manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Tektronix
  6. •
  7. Test Equipment
  8. •
  9. Tektronix 321 A User manual

Tektronix 321 A User manual

INSTRUCTION
MAN
UAL
TYPE
321A
OSCILLOSCOPE
Tektronix,
Inc.
S.W.
Millikan
Way
@
P.
O.
Box
500
@
Beaverton,
Oregon
97005
@
Phone
644-0161
@
Cables:
Tektronix
070-425
564
mentioned
above
should
be
taken
up
your
Tektronix
Field
Engineer.
ktronix
repair
and
replacement-part
is
geared
directly
to
the
field,
there-
p
all
requests
for
repairs
and
replace-
nt
parts
should
be
directed
to
the
Tek-
nix
Field
Office-or
Representative
in
your
.
This
procedure
will
assure
you
the
t
possible
service.
Please
include.
the
rument
Type
and
Serial
number
with
all
sts
for
parts
or
service.
ifications
and
price
change
priv-
reserved.
erton,
Oregon.
Printed
in
the
United
s
of
America.
All
rights
reserved.
tents
of
"any
form
wi
may
not
be
re-
luced
in
any
form:
without
permission
ie
copyright
owner.
INTeNsiTY
Focus
re
Ry
SCALE
Hive
ASTIGMATISM
——
HORIZONTAL
POSITION
VERTICAL
POSITION
*
A
POWER
on
..
VERTICAL
YARAne
TIME
AMPLIFIER
YOUS/OIV
Oa
THICKE
Co
FULL
VAMIAgLE
i
Aes
row
SATTeRES
x
TYPE
321A
srABKITY!
Ctinoscore]
TRIGGERING
Lever
tiny
ac
SLOPE
The
Type
321A
Oxcilloscope
Type
321A
Type
321A
SECTION
1
CHARACTERISTICS
Introduction
The
Tektronix
Type
321A
is
a
high-performance,
de-to-6
mc,
transistorized
oscilloscope.
Its
light
weight,
small
size
and
ability
to
operate
from
a
variety
of
power
sources
make
it
a
versatile
field
and
laboratory
instrument.
The
oscillo-
scope
can
operate
from
its
internally-contained
rechargeable
battery
pack,
an
external
de
source
or
from
a
115/230-volt
50-800
cycle
ac
line.
Regulated
power
supplies
in
the instru-
ment,
accurate
calibration,
and
precise
linearity
assure
exact
time
and
amplitude
measurements
despite
normal
voltage-
source
and
power-supply-load
changes
that
occur
under
actual
operating
conditions.
Operating
temperature
range derived
from
tests
indicates
optimum
performance
and
reliability on
its
self-contained
batteries
from
0°
to
+40°
C
at
altitudes
up
to
15,000
feet.
Temperature
range
without
batteries
when
operating
from
an
external
source
is
—-15°
C
to
+55°
C.
Non-operating
tem-
perature
range
is
—55°
C
to
+75°
C
without
batteries
and
—40°
C
to
+60°
C
with
batteries
at
altitudes
to
50,000
feet.
For
the
operator's
convenience,
a
front-panel
battery
light
indicates
when
the
internal
batteries
are
low.
If
external
de
or
ac
operation
is
being
used
instead
of
the
batteries,
the
light
turns
on
if
the
external
voltage
source
drops
too
low
for
proper
power
supply
regulation.
A
4-position
power
switch
on the
front
panel
permits
convenient
selection
of
charging
rate
and/or
power
source.
Vertical
Deflection
System
Bandpass—Dc
to
at
least
6
me
(3-db
down)
using
de
cou-
pling;
using
oc
coupling,
low-frequency
3-db
down
point
is
2cps
typical
from
a
1-ke
reference.
Sensitivity—0.01
v/div
to
20v/div
in
11
calibrated
steps;
accuracy
is
within
3%
of
front-panel
markings.
Con-
tinuously
variable
from
0.01
v/div
to
bout
50
v/div
uncalibrated.
Input
Impedance—35
pf
nominal
paralleled
by
1
megohm
{::1%),
8.2
pf
nominal
paralleled
by
10
megohms
(22%)
when
using
the
P6006
10X
Probe.
Maximum
Allowable
Input
Voltage
Rating—600
volts
com-
bined
de
and
peak
ac;
600
volts
(not
1200
volts)
peak-to-
peak
ac.
Triggering
Type—Automatic,
or
amplitude-level
selection
using
pre-
set
stability,
Mode—Ac-coupled
or
De-coupled.
Slope—Plus,
from
rising
slope
of
triggering
waveform,
or
minus
from
negative
slope
of
triggering
waveform.
Source—Internal
from
vertical
signal,
or
external
from
triggering
signal.
Signal
Requirements—Internal:
0.2
major
division
vertical
deflection
at
1
ke
increasing
to
1
major
division
at
6
me.
External:
1
volt
peak-to-peak
at
1
ke
increasing
to
3
volts
peak-to-peak
at
6mc.
Nominal
input
impedance:
5
pf
paralleled
by
100
kilohms
(-+-20%).
Sweep
Type—Miller
Integrator.
Sweep
Rates—0.5
ysec/div
to
0.5
sec/div
in
19
calibrated
steps.
Accurate
5X
sweep
magnifier
extends
calibrated
range
to 0.)
ysec/div,
Calibrated
sweep-rate
accuracy
is
-+3%.
Sweep
time
adjustable
between
steps
and
to
21.5
sec/div
uncalibrated.
External
Horizontal
Input
Bandpass—Dc
to at
least
1
me
(3-db
down).
Deflection
Factor—I
v/dv
+10%
with
5X
magnifier
on.
Input
Impedance—30
pf
typical
paralleled
by
100
kilohms
{5%).
Amplitude
Calibrator
Square
Wave—Frequency
about
2
kc.
Amplitude—500
mv
peak-to-peak.
Also
40
my
peak-to-
peak
internally
coupled
in
CAL
4
DIV
position
of
VOLTS/
DIV
switch.
Peak-to-peak
amplitude accuracy
is
+3%.
Cathode-Ray
Tube
Type—Special
Tektronix-manufactured
T3211,
3”
flat-face,
post-deflection
accelerator.
Low
heater
power.
Accelerating
Potential—4
kv.
Z-Axis
Modulation—External
terminal
permits
RC
coupling
to
crt
grid.
Unblanking—Deflection
unblanking.
Phosphor—Type
P31
normally
furnished;
Pl, P2, P7,
and
P11
phosphors
optional.
Other
phosphors
furnished
on
special
order.
Graticule
Ilumination—Variable
edge
lighting
when
operating
from
ac
line.
Display
Area—Marked
in
6-vertical
and
10-horizontal
1/4“
divisions.
Power
Requirements
Source—Operates
from
10
size
D
flashlight
cells,
or
10
size
D
rechargeable
cells
(approximately
3
hours
using
2.5
ampere-hour
cells;
approximately
5
hours
using
4
Characteristics
—-
Type
321A
ampere-hour
cells),
or
11.5
to
35 volts
de
(aircraft,
auto,
boat,
etc,),
or
103.5
to
126.5
volts
or
207
to
253
volts,
rms, 50
to
800
cycles,
single-phase
ac.
Power
Consumption—Approximately
700
ma
from
internal
batteries
or
external
de
source;
20
watts
nominal
at
115-volt
ac
fine.
Temperature
Protection—-Thermal
cutout
switch
interrupts
power
if
ambient
temperature
exceeds
131°
F
(55°
C).
Built-in
battery
charger
is
standard
equipment.
Environmental
Capabilities
Vibration
(operating)-—0.025"
peak-to-peak,
10
to
55
to
lO
cps
in
1
minute
sweeps
(4
G's)
for
15
minutes
on
each
axis.
Three-minute
vibration
at
resonance
or
55
cps
on
each
axis.
Shock
(operating}—20
G's,
Y2
sine,
11-msec
duration.
Two
shocks
each
direction
along
each
of
the
three
major
axis:
bottom,
top,
left
side,
right
side,
front
and
rear.
Total
of
12
shocks.
Shock
{non-operating)—60
G's,
¥%
sine,
11-msec
duro-
tion.
One
shock
each
direction
along
each
of
the
three
major
axis;
total
of 6
shocks.
Humidity
[non-operating)—Meets
Mil-Std-202B,
method
106A
(except
freezing
and
vibration)
through
5
cycles
(120
hours).
Transit
(non-operating)—Meets
National
Safe
Transit
test
when
factory
packaged.
Vibration
for
one
hour
at
slightly
greater
than
one
G,
Eighteen-inch
drop
in
any
orientation,
Mechanical
Specifications
Construction—Aluminum
alloy
chassis
and
cabinet.
Finish—Anodized
panel,
blue
vinyl-finish
cabinet.
Dimensions—8'/,"
high,
5%,”
wide,
16”
deep
overall.
ACCESSORIES
Information
on
accessories
for
use
with
this
instrument
is
included
at
the
rear
of
the
mechanical
