER MANUAL
BENUTZERHANDBUCH
MODE D’EMPLOI OPERATEUR
MANUALE UTENTE
ATM(1.009)-ATMI(1.001)_19.02_EN_DE_FR_IT

2
LANGUAGES
SPRACHEN
LANGUES
LINGUE
ENGLISH .................................................................................................................3
DEUTSCH..............................................................................................................18
FRANÇAIS.............................................................................................................33
ITALIANO...............................................................................................................48

3
ENGLISH
INDEX
1. INTRODUCTION..................................................................................................................................................................4
2. MAIN TECHNICAL SPECIFICATIONS................................................................................................................................4
3. INSTALLATION ...................................................................................................................................................................5
3.1 POWER SUPPLY AND START-UP .............................................................................................................................8
3.2 BATTERY POWERED ................................................................................................................................................8
4. FRONT PANEL KEYS AND INDICATORS .........................................................................................................................8
4.1 FUNCTION OF THE KEYS.........................................................................................................................................9
4.2 FUNCTION OF THE INDICATORS.............................................................................................................................9
5. SYMBOLS ON LCD DISPLAY.............................................................................................................................................9
6. OPERATION......................................................................................................................................................................10
6.1 AUTOMATIC ZERO UPON START-UP.....................................................................................................................10
6.2 SCALE ZERO FUNCTION........................................................................................................................................10
6.3 TARE FUNCTION .....................................................................................................................................................10
6.3.1 LIMITATION OF THE TARE FUNCTIONS........................................................................................................................ 10
6.4 OVERLOAD MESSAGE............................................................................................................................................10
6.5 COUNTING PIECES.................................................................................................................................................10
6.6 CHECKING QUANTITIES.........................................................................................................................................11
6.6.1 SETTING THE WEIGHT THRESHOLDS.......................................................................................................11
6.6.2 VISUAL INDICATION OF THE CHECK.........................................................................................................11
6.7 CHANGE OF THE UNIT OF MEASURE...................................................................................................................11
6.8 AUTO SWITCH-OFF FUNCTION ...............................................................................................................................11
6.9 VISUALISATION WITH SENSITIVITY TIMES 10 .......................................................................................................12
7. SET-UP ENVIRONMENT...................................................................................................................................................12
7.1 SET-UP ENVIRONMENT BLOCK DIAGRAM.............................................................................................................12
8. SEALING ACCESS TO THE BALANCE SETTINGS .......................................................................................................................... 15
9. ERROR MESSAGES .........................................................................................................................................................16
RECYCLING INSTRUCTION.................................................................................................................................................16
DECLARATION OF CONFORMITY ......................................................................................................................................17
WARRANTY ..........................................................................................................................................................................17

4
1. INTRODUCTION
This manual was created to help you install and learn all about the functional possibilities of the ATM and ATMI scales.
WARNINGS
The scales should be sited in a location that will not degrade the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The tables or floor must be rigid and not vibrate. Do not place near vibrating machinery.
Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.
Avoid air movement such as from fans or opening doors. Do not place near open windows.
Keep the scales clean.
Do not stack material on the scales when they are not in use.
FOR ATM MODEL: Avoid high humidity that might cause condensation. Avoid direct contact with water. Do not spray or
immerse the scales in water.
2. MAIN TECHNICAL SPECIFICATIONS
Model
Plate
dimensions
mm
Capacity
kg
Standard
Division
g
CE-M
Division
g
ATM6 / ATMI6
190x230
3/6
1/2
1/2
ATM15 / ATMI15
190x230
6/15
2/5
2/5
ATM25
190x230
15/25
5/10
5/10
Stabilisation Time
2 Seconds typical.
Operative Temperature:
0°C - 40°C / 32°F - 104°F.
Power supply:
with fitted 240 Vac mains power adapter or with fitted battery
Battery life:
up to 70 hours of continuous use.
Recharge time:
12 hours.
Parameter Setup:
Calibration fully digital and programmable from the keyboard.
Display :
LCD backlit with 5 digits 24 mm high.
Case:
ATM: ABS Plastic.
ATMI: IP65 stainless steel.
Zero range:
+/- 2% of max load capacity.
Auto zero at start-up:
+/- 10% of max load capacity.

