SICK CLV 41 Series User manual

OPERATING INSTRUCTIONS
CLV 41x Bar Code Scanner
Standard line

© SICK AG • Auto Ident • Germany • All rights reserved
2
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
These Operating Instructions describe the procedure for installing and setting up the
CLV 1x using the default settings defined at the factory and the steps required to
replace the device (transferring the application-specific parameter set). The CLV 1x
is simply referred to as the CLV in the sections below.
General
Software version
Software Function Version
CLV Setup User interface (Windows-based) from V 1.1.0
CLV Setup Help Online help (HTML) from V 1.0.1
FCC class
This appliance meets FCC class B, subject to the FCC-regulations, paragraph 15.
Remark: To meet the FCC-regulations, the cables have to shielded including
connections from shielding to the connector casings.

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 3
Operating Instructions
8 008 22 /I982/08-00
Section Page
1. Intended use................................................................................................................... 4
2. Sco e of delivery .......................................................................................................... 4
3. System requirements .................................................................................................. 4
4. Laser rotections ......................................................................................................... 4
5. Design ............................................................................................................................... 5
6. Bee er .............................................................................................................................. 5
7. Function of the LEDs ................................................................................................... 6
8. Electrical installation
8.1 Pin assignment of connector ...................................................................................... 6
8.2 Supply voltage .................................................................................................................. 7
8.3 Connecting the CLV to the AMV/AMS 0 .............................................................. 7
8. Connecting the PC to the CLV .................................................................................... 8
8.5 Connecting the interfaces ............................................................................................ 9
9. Assembly
9.1 Securing the CLV .......................................................................................................... 13
9.2 Mounting the polling reflector ................................................................................. 15
9.3 Mounting the external reading pulse sensor (alternative) ............................ 15
10. Power on and setu
10.1 Step-by-step setup instructions ............................................................................ 17
10.2 Default settings ............................................................................................................ 18
10.3 Profile programming ................................................................................................... 19
10. AutoSetup (code configuration) ............................................................................ 22
10.5 Parametrization with the CLV Setup software .................................................. 23
10.6 Operating modes/functions .................................................................................... 28
11. Re lacing a CLV 410
11.1 Transferring the application-specific parameter set ..................................... 30
11.2 Disassembly and environmentally-friendly disposal of the CLV ............... 31
12. Troubleshooting
12.1 Self-monitoring function ............................................................................................ 32
12.2 Error messages ........................................................................................................... 32
12.3 Causes of error and troubleshooting .................................................................. 32
13. Maintenance ................................................................................................................ 36
14. Scannable bar codes (sam les) ................................................................... 37
15. Additional safety instructions (laser rotection) ........................................... 37
16. Reading field diagrams ............................................................................................38
17. Technical data ............................................................................................................. 41
18. Dimensioned diagrams ............................................................................................ 42
19. Overview: su lementary technical documentation ....................................43
20. Index ................................................................................................................................ 44
Contents

© SICK AG • Auto Ident • Germany • All rights reserved
4
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
The CLV 1x with fixed focus is used to detect and decode bar codes in a
stationary reading station automatically, e.g. on objects on a conveyor belt. For this
purpose, it must be mounted securely on the station and supplied with an operating
voltage of .5...30 V DC to IEC 7 2. The CLV outputs the read result via its host
inteface to a higher-level computer for further processing.
1. Intended use
The CLV 1x is supplied with the following:
•two polling reflectors for automatic reading pulse generation
•a note on device (yellow) with power connection diagram
•an additional set of laser warning labels (Class 2) in German and French
One or more of the following, depending on the number of devices ordered:
•CLV 1x Operating Instructions, in German and English
•a CD-ROM with the CLV Setup program for Windows™, the online CLV Setup Help
(HTML) software and the browser Iview™
•a fold-out card with 12 printed profile bar codes
2. Sco e of delivery
The following are required to set up and operate the CLV:
•a SICK AMV 0-011 Connection Module (no. 1 017 132) or AMS 0-013
(no. 1 017 135)
Alternatively, an external power supply unit with an output voltage of .5 ... 30 V DC
to IEC 7 2 (functional extra-low voltage) and min. 3 W power output
•a PC with Windows-based GUI (Windows 3.1™, Windows 95™ or Windows NT™)
and serial port (COM x)
•an HTML browser (e.g. Netscape™) to use the online help
•an RS 232 data connection cable (e.g. no. 2 020 319).
Alternatively, the programming adapter for the LCV 10 (order no. available on
request) to connect the PC between the CLV 10 and AMV/AMS 0 as well as an
RS 232 data connection cable (e.g. no. 2 01 05 ) from the programming adapter
to the PC
•a 2 V DC ±20% operating voltage for the AMV 0 or 230 V AC (115V) ±10% for
the AMS 0
•a suitable sensor, e.g. a photoelectric switch, if the reading pulse is to be supplied
externally via the switching input on the CLV
3. System requirements
4. Laser rotection
Laser radiation!
The radiation emitted by the laser diode (red or IR light) on the CLV 10 is harmful to
the human eye. For this reason, do not look directly into the laser beam.
Opening the housing while the device is in operation does not prevent the laser diode
from being activated by the reading pulse.
The maximum output power of the laser beam at the reading window is 1
.
2 mW.
The CLV, therefore, is assigned to protection class 2.
Laser radiation, do not stare into the beam, class 2 laser roduct.
Max. out ut radiation: 1
. 2 mW (223 µµ
µµ
µs), emitted wavelength: 670 nm
See Section 15 for further safety instructions!
WARNING!