parts
list.
SECTION
2
PRELIMINARY
INSTRUCTIONS
Power
Requirements
The
regulated
power
supplies
in
the
Type
321A
will
operate
form
115-v
or
230-v
rms
ac
line,
from
an
external
de
source
(11.5
to
35
volts),
or
from
a
“battery
pack”
consisting
of
either
10
size
D
flashlight
cells or
10
size
D
rechargeable
cells.
Fuse
Data
Use
only
the
recommended
fuses
in
the
Type
321A
Oscil-
loscope. The
upper
fuse,
F621
(see
Fig.
2-1),
is
a
1.5-amp
3AG
Fast-Blo;
the
lower
fuse,
F601,
is
a
.25-amp
3AG
Fast-
Blo.
Neither
fuse
needs
to
be
changed
if
the
oscilloscope
is
converted
from
one
line
voltage
to
the
other
(I115v
and
230
v).
volt
operation,
so
these
connections
do
not
have
to
be
changed
when
converting
from
one
line
voltage
to
the
other.
When
wired
for
115-volt
operation,
terminals
1
and
2
are
joined
by
a
bare
bus
wire,
and
terminals
3
and
4
are
simi-
larly
joined,
as
shown
in
Fig.
2-2[a).
To
convert
to
230-
volt
operation,
remove
the
bare
bus
wires
between
these
terminals
and
substitute
a
single
connecting
wire
between
terminals
2
and
3,
as
shown
in
Fig.
2-2(b).
To
turn
on
the
Type
321A
when
the
power
cord
is
con-
nected
to
the
oscilloscope
power
connector
and
to
the ac
voltage
source,
set
the
POWER
switch
to
EXT
ON.
To
turn
off
the
oscilloscope,
set
the
POWER
switch
to
TRICKLE.
The
TRICKLE
position
can
be
regarded
as
the
normal
“off”
posi-
tion
for
the
instrument.
Fig.
2-1.
Location
of
fuses
and
Charger
switch
(left-side
view).
Ac
Operation
Unless
tagged
otherwise,
your
instrument
is
connected
at
the
factory
for
operation
at
103.5
to
126.5
volts,
50
to
800
cycles
ac
(115
volts
nominal),
However,
provisions
are
made
for
easy
conversion
to
operate
at
207
to
253
volts,
50
to
800
cycles
(230
volts
nominal).
The
power
transformer
T601
is
provided
with
split
input
windings
which
are
normally
connected
in
parallel
for
115-volt
operation,
but
which
can
be
connected
in
series
for
230-volt
operation.
The
primary
windings
are
marked
1, 2,
3,
and
4,
Terminals
1
and
3
are
connected
to
one
winding
and
terminals
2
and
4
are
connected
to
the
second
winding.
The
ac
input
leads
are
connected
to
terminals
1
and
4
for
both
115-volt
and
230-
®
To
115-v
AC
To
230-v
AC
Fig.
2-2.
(al
Transformer
connection
for
operation
from
103.5~
126.5
volt
ac
line;
(b)
connections
for
operation
from
207-253
volt
ae line.
As
long
as
the
ac
power
cord
is
connected
to
the
ac
line,
power
is
being
applied
to
the
power
transformer
T601,
ac-
rectifier
circuit,
and
graticule
lights
for
all
positions
of
the
POWER
switch.
Application
of
power
to
these
circuit
is
re-
quired
to
provide
battery-charger
operation
for
the
internal
batteries;
the
graticule
lights
provide
visual
indication
that
these
circuits
are
“on”.
Power
to
the
battery
charger
itself
is
controlled
by
a
Charger
switch
(see
Fig.
2-1).
If
the
switch
2-1
Preliminary
Instructions
—
Type
321A
is
set
to
LOW
or
HIGH,
the
ac
rectifier
is
connected
to
the
battery
charger
circuit
which,
in
turn,
provides
charging
cur-
rent
to
the
internal
rechargeable
batteries.
NOTE
If
dry
cells
are
used
instead
of
rechargeable
bat-
teries,
then
the
Charger
switch
must
be
set
to
DRY
CELLS
to
disconnect
the
charger
circuit.
For
further
information
about
battery
operation,
refer
to
the
topics
titled
“Battery
Operation”
and
‘'Battery
Charger"
appearing
in
this
section
of
the
manual.
If
the
Type
321A
is
being
operated
exclusively
from
the
ac
line
and
the
internal
batteries
are
removed,
the
Charger
switch
can
remain
in
the
DRY
CELLS
position
to
disconnect
the
charger
circuit.
If
you
prefer
to
completely
turn
off
all
power
to
the
Type
321A,
set
the
POWER
switch
to
TRICKLE
and
either
disconnect
the
power
cord from
the
ac
line
or
turn
off
the
power
at
a
wall
switch
(or
equiv-
alent}.
Battery
Operation
Operation
from
the
internal
battery source
can
be
accom-
plished
by
using:
1.
Ten
size
D
flashlight
cells
(approximately
'/,-hour
con-
tinuous
operation,
more
with
intermittent
operation),
or
Ey
Ten
size
D
Alkaline
cells
such
as
Eveready
E95,
Burgess
AL-2,
or
Mallory
MN-1300
(about
2'/;
hours
continuous
operation),
or
wo
.
Ten
size
D
nickel-cadmium
rechargeable
cells
(up
to
about
5
hours
continuous
operation,
depending
on
the
type
used).
The
nickel-cadmium
cells
are
the
most
practical
type
where
considerable
battery
operation
is
planned.
To
install
the
cells,
first
open
the
battery
cover
as
illus-
trated
in
Fig,
2-3,
Next,
install
the
batteries
by
following
the
procedure
given
in
Fig.
2-4.
CAUTION
Be
sure
to
observe
cell
polarity
indicated
on
the
battery
cover.
To
turn
on
the
Type
321A when
operating
from
the
internal
batteries,
set
the
POWER
switch
to
BATT
ON.
To
turn
off
the
oscilloscope,
set
the
POWER
switch
to
one
of
the
OFF
posi-
tions—FULL
or
TRICKLE.
These
OFF
positions
disconnect
the
batteries
from
the
oscilloscope
load
{10-volt
regulated
supply)
so
there
is
no
battery
drain.
To
charge
the
recharge-
able
batteries,
refer
to
the
next
topic
titled
“Battery
Charger”.
Loosen
thumb
screw
te
unlock
cover,
then
pull
cover
forward
to
remove.
Fig.
2-3.
Removing
the
battery cover
from
the
Type
321A
Oscilloscope.
Preliminary
Instructions
—
Type
321A
id
‘of
the
battery
Medak
Be
sure
to
observe
cell
polarity
as.
in-
dicated
on
the.
cover.
Fig.
2-4,
Procedure
for
installing
the
batteries.
Battery
Charger
As
mentioned
previously,
the
battery
charger
is
connected
to
the
internal
batteries
as
long
as
the
ac
power
cord
is
connected
to
an
ac
power
source
and
the
Charger
switch
{see
Fig.
2-1)
is
set
to
HIGH
or
LOW.
No
charging
occurs
if
the
Charger
switch
is
set to
DRY
CELLS.
Table
2-1
summarizes
the
charging
currents
that
can
be
obtained
by
using
various
line
voltages
in
combination
with
various
settings
of
the
POWER
and
Charger
switches.
The
table
shows
the
Type
321A
wired
for
115-volt
nominal
line
operation,
If
the
oscilloscope
is
wired
for
230-volt
nominal
line
operation,
the
charging
currents
will
still
be
the
same
but
the
ac
line
voltages
will
be
twice
the
amount
shown
in
the
table.
Use
Table
2-1
as
an
aid
in
determining
the
proper
posi-
tion
for
the
POWER
and
Charger
switches
for
the
particular
brand
or
type
of
rechargeable
batteries
being
used
in
the
Type
321A.
After
setting
the
switches
to
their
proper
posi-
tion,
the
line
voltage
can
then
be
set
to
the
charging
rate
recommended
by
the
manufacturer
of
the
cells.
An
auto-
transformer
having
a
rating
of
at
least
1
ampere
and
equipped
with
an
rms-reading
ac
voltmeter
can
be
used
to
set
the
line
voltage.
When
using
rechargeable
batteries,
sixteen
hours
of
charging
at
the
full
rated
charging
current
recommended
by
the
manufacturers
should
be
adequate
to
fully
charge
the
batteries.
Excess
charging
may
damage
the
batteries,
One
methed
for
determining
charge
conditions
of
the bat-
teries
is
to
set
the
POWER
switch
to
BATT
ON
and
then
measure
the
voltage
across
the
battery
terminals
(see
Fig,
2-4).
