Tenma 72-14400 User manual

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <1> V1.006/03/19
AC/DC Clamp Meter
Part Number: 72-14400
User Manual

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <2> V1.006/03/19
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer
to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
SAFETY NOTES
xDo not exceed the maximum allowable input range of any function
xDo not apply voltage to meter when resistance function is selected.
xSet the function switch OFF when the meter is not in use.
WARNINGS
xSet function switch to the appropriate position before measuring.
xWhen measuring volts do not switch to current/resistance modes.
xWhen changing ranges using the selector switch always disconnect the test leads from
the circuit under test.
xDo not exceed the maximum rated input limits.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating
the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
Remove the battery if the meter is to be stored for long periods.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
xVoltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be used
to ensure that the terminals are not "live".
xIf the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Input Limits
Function
Maximum Input
A
400A DC/AC
V DC, V AC
600V DC/AC
Frequency, Resistance, Diode, Continuity ,
Capacitance Test
250V DC/AC
Temperature (ć/
250V DC/ AC
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer
to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
SAFETY NOTES
xDo not exceed the maximum allowable input range of any function
xDo not apply voltage to meter when resistance function is selected.
xSet the function switch OFF when the meter is not in use.
WARNINGS
xSet function switch to the appropriate position before measuring.
xWhen measuring volts do not switch to current/resistance modes.
xWhen changing ranges using the selector switch always disconnect the test leads from
the circuit under test.
xDo not exceed the maximum rated input limits.
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery.
Inspect the condition of the test leads and the meter itself for any damage before operating
the meter. Repair or replace any damage before use.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
Remove the battery if the meter is to be stored for long periods.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
xVoltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of connection to the recessed electrical contacts. Other means should be used
to ensure that the terminals are not "live".
xIf the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Input Limits
Function
Maximum Input
A
400A DC/AC
V DC, V AC
600V DC/AC
Frequency, Resistance, Diode, Continuity ,
Capacitance Test
250V DC/AC
Temperature (
ć
/
250V DC/ AC

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <3> V1.006/03/19
Meter Description
1. Current clamp
2. Non-contact AC voltage indicator light
3. Clamp trigger
4. Rotary Function swith
5. Data Hold/ Back Light button
6. Relative button
7. LCD display
8. MODE select button
9. Range button
10. Hz% Hlod button
11. COM input jack
12. V ¡CAP 7(03+] jack
13. Battery Cover
1. AC DC AC (alternating current)
and DC (direct currrent)
2. Minus sign
3. 㻤㻚㻤㻚㻤㻚㻤 4000 count (0 to 3999)
measurement reading
4. AUTO AutoRange mode
5. REL Relative mode
6. Diode test mode
7. •))) Audible Continuity
8. HOLD Data Hold mode
9. ć,,P,m,V,A,K,M,¡,Units of measure list
10. Hz % Frequency/duty cycle test mode
Specifications
Function
Range & Resolution
Accuracy (% of reading)
AC Current
(50/60Hz)
40.00 AAC
± (2.5 % + 8 digits)
400 .0AAC
± (2.8 % + 5 digits)
DC Current 40.00 ADC
± (2.5 % + 5
digits
)
400 .0ADC ± (2.8 % + 5 digits)
DC Voltage
400.0 mVDC
± (0.8% + 2 digits)
4.000 VDC
± (1.5% + 2digits)
40.00 VDC
400.0 VDC
600.0 VDC
± (2 % + 2 digits)
AC Voltage
(50
-400Hz)
400.0mVAC
± (1.5% + 30 digits)
4.000 VAC
± (1.
5% + 5 digits)
40.00 VAC
400.0 VAC
600.0 VAC
± (2.0% + 5 digits)
¡
1
2
3
45
678
9
10

