Introduction User / Patient Safety
Accessories
This Instructions For Use manual is the most comprehensive source of
information for the safe and effective use of your product. This manual may be
used by surgeons, dentists, veterinarians, or other healthcare professionals.
Keep and consult this reference manual during the life of the product.
The following conventions are used in this manual:
A WARNING highlights a safety-related issue. ALWAYS comply with this
information to prevent patient and/or healthcare staff injury.
A CAUTION highlights a product reliability issue. ALWAYS comply with this
information to prevent product damage.
A NOTE supplements and/or clarifies procedural information.
If additional information, especially safety information, or in-service training is
required, contact your Enova sales representative or call Enova customer
service.
Indications for Use
LED Surgical Headlight Systems are adjunct light sources used for illumination
during clinical or surgical procedures in a medical setting.
Contraindications
None known.
Limitations of Use
None known.
For Use With
Rechargeable PP4 Power Pack and the Quantum Twist Cable.
See Accessories section.
This section describes components that may be ordered to replace original
equipment that is damaged, worn, or must be replaced.
WARNINGS:
Use only Enova-approved components and accessories, including power
packs, unless otherwise specified.
DO NOT modify any component or accessory, unless otherwise specified.
The following Enova-approved accessories are sold separately:
NOTE: For a complete list of accessories, contact your Enova's
sales representative or call Enova's customer service.
WWARNINGS:
ARNINGS:
Prior to operation, users of Enova LED Surgical Headlight Systems should
read and understand the operation manual for all equipment used in
conjunction with the LED surgical headlight. Contact or your local distributor
for service and warranty information.
DO NOT look directly into the LED surgical headlight when illuminated. DO
NOT shine the light directly into the patient eyes or face. Due to the intensity
of light, eye injury may result. Patient's eyes should be appropriately
protected in surgeries involving the face or head.
After a few minutes of operation, the headlamp housing may become warm to
the touch. This is normal and functions to dissipate heat from the LED. Always
use a surgical glove, or sterile cloth to adjust the angle of the lens when in use.
Use only Enova's SUPPLIED BASE STATION AND POWER PACKS. The
use of other base stations or power packs may result in fire, explosion, or
damage to the power packs or LED surgical headlight and will void warranty.
The power pack must be powered off when charging. DO NOT USE OR
STORE BASE STATION IN OPERATING ROOM OR SURGERY SUITE.
All Enova's LED surgical headlights are designed to be wireless. DO NOT USE
THE LED SURGICAL HEADLIGHT WHILE POWER PACKS ARE CHARGING.
DO NOT use the LED surgical headlight, power packs or base station in
explosive gas environments.
DO NOT autoclave, immerse or submerge in liquid, spray with liquids or leave
the LED surgical headlight "on" when inside the composite carrying case as it
will cause damage to the LED surgical headlight or may cause a fire.
Please consult local regulations for the disposal of electronic products.
DO NOT INTERCHANGE TWIST CABLES. Use specified twist cable for
correct LED surgical headlight otherwise, the LED surgical headlight will
function abnormally and may damage the circuitry causing the light to fail.
Trademarks not the property of Enova's Corporation are the property of
their respective owners.
NOTE: The user and/or patient should report any serious product-related
incident to both the manufacturer and the Competent Authority of the European
Member State where the user and/or patient is established.
NOTE: The expected service life of the headlight is 5 years.
PP4 Power Pack
Deluxe Comfort Pad Set
PP4 Power Pack Charging Base Station
Gray Twist Cable
NA C-13 Power Cord
Battery Belt
Composite Carrying Case
DESCRIPTION REF
80201-001
80403-001
80100-001
80302-001
80150-001
80500-001
80034-001