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 5
Operating Instructions
8 008 22 /I982/08-00
Fig. 1. CLV 1x design
5. Design
The beeper can be assigned a different result indicator function and the beeper
volume changed (default: low) by means of the EDIT RESULT OUTPUTS field on the
DEVICE CONFIGURATION card of the operating interface.
In these o erating instructions, it is assumed that the bee er is o erated
with the default settings.
TIP:
The beeper 6either provides an acoustic confirmation that a function was executed
successfully or indicates a malfunction. The sound opening is located below the LED
label on the rear, narrow side of the device (Fig. 1).
O erating mode/function Bee er function Sequence
Reading mode*) 1. The beeper confirms the start of reading mode when the device is switched on 1. 2 tones
after the wait time for teach-in mode has elapsed
2. The beeper confirms a good read (default setting) and output of the read result 2. 1 tone
Percentage Evaluation The beeper confirms output of the reading result after every 100 reads 1. 1 tone
AutoSetup*) 1. The beeper confirms successful read of the presented profile bar code 1. 1 tone
for the start or end of AutoSetup
2. The beeper confirms successful read of the application-specific bar code 2. 1 tone
3. The beeper signals unsuccessful read of the application-specific bar code 3. 3 tones
after approx. 35 seconds
Profile programming*) 1. The beeper confirms successful read of the presented profile bar code 1. 1 tone
2. The beeper confirms the start of reading mode 10 s after the last profile 2. 2 tones
bar code has been read
Self-test The beeper signals successful completion of the self-test after power-up 1 . 1 tone
*)After a laser timeout of 10 minutes (default), the beeper indicates with one long tone that the laser diode has been switched off
and that the reading pulse has been deactivated automatically.
Table 1
6. Bee er function
&
"
§
/
Key:
!D Sub HD connector, 15-pin
" Tapped blind hole M , depth: 6 mm
§Reading window
$Laser warning label
%Connection cable
& Sound opening of beeper (covered)
/Function LEDs
!
$
$%

© SICK AG • Auto Ident • Germany • All rights reserved
6
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
7. Function of the LEDs
The LED function indicators 7 are located on the rear, narrow side of the device (Fig. 1).
LED Color Function
Device Ready green Indicates that the the CLV is ready (Reading mode). Lights up when the device is switched on, after
the self-test has been successfully completed and when the wait time for teach-in mode has expired.
The LED extinguishes when the device is switched to a different operating mode.
Result green Function depends on the operating mode of the scanner:
•In Reading Mode or Teach-in Mode, the LED indicates a successful read (good read).
If the match code comparison is activated, it lights up if the bar code read matches the specified
match code(s).
The LED lights up constantly until the start of the next reading pulse
•In Percentage Evaluation with 100 scans per reading interval, the LED indicates the trend in the
read quality as follows:
LED off good read rate < 30%
LED blinks (frequency 2 Hz) good read rate 30%...70%
LED blinks (frequency 5 Hz) good read rate 70% ... 90%
LED lights up constantly good read rate > 90%
•The LED lights up briefly when set to AutoSetup before it starts to scan the presented bar code
Laser On green Lights up when the laser diode is activated for reading the bar code.
The laser diode is activated and deactivated by the reading pulse.
The laser diode is activated constantly in Percentage Evaluation and Free-running modes.
In Reflector Polling pulse mode, it is activated in accordance with the scanning frequency at every
20th scan (LED flickers).
The scan line(s) are dimmed (with red-light scanner only)
Data yellow Flickers when the CLV is transferring data to the host on the serial data interface
Table 2
Device Ready
Result
Laser On
Data
8. 1 Pin assignment of connector
8. Electrical installation
Pin Signal Function Wire colors
of cable
no. 6 010 1371)
1 DC + .5...+30 V Supply voltage red
2 Sensor 2 Switching input teach-in (match code 1) purple
3 Result 3 Switching output (to PLC) yellow
Term RS 22 Termination for data interface 1 red/black
5 GND Ground black
6 RD+ (RS 22/ 85) Data interface 1 (receiver) light blue
7 RD- (RS 22/ 85) Data interface 1 (receiver) blue
8 TD+ (RS 22/ 85) Data interface 1 (transmitter) gray/turquoise
9 TD- (RS 22/ 85) Data interface 1 (transmitter) green
10 RxD (RS 232) Data interface 2 (receiver) gray
11 TxD (RS 232) Data interface 2 (transmitter) pink
12 Result 1 Switching output (to PLC) brown
13 Result 2 Switching output (to PLC) orange
1 Sensor 1 Switching input for ext. reading pulse white
15 SensorGND Common ground (all inputs) black/white
– – Shield white/green
1) Cable with 15-pin D Sub HD socket and open cable end
5 6
11
10
15
1
Table 3