5
3. INSTALLATION
a) Unpack then product.
b) If present (depending on the model), it is necessary to unscrew and remove the locking screw for the transportation
placed under the scale, as shown in the figure below:
SCREW FOR WASHING THE SCALE (Only on ATMI)
Placed under the ATMI scale, there is a knob, which serve to make the scale sealed IP67 (knob closed) when needed
wash, or IP65 (knob open) when one need to weigh.(See drawings):
NB.: The knob closed may have effects on the weighing, so it must be left open when the scale is used for weighing.

6
c) Level the platform by adjusting the adjustable feet. The stability of the platform is very important.
ALL THE CORNERS MUST REST IN A UNIFORM WAY. Carefully check that all feet fully rest on the ground and that the
platform, loaded on the corner, is not unstable (if a corner is not resting on the ground its relative foot is easier to turn).

7
ATMI
1
Weighing plate
2
Locking screw
3
Carrier weighing plate
4
Display
5
Level bubbole
6
Keyboard
7
Footscrews
8
Battery cover
9
Screws for battery cover
1
2
3
4
5
7
8
6
9

8
3.1 POWER SUPPLY AND START-UP
The indicator can be powered with a fitted 240 Vac external power adapter or just with the fitted battery.
NOTE: it is advisable to completely recharge the battery (12 hours) with the first installation of the instrument.
TO POWER the instrument through the 240 Vac mains put the AC/DC power plug in the appropriate socket under the
scale and the power adapter in the 240 Vac mains socket.
TO TURN ON/TURN OFF the instrument press the ON/OFF key.
After the self-check the scale will show “hi rES” (in case of NOT approved instrument) or “LEGAL” (in case of approved
instrument).
3.2 BATTERY POWERED
The weighing indicator can be operated from the battery if desired.
When the battery needs charging the “LOW BAT”message on the weight display will appear. The battery should be
charged when the message appears. The scale will still operate for about 10 minutes after which it will automatically switch
off to protect the battery.
To charge the battery simply plug into the mains power. The scale does not need to be turned on.
The battery should be charged for 12 hours for full capacity.
You can find the “Charge”LED at the top of display to indicate the status of battery charging. When the scale is plugged
into the mains power the internal battery will be charged. If the LED is green the battery has a full charge. If it is Red
the battery is nearly discharged and yellow indicates the battery is being charged.
As the battery is used it may fail to hold a full charge. If the battery life becomes unacceptable then contact your distributor
4. FRONT PANEL KEYS AND INDICATORS

9
4.1 FUNCTION OF THE KEYS
- It turns on or turns off the scale
- Clears the displayed gross weight if it’s within +/- 2% of the total capacity.
- If pressed at length it allows to execute the reference operation (see section 6.5)
- A secondary function of incrementing the active digit or selecting when setting a
value for parameters or other functions.
- If pressed for an instant it executes the semiautomatic tare;
- If pressed at length it allows to convert from kg to g (see section 6.7).
-A secondary function of "Enter" key when setting parameters or other functions.
- Pressed together for an instant they allow to set the thresholds for the QUANTITY
CHECK function (see section 6.6).
4.2 FUNCTION OF THE INDICATORS
INDICATOR
FUNCTION
CHARGE
Indicates that the battery is going to recharge (see chapter 3.2).
5. SYMBOLS ON THE LCD DISPLAY
The LCD display has symbols which show the indicator’s functioning status; you will find the description for each symbol
below.
SYMBOL
FUNCTION
0
Indicates that the weight detected on the weighing system is near zero.
~
Indicates that the weight is not stable.
NET
Indicates that the displayed weight is a net weight.
PCS
Indicates that the number of sampled pieces is being displayed (see section 6.5).
kg
Indicates the unit of measure in use (kilogram).
g
Indicates the unit of measure in use (gram).