A
reading
higher
than
13
volts
indicates
that
the
batteries
are
fully
charged.
A
reading
lower
than
13
volts
indicates
that
more
charging
time
is
required.
As
mentioned
previously,
if
dry
cells
are installed
in
the
battery
holder
and
the
Type
321A
is
being
operated
from
the
ac
line,
the
Charger
switch
must
be
set
to
the
DRY
CELLS
position,
This
position
disconnects
the
charging
circuit
from
the
batteries.
TABLE
2-1
Charging
Currents
POWER*|
Charger
Aproximate
Charging
Switch
|
Switch
Current
in
Ma
Position
|
Position
|
103.5
v
109
v
115
v 121
v/126.5v
Ton
|_tOW
|
20.
2a
|
7]
1
8
EXT
ON
|—
ae
eennlan
ae
HIGH
|
229630
[33
|
35
TRICKLE
Low
|
31
34
|
38
4]
44
|
HIGH
|
3437
aaa
it
Low
|”
166
["
180
|
200220330”
HIGH
|""290_|
320
|
360
380” “aio
"The
BATT
ON
position
is
not
included
in
the
table
because
the
Type
321A
should
not
be
operated
from
the
ac
line
and
the
batteries
at
the
same
time,
Reoson:
Battery
drain
exceeds
charg-
ing
rote.
2-3
Preliminary
Instructions
—
Type
321A
De
Operation
Operation
from
an
external
de
source
is
acomplished
by
connecting
the
special
pigtail-type
de
power
cord
in
the
proper
manner.
For 11.5-
to
20-volt
operation,
the
black
(+]
and
white
(—)
leads
are
connected
to
the
voltage
source;
for
20-
to
35-volt
operation,
the
green
(+)
and
white
{--)
leads
are
connected
to
the
voltage
source.
When
the
connections
to
the
external
de
voltage
source
are
properly
made,
the
external
de
source
is
floating
with
respect
to
the
Type
321A
chassis.
Up
to
600
volts
difference
is
permissi-
ble,
if
necessary.
To turn
on
the
Type
321A,
set
the
POWER
switch
to
EXT
ON.
To
turn
off
the
oscilloscope,
set
the
POWER
switch
to
‘one
of
the
OFF
positions—TRICKLE
or
FULL.
When
operating
from
the
external
de
source,
the ac
line
cord
should
be
disconnected,
That
is,
only
one
of
the
exter-
nal
sources
should
be
used
rather
than both
at
the
same
time.
2-4
NOTE
The
internal
battery
charger
circuit
is
disconnected
during
external
dc
operation.
Therefore,
external
batteries
(if
used
as
the
dc
source)
cannot
be
charged
by
connecting
the
line
cord
to
the
ac
line.
Also,
an
external
de
source
cannot
be
used
to
charge
the
internal
batteries
since
the
POWER
switch
does
not
electrically
connect
the
two
sources
together.
LOW
BATTERIES
Light
The
LOW
BATTERIES
light
turns on
when
the
following
conditions
exist:
1.
The
POWER
switch
is
set
to
BATT
ON
and
the
internal
batteries
drop
to
11.5
v
(0.2
v)
or
lower.
2.
The
POWER
switch
is
set
to
EXT
ON
and
the
external
source
voltage
is
low
enough
to
cause
the
Type
321A
un-
regulated
voltage
to
drop
to
about
11.5v
or
lower.
SECTION
3
OPERATING
INSTRUCTIONS
General
Information
The
Type
321A
Oscilloscope
is
an
extremely
versatile
in-
strument,
adaptable
to
a
great
number
of
applications.
How-
ever,
to
make
full
use
of
the
instrument,
it
is
necessary
that
you
understand
completely
the
operation
of
each
front-panel
control.
This
portion
of
the
manual
is
intended
to
provide
you
with
the
basic
information
you
require.
If
you
are
familiar
with
other
Tektronix
oscilloscopes,
you
should
have
very
little
difficulty
in
understanding
the
operation
of
the
Type
321A.
The
function
of
many
controls
is
the
same
as
the
function
of
corresponding
controls
on
other
Tektronix
instruments.
A
front-panel
view
of
the
Type
321A
is
shown
in
Fig.
3-1.
Intensity
Control
The
INTENSITY
control
is
used
to
adjust
the
trace
bright-
ness.
This
permits
adjustment
of
trace
intensity
to
suit
the
ambient
light
conditions
and
changes
in
intensity
caused
by
changes
in
the
sweep
triggering
rate
(sweep
duty
cycle).
Clockwise
rotation
increases
the
intensity
and counterclack-
wise
rotation
decreases
the
intensity.
Focus
and
Astigmatism
Controls
The
FOCUS
and
ASTIGMATISM
controls
operate
in
con-
junction
with
each
other
to
allow
you
to
obtain
a
sharp,
clearly
defined
spot
or
trace.
To
adjust
these
controls:
1.
Adjust
the
INTENSITY
contral
for
the
most
pleasing
level.
2.
Set
the
ASTIGMATISM
control
to
midscale.
3.
Adjust
the
FOCUS
control
for
sharpest
det:
4,
Adjust
the
ASTIGMATISM
control
as
necessary
for
best
overall
focus.
Graticule
Illumination
Control
The graticule
used
with
the
Type
321A
is
accurately
marked
with
10
horizontal
and
6
vertical
divisions,
with
0.2-
division
markers
on
the
centerlines,
These
graticule
markings
allow
you
to
obtain
time
and
voltage
measurements
from
the
oscilloscope
screen.
Graticule
illumination
is
adjusted
by
the
SCALE
ILLUM
control,
located
just
to
the
right
of
the
oscilloscope
screen.
Rotating
the
control
clockwise
increases
the
brightness
of
the
graticule
markings
and
counterclockwise
rotation
decreases
the
brightness.
NOTE
The
graticule
is
illuminated
only
when
operating
from
an ac
line.
This
permits
longer
operation
when
on
batteries,
Positioning
Controls
Two
controls
are
used
with
the
Type
321A
Oscilloscope
to
position
the
trace
or
spot
on
the
screen.
The
HORIZONTAL
POSITION
control
moves
the
trace
to
the
right
when
it
is
rotated
clockwise and
to
the
left
when
it
is
rotated
counterclockwise.
This
control
has
a
positioning
range
of
approximately
15
divisions
with
the
sweep
magnifier
off,
and
approximately
75
divisions
with
the
sweep
magnifier
on.
The
HORIZONTAL
POSITION
control
is
a
combination
coorse/vernier
type
of
control.
Built-in
blacklash
between
its
two
sections
permits
30°
of
vernier
adjustment
for
a
given
coarse
setting.
If
a
30°
range
is
exceeded,
the
coarse
ad-
justment
takes
over
to
provide
fast
positioning
of
the
trace.
The
VERTICAL
POSITION
control
has
sufficient
range
to
position
the
trace
completely
off
the
top
or
bottom
of
the
screen,
or
to
any
intermediate
point.
The
trace
moves
up
when
the
control
is
rotated
clockwise
and
down
with
the
counterclockwise
rotation.
Intensity
Modulation
The
crt
display
of
the
Type
321A
Oscilloscope
can
be
in-
tensity
modulated
by
an
external
signal
to
display
additional
information.
This
is
accomplished
by
disconnecting
the
grounding
bar
from
the
CRT
GRID
connector
at
the
rear
of
the
instrument
and
connecting
the
external
signal
to
this
terminal.
A
negotive
signal
of
approximately
30
volts
peak
is
required
to
cut
off
the
beam
from
maximum
intensity,
less
with
lower
intensity
levels.
Negative-going
signals
as
low
as
5
volts
peak
will
accomplish
intensity
modulation.
HORIZONTAL
DEFLECTION
SYSTEM
Horizontal
Sweep
The
usual
oscilloscope
display
is
@
graphical
presentation
of
instantaneous
voltage
versus
time,
Voltage
is
represented
by
vertical
deflection
of
the
trace
and
time
is
represented
by
horizontal
deflection.
To
obtain
a
useful
display,
the
spot
formed
by the
electron
beam
is
deflected
horizontally
at
a
known
rate,
so
that
any
horizontal
distance
on
the
screen
represents
a
definite
known
period
of
time.
The
trace
formed
by
the
deflection
of
the
spot
across
the
screen
is
known
as
the
horizontal
sweep,
Since
the
horizontal
deflection
of