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 7
Operating Instructions
8 008 22 /I982/08-00
8. 2 Su ly voltage
The CLV 1x requires a supply voltage of .5 ... 30 V DC to IEC 7 2 (safety extra-low
voltage, listed class II power supply).
8. 3 Connecting the CLV to the AMV/AMS 40 (Connection Module)
Before it can be used in the reading station, the CLV must be connected to the
AMV/AMS 0 Connection Module and the host/PLC/sensor as shown in Fig. 2.
Fig. 2. Block diagram: connecting the CLV in the reading station via the AMV/AMS 0 Connection Module
AMV/AMS 40
Connection
Module
Sensor 1
Result 1
Scanner
Decoder
Interface
CLV 41x
Result 2
Result 3
Data
DC 2 V 2 V DC (AMV)
230 V AC (AMS)
HOST/PLC
Data
Result 1
Result 2
Result 3
PLC
Sensor 1
Photoelectric
switch
Optional
cable
A description of the basic wiring and configuration of the Connection Module is
provided in the AMV/AMS 0 Operating Instructions (no. 8 008 292) supplied with
the module.
ØMount the AMV/AMS 0 as close as possible to the CLV
ØConnect the CLV to the 15-pin D Sub HD socket of the AMV/AMS 0
The connection cable can be extended with extension cable no. 6 010 075 (2 m).
ØDo not exceed the max. cable lengths (host interface) between the CLV and the
host/PLC (see 8.5.1 Connecting the host interface)
ØConnect a suitable, external reading pulse sensor via the AMV/AMS 0 if reflector
polling (default setting) is not to be used for automatic reading pulse generation
ØIf an external power supply unit is used instead of the AMV/AMS 0, it must be
able to provide a continuous output of at least 3 W and must comply with IEC 7 2
(functional extra-low voltage). The core cross-section for the CLV power supply
(pins 1 and 5) must be at least 0.09 mm2.
Use cable no. 6 010 137 (open wire ends) to connect the CLV. See Table 3 for
wire color assignment
TIP:

© SICK AG • Auto Ident • Germany • All rights reserved
8
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
8. 4 Connecting the PC to the CLV
A PC with Windows™ must be connected to the RS 232 port on the CLV to operate
and parametrize the device. There are two ways of connecting the PC:
1. Connect the PC COM x ort to the internal terminals on the AMV/AMS 40
ØSwitch off the power supply to the Connection Module and the PC and connect
the PC and Connection Module as described in the AMV/AMS 0 Operating
Instructions
ØDisconnect the CLV from the host if they are already connected (RS 232
interface)
ØSwitch on the PC and AMV/AMS
ØCopy the operating and parametrization software CLV Setup and the online help
CLV Setup Help to the hard disk of the PC and install the software accordingly
ØSet the communication parameters in CLV Setup for port COM x (SERIAL PORT
under OPTIONS) as shown in Table (Default setting on CLV)
2. Connecting the PC ort COM x via the o tional rogramming ada ter
Fig. 3. Connecting the PC to the optional programming adapter
PC with
Windows
AMV/AMS 40
Connection
Module
CLV 41x
Host/PLC
Programming adapter
(order no. available on request)
e.g. cable
order no.
2 01 05
1 10
15
6
11
5
5
1
9
6
to CLV 41x
to AMV/AMS 40
to PC
5 6
11
10
15
1
11
15
510 61
Host
Term
1
2
3
Clock
Button
LED for
switching
outputs
Programming
adapter
ØSwitch off the power supply to the Connection Module and PC
ØDisconnect the CLV connector from the Connection Module (Fig. 3)
ØConnect the 15-pin cable connector of the programming adapter to the socket on
the AMV/AMS 0. Connect the free 15-pin socket of the programming adapter to
the CLV connector. The adapter is supplied with power from the Connection
Module
ØConnect the PC port COM x to the 9-pin connector of the programming adapter via
an RS 232 cable (e.g. no. 2 01 05 )
ØSet the slide switch on the programming adapter to TERM (the CLV is then
disconnected from host and communicates with the PC only). The PC is
disconnected from the CLV when the switch is set to HOST
ØSwitch on the PC and AMV/AMS