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <4> V1.006/03/19
Resistance
400.0 ¡
± (1.0% + 4 digits)
4.000K
¡
± (1.5% + 2 digits)
40.00K¡
400.0K¡
4.000M¡
± (2.5% + 3 digits)
40.00M
¡
± (3.5% + 5 digits)
Capacitance
40.00nF
r(4.0% reading + 20 digits)
400.0nF
r(3% reading + 5 digits)
4.000
F
40.00
F
100.0
F
r(4.0% reading + 10 digits)
Frequency
10-10kHz
Sensitivity:100V(<50Hz);50V(50 to
400Hz;15V(401Hz to 10kHz)
r(1.5% reading + 2 digits)
Temp
(type-K)
(probe accuracy
not included)
-20.0 to 760.0qCr(3%rdg+5qC)
-4.0 to1400.0qFr(3%rdg+9qF)
Clamp size Opening 1.2" (30mm) approx
Diode Test Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check Threshold <150:; Test current < 0.5mA
Low Battery Indication “ ” is displayed
Overrange Indication “OL”” is displayed
Measurements Rate 2 per second, nominal
Input Impedance 10M:(VDC and VAC)
Display 4000 counts LCD
AC Current 50-60Hz (AAC)
AC Voltage bandwidth 50-400Hz (VAC)
Operating Temperature 41 to 104oF (5 to 40oC)
Storage Temperature -4 to 140oF (-20 to 60oC)
Operating Humidity Max 80% up to 87qF (31qC) decreasing linearly to 50% at
104qF(40qC)
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Over voltage Category III 600V
Battery One 9V Battery
Auto OFF approx. 30 minutes
Dimensions/Weight 197x70x40mm/183g
Safety For indoor use and in accordance with Overvoltage Category
II, Pollution Degree 2. Category II includes local level,
appliance, portable equipment, etc., with transient
overvoltages less than Overvoltage Cat. III
Resistance
400.0 ¡
± (1.0% + 4 digits)
4.000K¡
± (1.5% + 2 digits)
40.00K¡
400.0K¡
4.000M¡
± (2.5% + 3 digits)
40.00M
¡
± (3.5% + 5 digits)
Capacitance
40.00nF
r(4.0% reading + 20 digits)
400.0nF
r
(3% reading +
5 digits)
4.000F
40.00F
100.0F
r(4.0% reading + 10 digits)
Frequency
10-10kHz
Sensitivity:100V(<50Hz);50V(50 to
400Hz;15V(401Hz to 10kHz)
r(1.5% reading + 2 digits)
Temp
(type
-K)
(probe accuracy
not included)
-20.0 to 760.0qC
r(3%rdg+5qC)
-4.0 to1400.0qF
r(3%rdg+9qF)
Clamp size Opening 1.2" (30mm) approx
Diode Test Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check Threshold <150:; Test current < 0.5mA
Low Battery Indication “ ” is displayed
Overrange Indication “OL”” is displayed
Measurements Rate 2 per second, nominal
Input Impedance 10M:(VDC and VAC)
Display 4000 counts LCD
AC Current 50-60Hz (AAC)
AC Voltage bandwidth 50-400Hz (VAC)
Operating Temperature 41 to 104oF (5 to 40oC)
Storage Temperature -4 to 140oF (-20 to 60oC)
Operating Humidity Max 80% up to 87qF (31qC) decreasing linearly to 50% at
104qF(40qC)
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Over voltage Category III 600V
Battery One 9V Battery
Auto OFF approx. 30 minutes
Dimensions/Weight 197x70x40mm/183g
Safety For indoor use and in accordance with Overvoltage Category
II, Pollution Degree 2. Category II includes local level,
appliance, portable equipment, etc., with transient
overvoltages less than Overvoltage Cat. III
Meter Description
1. Current clamp
2. Non-contact AC voltage indicator light
3. Clamp trigger
4. Rotary Function swith
5. Data Hold/ Back Light button
6. Relative button
7. LCD display
8. MODE select button
9. Range button
10. Hz% Hlod button
11. COM input jack
12. V ¡CAP 7(03+] jack
13. Battery Cover
1. AC DC AC (alternating current)
and DC (direct currrent)
2. Minus sign
3. 㻤㻚㻤㻚㻤㻚㻤 4000 count (0 to 3999)
measurement reading
4. AUTO AutoRange mode
5. REL Relative mode
6. Diode test mode
7. •))) Audible Continuity
8. HOLD Data Hold mode
9. ć,,P,m,V,A,K,M,¡,Units of measure list
10. Hz % Frequency/duty cycle test mode
Specifications
Function
Range & Resolution
Accuracy (% of reading)
AC Current
(50/60Hz)
40.00 AAC
± (2.5 % + 8 digits)
400 .0AAC
± (2.8 % + 5 digits)
DC
Current
40.00 ADC
± (2.5 % + 5 digits)
4
00 .0ADC
± (2.8 % + 5 digits)
DC Voltage
400.0 mVDC
± (0.8% + 2 digits)
4.000 VDC
± (1.
5% + 2digits)
40.00 VDC
400.0 VDC
600.0 VDC
± (2 % + 2 digits)
AC Voltage
(50-400Hz)
400.0mVAC
± (1.5% + 30 digits)
4.000 VAC
± (1.5% + 5 digits)
40.00 VAC
400.0 VAC
600.0 VAC
± (2.0% + 5 digits)
¡
1
2
3
45
678
9
10