the
spot
bears
a
definite
relationship
to
time,
and
provides
the
means
for
making
time
measurements
from
the
screen,
the
horizontal
sweep
is
also
known
as
the
time
base.
(See
Fig.
3-2),
The
rate
at
which
the
spot
is
deflected
across
the
screen
is
accurately
controlled
by
the
setting
of
the
TIME/DIV
con-
trol.
The
setting
of
the
TIME/DIV
control
determines
the
sweep
rate,
and
thus
the
number
of
cycles
displayed
on
the
ert
screen,
The control
is
set
to
display
the
portion
of
the
waveform
you
wish
to
observe.
3-1
Operating
Instructions
—
Type
321A
CRT
CONTROLS
FOCUS
~-
Controls
sharpness
of
INTENSITY.Controls
brightness
of
spot
or
trace. trace.
ASTIGMATISM—Used
in
conjune-
tion
with
FOCUS
to
obtain
overall
focus.
SCALE
ILLUM—Adjusts
brightness
of
graticule
markings
(when
oper-
ating
from
AC
line).
|
HORIZONTAL
CONTROLS
HORIZONTAL
POSITION—Coarse-
vernier
type
of
control
that
posi-
tions
trace
horizontally.
VERTICAL
CONTROLS
=|
=
POWER—Switch
turns
regulated
10-volt
power
on
and
off.
Also,
ects
charging
rate.
TIME/DIV
and
VARIABLE—Selects
sweep
rate
and
external
horizontal
input.
EXT
HORIZ
INPUT-——Terminal
for
accepting
external
horizontal
sig-
nal.
VERTICAL
POSITION
—
Positions
VERTICAL.
trace vertically.
AMPIRER
AC-DC-GND—Selects
either
AC
or
DC
input
coupling.
The
GND
posi-
j
tion
connects
the
Vertical
Amplifier
to
ground
but
does
not
ground
the
input
signal.
YOLTS/DIV
and
VARIABLE
—
Se.
lects
vertical
deflection
factor
and
calibrator
signal.
TRIGGERING
CONTROLS
LEVEL—Selects
point
on
trigg
ing
i
|
at
which
sweep
is
tri
I.
INPUT
——
Terminal
for
accepting
signal
at
which
sweep
is
triggered.
waveforms
to
be
displayed
on
ert.
SLOPE—Determines
whether
sweep
is
triggered
on
++ or
—
slope
of
triggering
signal.
CAL
OUT
$00
MV—Terminol
pro-
vides
500-my
square
wave
for
compensating
probe.
AC-DC—Selects
AC
or
DC
coupling
for
triggering
signal.
.
INT-EXT—Selects
either
internal
or
DC
BAL—Potentiometer
for
setting
external
triggering
signal.
de
balance
of
Vertical
Amplifier.
STABILITY—Potentiometor
for
sot
ting de level
of
sweep
generator.
INPUT—Terminal
for
accepting
ex-
ternal
triggering
signal.
Fig.
3-1.
Functions
of
the
Type
321A
Oscilloscope
front-panel
controls.
3-2
Operating
Instructions
—
Type
321A
Voltage
FS
7
G
Time
—$—$
$$»
Fig.
3-2.
The
oscilloscope
plots
instantaneous
voltage
versus
time,
thereby
serving
both
as
a
voltm
and
a
timer.
The
Time
Base
has
19
accurately
calibrated
sweep
rates
ranging
from
Sysec/div
to
Ssec/div.
These
calibrated
sweep
rates
are
obtained
when
the
VARIABLE
(TIME/DIV)
control
is
fully
clockwise
in
the
CALIB
position.
The
TIME/
DIV
switch
selects
the
calibrated
sweep
rates
and
can
be
rotated
360°
since
there
are
no
mechanical
stops,
The
VARI-
ABLE
control
permits
you
to
vary
the
sweep
rate
continuously
between
.Sysec/div
and
approximately
1.5sec/div.
All
sweep
rates
obtained
with
the
VARIABLE
control
in
any
position
other
than
fully
clockwise
are
uncalibrated,
Sweep
Magnifier
The
sweep
magnifier
allows
you
to
expand
any
two-
division
portion
of
the
displayed
waveform
to
the
full
ten-
division
width
of
the
graticule.
This
is
accomplished
by
first
using
the
HORIZONTAL
POSITION
control
to
move
the
por-
tion
of
the
display
you
wish
to
expand
to
the
center
of
the
graticule,
then
placing
the
5
MAG
switch
in
the
"on"
posi-
tion
(pull
out
the red
VARIABLE
TIME/DIV
knob;
see
Fig.
3-3).
Any
portion
of
the
display
magnified
by
the
horizontal
sweep
can
then
be
observed
by
rotating
the
HORIZONTAL
POSITION
control.
In
magnified
sweep
operation,
the
sweep
rate
indicated
by
the
position
of
the
TIME/DIV
switch
must
be
divided
by
5
to
obtain
the
actual
time
required
for
the
spot
to
move
‘one
division.
For
example,
if
the
TIME/DIV
switch
is
set
to
5
MILLI
SEC,
the
actual
time
per
division
is
5
milliseconds
divided
by
5,
or
1
millisecond
per
division.
The
actual
time-
per-division
must
be
used
for
all
tithe
measurements.
External
Horizontal
Input
For
special
applications
you
can
deflect
the
trace
hori-
zontally
with
some
externally
derived
waveform
rather
than
by
means
of
the
internal
sweep
sawtooth.
This
allows
you
to
use the
oscilloscope
to
plot
one
function
versus
another,
To
use
the
external
horizontal
input,
connect
the
externally
derived
waveform
to
the
EXT
HORIZ
INPUT
connector
and
place
the
TIME/DIV
switch
in
the
EXT
position.
The
hori-
zontal
deflection
factor
is
approximately
1
volt/division
with
the
5
MAG
on.
TIME/DIV
VARIABLE
Unmagnified
Woveform
VARIABLE
5x
MAG
ON
(pull)
|}
4+
Fig.
3-3.
Operotion
of
the
sweep
magnifier.
Sweep
Triggering
The
oscilloscope
display
is
formed
by
the
repetitive
sweep
of
the
spot
across
the
oscilloscope
screen.
If
the
sweeps
are
allowed
to
occur
at
random,
or
a
rate
unrelated
to
the
input
waveform,
the
displayed
waveform
will
be
traced
out
at
a
different
point
on
the
screen
with
each
sweep.
This
will
either
cause
the
waveform
to
drift
across
the
screen
or
to
be
indistinguishable.
In
most
cases
it
is
desirable
for
repetitive
waveforms
to
appear
stationary
on
the
oscilloscope
screen
so
that
the
characteristics
of
the
waveform
can
be
examined
in
detail.
As
a
necessary
condition
for
this
type
of
display,
the
start
of
the
sweep
must
bear
a
definite,
fixed-time
relationship
to
the
observed
waveform.
This
means
that
each
sweep
must
stort
at
the
same
time,
relative
to
some
point
on
the
observed
waveform.
In
the
Type
321A,
this
is
accomplished
by
start-
ing
or
triggering
the
sweep
with
the
displayed
waveform,
or
with
another
waveform
bearing
a
definite
time
relationship
to
the
displayed
waveform.
The
waveform
used
to
start
the
horizontal
sweep
is
called
a
“triggering
signal”
(whether
it
is
the
waveform
being
ob-
served,
or
some
other
waveform).
The
following
instructions
tell
you
how
to
select
the
triggering
signal
source.
Selecting
the
Triggering
Source
In
preparing
the
Type
321A
Oscilloscope
for
triggered
operation
of
the
sweep,
it
is
first
necessary
to
select
the
triggering
signal
source
which
will
provide
the
best
display
for
the
particular
application.
The
sweep
can
be
triggered
by the
displayed
waveform,
or
by an
externally
derived
waveform.
This
selection
is
made
by the
setting
of
the
INT-
EXT
switch
(see
Fig.
3-4).
Each
type
of
triggering
has
certain
advantages
for
some
applications.
Triggering
from
the
disployed
waveform
is
the
method
most
commonly
used.
The
displayed
waveform
is
selected
when
the
INT-EXT
switch
is
in
the
INT
position.
Internal
trig-
3-3
Operating
Instructions
—
Type
321A
gering
is
convenient
since no
external
triggering
connections
are
required.
Satisfactory
results
are
obtained
in
most
appli-
cations.
To
trigger
the
sweep
from
some
external
waveform,
con-
nect
the
triggering
waveform
to
the
{TRIGGERING)
INPUT
connector
and
place
the
INT-EXT
switch
in
the
EXT
position.
{External
triggering
provides
definite
advantages
over
in-
ternal
triggering