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 9
Operating Instructions
8 008 22 /I982/08-00
Parameter Value
COM port freely selectable
Baud rate 9600 bit/s
Data bits 8
Parity none
Stop bit 1
Table
ØCopy the operating and parametrization software CLV Setup and the online help
CLV Setup Help to the hard disk of the PC and install the software accordingly
ØSet the communication parameters in CLV Setup for port COM x (SERIAL PORT under
OPTIONS) as shown in Table (default setting of the CLV)
With the CLV pulse mode “Sensor Input“, the reading pulse can be triggered with the
Clock button on the programming adapter. For this purpose, choose the pulse mode
via the EDIT READING PULSE field on the DEVICE CONFIGURATION card of the user interface
and download it to the CLV.
8. 5 Connecting the interfaces
All of the interfaces are routed via the 15-pin D Sub HD connector À (Table 3).
8. 5. 1 Connecting the host interface
The CLV is equipped with two host interfaces (RS 232 and RS 22/ 85). One of
these must be selected for data communication with the host/PC. If the RS 232
interface is not connected to the host, it can be used exclusively to monitor data
traffic on the parallel RS 22/ 82 interface.
TIP:
The internal modules of the CLV may be damaged irreparably if the data interface
is connected incorreclty.
ØConnect the host interface of the CLV to the host in accordance with the EMC
requirements using shielded cables(Fig. )
ØTo prevent interference, do not lay the data cables over a long distance parallel
to the power supply and motor cables, e.g. in cable ducts
ØApply the shield on one side (SICK recommendation)
Recommended max. cable lengths between the CLV and host:
CAUTION
Interface ty e Transmission rate Distance from host
RS 232 up to 19 200 bits/s max. 10 m
38 00 ... 57 600 bits/s max. 3 m
RS 22/ 851) max. 38 00 bits/s max. 1 200 m
max. 57 600 bits/s max. 500 m
1) with appropriate line termination
Table 5

© SICK AG • Auto Ident • Germany • All rights reserved
10
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
Fig. . Connecting the host interface
The default communication parameters of the CLV host interface are as follows:
Parameter Value
Data transmission rate 9600 bit/s
Data bits 8
Parity none
Stop bit 1
Protocol Sick (start character: STX, stop character: ETX, no
repeat request: none, timeout: 50 ms)
Table 6
The parameters can be changed in the DATA FORMAT and INTERFACE PROTOCOL fields on
the HOST INTERFACE card of the user interface.
By reading in profile bar code nos. 11 and 12 (Section 10.2 Default setting), you can
define the above default settings temporarily with a different parameter set to allow
the PC to communicate with the CLV.
8. 5. 2 Connecting the functional interfaces
Sensor 1 switching in ut
In order to trigger a reading process, the CLV requires a suitable clock pulse in
Reading mode that signals the presence of a conveyor object in the reading field.
The clock pulse generates the reading interval (time window for evaluation) internally
in the CLV. In the default setting, the reading pulse is supplied by means of internal,
continuous reflector polling (see Section 9.2).
If an external sensor (e.g. photoelectric switch, switch) is to notify the CLV of the
presence of a conveyor object, the following steps must be performed:
ØConnect the sensor to the CLV as shown in Fig. 5
TIP:
IMPORTANT
RxD
TxD
GND
RD+
RD–
TD+
TD–
GND
Term
TxD
RxD
GND
TD+
TD–
RD+
RD–
GND
10
11
5
6
7
8
9
5
4
CLV 410 Host
RS 422
CLV 410 Host
RS 232
RS 485 (SICK Network): Connection diagram available on request
Terminal assignment
AMV 40-011/AMS 40-012, -013:
T. 6
T. 7
R+
R-
T. 8
T. 9
T+
T-
T. 5
GND
T. 10
T. 11
RxD
TxD
GND T. 5

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 11
Operating Instructions
8 008 22 /I982/08-00
ØSet the trigger source to SENSOR INPUT via the EDIT READING PULSE field on the DEVICE
CONFIGURATION card of the user interface and download the parameter settings to
the CLV
An external clock pulse is not required for Percentage Evaluation operating mode.
Fig. 5. Connecting the Sensor 1 input to an external reading pulse sensor
Sensor 1 switching in ut
Switching Start internal reading interval when input
characteristics live (active high; default setting)
Pro erties opto-decoupled, non-interchangeable
Electrical values Low: -1V ≤ Vin ≤ +1 V High: -28 V ≤ Vin ≤ -3 V
-0.3 mA ≤ Iin ≤ +0.3 mA -18 mA ≤ Iin ≤ -1. mA
+3 V ≤ Vin ≤ +28 V
+1. mA ≤ Iin ≤ +18 mA
Table 7
Sensor 2 switching in ut
This input is used exclusively to teach in match code 1. The teach-in procedure is
an alternative to entering the match code on the keyboard via the CODE COMPARISON
field on the DEVICE CONFIGURATION card of the user interface.
In order to teach in the match code, the input is connected briefly to the power
supply via a switch (Fig. 6). This can also take place in the Connection Module.
Fig. 6. Connection of Sensor 2 input for teaching in match code 1
Switch
CLV 41x PNP sensor
DC 24V
6K8
1
14
15
5
Sensor
GND
GND
DC 24V
DC 24V
Sens GND
Sensor 1
1K5
1K5
VS= + .5...+30 V DC
VS
VS
VS
15
5
1
14
Terminal assignment
AMV 40-011/AMS 40-012 ,-013:
T. 18
T. 1
T. 15
VS
SENS
SGND
T. 5
GND
CLV 41x
DC 24V
6K8
1
14
15
5
Sensor
GND
GND
DC 24V
DC 24V
Sens GND
1K5
1K5
VSSensor 2
VS
SensGND
GND
2Switch S
Electrical values and properties:
as Sensor 1 (see Table 7)
Terminal assignment
AMV 40-011/AMS 40-012, -013:
T. 18
T. 2
T. 15
VS
Sensor 2
SGND
T. 5
GND
VS= + .5...+30 V DC
Note:
Instead of the bridge the “GND-SGND“
Jumper can be set in the AMV/S 0.