10
6. OPERATION
6.1 AUTOMATIC ZERO UPON START-UP
The indicator has an “auto zero at start-up” function: in other words it means that if at start-up a weight within +/- 10%
of the capacity is detected, it will be zeroed; if the weight is not within this tolerance, the display shows “Err 4” and with
a non approved instrument the present weight after a few instants, while with an approved instrument “Err 4” is shown
continuously.
6.2 SCALE ZERO FUNCTION
You can press the ZERO key at any time to set the zero point from which all other weighing and counting is measured,
within +/- 2% of power up zero. This will usually only be necessary when the platform is empty. When the zero point is
obtained the display will show the indicator for zero.
The scale has an automatic rezeroing function to account for minor drifting or accumulation of material on the platform
(see step “F7 AZn”, section 7.1). However you may need to press the ZERO key to rezero the scale if small amounts
of weight are shown when the platform is empty.
6.3 TARE FUNCTION
Zero the scale by pressing the ZERO key if necessary. The zero symbol will be on.
Place a container on the platform, a value for its weight will be displayed.
Press the TARE key to tare the scale when the weight reading is stable. The weight that was displayed is stored as
the tare value and that value is subtracted from the display, leaving zero on the display. The "NET" indicator will be on.
As product is added only the weight of the product will be shown. The scale could be tared a second time if another
type of product was to be added to the first one. Again only the weight that is added after taring will be displayed.
When the container is removed a negative value will be shown.
To remove the tare, unload the scale and press again TARE.
6.3.1 LIMITATION OF THE TARE FUNCTIONS
It is possible to limit the tare functions, selecting "oFF" in the F8 step of the set-up environment (see section 7)
The SEMI AUTOMATIC TARE could be made only if the tare is equal to 0.
NOTE: With approved instrument, the F8 step is not settable.
6.4 OVERLOAD MESSAGE
Please do not add item that is over the maximum capacity. When reading “- - - - - -“ and hear beeping sound, remove
the item on the platter to avoid damage to the load cell.
6.5 COUNTING PIECES
Through the reference function it’s possible to use the scale as a piece counter.
COUNTING PROCEDURE
The operations to be carried out are the following:
1) Place the empty container on the scale and press TARE to tare it.
2) Check that the zero is on the display and put the quantity of pieces chosen for the SAMPLE on the scale and
press the ZERO key at length.
3) The display suggests a REFERENCE QUANTITY. The possible options are: 10, 20, 50, 100, and 200.
4) Press ZERO the number of times needed to reach the desired sample size.
5) Press TARE; the display will indicate "- - - - - -" and the indicator will calculate the Average Piece Weight (APW).
After a few instants the display will indicate the quantity selected put on the platform.
6) Add the rest of the items to count in the container and whose value will appear on the display.
7) Unload the scale, the APW will remain stored in memory for the next counting of similar pieces, without having to
repeat the REFERENCE operation.
8) Per cancel or to carry out a new reference operation, press the ZERO key at length and repeat the operations
as describe in point 1).

