
12 December 2012
1. Press the ACCESS key to enter the Calibrate mode and wait for the MC-I Portable Rate Meter to
complete a diagnostic test.
2. When the prompt tot.Eng appears in the upper display, press INCREMENT to scroll through the
available volume units of measurement (bbl, gal, m³, liter). Select “gal” from the unit menu. Press
ENTER/STEP to accept the selection.
3. When Pu.P.gAL appears in the upper display, enter the calculated divisor of 566.79 according to
the steps below. The lower display will show the previously entered meter factor. The factory
default is 900.00. The right-most digit, the hundredths position, will blink, indicating it is the
digit currently selected for editing. All digits must be entered from right to left.
a. Press DEC-POINT till the decimal point is in the hundreds position.
b. Since 9 is the first digit to be entered (in the hundredths position for the factor 566.79), press
INCREMENT until 9 is displayed.
Press ENTER/STEP to accept the selection of the digit 9. The digit to the left of the 9 (the
tenths position) will begin blinking.
c. Since 7 is to be entered in the tenths position (for the factor 566.79), press INCREMENT
until 7 is displayed.
Press ENTER/STEP to accept the selection of digit 7. The digit to the left of the 7 (the ones
position) will begin blinking.
d. Press INCREMENT until 6 is displayed. Press ENTER/STEP to accept the selection and
proceed to the tens position.
e. Press INCREMENT until 6 is displayed. Press ENTER/STEP to proceed to the hundreds
position.
f. Press INCREMENT until 5 is displayed. Press ENTER/STEP to accept the selection.
g. Press INCREMENT until 0 is displayed. Press ENTER/STEP to accept the selection.
4. The upper display will show rAt.Eng, prompting you to enter the units of measure for the flow
rate. Pressing INCREMENT allows you to scroll through the engineering flow rate units:
• barrels per day (bpd) - factory default
• gallons per minute (g.P.nn)
• cubic meters per day (nn3.P.d)
• liters per minute (Lit.P.nn)
• <volume unit> per day (PEr.dAY)
• <volume unit> per hour (PEr.Hr)
• <volume unit> per minute (Pr.nn in)
• <volume unit> per second (PEr.SEC)
5. Press INCREMENT until PEr.dAY is shown on the bottom display. Press ENTER/STEP.
The top display will be empty, and the bottom display will show SAVing momentarily while the
MC-I Portable Rate Meter saves the calibration settings to nonvolatile memory. The MC-I
Portable Rate Meter will automatically return to the Run mode.
6. In Run mode, press the DEC. POINT key to change the position of the decimal point. Select the
decimal point for the resolution desired.