Storz 11301 BNX Series User manual

Flexible Intubation Video Endoscope
Series 11301 BNX, Monitor 8403 ZX
Instruction
Manual

Thank you for your expression of confidence in the KARL STORZ brand name. Like all of our other products, this
product is the result of years of experience and great care in manufacture. You and your organization have decided
in favor of a modern, high-quality item of equipment from KARL STORZ.
This instruction manual is intended to serve as an aid in the proper handling, cleaning, disinfection and, where
required, sterilization of the flexible intubation video endoscope. All essential details of the equipment and all actions
required on your part are clearly presented and explained. We thus ask that you read this manual carefully before
proceeding to work with the instrument. Keep this manual available for ready reference.
Warning: KARL STORZ instruments are provided non-sterile, and must be cleaned and high-level
disinfected/sterilized prior to the initial use and before each subsequent use.
KARL STORZ is constantly working on the further development of all products. Please appreciate that changes to the
scope of supply in form, equipment and technology are possible for this reason. Therefore, no claims may be
deduced from the information, figures and descriptions in this manual.
CA TION: Federal (USA) law restricts this device to sale by or on the order of a physician.
Important Information
For sers of KARL STORZ
Instruments
2

ymbol Description
Follow the instruction manual
White balance
Single image capture
Video recording
This device has been marked in accordance with the European irective 2012/19/EU on waste
electrical and electronic equipment (WEEE)
evice type BF
Electronic information product pollution control (China RoHS)
Manufacturer
Symbol Description
3

Controls
Illustrated Description
4
13242 XL Leakage Tester
8401 YZ
Z11457 Pressure Compensation Cap
l
2^
2%
2$
g
f
a
2)
2!
;
d
s
h
j
k
2&
o
i
u
w
e
t
y
r
q
2# 2@
2*

qConnecting cord and cable connector
wSuction valve 11301 CE1/11301 C 1
eConnection for Leakage Tester (P/N 13242XL) or pressure compensation cap (vent port)
rInstrument channel port
tShaft
yHandle
ueflection lever
iistal tip
oLeakage Tester (P/N 13242XL)
aMonitor/CCU
sConnection 2
dConnection 1
fMains connection
gReset button
hHolder for S memory card
jStatus bar
kPilot lamp
lOn/Off switch
;Pilot lamp charging (orange)
2) Control keys
2! Function symbols
2@ USB port
2# H MI port
2$ Video recording Start/Stop
2% White balance
2^ Still image recording
2& Pressure compensation cap Z11457
2* Cleaning cap 8401 YZ
Controls, Displays, Connectors
And Their ses
5

Important information for users of KARL STORZ instruments 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬2
Symbol description 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬3
Illustrated description 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬4
Controls㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬4
Controls, isplays, Connectors and their uses 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬5
Safety instructions 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬9
Indications for use 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬9
User qualification㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬9
Warnings and cautions 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬10
Safety precautions when using the Flexible Intubation Video Endoscope 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬10
Handling 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬13
Unpacking the device㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬13
Basic equipment and accessories 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬13
Operating conditions 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬14
Electromagnetic compatibility㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬14
External system components 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬14
Installation and operating instructions㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬15
Handling 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬15
Operating㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬15
Holding correctly 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬16
Installing the suction valve㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬16
Installing the tube holder㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬16
Handling the shaft㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬17
Handling the deflection mechanism 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬18
Handling the working channel 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬19
Connection to the monitor 8403 ZX 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬20
Use of the H MI output 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬20
Functions of the monitor 8403 ZX㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬21
Focus 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬21
Function buttons of the Flexible Intubation Video Endoscope 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬21
Inspecting the suction function㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬22
Inspecting the insufflation㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬22
Inspecting the surfaces㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬23
Checking the deflection 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬23
Switching the Flexible Intubation Video Endoscope on/off 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬24
Contents
6

Video recording㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬24
Single image capture 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬24
Inspecting the image quality 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬25
Switching off the monitor 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬25
System reset 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬26
Monitor bag 8403 Y 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬26
Cleaning and Sterilization/ isinfection instructions 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬 27
General information㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬27
Water quality requirements 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬28
Cleaning equipment and accessories㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬29
Bedside pre-cleaning (at point of use)㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬30
Leakage Test㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬 31
Manual cleaning 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬33
etergent cleaning㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬33
Water rinsing 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬35
rying 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬36
Inspection㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬36
Sterilization㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬37
Sterilization methods㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬38
Ethylene oxide (EtO) sterilization㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬38
STERIS System 1E (SS1E) liquid chemical sterilant processing system 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬39
High-level disinfection 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬39
Reference for cleaning and sterilization㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬40
Storage 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬41
Endoscope return instructions㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬42
escription of the unit㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬43
System description 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬43
Technical data 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬43
Maintenance㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬44
Servicing and repair㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬44
Maintenance㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬44
irective compliance㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬44
Contents
7