11
6.6 CHECKING QUANTITIES
The scale is fitted with the quantity check function with 2 thresholds. Messages on the display are enabled on the
basis of the set thresholds and the weighing result.
The check can be made on the weight or on the pieces.
To check the pieces one should execute the reference operation (see section 6.5) and then configure the thresholds.
6.6.1 SETTING OF THE WEIGHT THRESHOLDS
Press TARE and ZERO together:
-The display shows “LOW” and after “000.000”, with the first blinking digit. In this step one configures the lower
threshold.
-Use the TARE key to increase the value of the digit and the ZERO key to go on the following digit.
-Confirm the entered value with ZERO.
-The display shows “HIGH” and after “000.000”, with the first digit blinking. In this step one configures the upper
threshold.
-Use the TARE key to increase the value of the digit and the ZERO key to go on the following digit.
-Confirm the entered value with ZERO.
-The thresholds are configured and the indicator returns to weighing.
6.6.2 VISUAL INDICATION OF THE WEIGHT CHECK
Put a weight on the plate of the scale.
The indication message will enable depending on which thresholds have been configured:
WITH BOTH THRESHOLDS CONFIGURED
-The HI message of the LCD display will appear when the detected weight exceeds the UPPER THRESHOLD.
-The OK message of the LCD display will appear when the detected weight is between the LOWER THRESHOLD
and the UPPER WEIGHT THRESHOLD.
-The LO message of the LCD display will appear when the detected weight is less than the LOWER
THRESHOLD.
WITH THE UPPER THRESHOLD CONFIGURED AND THE LOWER THRESHOLD = 0
-The OK message of the LCD display will appear when the detected weight exceeds the UPPER THRESHOLD.
-The LO message of the LCD display will appear when the detected weight is less than the UPPER THRESHOLD.
WITH THE UPPER THRESHOLD = 0 AND THE LOWER THRESHOLD CONFIGURED
-The HI message of the LCD display will appear when the detected weight exceeds the LOWER THRESHOLD.
-The OK message of the LCD display will appear when the detected weight is less than the LOWER
THRESHOLD.
NOTES
-The weight check function is enabled when the limit values are entered. To disable the function one should press
again TARE and ZERO key together.
-The function is not valid when one enters a lower limit greater than the upper limit.
-The weight check is enabled with a weight greater than the minimum weight (20d).
6.7 UNIT OF MEASURE CONVERSION
By pressing at length the TARE key the conversion from kg to g takes place; the relative symbol will turn on in the
display (see section 5).
NOTE:
- The function is not available if the counting function is enabled.
- With APPROVED instrument the weight in secondary units is displayed for 5 seconds, after which the display goes
to the scale unit of measure.
6.8 FUNCTION OF AUTOMATIC TURN-OFF
The scale is fitted of a programmable automatic turn-off function, which allows to save energy in case of not using it
temporarily; the automatic turn-off starts functioning when the scale is unloaded (gross weight = 0) for the time (in
minutes) programmed (see step “F4 oFF”, section 7.1).

12
The values vary among:
00 DISABLED
03 3 minutes
05 5 minutes
15 15 minutes
30 30 minutes
6.9 VISUALISATION WITH SENSITIVITY TIMES 10
It’s possible to use this function as a test in the calibration phase.
Keep the TARE and ZERO keys pressed for three seconds and release: the display will show the weight with the
minimum division times 10 (a blinking digit is added).
After about 5 seconds the display will return in the standard mode.
7. SET-UP ENVIRONMENT
The scale has various parameters that can be configured in the SET-UP environment.
Enter in the SET-UP environment by keeping pressed TARE key at start –up until display shows the first step “F0 CAL”.
- Pressing ZERO will cycle through the other functions.
- Pressing TARE will allow you to get into the function.
- Select a value by pressing the ZERO key or
- Set a numeric value, using the TARE key to increase the value of the digit, the ZERO key to go on the following digit and
confirm on last digit with the ZERO key.
- Press the TARE key to confirm the modification and pass to the following step.
- To exit from the set-up, press the ON/OFF key.
7.1 SET-UP ENVIRONMENT BLOCK DIAGRAM
F0 CAL
F2 inP
F3 bK
F1 inC
F4 oFF
F5 CAP
F7 SPEED 7.5, 15, 30
F8 rSt
SETUP
ENVIRONMENT
r 3000, r 6000,
r dual
bk of, bk on,
bk AU
0, 3, 5, 10, 20
3, 6, 15, 25
F9 GrAV 9,XXXXX
F6 rE ZEro 0.5d, 1d,
2d, 4d
Pn---
In the parameter description:
- The METRIC parameters are shown with the (*) symbol, and, with approved instrument, these are read only.