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <5> V1.006/03/19
Resistance
400.0 ¡
± (1.0% + 4 digits)
4.000K¡
± (1.5% + 2 digits)
40.00K¡
400.0K¡
4.000M¡
± (2.5% + 3 digits)
40.00M
¡
± (3.5% + 5 digits)
Capacitance
40.00nF
r(4.0% reading + 20 digits)
400.0nF
r
(3% reading +
5 digits)
4.000F
40.00F
100.0F
r(4.0% reading + 10 digits)
Frequency
10-10kHz
Sensitivity:100V(<50Hz);50V(50 to
400Hz;15V(401Hz to 10kHz)
r(1.5% reading + 2 digits)
Temp
(type
-K)
(probe accuracy
not included)
-20.0 to 760.0qC
r(3%rdg+5qC)
-4.0 to1400.0qF
r(3%rdg+9qF)
Clamp size Opening 1.2" (30mm) approx
Diode Test Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check Threshold <150:; Test current < 0.5mA
Low Battery Indication “ ” is displayed
Overrange Indication “OL”” is displayed
Measurements Rate 2 per second, nominal
Input Impedance 10M:(VDC and VAC)
Display 4000 counts LCD
AC Current 50-60Hz (AAC)
AC Voltage bandwidth 50-400Hz (VAC)
Operating Temperature 41 to 104oF (5 to 40oC)
Storage Temperature -4 to 140oF (-20 to 60oC)
Operating Humidity Max 80% up to 87qF (31qC) decreasing linearly to 50% at
104qF(40qC)
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum.
Over voltage Category III 600V
Battery One 9V Battery
Auto OFF approx. 30 minutes
Dimensions/Weight 197x70x40mm/183g
Safety For indoor use and in accordance with Overvoltage Category
II, Pollution Degree 2. Category II includes local level,
appliance, portable equipment, etc., with transient
overvoltages less than Overvoltage Cat. III
Operation
NOTICES: Read and understand all warning and precaution statements listed in the
safety section of this operation manual prior to using this meter. Set the function
select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements
WARNING: Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1. Set the Function switch to
the 400ADC,40ADC,400AAC or40AAC range.
2. If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
3. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive Vterminal.
2. Set the function switch to the Vposition.
3. Select AC or DC with the MODE button.
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
Resistance Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive terminal.
2. Set the function switch to the :•))) CAP position.
3. Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
4. For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative COM jack and the red test lead
banana plug into the positive diode jack.
2. Turn the rotary switch to the :•))) CAP position.
3. Press the MODE button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V.
Reverse voltage will indicate “OL”. Shorted devices will indicate near 0mV and an open
device will indicate “OL” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
¡
¡
Operation
NOTICES: Read and understand all warning and precaution statements listed in the
safety section of this operation manual prior to using this meter. Set the function
select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements
WARNING: Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1. Set the Function switch to
the 400ADC,40ADC,400AAC or40AAC range.
2. If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
3. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive Vterminal.
2. Set the function switch to the Vposition.
3. Select AC or DC with the MODE button.
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
Resistance Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive terminal.
2. Set the function switch to the :•))) CAP position.
3. Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
4. For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative COM jack and the red test lead
banana plug into the positive diode jack.
2. Turn the rotary switch to the :•))) CAP position.
3. Press the MODE button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V.
Reverse voltage will indicate “OL”. Shorted devices will indicate near 0mV and an open
device will indicate “OL” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
¡
¡

