TESTO 310 User manual

testo 310 · Flue gas analyzer
Instruction manual

2

1 Contents
3
Pos: 1 /TD/Überschr iften/1. Inhalt @ 0\mod_1177 587817070_79.doc x@ 124 3 @ 1@ 1
1Contents
1Contents ...................................................................................................3
2Safety and the environment....................................................................5
2.1. About this document........................................................................5
2.2. Ensure safety...................................................................................6
2.3. Protecting the environment..............................................................7
3Specifications ..........................................................................................8
3.1. Use ..................................................................................................8
3.2. Technical data .................................................................................8
3.2.1. Measurement ranges and resolution................................................................8
3.2.2. Other instrument data ......................................................................................9
4Product description...............................................................................11
4.1. Measuring instrument ....................................................................11
4.1.1. Front view ......................................................................................................11
4.1.2. Keypad ..........................................................................................................11
4.1.3. Display...........................................................................................................12
4.1.4. Connections...................................................................................................14
4.1.5. Condensate outlet/interface ...........................................................................14
4.1.6. Rear view.......................................................................................................15
4.2. Flue gas probe...............................................................................16
4.3. Area versions.................................................................................17
4.4. Reading display .............................................................................18
5Using the product ..................................................................................20
5.1. Mains unit / rechargeable battery ..................................................20
5.1.1. Charging the rechargeable battery.................................................................20
5.1.2. Mains operation .............................................................................................21
5.2. Performing settings........................................................................21
5.2.1. Instrument configuration menu / commissioning ............................................21
5.2.2. Measurements configuration menu................................................................23
5.3. Measuring......................................................................................24
5.3.1. Preparing for measurement ...........................................................................24
5.3.1.1. Zeroing phases................................................................................24
5.3.1.2. Using the flue gas probe..................................................................25
5.3.1.3. Setting fuel ......................................................................................25
5.3.2. Flue gas.........................................................................................................26
5.3.3. Ambient CO ...................................................................................................26
5.3.4. Draught measurement ...................................................................................26
5.3.5. Differential pressure.......................................................................................27

1 Contents
4
6Maintaining the product ....................................................................... 30
6.1. Cleaning the measuring instrument .............................................. 30
6.2. Cleaning the flue gas probe .......................................................... 30
6.3. Draining the condensate container ............................................... 30
6.4. Checking / replacing the particle filter ........................................... 31
7Tips and assistance .............................................................................. 32
7.1. Questions and answers ................................................................ 32
7.2. Accessories and spare parts......................................................... 33
Pos:2 /TD/--- Seitenwechsel--- @ 0\mod_1173774430601_0. docx @ 283 @ @ 1

2 Safety and the environment
5
Pos: 3 /TD/Übersc hriften/2. Sic herheit und Umwelt @ 0\mod_11737 74719351_79.docx @ 292 @ 1 @ 1
2Safety and the environment
Pos: 4 /TD/Übersc hriften/2.1 Zu diese mD okument @ 0\mod_117377525 2351_79.docx @ 346 @ 2 @ 1
2.1. About this document
Pos: 5 /TD/Sic herheit und Umwelt/Zu diese m Dokument/Verwendung/V erwendung (Standar d) @ 0\mod_1173775068554_ 79.docx @ 337 @ 5 @ 1
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Pos: 6 /TD/Sic herheit und Umwelt/Zu diese m Dokument/Symbole und Sc hreibkonventionen/ Tabellenkopf W arnhinweise @ 2\mod_120764 5198296_79.docx @ 14334 @ 5 @ 1
Warnings
Always pay attention to information that is marked by the following
warnings with warning pictograms. Implement the specified
precautionary measures.
Representation Explanation
Pos: 7 /TD/Sic herheit und Umwelt/Zu diese m Dokument/Symbole und Sc hreibkonventionen/ Warnhinweis VORSIC HT@ 2\ mod_1207651416515_79. docx @ 14416 @ @ 1
CAUTION indicates potential minor injuries
Pos: 8 /TD/Sic herheit und Umwelt/Zu diese m Dokument/Symbole und Sc hreibkonventionen/ Warnhinweis ACHTUNG (Pr oduktschaden) @ 2\ mod_1207651536812_79. docx@ 14434 @ @ 1
NOTICE indicates circumstances that may lead to
damage to the products