13
F0 CAL CALIBRATION (*)
Premise: The unit of measure of the calibration is fixed at kilogram (“kg”).
-The displays show “F0 CAL”: press the TARE key to get into the step.
-The displays show the acceleration value set in the F9 step: enter the gravitational acceleration value of the calibration
zone if different from the use zone, by pressing the TARE key to increase the value of the digit and the ZERO key to go
on the following digit; press the ZERO key on last digit to confirm; one must modify all the 5 decimal digits of the
gravitational acceleration. The value will be set also in the F9 step.
-The display shows “UnLd”;
-Empty the platter, after instability indicator off, press TARE key to sure;
-The display shows “AX Y”; in which X is the number and Y the value of the sample weight to be used;
-Select the sample weight to be used by pressing the ZERO key: the indicator scroll values defined in the table below;
press TARE to confirm;
-Put the sample weight on the scale and press TARE when the instability indicator is off
-The indicator returns to weighing.
CALIBRATION WEIGHTS
MODEL
ATM6 / ATMI6
ATM15 / ATMI15
ATM25
A0
2 kg
5 kg
10 kg
A1
4 kg
10 kg
20 kg
A2
6 kg
15 kg
25 kg
IN CASE THE ZONE OF USE IS DIFFERENT FROM THE CALIBRATION ZONE ONE SHOULD:
1) Carry out the calibration as previously described.
2) Enter the F9 step and set the gravity acceleration value for the ZONE OF USER, by pressing the TARE key to
increase the value of the digit and the ZERO key to go on the following digit; press the ZERO key on last digit to
confirm; one must modify the 5 decimal digits of the gravitational acceleration.
3) Save and exit the SET-UP environment (press ON/OFF key).
4) The weight error, caused by the different gravitational attraction value between the calibration zone and the zone
of use, is corrected automatically.
F1 inC SCALE DIVISIONS (*)
Once entered in the step with the TARE key it is possible to set the scale total divisions.
By pressing ZERO one can select the suggested values:
Value
Divisions
ATM
3000
3000
6000
6000
dUAL
3000 + 3000
ATMI
3000
3000
6000
6000
dUAL1
3000 + 3000
30000
30000
60000
60000
dUAL2
30000 + 30000
F2 inP VISUALISATION OF A/D CONVERTER POINTS
By pressing TARE the instrument shows the A/D converter points relative to the weight on the scale.
Press TARE to exit the step.
F3 bK DISPLAY BACKLIGHTING
Once entered in the step with the TARE key it is possible to set the backlight functioning mode.
By pressing ZERO one can select the suggested values:
EL oFF backlighting always off
EL on backlighting always on.

14
EL AU backlighting on only when the weight is greater than 20d or when a key is pressed. When a key is pressed, if the
weight is less than 20d, the backlighting remains active for 5 seconds.

15
F4 oFF AUTO SWITCH-OFF
Once entered in the step with the TARE key it is possible to set the minutes of not being used for the auto switch-off (see
section 6.5); by pressing ZERO one can select the suggested values:
oF 0 (disabled)
oF 3 (3 minutes of not being used)
oF 5 (5 minutes of not being used)
oF 10 (10 minutes of not being used)
oF 20 (20 minutes of not being used)
F5 CAP SCALE CAPACITY (*)
By pressing TARE key the display shows “p n - - -” and wait for password input:
- Enter “315”, by pressing the TARE key to increase the value of the digit and the ZERO key to go on the following digit;
- Press the ZERO key on last digit to confirm.
After input correct password, it will be possible to set the capacity by pressing TARE key:
33kg
66kg
15 15kg
25 25kg
Press the ZERO key to confirm.
F6 AZn ZERO TRACKING (*)
Once entered in the step with the TARE key it is possible to set the divisions' number for the zero tracking, in other words,
scale’s thermal drift compensation parameter; the set value corresponds to the number of divisions which is cleared.
By pressing the ZERO key one can select the suggested divisions: 0.5d, 1d, 2d, 4d.
F7 SPd SETTING ADC CONVERTER READING SPEED
Once entered in the step with the TARE key it is possible to set the reading speed of the ADC converter; the set value
corresponds to the number of readings made in a second.
By pressing the ZERO key one can select the suggested reading speeds: 7.5, 15, 30.
F8 TArE LIMITATION OF THE TARE FUNCTIONS (*)
on limitation disabled
oFF limitation enabled
Refers to the 6.3.1 section for the functioning mode.
F9 Grav GRAV (*)
Through this step one selects the acceleration value of use of the instrument:
Manual entry of the g value: one may manually enter the gravitational acceleration value, by pressing the TARE key to
increase the value of the digit and the ZERO key to go on the following digit; press the ZERO key on last digit to confirm;
one must modify the 5 decimal digits of the gravitational acceleration.
In case one enters a wrong g value: the minimum decimal value is suggested (9,75001); a wrong value is any decimal
number that is not between 9,75001 and 9,84999 (inclusive).
F11 rSt RESET (*)
For use of the manufacturer.