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <6> V1.006/03/19
For Continuity tests, if the resistance is < 150:, a tone will sound.
Capacitance Measurements
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.
1. Set the rotary function switch to the :•))) CAP position.
2. Insert the black test lead banana plug into the negative (COM) jack.
Insert the red test lead banana plug into the positive (V) jack.
3. Touch the test leads to the capacitor to be tested.
4. Read the capacitance value in the display
Frequency or % duty cycle measurements
1. Set the rotary function switch to the “VDC/AC,Hz” position.
2. Insert the black lead banana plug into the negative COM jack and the red test lead
banana plug into the positive V jack.
3. Select Hz or % duty with the Hz/% button.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
Temperature Measurements
WARNING: To avoid electric shock, disconnect both test probes from any source of
voltage before making a temperature measurement.
1. Set the function switch to ć.
2. Insert the Temperature Probe into the negative (COM) and the V jacks, making sure to
observe the correct polarity.
3. Touch the Temperature Probe head to the part whose temperature you wish to
measure. Keep the probe touching the part under test until the reading stabilizes (about 30
seconds).
4. Read the temperature in the display. The digital reading will indicate the proper decimal
point and value.
WARNING: To avoid electric shock, be sure the thermocouple has been removed before ch
anging to another measurement function
Non-Contact AC Voltage Measurements
1. Touch the probe tip to the hot conductor or insert into the hot side of the electrical outlet.
2. If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results, rub the probe tip
along a length of the cord to assure placing the tip in close proximity to the live conductor.
WARNING: Risk of Electrocution. Before use,
always test the Voltage Detector on a known live
circuit to
verify proper operation
Operation
NOTICES: Read and understand all warning and precaution statements listed in the
safety section of this operation manual prior to using this meter. Set the function
select switch to the OFF position when the meter is not in use.
AC/DC Current Measurements
WARNING: Ensure that the test leads are
disconnected from the meter before making current
clamp measurements.
1. Set the Function switch to
the 400ADC,40ADC,400AAC or40AAC range.
2. If the range of the measured is not known,
select the higher range first then move to the lower
range if necessary.
3. Press the trigger to open jaw. Fully enclose one conductor
to be measured.
The clamp meter LCD will display the reading.
AC/DC Voltage Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive Vterminal.
2. Set the function switch to the Vposition.
3. Select AC or DC with the MODE button.
4. Connect the test leads in parallel to the circuit under test.
5. Read the voltage measurement on the LCD display.
Resistance Measurements
1. Insert the black test lead into the negative COM terminal and the
red test lead into the positive terminal.
2. Set the function switch to the :•))) CAP position.
3. Touch the test probe tips across the circuit or component under
test. It is best to disconnect one side of the device under test so
the rest of the circuit will not interfere with the resistance reading.
4. For Resistance tests, read the resistance on the LCD display.
Diode and Continuity Measurements
1. Insert the black test lead banana plug into the negative COM jack and the red test lead
banana plug into the positive diode jack.
2. Turn the rotary switch to the :•))) CAP position.
3. Press the MODE button until “ “ appears in the display.
4. Touch the test probes to the diode under test. Forward voltage will indicate 0.4V to 0.7V.
Reverse voltage will indicate “OL”. Shorted devices will indicate near 0mV and an open
device will indicate “OL” in both polarities.
Red Black Black Red
Probe Probe Probe Probe
Forward test Reverse test
¡
¡