2 Safety and the environment
6
Pos: 9 /TD/Sic herheit und Umwelt/Zu diese m Dokument/Symbole und Sc hreibkonventionen/ Tabellenkopf Sy mbole und Schreibkonv. @ 2\mod_120 7652481703_79.doc x @ 14452 @ 5 @ 1
Symbols and writing standards
Represen-
tation
Explanation
Pos: 10 /TD/Sic herheit und Umwelt/Zu die sem Dokument/Symbole und Sc hreibkonventione n/Symbole und Schreibkon v. [Standard_klei n] @ 2\mod_1207652632140_79. docx @ 14470 @ @ 1
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Pos: 11 /TD/Sic herheit und Umwelt/Zu die sem Dokument/Symbole und Sc hreibkonventione n/Schreibkonv. [ OK] @ 2\mod_120765548921 8_79.docx @ 14506 @ @ 1
[OK] Control keys of the instrument or buttons of the
program interface.
Pos: 12 /TD/Übersc hriften/2.2 Sic herheit gewährleisten @ 0\ mod_1173780783960_ 79.docx @ 366 @ 2 @ 1
2.2. Ensure safety
Pos: 13 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/Produkt bestimmungs gemäß verwenden @ 0\mod_1173 781261848_79.doc x @ 386 @ @ 1
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
Pos: 14 /TD/Sic herheit und Umwelt/Si cherheit gewähr leisten/Gerät bei Beschädigunge n nicht in Betrieb nehmen @ 0\mod_1 186985945375_79.doc x @ 2252 @ @ 1
> Do not operate the instrument if there are signs of damage at
the housing, mains unit or feed lines.
Pos: 15 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/Keine Messung an spann ungsführenden Teilen @ 0\mod_11756925 64164_79.docx @ 592 @ @ 1
> Do not perform contact measurements on non-insulated, live
parts.
Pos: 16 /TD/Sic herheit und Umwelt/Si cherheit gewährlei sten/Nicht mit Lö sungsmitteln la gern @ 0\mod_11756923751 79_79.docx @ 583 @ @ 1
> Do not store the product together with solvents. Do not use any
desiccants.
Pos: 17 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/Nur beschriebene W artungsarbeiten durc hführen @ 0\mod_11756927 05195_79.docx @ 601 @ @ 1
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
Pos: 18 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/testo 350/testo 350 Si cherheit @ 5\mod_1261385 845735_79.doc x@ 53306 @ @ 1
> Any further or additional work must only be carried out by
authorised personnel. Testo will otherwise refuse to accept
responsibility for the proper functioning of the measuring
instrument after repair and for the validity of certifications.
Pos: 19 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/Nur in geschlossenen, tr ockenen Räumen betrei ben @ 0\mod_11869857978 28_79.docx @ 2243 @ @ 1
> Only use the device in closed, dry rooms and protect it from rain
and moisture.
Pos: 20 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/Temperaturangaben auf Sonden/Füh lern @ 0\mod_117569329307 0_79.docx @ 610 @ @ 1
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and
feed lines to any temperatures in excess of 70 °C unless they
are expressly permitted for higher temperatures.

2 Safety and the environment
7
Pos: 21 /TD/Sic herheit und Umwelt/Sicher heit gewährleiste n/vor Ort gültige Sic herheitsbestimmungen be achten @ 0\mod_11869971 07328_79.docx @ 2298 @ @ 1
> The objects to be measured or the measurement environment
may also pose risks: Note the safety regulations valid in your
area when performing the measurements.
Pos: 22 /TD/Sic herheit und Umwelt/Si cherheit gewähr leisten/Akku-Sic herheitshinweise ( LiIon) @ 0\mod_11 87184666109_79.doc x @ 2363 @ @ 1
> Improper use of rechargeable batteries can lead to destruction
or injuries by means of current surges, fire or escaping
chemicals. The following instructions must be observed to avoid
such hazards:
• Only use in accordance with the directions in the instruction
manual.
• Do not short, take apart or modify.
• Do not expose to heavy impacts, water, fire or temperatures
above 60 °C.
• Do not store in the proximity of metal objects.
• Do not use leaky or damaged rechargeable batteries. In the
event of contact with battery acid: Thoroughly wash affected
area with water and consult a doctor, if necessary.
• Only charge in the instrument or the recommended charging
station.
• Immediately stop the charging process if this is not
completed in the given time.
• In the event of improper function or signs of overheating,
immediately remove the rechargeable battery from the
measuring instrument/charging station. Caution:
Rechargeable battery may be hot!
Pos: 23 /TD/Übersc hriften/2.3 Umwel tsc hützen @ 0\mod_117378084 3645_79.docx @ 375 @ 2 @ 1
2.3. Protecting the environment
Pos: 24.1 /TD/Si cherheit und Umwelt/ Umwelt schützen/Akkus/ Batterien entsorgen @ 0\mod_117569 3637007_79.docx @ 619 @ @ 1
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
Pos: 24.2 /TD/Si cherheit und Umwe lt/Umwelt schützen/ Produkt entsorgen @ 0\mod_1173780 307072_79.doc x@ 357 @ @ 1
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
Pos: 25 /TD/Übersc hriften/3. Leistu ngsbeschreibung @ 0\mod_1 173774791554_79.d ocx @ 301 @ 1 @ 1