in
certain
cases,)
With
external
triggering,
the
triggering
signal
usually
remains
constant
in
amplitude
ond
shape,
It
is
thereby
possible
to
observe
the
shaping
and
amplification
of
a
signal
in
an
external
circuit
without
reset-
ting
the
oscilloscope
triggering
controls
for
each
observation.
Also,
time
and
phase
relationships
between
the
waveforms
at
different
points
in
the
circuit
can
be
seen.
If,
for
example,
the
external triggering
signal
is
derived
from
the
waveform
at
the
input
to
a
circuit,
the
time
relationship
and
phase
of
the
waveforms
ot
each
point
in
the
circuit
are
compared
to
the
input
signa!
by
the
display
presented
on
the
oscilloscope
screen.
Selecting
the
Triggering
Slope
The
horizontal
sweep
can
be
triggered
on
either
the
rising
or
falling
portion
of
the
triggering
waveform,
When
the
SLOPE
switch
is
in
the
4.
position,
the
sweep
is
triggered
on
the
rising
portion
of
the
triggering
waveform;
when
the
SLOPE
switch
is
in
the
—
position,
the
sweep
is
triggered
on
the
falling
portion
of
the
waveform
(see
Fig.
3-5).
In
many
applications
the
triggering
slope
is
not
important
since
triggering
on
either
slope
will
provide
a
display
suitable
to
the
application.
Selecting
the
Triggering
Mode
Automatic
Mode
Automatic
triggering
is
obtained
by
rotating
the
(TRIG-
GERING)
LEVEL
control
fully
counterclockwise
to
the
AUTO
position.
This
mode
allows
triggering
at
the
average
voltage
point
of
the
applied
waveform.
Also,
the
sweep
runs
at
approxi-
mately
a
50-cycle
rate
when
no
triggering
signals
are
ap-
plied;
this
produces
a
reference
trace
or
baseline
on
the
screen,
Automatic
triggering
can
be
used
with
both
internal
and
external
triggering
signals,
but
for
most
waveforms
it
is
useful
only
for
triggering
at
frequencies
above
50
cycles.
Automatic
triggering saves
considerable
time
in
observing
@
series
of
waveforms
since
it
is
not
necessary
to
reset
the
triggering
level
for
each
observation.
For
this
reason
it
is
the
mode
that
is
normally
used.
Other
modes
are
generally
used
only
for
special
applications,
or
where
stable
trigger
ing
is
not
attainable
in
the
automatic
mode,
Displayed
Waveform
External
Triggering
Waveform
INPUT
Switch
Oscilloscope
Sweop-Starting
eee
wits
Fig.
3-4
The
triggering
signal
is
selected
from
two
possible
sources
with
the
INT-EXT
switch.
3-4
Operating
Instructions
—
Type
321A
LeveL
LEVEL
a
+
Sweep
Triggers
-
On
+
Slope
Waveforms
obtained
with
the
trig-
geting
level
control
set
in
the
—
region.
“
SLOPE
\
> L
j
\
Sweep
Triggers
On
—
Slope
SLOPE
+
oP
/
Sweep
Triggers
on
+
Slope
SLOPE
Waveforms
obtained
with
the
trig-
gering
level
control
set
in
the
+
region.
N
(\
Meade
+
~
>
Sweep
Triggers
On
-~
Slope
Fig.
3-5.
Effects
on
the
oscilloscope
display
produced
by
+
and
—
settings
of
the
SLOPE
and
LEVEL
controls.
Operating
Instructions
—
Type
321A
Ac
Mode
Ac-mode
triggering
is
obtained
by
setting
the
AC-DC
switch
to
the
AC
position.
This
mode
provides
stable
trig-
gering
on
virtually
all
types
of
waveforms,
As
a
general
rule,
however,
the
ac
mode
is
unsatisfactory
for
triggering
with
low
amplitude
waveforms
at
frequencies
below
approxi-
mately
15
cycles,
This
figure
will
vary
depending
upon
the
amplitude
and
shape
of
the
triggering
waveform
and
should
not
therefore
be
set
as
an
absolute
standard.
Triggering
at
frequencies
below
15
cycles
can
be
accomplished
when
higher
amplitude
triggering
signals
are
used.
In
the
ac
mode,
the
triggering
point
depends
on
the
average
voltage
level
of
the
triggering
signals.
If
the
trig-
gering
signals
occur
at
random,
the
average
voltage
level
will
vary
causing
the
triggering
point
to
vary
also.
This
shift
of
the
triggering
point
may
be
enough
so
that
it
is
impossible
to
maintain
a
stable
display.
In
such
cases
you
should
use
the
de
mode,
Dc
Mode
De
mode
triggering
is
obtained
by
setting
the
AC-DC
switch
to
the
DC
position.
This
mode
of
triggering
is
por-
ticularly
useful
in
triggering
from
waveforms
which
are
not
adaptable
to
the ac
mode,
such
as
random
pulse
trains
or
very
low-frequency
waveforms.
Random
pulse
trains
pose
a
special
problem
in
the
ac
mode
since
the
random
occur-
rence
of
the
input
waveforms
causes
the
average
voltage
level
to shift.
This
in
turn
may
cause
the
triggering
level
to
shift
to
an
unstable
point.
This
problem
is
not
encountered
in
the de
mode
since
the
triggering
point
is
determined
only
by
instantaneous
voltages.
In
the
de
mode,
when
the
triggering
signal
is
obtained
from
the
Vertical
Amplifier,
varying
the
VERTICAL
POSITION
control
will
change
the
triggering
point.
For
this
reason,
you
may
find
it
necessary
to
readjust
the
LEVEL
control
when
you
change
the
vertical
position
of
the
trace.
To
eliminate
this
effect,
you
can
use
the
ac
mode
provided
the
triggering
signal
is
otherwise
suitable
for
this
mode
of
operation.
In
the de
mode,
the
de
level
of
the
external triggering
signals
will
also
effect
the
triggering
point.
Generally,
when
the
triggering
signa!
is
small
compared
to
its
de
level,
the
ac
mode
should
be
used.
How
to
Set the
Triggering
Level
In
the ac
and
de
triggering
modes,
the
LEVEL
control
determines
the
voltage
level
on
the
triggering
waveform
at
which
the
sweep
is
triggered.
Using
this
control,
the
sweep
can
be
continuously
triggered
at
any
point
on
the
waveform
so
long
as
the
slope
of
the
waveform
is
great
enough
to
provide
stable
triggering.
In
the
de
mode,
the
sweep
cannot
be
triggered
with
any
degree
of
stability
at
the
top
of
a
square
wave,
for
example,
because
the
time
that the
voltage
remains
constant
is
comparatively
long.
As
o
result,
the
sweep
triggers
at
random
points
along
the
top
of
the
square
wave,
producing
considerable
trace
jitter.
You can
use
the
same
method
to
set
the
LEVEL
control
for
either
the
ac
or
dc
mode.
After
selecting
the
triggering
slope,
rotate
the
LEVEL
control
fully
counterclockwise
to
the
AUTO
position.
Then
rotate
the
LEVEL
control
clockwise
until
the
sweep
no
longer
triggers.
Continue
to
rotate
the
3-6
control
in
the
clockwise
direction
until
the
sweep
again
trig-
gers
and
a
stable
display
is
obtained.
Further
rotation
of
the
control
in
the
clockwise
direction
causes
the
sweep
to
trigger
at
more
positive
points
on
the
triggering
waveform.
In
the
fully
clockwise
direction
the
trace
will
free
run
(Fig.
3-5).
FREE-RUNNING
OPERATION
With
the
Type
321A,
you
can
get
a
periodic,
free-running
sweep,
independent
of
any
external
triggering
or
synchroniz-
ing
signal,
by
rotating
the
LEVEL
control
fully
clockwise
to
the
FREE
RUN
position.
This
permits
you
to
observe
the
trace
without
an
input
signal.
This
trace
can
then
be
used
to
posi-
tion
the
sweep
or
to
establish
a
voltage
reference
line.
The
difference
between
the
traces
produced
in
the
AUTO
position
and
the
FREE
RUN
position
is
the
repetition
rate.
The
repeti-
tion
rate
in
the
FREE
RUN
position
is
dependent
upon
the
setting
of
the
