IC10/20FR Page 6
Instructions for use
Start-up 5- Wait for temperature of the machine to be stable.
Once is stable, the indicator of the desired temperatu-
re will remain steady.
Do not put any biological indicator in the rea-
ding positions until the auto-reader does not
conrm temperature stability.
The auto-reader will keep the selected incubation pro-
gram during the reading process. The system does not
allow to modify the incubation program once the rea-
ding process has begun, unless the reading is nished.
6-Start readings once the blue temperature selec-
tion’s light remains stable, this indicates that incu-
bation temperature was reached. Once the selected
temperature has been reached and remains stable,
after turning on the equipment, a one-time automa-
tic check of the Biological Indicators incubation posi-
tions will be carried out. Indicating lights of each well
will turn on bringing the result, a green light
will show a satisfactory result while a red light
will indicate error.
When the equipment detects an error in any position,
a ticket indicating the day, time and position involved
will be printed. To guarantee the reliability of the re-
sults, these positions will be deactivated, thus preven-
ting their use. To evidence this, the red light of
that position will remain blinking indenitely, in order
to indicate that the position has been disabled.
For a correct automatic check, the protective
cover of the auto-reader should be in its place.
Read BIs instructions for use. Before placing
the vials in the auto-reader, press the top to
seal the tube. Crush the ampoule (using the
ampoule crusher within the box of BIs or the
one located within the incubation area of the
auto-reader) and make sure that the liquid has
completely wet the spore carrier placed at the
bottom of the biological indicator plastic tube.
Do not remove or change placement of BI once
the reading process has begun. If this occurs,
results may be invalidated.
Use a non-sterilized biological indicator as a
positive control every time a processed indica-
tor is incubated. For more information refer to
the Positive Control section of this user guide.
7-Place the rapid BI in a reading position and wait for
the result. Reading will start automatically. Immediate-
ly after placing the BI in a position, an alarm will sound
indicating that reading has successfully began and the
position LED will blink until the result is informed.
NOTE: If the auto-reader does not run automatically, it
might indicate that the auto-reader temperature is not
steady yet (check the temperature indicating light to
remain steady) or that the incubation position is disa-
bled due to an error in the initial autotest.
All equipment is manufactured with the fo-
llowing settings:
- Time zone: UTC +0:00
- Printing language: English
For changing time zone, refer to Time zone change
mode section.
For changing printing language, refer to Printing lan-
guage section.
1-Place the auto-reader on a rm surface, free from
vibrations, away from sunlight, currents of hot or cold
air, chemicals and corrosive or ammable substances.
Do not place the equipment in a way that disconnec-
tion of the plug from the power supply could be di-
cult. Leave a gap of at least 10 cm from the wall. Do
not move the auto-reader periodically or during its
use. Connect the auto-reader to a secure and stable
electrical power source.
Do not wet or heat the auto-reader. If liquid is
spilled on the auto-reader, disconnect it and
dry it immediately. Before turning on, verify
that all reading positions are empty.
2-Turn on the machine by connecting the end of the
AC power supply to the mains and then connect the
other end of the source (plug) on the rear of the IC10/
20FR unit. The printer’s paper indicator light will turn
on, if this light becomes intermittent, this means that
the thermal paper must be loaded, or that the printer
cover is not tightly closed. Otherwise, the light will
remain stationary.
The machine will start running the last time and tem-
perature set-up.
3-Synchronize IC10/20FR with local time using Biono-
va® software available at http://www.terragene.com.
ar/productos/infection-control/incubators/softwa-
re-downloads/ ?lang=en. For this, connect a PC to the
USB plug of the auto-reader and start the program.
The Bionova® IC10/20FR will be synchronized with the
time in the PC. Synchronization can be done as long
as there is not an ongoing reading.
4- The incubator will initiate with the last setup pro-
gram for temperature and time. To change the in-
cubation temperature, press the button for 3
seconds. Lights corresponding to 60 °C will start blin-
king, indicating that a temperature program must be
selected. For temperature selection, press the button
to select 60 ºC. After 4 seconds the temperature
modication will be accepted.
For 60 ºC, 30 min., 1 or 3 hrs. program can be selected.
It’s possible to select dierent programs for each half
or the same for both, depending on the indicator's -
nal reading characteristics. Check that the time setting
is appropriate for the BIs to be incubated in each zone.
To change the time program, press the button
of the correspondent zone as many times as needed
to choose the right setting.