4 DePuy Synthes Battery Power Line II Surgical Technique
Safety instructions
The surgeon has to evaluate if the machine is suitable
for an application, based on power limitation of the
machine, attachment and cutting tool regarding bone
strength/anatomical situation as well as handling of the
machine, attachment and cutting tool regarding bone
size. In addition, the contraindications of the implant
have to be respected. Please refer to the corresponding
surgical techniques of the implant system used.
The Battery Power Line II system is only to be used for
patient treatment after careful consultation of the in-
structions for use. It is recommended that an alternative
system is available to use during application, as technical
problems can never be completely ruled out.
The Battery Power LineII is designed for use only by
physicians and trained medical personnel.
DO NOT use any apparently damaged components.
DO NOT use any component if the packaging is dam-
aged.
DO NOT use this equipment in the presence of oxygen,
nitrous oxide or a mixture consisting of ammable anes-
thetic and air.
To ensure proper operation of the tool, only use Synthes
original accessories.
Before rst and every use and prior returning for service
power tools and their accessories/attachments, excluding
the battery, have to be run through the complete repro-
cessing procedure. Protective covers and lms must be
fully removed before sterilization.
For the tool to function properly, Synthes recommends
that it is cleaned, disinfected and serviced after each use
in accordance with the process dened in the “Care and
Maintenance” section. Compliance with these specica-
tions can considerably extend the service life of the tool
and reduce the risk of malfunction or harm to the user
and patient. Only use Synthes Special Oil (519.970) to lu-
bricate the tool.
Introduction
General Information
We recommend using new Synthes cutting tools for
every surgical procedure. Efciently working cutting
tools are the basis for successful surgery. Therefore,
check used cutting tools after every use for wear and/or
damage and replace them if necessary.
Cutting tools must be cooled with irrigation uid to
prevent heat necrosis.
The user of the product is responsible for proper use of
the equipment during surgery.
Check proper operation of the tool before using it on
the patient.
Unusual Transmissible Pathogens
Surgical patients identied as at risk for Creutzfeldt-
Jakob disease (CJD) and related infections should be
treated with single-use instruments. Dispose of the
instruments used or suspected of use on a patient with
CJD after surgery and/or follow the current national
recommendations.
To prevent overheating, always respect the specied
duty cycles on page 76. High torque of the powerful
Reamer/Drill (530.705) must always be observed.
For important information regarding electromagnetic
compatibility (EMC) please refer to the “Electromagnetic
Compatibility” section in this manual.
The tool is classied as type BF against electrical shock
and leakage current. The tool is suitable for use on
patients in accordance with IEC 60601-1.
Servicing
This system requires regular maintenance service, at least
once a year, in order to maintain its functionality. This
service has to be performed by the original manufacturer
or an authorized site.
The manufacturer assumes no responsibility for damage
resulting from improper operation, neglected or unau-
thorized maintenance of the tool.