Important information 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬44
isposal 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬45
Troubleshooting 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬46
Accessories㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬47
Electromagnetic compatibility information for KARL STORZ Flexible Intubation Video Endoscope 11301 BNX
Monitor 8403 ZX 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬49
Shipping 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬54
Return Policy 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬54
Repair Program 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬55
Warranty Policy 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬55
Limitation of liability 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬56
Software ownership and licensing 㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬㛬57
Contents
8

Please read these safety instructions carefully before using the device. Before using the instrument on the patient it is
imperative that you be acquainted with how the instrument operates and is controlled.
Indications for Use
The Flexible Intubation Video Endoscope is intended to be used for examination and visualization of a patient's upper
airway and for aiding placement of a tracheal tube.
Warning: Flexible Intubation Video Endoscope are applied parts, which should only be used with supply units
specified for this purpose by KARL STORZ.
User Qualification
The KARL STORZ Flexible Intubation Video Endoscope may only be used by persons with an appropriate medical
qualification and who are acquainted with the laryngoscopic technique. The information given in these instructions
only serves to instruct in the correct handling and preparation of the Flexible Intubation Video Endoscope. It is not
intended as an introduction to the technique of laryngoscopy.
Safety Instructions
9

Warnings and Cautions
Please read this manual carefully and follow the instructions exactly. The words Warning, Caution, and Note convey
special meanings. Wherever they are used in this manual, they should be carefully reviewed to ensure the safe and
effective operation of this product. To further emphasize the words Warning, CA TION and Note, they are
preceded by a pictogram.
Warning: A Warning indicates that the personal safety of the patient or physician may be involved. isregarding a
Warning could result in injury to the patient or physician.
CA TION: A Caution indicates that the personal safety of the patient or physician may be involved. isregarding a
Warning could result in injury to the patient or physician.
Note: A Note indicates special information about operating the product, or clarifies important information.
afety Precautions When Using the Flexible Intubation Video Endoscope
The Flexible Intubation Video Endoscope must be used according to the recognized medical rules and procedures of
endoscopy.
Warning: Every time the Flexible Intubation Video Endoscope is used, you must first check that it, and any
accessories used in combination with it, is in perfect condition. amaged Flexible Intubation Video Endoscopes or
damaged accessories must not be used.
Warning: Combinations of medical devices are only then assured to be safe if
• they are identified as such in the respective instruction manual or
• the intended purpose and the interface specifications of the devices used in combination permit this.
Pay careful attention to the instructions and interface specifications of medical products used in combination.
Warning: The Flexible Intubation Video Endoscope is delivered non-sterile and must be cleaned and high-level
disinfected/sterilized before initial and subsequent use.
Warning:: The pressure compensation cap (P/N Z11457) 2& must be removed from the vent port efor:
• endoscopic interventions
• immersing in liquid, e.g. disinfectant solution.
Warning: The pressure compensation cap (P/N Z11457) 2& must be in place on the vent port wfor:
• gas/plasma sterilization
• shipping
Warning: o not point the light beam of the Flexible Intubation Video Endoscope into the patient and/or other
healthcare provider's eye. Looking into the light beam may cause eye damage.
Safety Instructions
10

Warning: anger of explosion if flammable gases are used in the immediate proximity of the Flexible Intubation
Video Endoscope. The Flexible Intubation Video Endoscope must not be used in the presence of flammable or
combustible media.
Warning: The Flexible Intubation Video Endoscope may only be repaired by personnel authorized by KARL STORZ.
Unauthorized repairs may impair functioning and safety, and are therefore strictly forbidden.KARL STORZ gives no
warranty for Flexible Intubation Video Endoscopes that are repaired by unauthorized personnel.
Warning: amage to the Flexible Intubation Video Endoscope and injury to the patient or user resulting from
incorrect operation are not covered by the manufacturer’s warranty.
Warning: Always unplug the Flexible Intubation Video Endoscope before carrying out any maintenance and
cleaning work on it.
Warning: The Flexible Intubation Video Endoscope must not be used during the discharge of a cardiac defibrillator.
Remove the Flexible Intubation Video Endoscope from the operation area before discharge takes place.
Warning: Switch off the Flexible Intubation Video Endoscope before putting it down and do not place it in the
proximity of the patient (as per IEC 60601-1-1).
Warning: Only KARL STORZ devices (monitor 8403 ZX) may be used. isregarding this warning can endanger the
patient and user.
Warning: If the Flexible Intubation Video Endoscope should malfunction during application on a patient, stop the
application immediately. Place the distal tip of the Flexible Intubation Video Endoscope in the neutral, nondeflected
position and remove the Flexible Intubation Video Endoscope slowly and carefully from the patient.
CA TION: o not steam sterilize (autoclave) the Flexible Intubation Video Endoscopes under any conditions.
CA TION: The maximum permissible temperature for decontamination, sterilization, transport and storage is
65° C (149° F).
CA TION:: o not allow any electrically live components or instruments to come into contact with the Flexible
Intubation Video Endoscope.
CA TION: o not place the Flexible Intubation Video Endoscope on devices that become very hot. o not expose
it to direct sunlight or excessive heat, to avoid damage to the device.
CA TION: Flexible Intubation Video Endoscopes must never be exposed to X-rays, to avoid damage.
CA TION: o not hit the distal tip of the Flexible Intubation Video Endoscope against hard objects, and never use
a sharp object to remove dirt.
CA TION: Never lay heavy objects on the Flexible Intubation Video Endoscope.
Safety Instructions
11