© SICK AG • Auto Ident • Germany • All rights reserved
12
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
ØProcedure for teaching in match code 1
Prerequisite: CLV is in Reading mode and the current parameter set (if
application-specific) is stored permanently.
Notes:
1. In order to replace the match code taught in in this way with a different match
code, you can repeat steps 3 to 6 without having to switch off the CLV.
2. If match code 1 is set using the teach-in method, it must not be modified with the
user interface (or command strings) until the CLV is switched off again. The
methods used to set the match code cannot be mixed. This does not affect
match code 2.
Result 1 to Result 3 switching out uts
The three outputs (Fig. 7) can each be assigned different result functions. When the
associated result occurs, the corresponding output becomes live for the selected
pulse duration at the end of the reading interval. The pulse duration can be set
separately for each output.
Ste Action Res onse from CLV1)
1 Switch off CLV –
2 Close switch S and CLV switches to Reading mode and activates
switch on CLV again teach-in mode for the match code.
3 Open switch S CLV in teach-in mode and searches match
code.
Position match code at appropriate –
distance (see 16. Reading field
diagrams)
5 Trigger set reading pulse CLV reads the match code.
The Result LED lights up to confirm
successful read (good read; default setting).
Beeper emits a short acoustic signal.
6 Close switch S again CLV stores the match code and its code type
in the valid parameter set and permanently
in the EEPROM.
From this point on, it compares each match
code read with the match code taught in.
1) We recommend that you connect the PC with the CLV Setup interface to check the read result
Table 8
Fig. 7. Connection example: Result 1 output
Configuration of Result 2 (pin 13/terminal 13) and Result 3 (pin 3/terminal 3): as Result 1 output
VS= depending on
operating voltage applied
(+ .5...+30 V DC)
„Result 1“
VS
Vo
CLV 41x
Pulse duration:
10 ms ... 9.9 s or static
(to end of next reading
pulse)
12
5
Terminal assignment
AMV 40-011/AMS 40-012, -013:
T. 12
T. 5
Result 1
GND

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 13
Operating Instructions
8 008 22 /I982/08-00
The function assignment, pulse duration and polarity of the signals can be modified
via the EDIT RESULT OUTPUTS field on the DEVICE CONFIGURATION card of the user interface.
The Result LED is not affected by this and lights up in Reading mode for a Good
Read until the end of the next reading pulse.
TIP:
Result 1 to Result 3 switching out uts
Switching PNP-switching to supply voltage VS
characteristics
Pro erties short-circuit + temperature-protected, not electrically isolated from VS
Function assignment Result 1: Device Ready, polarity: not inverted
(default setting) Result 2: Good Read, polarity: not inverted
Result 3: No Read, polarity: not inverted
Pulse duration: 100 ms
Electrical values 0 V ≤ VOut ≤ VSGuaranteed: VOut ≤ VS − 1.5 V at IOut ≤ 600 mA
IOut ≤ 600 mA
Table 9
9. 1 Securing the CLV
The two fastening threads (2, Fig. 1) on the bottom, narrow side of the housing are
used to secure the CLV. The thread dimensions are shown in Section18. Dimensioned
diagrams.
9. Assembly
Mounting accessories
The CLV can be easily secured to the base to suit the relevant application using one
of the two optional mounting brackets (Fig. 8). This facilitates precise adjustment in
two planes. The dimensions of these brackets are shown in Section 18.
Dimensioned drawings. The elongated holes in the brackets allow the CLV to be
adjusted with a freedom of rotation of approx. ± 15°.
Risk of damage to the housing
The max. depth of engagement of the two M tapped blind holes is 6 mm from
the housing surface. Do not exceed this depth.
Fig. 8. Mounting example: securing the CLV using the mounting bracket (accessory)
Aligning the CLV with the bar code
The alignment between the CLV and line or raster scanner depends on the
application. The basic alignment between the two scanning methods and the
conveyor object is shown in Fig. 9.
Mounting bracket (small)
no. 2 020 077
CAUTION