timing
switch.
The
repetition
rate
in
the
AUTO
position
is
fixed
at
approximately
50
cycles.
At the
faster
sweep
rates,
the
trace
in
the
AUTO
position
will
appear
to
be
dim.
In
the
FREE
RUN
position
the
trace
intensity
remains
essentially
constant
for
all
sweep
rates.
VERTICAL
DEFLECTION
SYSTEM
Input
Coupling
Input
signals
to
the
Vertical
Amplifier
can
be
either
ac-
or
de-coupled
by
placing
the
AC-DC-GND
switch
in
the
ap-
propriate
AC
or
DC
position.
De
coupling
applies
both
the
ac
and
de
components
of
the
input
signal
to
the
vertical
amplifier
circuit.
This
permits
measurement
of
the
de
voltage
level
as
well
as
the
amplitude
of
the
ac
component,
It
is
sometimes
neither
necessary
nor
desirable
to
display
the dc
component,
however,
and
in
such
cases
as
coupling
should
be used.
This
is
accomplished
by
setting
the
AC-DC-GND
switch
to
AC.
With
oc
coupling,
a
capacitor
is
placed
in
series
with
the
input
connector
to
block
the
de
component
while
allowing
the
ac
component
to
be
displayed.
Placing
the
AC-DC-GND
switch
to
the
GND
position
grounds
the
input
circuit
of
the
vertical
amplifier
to
provide
a
de
zero
reference.
In
this
position
the
switch
internally
dis-
connects,
but
does
not
ground,
the
applied
signal
to
the
input
connector.
Thus,
the
GND
position
eliminates
the
usual
need
for
externally
grounding
the
(Vertical
Amplifier)
INPUT
connector
of
the
Type
321A
or
the
probe
tip
to
establish
a
ground
reference.
Deflection
Factor
The
electrical
waveform
to
be
observed
is
applied
to
the
(Vertical
Amplifier)
INPUT
connector.
The
waveform
is
then
applied through
the
vertical-deflection
system
to
cause
the
spot
to
be
deflected
vertically
lo
trace
out
the
waveform
on
the
screen
of
the
crt.
The
VOLTS/DIV
switch
controls
the
vertical
deflection
factor
in
accurately
calibrated
steps.
The
VARIABLE
control
provides
uncalibrated
variable
deflection
factors
between
the
fixed
steps
of
the
VOLTS/DIV
switch.
The
VARIABLE
control
has
360°
rotation
range
and
a
detent
position
when
the
control
is
set to
CALIB,
NOTE
To
make
the
deflection
factor
equal
to
that
indi-
cated
by
the
VOLTS/DIV
switch,
set
the
VARIABLE
control
to
the
CALIB
detent
position.
Dc
Balance
Adjustment
The
need
for
adjustment
of
the
DC
BAL
control
is
indi-
cated
by
a
vertical
shift
in
the
position
of
the
trace
as
the
VARIABLE
(VOLTS/DIV)
control
is
rotated,
This
adjustment
should
be
made
as
follows:
1.
Set
the
AC-DC-GND
switch
to
GND.
2.
Set
the
oscilloscope
controls
for
a
free-running
trace.
3.
Rotate
the
VARIABLE
(VOLTS/DIV)
control
back
and
forth,
and
adjust
the
DC
BAL
control
simultaneously
until
the
trace
position
is
no
longer
affected
by
rotation
of
the
VARIABLE
control.
Input
Signal
Connections
Certain
precautions
must
be
observed
when
you
are
con-
necting
the
oscilloscope
to
an
input
signal
source.
This
is
to
insure
that
accurate
information
is
obtained
from
the
oscilloscope
display.
This
is
particularly
true
when
you
are
observing
low-level
signals,
or
waveforms
containing
high-
or
extremely
low-frequency
components.
For
opplica-
tions
where
you
are
observing
low-level
signals,
shielded
cables should
be
used
whenever
possible,
with
the
shield
connected
to
the
chassis
of
both
the
oscilloscope
and
the
signal
source.
Unshielded
input
leads
are
generally
unsatis-
factory
due
to
their
tendency
to
pick
up
stary
signals
which
produce
erroneous
oscilloscope
displays.
Regardless
of
the
type
of
input
used,
the
leads
should
be
kept
as
short
as
possible.
Distortion
of
the
input
waveform
may
result
if:
Very
low-frequency
input
signals
are
ac-coupled
to
the
oscilloscope
n
.
High-frequency
waveforms
are
not
properly
terminated
»
.
The
input
waveform
contains
high-frequency
components
which
exceed
the
bandpass
of
the
oscilloscope.
You
must
be
aware
of
the
limitations
of
the
instrument.
In
analyzing
the
displayed
waveform,
you
must
consider
the
loading
effect
of
the
oscilloscope
on the
input
signal
source,
In
most
cases
this
loading
effect
is
negligible;
how-
ever,
in
some
applications
loading
caused
by the
oscilloscope
may
materially
alter
the
results
obtained.
In
such
cases
you
may
wish
to
reduce
the
amount
of
loading
to
a
negligible
amount
through
the
use
of
a
probe.
Use
of
Probes
Occasionally
connecting
the
input
of
an
oscilloscope
to
a
signal
source
loads
the
source
sufficiently
to
adversely
affect
both
the
operation
of
the
source
and
the
waveform
displayed
on
the
oscilloscope.
In
such
cases
an
attenuator
Operating
Instructions
—
Type
321A
probe
may
be
used
to
decrease
both
the
capacitive
and
resis-
tive
loading
caused
by
the
oscilloscope
to
a
negligible
value.
In
addition
to
providing
isolation
of
the
oscilloscope
from
the
signal
source,
an
attenuator
probe
also
decreases
the
amplitude
of
the
displayed
waveform
by
the
attenuction
factor
of
the
probe.
Use
of
the
probe
allows
you
to
increase
the
vertical-deflection
factors
of
the
oscilloscope
to
observe
large-amplitude
signals
beyond
the
normal
limits
of
the
oscilloscope.
Signal
amplitudes,
however,
must
be
limited
to
the
maximum
allowable
value
of
the
probe
used.
When
making amplitude
measurements
with
an
attenuator
probe,
be
sure
to
multiply
the
observed amplitude
by
the
attenua-
tion
of
the
probe.
If
the
waveform
being
displayed
has
rapidly
rising
or
falling
voltages,
it
is
generally necessary
to
clip
the
probe
ground
lead
to
the
chassis
of
the
equipment
being
tested.
Select
a
ground
point
near
the
point
of
measurement,
as
shown
in
Fig.
3-6,
Probe
ae
Fig.
3-6.
Connecting
a
probe
to
the
input
signal
source
Before
using
a
probe
you
must
check
(and
adjust
if
neces-
sary]
the
compensation
of
the
probe
to
prevent
distortion
of
the
applied
waveform.
The
probe
is
compensated
by
adjust-
ing
the
control
located
in
the
body
of
the
probe.
Adjustment
of
the
probe
compensates
for
variations
in
input
capacitance
from
one
instrument
to
another.
To
insure
the
accuracy
of
pulse
and
transient
measurements,
this
adjustment
should
be
checked
frequently.
To
adjust
the
probe
compensation,
set
the
VOLTS/DIV
con-
trol to
the
.01
position
and
the
LEVEL
control
to
the
AUTO
position.
Set the
SLOPE
switch
to
+
and
the
INT-EXT
switch
to
INT.
Connect
the
probe
tip to
the
CAL
OUT
500
MV
con-
nector.
Set
the
TIME/DIV
switch
to
.5
MILLI
SEC
and
adjust
the
probe
to
obtain
flat
tops
on
the
displayed
square
wave-
form
(see
Fig.
3-7.)
3-7
Operating
Instructions
—
Type
321A
Voltage
Measurements
The
Type
321A
Oscilloscope
can
be
used
to
measure
the
voltage
of
the
input
waveform
by
using
the
calibrated
ver-
tical-deflection
factors
of
the
oscilloscope.
The
method
used
for
all
voltage
measurements
is
basically
the
same
although
the
actual
techniques
vary
somewhat
depending
on
the
types
of
voltage
measurements,
i.e.,
a¢-component
volt-
age
measurements,
or
instantaneous
voltage
measurements
with
respect
to
some
reference
potential.
Many
waveforms