CA TION: To avoid damaging the monitor, do not remove the S card or connecting cord while a video is
recording.
CA TION: To avoid damaging the monitor, always switch it off before replacing the S card.
Safety Instructions
12

Unpacking the Device
Warning: The Flexible Intubation Video Endoscope is delivered unsterilized, so it has to be disinfected/sterilized
prior to initial use and before each subsequent reuse.
Check for missing items and evidence of shipping damage.
File any complaints immediately with KARL STORZ or the supplier.
Basic Equipment and Accessories
The set 11301BNX-KT comprises:
1 11301 BNX Flexible Intubation Video Endoscope 5.5 x 65 mm
1 Carrying case 27677 FV
1 Instruction manual Z17845US
1 Leakage Tester 13242 XL
1 Pressure compensation cap Z11457
1 Sterile suction valve 11301 CE1
1 Luer plug 29100-BK
2 Cleaning adaptor 11301 C 1
1 Flat cleaning brush 27652/3
1 Short cleaning brush 9990/50
1 Long cleaning brush 110910/50
1 LIPP tube holder (only intubation endoscopes) 11301 CFX
1 Cleaning cap (P/N 8401 YZ)
Handling
13

Operating Conditions
Flexible intubation video endoscope:
Ambient temperature: 15° C– 37° C (59° F–98.6° F)
Humidity: 15%– 80% relative humidity
Monitor 8403 ZX:
Ambient temperature: 15° C– 37° C (59° F–98.6° F)
Humidity: 15%–80% relative humidity
Electromagnetic Compatibility
This device has been tested and is found to comply with the EMC limits of EN/IEC 60601-1-2 :2001 CISPR 11 Class B.
External ystem Components
Charger
Using the medically approved charger (P/N 40150031), the Li-Ion battery of the monitor can be recharged extremely
quickly.
Moreover, this charger enables the monitor to be operated with a discharged Li-Ion battery.
The charger has a green LE display which indicates that the output voltage is correct.
Country specific configuration
The charger features interchangeable primary adapters, which allow the system to be adapted to the country-specific
power supply outlets.
a Primary adaptor for EUROPE: ET27-30-0004 516
b Primary adaptor for UK: ET27-30-0004 5178
c Primary adaptor for US/Japan: ET27-30-0004 520
d Primary adaptor for Australia: ET27-30-0004 519
e Primary adaptor USA ET27-30-0004369
f Adaptor cable ET27-30-0004 370
Handling
14
abc
d
ef

Handling
The Flexible Intubation Video Endoscope must be used according to the medical rules and procedures recognized
for the laryngoscopic methods.
CA TION: o not store or reprocess the Flexible Intubation Video Endoscope together with sharp objects in order
to avoid causing damage to the outer casing.
Operating
The following instructions outline the correct operation of the Flexible Intubation Video Endoscope and the
corresponding accessories. This is not an introduction to the techniques of laryngoscopy. Instructions on the
endoscopic technique can be found in the relevant medical literature.
The Flexible Intubation Video Endoscope is used for endotracheal intubation and inspection of the oropharynx.
Installation and Operating
Instructions
15

Holding Correctly
Grip the handle of the Flexible Intubation Video Endoscope in your left or right hand, as shown in the picture. With
this kind of control grip, the Flexible Intubation Video Endoscope can be operated with one hand.
Installing the uction Valve
Place the suction valve (P/N 11301CE1) wonto the valve connection and turn it until the lateral projections on the
insert are located above the recesses on the connection.
Click the suction valve (P/N 11301CE1) insert into the valve connection by pressing it gently.
Installing the Tube Holder
Push the tube holder onto the Flexible Intubation Video Endoscope from distal until it slides into place.
Installation and Operating
Instructions
16
w