© SICK AG • Auto Ident • Germany • All rights reserved
14
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
Fig. 9. Scanning method alignment relative to the bar code and conveyor direction
Line scanner Raster scanner
Reading angle and distance
When mounting the device, ensure that all of the possible reading angles at which the
bar code can be positioned relative to the scan line are taken into account. Fig. 10
shows this using a line scanner as an example. The reading window and code should
be almost parallel, whereby total surface reflection (see below) should be avoided.
Fig. 10. Mounting example of CLV with line scanner: reading angles taken into account
α: Tilt
β: Pitch
γ: Skew
Reading
distance
Reading
range
βα
γ
The following angle limits should not be exceeded (CLV):
Angle Tilt αPitch βSkew γ
Limit value 15° (depening on code print) 5° 5°
Table 10
When selecting the mounting location for the device, the limit values of the reading
range for the selected bar code resolution and the useful aperture angle (max. 60°
with front light emission) must be taken into account. See also Section 16. Reading
field diagrams.
Avoiding surface reflection
If the reading beam strikes the surface of the bar code vertically, reflection
interference may be caused when the returned light is received.
ØMount the CLV with front light emission (a) in such a way that the scan line(s) strike
the bar code at an angle less than approx. 15° from the vertical reference (Fig. 11).
ØThe CLV with lateral light emission (b) must be mounted flush so that the scan line
is emitted at an angle less than 105° relative to the housing.

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 15
Operating Instructions
8 008 22 /I982/08-00
Fig. 11. Preventing surface reflection by rotating the CLV from the perpendicular
Line scanner (To view)
a) CLV with front light emission b) CLV with lateral light emission
(flush mounting)
105°
105°
Count direction of code osition CP
The CLV determines the position of the bar code within the scan line (CP value),
Fig. 12. This can be used to separate several identical bar codes (identical code
type, length and content) and output them accordingly.
Fig. 12. CLV 1x: Count direction of the code position CP
9. 2 Mounting the olling reflector
In order to generate the reading pulse automatically using internal reflector polling
(default setting), one of the two polling reflectors supplied (special bar code) must
be mounted opposite the CLV behind the conveyor section.
ØNote the maximum ranges of the two reflectors (Fig. 13)
ØMount the polling reflector (self-adhesive) opposite the CLV (in the alignment line)
in such a way that the CLV can view the reflector freely in the gaps between the
conveyor objects and can scan it reliably (center of scan line).
The scan line of the CLV should scan the code lines of the polling reflector
vertically.
ØCheck the polling reflector installation for 100% read quality using the
Percentage Evaluation mode in the TERMINAL option of the user interface (see
Section 10.6). The CLV outputs the code content of the polling reflector and
signals REPO as the code type.
Function of reflector olling
In Reflector Polling pulse mode, the CLV activates the laser diode with every 20th
scan and checks whether it detects the reflector:
- if the reflector is still detected, the CLV repeates the procedure after the 20th
scan
0
CP
100

© SICK AG • Auto Ident • Germany • All rights reserved
16
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
Fig. 13. Mounting the polling reflector (REPO)
white blue
50
50
1
1
Polling
reflector
- the CLV starts the reading interval when the reflector is concealed by a conveyor
object in the reading field. The interval ends when the CLV detects the reflector
again. It then repeats the procedure above.
Notes:
This clock mode (max. delay approx. 100 ms at a scan rate of 200 Hz) is only
suitable for applications that are not critical with respect to time.
When the CLV is switched on, it must be able to detect the polling reflector. The first
reading interval can only start if this is the case.
In Reading mode, the CLV does not output the content of the polling reflector.
9. 3 Mounting the external reading ulse sensor (alternative)
If an external sensor (e.g. photoelectric switch) is used to generate the reading
pulse, it must be mounted on the CLV in a suitable manner. Ensure that objects of
different size carrying bar codes trigger the internal reading pulse in such a way that
sufficient time is provided to read the code. Examples of the CLV with line scanner
are provided in Fig. 1 .
Fig. 1 . Example of the position of the external reading pulse generator
Bar code in the center or at the
end of the conveyor object
Bar code at the start of
the conveyor object
b
b
b < a b < a
ØChoose the external sensor as the trigger source via the SENSOR INPUT field under
EDIT READING PULSE on the DEVICE CONFIGURATION card of the user interface and then
download the parameters to the CLV
a
a
• Reading distances (REPO)
Other distances must be respected
for CLV 12 and CLV 1 . Only blue
REPO is allowed.
• Reading distances (REPO) CLV 12: 150 ... 350 mm
• Reading distances (REPO) CLV 1 : 230 ... 650 mm
Distance:
500 ... 800 mm
Distance:
0 ... 500 mm