16
8. SEALING ACCESS TO THE BALANCE SETTINGS
For legal for trade applications, the balance must be sealed to prevent access to the metrological parameters.
To seal the balance, press the recessed Lock switch at the bottom of the balance, momentarily during power up (the
message “LEGAL” appears an instant), then cover the Lock switch access hole. A cover seal and paper seal may be
installed as shown.
To regain access to the locked balance settings, break the seal and press the recessed Lock switch
momentarily during power up (the message “HIRES” appears an instant).
ATM SEALING

17
ATMI SEALING
9. ERROR MESSAGES
ERROR MESSAGES
DESCRIPTION
RESOLUTION
- - - - - -
Over range
Remove weight from the scale.
Err 4
Zero Setting
Error
Upon start-up or when the ZERO key is pressed, the weight on the scale
exceeds the percentage programmed on the max. capacity.
Remove the excess weight and retry.
Err 5
Keyboard Error.
The keyboard could be damaged.
Err 6
Input signal
greater than
3mV/V
Indicates that the A/D converter value is greater than the maximum value:
- Remove the weight from the scale if it’s exceeding.
- The load cell or the electronics could be damaged.
RECYCLING INSTRUCTION
The crossed-out wheeled bin on the product means that at the product end of life, it must be taken to
separate collection or to the reseller when a new equivalent type of equipment is purchased. The
adequate differentiated refuse collection in having the product recycled, helps to avoid possible
negative effects on the environment and health and supports the recycling of the materials of which
the equipment is made. The unlawful disposal of the product by the user will entail fines foreseen by
the current regulations.

18
DECLARATION OF CONFORMITY
This device conforms to the essential standards and norms relative to the
applicable European regulations. The Declaration of conformity is available in the
web site www.scalehouse.it.
WARRANTY
Scale House products are guaranteed for a period of twelve months from delivery,
excluding the parts classified as expendable materials such as mechanical printing
heads, batteries, electric motors and wheels. The warranty for these expendable
materials is three months. The warranty refers to breakdowns resulting from any
construction defect or material defect of the product supplied and covers the cost
of labor and spare parts. The product must be returned to the Seller address in its
original packaging with shipping paid by the sender. The warranty does not apply
to breakdowns due to improper use or non-observance of the operating
instructions, electrical phenomenon, unauthorized repair attempt, connection to
other equipment or removal of any product identification elements (serial number,
label, etc.). This warranty does not provide for any compensation for damages,
direct or indirect, incurred by the user due to complete or partial failure of
instruments, even during the warranty period. The warranty for the load cells
excludes the damages caused by knocks and overloads.
AUTHORISED SERVICE CENTRE STAMP