www.element14.com
www.farnell.com
www.newark.com
www.cpc.co.uk
TM
TM
Page <7> V1.006/03/19
For Continuity tests, if the resistance is < 150:, a tone will sound.
Capacitance Measurements
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.
1. Set the rotary function switch to the :•))) CAP position.
2. Insert the black test lead banana plug into the negative (COM) jack.
Insert the red test lead banana plug into the positive (V) jack.
3. Touch the test leads to the capacitor to be tested.
4. Read the capacitance value in the display
Frequency or % duty cycle measurements
1. Set the rotary function switch to the “VDC/AC,Hz” position.
2. Insert the black lead banana plug into the negative COM jack and the red test lead
banana plug into the positive V jack.
3. Select Hz or % duty with the Hz/% button.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
Temperature Measurements
WARNING: To avoid electric shock, disconnect both test probes from any source of
voltage before making a temperature measurement.
1. Set the function switch to ć.
2. Insert the Temperature Probe into the negative (COM) and the V jacks, making sure to
observe the correct polarity.
3. Touch the Temperature Probe head to the part whose temperature you wish to
measure. Keep the probe touching the part under test until the reading stabilizes (about 30
seconds).
4. Read the temperature in the display. The digital reading will indicate the proper decimal
point and value.
WARNING: To avoid electric shock, be sure the thermocouple has been removed before ch
anging to another measurement function
Non-Contact AC Voltage Measurements
1. Touch the probe tip to the hot conductor or insert into the hot side of the electrical outlet.
2. If AC voltage is present, the detector light will illuminate.
NOTE: The conductors in electrical cord sets are often twisted. For best results, rub the probe tip
along a length of the cord to assure placing the tip in close proximity to the live conductor.
WARNING: Risk of Electrocution. Before use,
always test the Voltage Detector on a known live
circuit to verify proper operation
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy
may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP,
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is
located on the left side of the meter (top button). While data hold is active, the HOLD
display icon appears on the LCD. Press the data hold button again to return to normal
operation.
REL BUTTON
For DCA and Capacitance Zero & Offset adjustment.
RANGE BUTTON
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the
best range for the measurements being made and is generally the best mode for most measurements.
For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range”
display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to
exit the ManualRanging mode and return to AutoRanging.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy
may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP,
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is
located on the left side of the meter (top button). While data hold is active, the HOLD
display icon appears on the LCD. Press the data hold button again to return to normal
operation.
REL BUTTON
For DCA and Capacitance Zero & Offset adjustment.
RANGE BUTTON
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the
best range for the measurements being made and is generally the best mode for most measurements.
For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range”
display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to
exit the ManualRanging mode and return to AutoRanging.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy
may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP,
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is
located on the left side of the meter (top button). While data hold is active, the HOLD
display icon appears on the LCD. Press the data hold button again to return to normal
operation.
REL BUTTON
For DCA and Capacitance Zero & Offset adjustment.
RANGE BUTTON
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the
best range for the measurements being made and is generally the best mode for most measurements.
For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range”
display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to
exit the ManualRanging mode and return to AutoRanging.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy
may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP,
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is
located on the left side of the meter (top button). While data hold is active, the HOLD
display icon appears on the LCD. Press the data hold button again to return to normal
operation.
REL BUTTON
For DCA and Capacitance Zero & Offset adjustment.
RANGE BUTTON
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the
best range for the measurements being made and is generally the best mode for most measurements.
For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range”
display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to
exit the ManualRanging mode and return to AutoRanging.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy
may randomly trip the sensor. This is normal operation
MODE BUTTON
To select DC/ACV,OHM/ Diode/Continuity/CAP,
Data Hold Button
To freeze the LCD meter reading, press the data hold button. The data hold button is
located on the left side of the meter (top button). While data hold is active, the HOLD
display icon appears on the LCD. Press the data hold button again to return to normal
operation.
REL BUTTON
For DCA and Capacitance Zero & Offset adjustment.
RANGE BUTTON
When the meter is first turned on, it automatically goes into AutoRanging. This automatically selects the
best range for the measurements being made and is generally the best mode for most measurements.
For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE button. The “Auto Range” display indicator will turn off, The “Manual Range”
display indicator will turn on
2. Press the RANGE button to step through the available ranges until you select the range you want.
3. Press and hold the RANGE button for 2 seconds to
exit the ManualRanging mode and return to AutoRanging.
Battery Replacement
1. Remove the one rear Phillips head screw
2. Open the battery compartment
3. Replace the Requires One 9V battery
4. Re-assemble the meter
Important Notice : This data sheet and its contents (the “Information”) belong to the members of the Premier Farnell group of companies (the “Group”) or are licensed to it. No licence is granted
for the use of it other than for information purposes in connection with the products to which it relates. No licence of any intellectual property rights is granted. The Information is subject to change
without notice and replaces all data sheets previously supplied. The Information supplied is believed to be accurate but the Group assumes no responsibility for its accuracy or completeness, any
error in or omission from it or for any use made of it. Users of this data sheet should check for themselves the Information and the suitability of the products for their purpose and not make any
assumptions based on information included or omitted. Liability for loss or damage resulting from any reliance on the Information or use of it (including liability resulting from negligence or where the
Group was aware of the possibility of such loss or damage arising) is excluded. This will not operate to limit or restrict the Group’s liability for death or personal injury resulting from its negligence.
Tenma is the registered trademark of the Group. © Premier Farnell Limited 2016.
Table of contents
Other Tenma Measuring Instrument manuals

Tenma
Tenma 72-2985 User manual

Tenma
Tenma 72-8170 User manual

Tenma
Tenma 72-9395 User manual

Tenma
Tenma 72-3515 User manual

Tenma
Tenma 72-8150 User manual

Tenma
Tenma Smart Tweezer ST-5S-BT2 User manual

Tenma
Tenma 72-3545 User manual

Tenma
Tenma 72-7712 User manual

Tenma
Tenma TEN01056 User manual

Tenma
Tenma 72-14415 User manual

Tenma
Tenma 72-10465 User manual

Tenma
Tenma TEN01073 User manual

Tenma
Tenma 72-14435 User manual

Tenma
Tenma 72-7730A User manual

Tenma
Tenma 72-7935 User manual

Tenma
Tenma 72-9490 User manual

Tenma
Tenma 72-3500 User manual

Tenma
Tenma 72-942 User manual

Tenma
Tenma 72-10465 User manual

Tenma
Tenma 72-7222 User manual