Handling the haft
CA TION: o not carry the Flexible Intubation Video Endoscope by the shaft. o not pull, clamp or twist the shaft
of the Flexible Intubation Video Endoscope.
CA TION: The shaft of the Flexible Intubation Video Endoscopes must never be kinked or coiled up too tightly. It is
best to store the Flexible Intubation Video Endoscope in a suspended position or to keep it sterilized in a tray
specifically designed for sterilizing and storage.
Warning: o not apply lubricant to the lens, to avoid clouding the image and thus posing a risk to the patient.
Use a suitable water-soluble lubricant (e.g. Surgilube) for the distal section of the shaft.
17
Installation and Operating
Instructions

Handling the Deflection Mechanism
Note: The neutral position of the distal tip is straight, without any deflection whatsoever. An audible click is heard to
indicate the distal tip is in neutral position.
Warning: The distal tip of the Flexible Intubation Video Endoscope must be straight, not angled, when the Flexible
Intubation Video Endoscope is inserted into the patient.
Warning: The deflection lever must never be operated with force or abruptly to avoid perforation and injury to the
patient or damaging the device irreparably..
Warning: Before removing the Flexible Intubation Video Endoscope from the patient, the distal tip must be put into
the neutral position, as otherwise injuries to the patient or damage to the instrument may occur.
Warning: If the distal tip can no longer be properly controlled (sticking etc.) via the deflection lever, or if the
deflection mechanism develops any other fault, the Flexible Intubation Video Endoscope must not be used.
Warning: If a functional defect occurs with the Flexible Intubation Video Endoscope during use on the patient,
discontinue use immediately. Put the distal tip of the Flexible Intubation Video Endoscope into the neutral, non-
deflected position and remove the Flexible Intubation Video Endoscope slowly and carefully from the patient..
CA TION: The resistance that can be felt when deflecting the tip using the deflection lever increases as the angle
of deflection increases. Actuate the deflection gently and carefully, without any sudden or jerky movements, as far as
the limit stop.
The distal tip is deflected by actuating the deflection lever slowly. The distal end will deflect in accordance with the
direction of actuation (forward or backward).
Installation and Operating
Instructions
18

Handling the Working Channel
o not insert cleaning brushes through the channels forcefully. This can damage the cladding, especially if the tip of
the instrument is angled.
When connecting and operating light sources and suction/irrigation devices, pay attention to the instruction manuals
for the particular devices. The irrigation devices recommended by KARL STORZ are set to a maximum pressure of 54
kPa (400 mmHg/0.54 bar/7.8 psi).
Installation and Operating
Instructions
19

Connection to the Monitor 8403 ZX
The Flexible Intubation Video Endoscope can be used together with the KARL STORZ MONITOR 8403 ZX.
Warning: It must be ensured that the electrically non-insulated parts of the Flexible Intubation Video Endoscope do
not come into contact with conductive surfaces, or voltage-carrying elements of other units, at any time.
Insert the connecting cord qof the Flexible Intubation Video Endoscope into either connection socket sor don the
monitor.
The connecting cord qcan be connected or disconnected during operation. Insert the cable from the power supply
into the connection socket fof the monitor and connect the power supply with the mains socket. The power cord
can be fixed to the connection using the union nut.
The battery is charged automatically as soon as the monitor is connected to the power supply. This will be indicated
by the lighting up of the orange colored LE ;.
Use of the HDMI Output
The device features a port for an external monitor. An H MI- VI cable can be used to send the video signal to an
external monitor.
CA TION: The C-MAC®Monitor must be switched off when an external monitor is connected.
Note: The connected monitor must support a resolution of 1024 × 768 or 1280 × 720.
Note: To connect the C-MAC®Monitor to a KARL STORZ monitor, the H MI- VI adaptor with article number
20 9190 10 is required.
Installation and Operating
Instructions
20
fd
s
fd
This manual suits for next models
1
Table of contents
Other Storz Analytical Instrument manuals

Storz
Storz OR1 .avm User manual

Storz
Storz 11272 VN User manual

Storz
Storz VIDEOSCOPE User manual

Storz
Storz 11272 VA User manual

Storz
Storz 11302 BDX User manual

Storz
Storz C-MAC S User manual

Storz
Storz OR1 FUSION WO300 User manual

Storz
Storz C-MAC 8401 Series User manual

Storz
Storz 112 Series User manual