
Examples: 1s keys 0, 1 or 1, ENTER
15s keys 1, 5
3.3.3.3 Vacuum level setting (sensor enabled)
For a selected program set the vacuum level, starting with the values; the
decimal point is automatically inserted following the second digit entry and the
validation is automatically performed following the third digit entry (the new
vacuum level is blinking). The vacuum level is rounded off to the nearest half
value. In the middle of an acquisition, use key "ENTER" to validate the vacuum
level and key "ESC" to come backward and start over with a new acquisition
(the old vacuum level is blinking). Set vacuum level to zero to bypass the
pressure transducer and proceed only using the vacuum plus time.
Examples: 90.0% keys 9, 0, 0 or 9, 0, ENTER or
keys 9, 0, 1 or 9, 0, 2 or 9, 0, 3 or 9, 0, 4
97.5% keys 9, 7, 5 or
keys 9, 7, 6 or 9, 0, 7 or 9, 0, 8 or 9, 0, 9
0.0% keys 0, 0, 0 or 0, ENTER
3.3.3.4 Vacuum plus time setting (sensor enabled)
For a selected program set the vacuum plus time, in seconds; the validation is
automatically performed following the second digit entry (the new vacuum plus
time is blinking). In a middle of an acquisition, use key "ENTER" to validate the
vacuum plus time and key "ESC" to come backward and start over with a new
acquisition (the old vacuum plus time is blinking).
Examples: 1s keys 0, 1 or 1, ENTER
15s keys 1, 5
3.3.3.5 Gas time setting (sensor disabled)
For a selected program set the gas time setting following the same procedure as
for the vacuum time. Keep in mind that increasing gas time decrease sealing
pressure. Some vacuum must be kept inside to assure proper functioning.
3.3.3.6 Gas flush level setting: (sensor enabled)
For a selected program set the gas flush level following the same procedure as
for the vacuum level; the maximum gas flush level setting is 10% below the
vacuum setting.
3.3.3.7 Sealing time setting:
For a selected program set the sealing, starting with the seconds; the decimal point is