19
DEUTSCH
INHALT
1. EINLEITUNG....................................................................................................................................................................19
2. TECHNISCHE SPEZIFIKATIONEN.................................................................................................................................19
3. INSTALLATION ...............................................................................................................................................................20
3.1 SPANNUNGSVERSORGUNG UND INBETRIEBNAHME ......................................................................................23
3.2 BATTERIEBETRIEB................................................................................................................................................23
4. FRONTPANEEL TASTEN UND ANZEIGEN ...................................................................................................................23
4.1 TASTENFUNKTIONEN...........................................................................................................................................24
4.2 ANZEIGENFUNKTIONEN.......................................................................................................................................24
5. SYMBOLE DES LCD DISPLAYS ....................................................................................................................................24
6. FUNKTIONEN..................................................................................................................................................................25
6.1 AUTOMATISCHE NULLUNG BEIM START............................................................................................................25
6.2 WAAGE NULLEN....................................................................................................................................................25
6.3 TARAFUNKTION ....................................................................................................................................................25
6.3.1 BESCHRÄNKUNG DER TARAFUNKTIONEN...............................................................................................25
6.4 ÜBERLAST-ANZEIGE ............................................................................................................................................25
6.5 STÜCKZÄHLUNG...................................................................................................................................................25
6.6 KONTROLLE DES GEWICHTS ODER DER STÜCKANZAHL...............................................................................26
6.6.1 EINSTELLEN DER SCHWELLENWERTE.....................................................................................................26
6.6.2 VISUELLE ANZEIGEN BEIM GEWICHTS-CHECK .......................................................................................26
6.7 AUSWAHL DER GEWICHTSEINHEIT ..................................................................................................................26
6.8 AUTOMATISCHE ABSCHALTUNG.........................................................................................................................27
6.9 ANZEIGE MIT „EMPFINDLICHKEIT X 10“..............................................................................................................27
7. SETUP UMGEBUNG .......................................................................................................................................................27
7.1 SETUP BLOCKSCHALTBILD .................................................................................................................................27
8. ZUGRIFF AUF DIE WAAGENEINSTELLUNGEN VERSIEGELN....................................................................................30
9. FEHLERMELDUNGEN....................................................................................................................................................31
RECYCLINGHINWEIS.........................................................................................................................................................31
KONFORMITÄTSERKLÄRUNG..........................................................................................................................................32
GARANTIE...........................................................................................................................................................................32

20
1. EINLEITUNG
Diese Beschreibung erklärt Ihnen den Einsatz und die Funktionen der ATM / ATMI Waage.
HINWEISE
Die Waage sollte an einem Ort eingesetzt werden, welcher die Genauigkeit nicht beeinträchtigt.
Extreme Temperaturen vermeiden. Nicht in direktem Sonnenlicht oder bei Ventilatoren betreiben.
Achten Sie auf stabile eine Auflage. Tisch oder Boden muss eben und frei von Vibrationen sein. Nicht in der Nähe von
vibrierenden Maschinen betreiben.
Achten Sie auf ein sauberes Netz. Nicht an der gleichen Steckdose mit großen Geräten oder Motoren anschließen.
Vermeiden Sie Zugluft von offenen Fenstern oder Ventilatoren. Nicht in der Nähe von Türen platzieren.
Halten Sie die Waage sauber.
Stapeln Sie keine Waren auf die unbenutzte Waage.
ATM MODELL: Nicht bei hoher Luftfeuchtigkeit betreiben und direkten Kontakt mit Wasser vermeiden. Nicht in Wasser
eintauchen oder ansprühen.
2. TECHNISCHE HAUPTEIGENSCHAFTEN
Version
Schale
(Abmessungen
mm)
Wägebereich
kg
Ziffernschritt [g]
CE-M
Ziffernschritt
g
ATM6 / ATMI6
190x230
3/6
1/2
1/2
ATM15 / ATMI15
190x230
6/15
2/5
2/5
ATM25
190x230
15/25
5/10
5/10
Stabilisierungszeit:
2 Sekunden.
Umgebungstemperatur:
0°C - 40°C / 32°F - 104°F.
Stromversorgung:
Mit 240 V~ Netzteil oder mit interner Batterie
Betriebszeit Batterie:
Bis zu 70 Stunden ununterbrochene Anwendung.
Ladezeit:
12 Stunden.
Parameter Setup:
Digitale Kalibrierung und programmierbar über Tastatur.
Display :
Hintergrundbeleuchtetes LCD-Display mit 5 Ziffern mit 25 mm Höhe.
Gehäuse:
ATM: ABS Plastic.
ATMI: IP65 Edelstahlgehäuse.
Nullstellbereich:
+/- 2 % des Max. Wägebereich.
Nullstellung beim Start:
+/- 10 % des Max. Wägebereich.
This manual suits for next models
6
Table of contents
Languages:
Other Scale House Scale manuals

Scale House
Scale House KS Series User manual

Scale House
Scale House MB1 User manual

Scale House
Scale House FPW Series User manual

Scale House
Scale House DHS Series User manual

Scale House
Scale House NCS3K User manual

Scale House
Scale House OCSF-50 User manual

Scale House
Scale House ASB6 User manual

Scale House
Scale House GAI420 User manual

Scale House
Scale House PSL Series User manual