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 17
Operating Instructions
8 008 22 /I982/08-00
10. Power on and setu
10. 1 Ste -by-ste setu instructions
ØSetu with the default settings
1. Connect the CLV to the 4.5...30 V DC ower su ly
(see also Section 8.1 Pin assignment of the connector)
2. Align the CLV with the white olling reflector su lied (s ecial code),
distance a rox. 300 mm
(see also Section 9.2 Mounting the polling reflector)
3. Present the bar code sam le above at a distance of a rox. 200 mm and
cover the olling reflector
(see also Section 10.6.1 Reading mode )
4. Uncover the olling reflector again
The CLV confirms that the read was successful with acoustic signal from the
beeper and the Result LED lights up
5. The CLV can now be o erated with the default settings!
(see Section 10.2 Default settings of the CLV for the default settings)
ØSetu with rofile bar codes (change arameter set)
1. Present the required rinted rofile bar code(s) on the card su lied
to the CLV
The CLV confirms every good read of the profile bar code with an acoustic signal
from the beeper. If a further bar code is not presented to the CLV within 10 s, the
CLV switches to Reading mode and outputs two consecutive acoustic signals
(see also Section 10.3 Profile programming)
2. The CLV can now be o erated with the modified settings!
ØAutoSetu (modified code configuration)
1. Present bar code no. 10 to the CLV
The CLV confirms the start of AutoSetup by means of an acoustic signal from the
beeper and the Result LED lights up for an extended period of time.
(see also Section 10. AutoSetup)
2. Present the a lication-s ecific bar code to the CLV at the maximum
reading distance in the a lication
The CLV confirms the good read by means of an acoustic signal from the beeper
and the Result LED lights up. It then switches to Reading mode and outputs two
consecutive acoustic signals.
3. The CLV can now be o erated with the modified code configuration!
The code ty e resented to the CLV is then evaluated
Bar code sam le
Code type: Code 39
Module width: 0.5 mm
Quick start
Profile
programming
Teaching in application-
specific bar codes
UVWXYZ

© SICK AG • Auto Ident • Germany • All rights reserved
18
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
ØSetu with CLV Setu software (change arameter set)
1. Connect the PC to the CLV
(see also Section 8. Connecting the PC to the CLV)
2. Start the CLV Setu rogram on the PC
(see also Section 10.5 Parametrization with the CLV Setup software)
3. U load the configuration from the CLV
4. Change the configuration cards as required.
The arameters and their functions are ex lained in the online hel .
5. Download the configuration to the CLV and store the values ermanently
6. Save the modified arameter set on the PC as a configuration file!
Alternative:
5. Print out the rofile bar codes and resent them to the CLV in the s ecified
sequence
(see also Section 10.3 Profile programming)
ØSetu with command strings from the host
(not described here, further information is available on request)
Alternative
procedure
10. 2 Default settings
A concise overview of the default CLV settings that are effective when the device is
switched on for the first time after delivery is shown in Table 11.
The tried-and-tested default parameters have been selected so that CLV can be
used directly in many applications without requiring modification.
Parameter Default setting (CLV 410)
Code types (active) Code 39, Code 128, Code 93, Codabar, EAN, EAN 128, UPC, 2/5
Interleaved
Code length Freely definable (all code types, except EAN and UPC, 2/5 Interleaved: ...50
characters)
Multiple read 3 (all code types)
Min./max. no. of codes 1
Resolution 0.5 mm
Scanning frequency 500 Hz
Min. reading distance 50 mm
Reading pulse generation Start: polling reflector End: polling reflector
Beeper Result function: Good Read, volume: low
Switching outputs Result 1: Device Ready; Result 2: Good Read; Result 3: No Read
Protocol (Host) Start character: STX, stop character: ETX
Transmission rate (host) 9600 bits/s
Data format (host) 8 data bits, no parity, 1 stop bit
Output format (host) Header: blank, separator: blank, terminator: blank, error string: NOREAD+separator
Output sequence Code position
Output time Read result: end of pulse Separator: after code
Test string Inactive
Table 11
TIP: You can call up the entire default settings from the CLV Setup program as follows:
ØChoose FACTORY DEFAULTS in the FILE menu or
ØClick the symbol.