contain
both
ac
and
de
voltage
components,
and
it
is
often
necessary
to
measure
one
or
both
of
these
components.
When
making
voltage
measurements,
you
should
display
the
waveform
over
as
large
a
vertical
portion
of
the
screen
as
possible
for
maximum
accuracy.
Also,
it
is
important
that
you
do
not
include
the
width
of
the
trace
in
your
measure-
ments.
You
should
consistently
make
all
measurements
from
one
side
of
the
trace.
If
the
bottom
side
of
the
trace
is
used
for
one
reading,
it
should
be
used
for
all
succeeding
read-
ings.
The
VARIABLE
(VOLTS/DIV)
control
must
be
in
the
CALIB
detent
position.
Ac
Component
Voltage
Measurements
Te
measure
the ac
component
of
a
waveform,
the
AC-DC.
GND
switch
should
be
set
to
the
AC
position.
In
this
position
only
the
ac
components
of
the
input
waveform
are
displayed
on
the
oscilloscope
screen,
However,
when
the
ac
component
of
the
input
waveform
is
very
lew
in
frequency,
it
will
be
necessary
for
you
to
make
voltage
measurements
with
the
AC-DC-GND
switch
in
the
DC
pasition,
To
make
a
peak-to-peak
voltage
measurement
on
the
ac
component
of
a
waveform,
perform
the
foilewing
steps
(see
Fig.
3-8].
1,
With
the aid
of
the
graticule,
measure
the
vertical
dis-
tance
in
divisions
from
the
positive
peak
to
the
negative
peak,
2.
Multiply
the
setting
of
the
VOLTS/DIV
control
by
the
dis-
tance
measured
to
obtain
the
indicated
voltage.
3.
Multiply
the
indicated
voltage
by
the
attenuation
factor
of
the
probe
you
are
using
to
obtain
the
true
peak-to-peak
voltage.
As
an
example
of
this
method,
assume
that
using
the
P6006
Probe
and
a
deflection
factor
of
1
volt
per division,
you
measure
a
vertical
distance
between
peaks
of
4
divisions.
In
this
case,
then,
4
divisions
multiplied
by
1
volt
per
divi-
sion
gives
you
an
indicated
voltage
of
4
volts
peak-to-peak.
The
indicated
voltage
multiplied
by
the
probe's
attenuation
factor
of
10
then
gives
you
the
true
peak-to-peak amplitude
of
40
volts,
When
sinusoidal
waveforms
are
measured,
the
peak-to-
peak
voltage
obtained
can
be
converted
to
peak,
rms,
or
average
yoltage
through
use
of
standard
conversion
factors,
Instantaneous
Voltage
Measurements
The
method
used
to
measure
instantaneous
voltages
is
virtually
identical
to
the
method
described
previously
for
the
measurement
of
the ac
components
of
a
waveform.
How-
ever,
for
instantaneous
voltage
measurements
the
AC-DC-
GND
switch
must
be
placed
in
the
DC
position.
Also,
since
instantaneous
voltages
are
measured
with
respect
to
some
potential
[usually
ground],
a
reference
line
must
be
established
on the
oscilloscope
screen
which
corresponds
to
that
potential.
If,
for
example,
voltage
measurements
are
to
be
made
with
respect
to
+100
volts,
the
reference
line
would
correspond
to
+100
volts,
In
the
following
procedure
a
method
is
presented
for
establishing
this
reference
line
at
ground,
since
measurements
with
respect
to
ground
are
the
Adjusted
Correctly
Incorrect
ems
ice
t
and
carefully
tighten locking
Fig.
3-7.
The
probe
is
adjusted
to
obtain
an
undistored
presentation
of
the
calibrator
squarewave.
3-8
Operating
Instructions
—
Type
321A
Probe
Attenuation
Factor
|
x
|
VOLTS/DIV
Switch
Setting
|
x
|
Vertical
Deflection
=
Volts,
peak-to-pack
Fig.
3-8.
Measuring
the
peak-to-peak
ac
voltage
of
an
applied
waveform.
most
common
type.
The
same
general
method
may
be
used
to
measure
voltage
with
respect
to
any
other
potential,
how-
ever,
so
long
as
that
potential
is
used
to
establish
the
ref-
erence
line.
To
obtain
an
intantaneous
voltage
measurement
with
res-
pect
to
ground,
or
some
other
voltage,
perform
the
following
steps
(see
Fig.
3-9).
1.
To
establish
a
ground
reference
line,
set
the
AC-DC-
GND
switch
to
GND.
Or,
to
establish
a
reference
line
which
represents
a
voltage
other
than
ground,
touch
the
probe
tip
to
the
voltage
and
leave
the
AC-DC-GND
switch
at
DC,
Then
adjust
the
oscilloscope
controls
to
obtain
a
free-running
sweep.
Vertically
position
the
trace
to
a
convenient
point
on
the
oscilloscope
screen.
This
point
will
depend
on
the
polarity
and
amplitude
of
the
in-
put
signal,
but
should
always
be
chosen
so
that
the
trace
lies
along
one
of
the
major
divisions
of
the
graticule.
The graticule
division
corresponding
to
the
position
of
the
trace
is
the
voltage
reference
line
and
all
voltage
measurements
must
be
made
with
respect
to
this
line.
(Do
not
adjust
the
VERTICAL
POSITION
contro!
after
the
reference
line
has
been
established.}
.
If
ground
reference
was
established,
set
the
AC-DC-GND
switch
to
DC;
if
a
reference
line
other
than
ground
was
established,
remove
the
probe
tip
from
this
voltage
and
connect
it
to
the
signal
source.
Adjust
the
LEVEL
control
for
a
stable
display.
r
(usually
ground)
Instantaneous
|
Probe
Attenuation
Factor
x
|
VOLTS/DIV
Switch
Setting
| x
Vertical
Deflection
From
Reference
Lino
Fig. 3-9.
Measuring
1
instantaneous
voltage
with
respect
to
ground
(or
some
other
reference
voltage)
Operating
Instructions
—
Type
321A
3.
Measure
the
vertical
distance
in
divisions
from
the
desired
point
on
the
waveform
to
the
voltage
reference
line.
>
Multiply
the
setting
of
the
VOLTS/DIV
control
by the
dis-
tance
measured
to
obtain
the
indicated
voltage.
w
Multiply
the
indicated
voltage
by
the
attenuation
factor
of
the
probe
you
are using
to
obtain
the
actual
voltage
with
respect
to
ground
[or
other
reference
voltage).
As
an
example
of
this
method,
assume
you
are
using
the
P6006
Probe
and
a
deflection
factor
of
0.2
volt
per
division.
After
setting
the
voltage
reference
line
at
the
second
from
bottom
division
of
the
graticule,
you
measure
a
distance
of
3
divisions
to
the
point
you
wish
to
check.
In
this
case,
3
divisions
multiplied
by
0.2
volt
per
division
gives
you
an
indicated
0.6
volt.
Since
the
voltage
point
is
above
the
voltage
reference
line
the
polarity
is
indicated
to
be
posi
tive.
The
indicated
voltage
multiplied
by
the
probe
attenua-
tion
factor
of
10
then
gives
you
the
actual
voltage
of
+6
volts.
Time
Measurements
The
calibrated
sweep
of
the
Type
321A
Oscilloscope
causes
any
horizontal
distance
on
the
screen
to
represent
a
definite
known
interval
of
time.
Using
this
feature
you
can
accurately
measure
the
time
lapse
between
two events
displayed
on
the
oscilloscope
screen.
One
method
which
produces
suf-
ficient
accuracy
for
most
applications
is
as
follows
(see
Fig.
3-10).
1,
Measure
the
horizontal
distance
between
the
two
dis-
played
events
whose
time
interval
you
wish
to
find.
2.
Multiply
the
distance
measured
by
the
setting
of
the
TIME/DIV
control
to
obtain
the
apparent
time
interval.
(The
VARIABLE
TIME/DIV
control
must
in
the
CALIB
posi-
tion.)
NOTE
Divide
the
apparent
time
interval
by
5
if
the
mag-
nifier
is
on.
[
Appropriate
TIME/DIV
Switch
Setting
|
x
|
Horizontal
Distance
=
Time
Sweep
Magnification
Fig.
3-10.
Measuring
the
interval
between
events
displayed
on
the
oscilloscope
sert