CLV 1x Bar Code Scanner
© SICK AG • Auto Ident • Germany • All rights reserved 19
Operating Instructions
8 008 22 /I982/08-00
The CLV must not be connected with the PC (offline mode). The default setting
values are stored permanently in the user interface and on the CLV. The parameter
values are displayed in the menu options of the individual configuration cards. You
can print out the entire default settings by clicking the icon.
Note:
When you make changes to the default settings, we recommend that you save them
as a different configuration file (*.scl) on the PC using the SAVE AS option in the
program.
IMPORTANT
Tem orary host interface defaults
In order to allow you to access the CLV with the PC at any time, you can set the
host interface temporarily to the default setting (9600 bits/s, 8 data bits/1 stop
bit, no parity) using the profile bar codes:
ØPresent rofile bar code no. 11 to the CLV in Reading mode and su ly
an a ro riate clock ulse to the device.
The CLV resets the host interface to the default values and confirms this with
an acoustic signal from the beeper. It then resumes Reading mode
immediately
ØTo restore the revious host interface configuration on the CLV, resent
rofile bar code no. 12 to the CLV.
The CLV confirms that the values have been reset with an acoustic signal and
resumes Reading mode immediately.
Alternative:
ØSwitch the CLV off and on again to return to the original configuration
10. 3 Profile rogramming
Profile programming uses special bar codes to change the CLV parameters directly
without having to connect the PC with user interface. The profile bar codes are
simply presented to the CLV. The device then transfers the coded parameter values
to the current parameter set and stores them permanently. The changes remain valid
after the programming procedure has been completed and affect the further reading
procedure. The CLV waits 10 s after each profile bar code. If no more profile bar
codes are presented, it resumes standard Reading mode and confirms with two
consecutive acoustic signals.
Since the CLV overwrites important parameter values (e.g. those generated with
AutoSetup) without requiring confirmation, we recommend that you upload the valid
parameter set to the PC and store it as a configuration file (*.scl) with the SAVE AS
option in the user menu before you change the parameter settings.
You can program the CLV at any time in Reading mode. Alternatively, you can present
one or more profile bar codes to the CLV for reading within 5 seconds after the
device has been switched on and the self-test has been completed (free-running).
The CLV does not output any reading results during profile programming and does
not respond to an external reading pulse.
IMPORTANT
Printing out the
default settings
Parametrization
without a PC

© SICK AG • Auto Ident • Germany • All rights reserved
20
Operating Instructions
8 008 22 /I982/08-00
CLV 1x Bar Code Scanner
Printed rofile bar codes
A total of 12 selected profile bar codes are provided on the „Configuration Profiles
for CLV 1x Bar Code Scanner“ card to make important changes (from the default
settings or the application-specific parameter set respectively). By folding the
individual segments of the card, you can ensure that the CLV only reads the
required profile bar code. The function of the profile bar codes is explained in
Table 12.
If you cannot access the host interface of the CLV from the “CLV Setup” user
interface (e.g. baud rate or data format not in line with default settings), the profile
bar code no. 11 can be used temporarily to set the CLV to the default settings
(host interface). To reset the system, use profile bar code no. 12 or switch off CLV.
IMPORTANT
Barcode Function
Profile 1 Default settings for all parameter values on the CLV
(Code configuration: all code types are enabled for evaluation).
Overwrites the code configuration generated by AutoSetu !
Profile 2 Changes the configuration on the host interface only:
- Data transmission rate: 38 00 Bits/s
- Data format: 8 data bits, no parity, 1 stop bit
- Repeat request: none
Profile 3 Changes the configuration on the host interface only:
- Activates RS 22/ 85 interface
- SICK network protocol (standard)
- Data transmission rate: 38 00 Bits/s
- Data format: 7 data bits, odd parity, 1 stop bit
- Device number: 99
Profile Changes the configuration on the host interface only:
Outputs the read result immediately after the evaluation criteria have been fulfilled
Profile 5 Changes the configuration on the host interface only:
- Outputs the error status ST in the separator
- CLV outputs: CR LF in terminator (at the start of the line after every code)
Profile 6 Parameterizes the CLV as Master in master/slave configuration
- No. of slaves: 1
- Timeout: automatic
Profile 7 As profile 6, but no. of slaves: 2
Profile 8 Parameterizes the CLV as Slave in master/slave configuration
Profile 9 Parameterizes the CLV for pass-through mode:
Outputs “ / “ and the device number in the header
Profile 10 Starts or ends AutoSetup
Profile 11 Resets the host interface parameters to the defaults temporarily
Profile 12 Returns the host interface parameters to the original values
Table 12
The function of the individual parameters is explained in the online CLV Setup Help.TIP:
Table of contents
Other SICK Barcode Reader manuals

SICK
SICK CLV65 Series User manual

SICK
SICK CLV 490 Series User manual

SICK
SICK CLV62 Series Manual

SICK
SICK Lector 632 User manual

SICK
SICK CLV61 DualPort Series User manual

SICK
SICK IDM series User manual

SICK
SICK CLV 410 Operating instructions

SICK
SICK CLV 420 User manual

SICK
SICK CLV 45 Series User manual

SICK
SICK IDM series User manual

SICK
SICK CLV60 series User manual

SICK
SICK CLV63x User manual

SICK
SICK Lector85x User manual

SICK
SICK CLV62x Reference guide

SICK
SICK CLV62 Series User manual

SICK
SICK IDM140BT Series User manual

SICK
SICK Lector61x User manual

SICK
SICK CLV69x User manual

SICK
SICK Lector620 User manual

SICK
SICK CLV 490 User manual