Other manuals for 321 A

5

Other Tektronix Test Equipment manuals

Tektronix TDS 500D series Owner's manual

Tektronix

Tektronix TDS 500D series Owner's manual

Tektronix TAS 465 User manual

Tektronix

Tektronix TAS 465 User manual

Tektronix MDO3000 Series User manual

Tektronix

Tektronix MDO3000 Series User manual

Tektronix WFM6100 Opt. MB Use and care manual

Tektronix

Tektronix WFM6100 Opt. MB Use and care manual

Tektronix A605 User manual

Tektronix

Tektronix A605 User manual

Tektronix 222 User manual

Tektronix

Tektronix 222 User manual

Tektronix TDS7000 Series Manual

Tektronix

Tektronix TDS7000 Series Manual

Tektronix TPS2000B Series User manual

Tektronix

Tektronix TPS2000B Series User manual

Tektronix Keithley S500 User manual

Tektronix

Tektronix Keithley S500 User manual

Tektronix THDP0100 User manual

Tektronix

Tektronix THDP0100 User manual

Tektronix 2247A User manual

Tektronix

Tektronix 2247A User manual

Tektronix TEK 222 User manual

Tektronix

Tektronix TEK 222 User manual

Tektronix 5 Series User guide

Tektronix

Tektronix 5 Series User guide

Tektronix 2336YA User manual

Tektronix

Tektronix 2336YA User manual

Tektronix TBS1000C Series User manual

Tektronix

Tektronix TBS1000C Series User manual

Tektronix TDS 520D User manual

Tektronix

Tektronix TDS 520D User manual

Tektronix P6101B User manual

Tektronix

Tektronix P6101B User manual

Tektronix 442 User manual

Tektronix

Tektronix 442 User manual

Tektronix 314 User manual

Tektronix

Tektronix 314 User manual

Tektronix 7B53A User manual

Tektronix

Tektronix 7B53A User manual

Tektronix P6455 TTL/MOS User manual

Tektronix

Tektronix P6455 TTL/MOS User manual

Tektronix WVR5250 User manual

Tektronix

Tektronix WVR5250 User manual

Tektronix 502 series User manual

Tektronix

Tektronix 502 series User manual

Tektronix P6245 User manual

Tektronix

Tektronix P6245 User manual

Popular Test Equipment manuals by other brands

Apera Instruments ZenTest PH60-Z user manual

Apera Instruments

Apera Instruments ZenTest PH60-Z user manual

ASG DTT-5 user manual

ASG

ASG DTT-5 user manual

The Diagnostic Box TDB003 operating manual

The Diagnostic Box

The Diagnostic Box TDB003 operating manual

DIY DSO 138 user manual

DIY

DIY DSO 138 user manual

Topdon ArtiPad I user manual

Topdon

Topdon ArtiPad I user manual

VOLTCRAFT VC40 operating instructions

VOLTCRAFT

VOLTCRAFT VC40 operating instructions

ATD Tools ATD-5494 user manual

ATD Tools

ATD Tools ATD-5494 user manual

Aeroflex TCAS-201-2 Maintenance manual

Aeroflex

Aeroflex TCAS-201-2 Maintenance manual

ATD Tools ATD-5614 owner's manual

ATD Tools

ATD Tools ATD-5614 owner's manual

Berkeley Varitronics Systems Cicada manual

Berkeley Varitronics Systems

Berkeley Varitronics Systems Cicada manual

ATEQ H6000 user manual

ATEQ

ATEQ H6000 user manual

Bosch DAS 3000 manual

Bosch

Bosch DAS 3000 manual

Rothenberger ROTEST GW 150/4 Instructions for use

Rothenberger

Rothenberger ROTEST GW 150/4 Instructions for use

YOKOGAWA SL1400 ScopeCorder LITE Operation guide

YOKOGAWA

YOKOGAWA SL1400 ScopeCorder LITE Operation guide

Water I.D. Pool Lab 1.0 user manual

Water I.D.

Water I.D. Pool Lab 1.0 user manual

Compliance West MegaPulse 1.2x50/8x20-2PF-HR instruction manual

Compliance West

Compliance West MegaPulse 1.2x50/8x20-2PF-HR instruction manual

SIGLENT SDS2104X user manual

SIGLENT

SIGLENT SDS2104X user manual

barfield TT-1000A User instruction manual

barfield

barfield TT-1000A User instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.