3 Specifications
8
3Specifications
Pos: 26 /TD/Übersc hriften/3.1 Verwendun g @ 0\mod_11762110164 37_79.docx @ 695 @ 2 @ 1
3.1. Use
Pos: 27 /TD/Leis tungsbeschreibung/V erwendung/testo 3 xx/testo 310_Verwendun g @ 12\mod_1332405181369_ 79.docx @ 113712 @ @ 1
The testo 310 is a handheld measuring instrument for the
professional flue gas analysis of combustion plants:
• Small combustion plants (oil, gas, wood, coal)
• Low-temperature and condensing boilers
• Gas water heaters
These systems can be adjusted using the testo 310 and checked
for compliance with the applicable limit values.
The following tasks can also be carried out using the testo 310:
• Regulating the O2, CO and CO2 values in combustion plants
for the purpose of ensuring optimal operation.
• Draught measurement.
• Measuring and regulating the gas flow pressure in gas water
heaters.
• Ambient CO measurement.
The testo 310 must not be used:
• as a safety (alarm) device
Pos: 28 /TD/Übersc hriften/3.2 Techni sche Daten @ 0\mod_11762110 88437_79.docx @ 704 @ 2 @ 1
3.2. Technical data
Pos: 29 /TD/Leis tungsbeschreibung/ Technische Daten/t esto 310 Technische Daten/ Messbereiche_Gen auigkeiten testo310 @ 12\ mod_1332405668932_7 9.docx @ 113821 @ 3 @ 4
3.2.1. Measurement ranges and resolution
Measure-
ment
parameter
Measurement
range
Resolution Accuracy Response time
t90
O20.0 to 21.0 Vol.% 0.1 vol.% ±0.2 vol.% 30s
CO 0 to 4000 ppm 1 ppm ± 3 ppm (<30 ppm)
± 10% of meas. val. (31 to
200 ppm)
± 20 ppm (201 to 400 ppm)
± 5% of meas. val. (401 to
2000 ppm)
± 10% of meas. val. 2001 to
4000 ppm
60s

3 Specifications
9
Measure-
ment
parameter
Measurement
range
Resolution Accuracy Response time
t90
COamb 0 to 4000 ppm 1 ppm ± 3 ppm (<30 ppm)
± 10% of meas. val. (31 to
200 ppm)
± 20 ppm (201 to 400 ppm)
± 5% of meas. val. (401 to
2000 ppm)
± 10% of meas. val. 2001 to
4000 ppm
60s
Draught -20.00 to 20.00
hPa
0.01 hPa ± 0.03hPa (-3.00 to 3.00 hPa)
±1.5% of meas. val. (rest of range)
-
∆P -40.0 to 40.0 hPa 0.1 hPa ± 0.5hPa
-
Flue gas
temperature
0.0 to 400.0 °C 0.1°C
± 1°C (0.0 to 100.0°C)
± 1.5% of meas. val. (>100°C)
<50s
Ambient
temperature
-20.0 to 100.0 °C 0.1°C
± 1°C
<50s
Pos: 30 /TD/Leis tungsbeschreibung/ Technische Daten/t esto 310 Technische Daten/ weitere Gerätedat en testo 310 @ 12\mod_133240916 6268_79.docx @ 113855 @ 35 @ 1
3.2.2. Other instrument data
Flue gas analyser
Feature Values
Storage and
transport
temperature
-20 to 50 °C
Operating
temperature
-5 to 45 °C
Power supply Rechargeable battery: 1500mAh
Mains unit: 5 V/1 A
Protection class IP40
Weight incl. probe approx. 700g
Dimensions 201 x 83 x 44 mm
Battery charge time Approx. 5-6 hours
Rechargeable
battery life
> 8h (pump on, 20°C ambient temperature)
EU Directive 2004/108/EC

3 Specifications
10
Feature Values
Warranty Measuring instrument, flue gas probe: 24
months
Thermocouple: 12 months
Sensors O2, CO: 24 months,
Rechargeable battery: 12 months
Terms of warranty Terms of warranty: see website
www.testo.com/warranty
Pos: 31 /TD/Übersc hriften/4. Produkt beschreibung @ 0\mod_1 173774846679_79.doc x@ 310 @ 1 @ 1

4 Product description
11
4Product description
Pos: 32 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Übersicht @ 12\mod_13 32410558502_79.doc x @ 113960 @ 23 @ 1
4.1. Measuring instrument
4.1.1. Front view
1 Display
2 Function keys
3 Keypad
Pos: 33 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Tastatur @ 12\mod_133 2410926380_79.doc x @ 113994 @ 3 @ 1
4.1.2. Keypad
Button Functions
[] Switch measuring instrument on / off
[OK]
Example
Function key (orange, 3x), relevant function is shown on
the display
[▲]Increase value, select parameter
[▼]Decrease value, select parameter

4 Product description
12
Button Functions
[esc] Back, cancel print process
[] Switch display light on/off
[] Transmit data to the Testo protocol printer.
Pos: 34 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Disp lay @ 12\mod_133241163962 8_79.docx @ 114028 @ 3 @ 1
4.1.3. Display
1 Measurement type (an arrow marks the measurement type that
is activated):
Icon Measurement
Flue gas (icon visible when the instrument is
switched off)
Ambient CO (icon visible when the instrument is
switched off)
Draught (icon visible when the instrument is switched
off)
Differential pressure (icon visible when the
instrument is switched off)

4 Product description
13
2 Status:
Icon Meaning
Measuring gas pump (icon visible when the
instrument is switched off)
The inner segments light up alternately when the
measuring gas pump is running.
Error
Flashes when an error occurs, an error code is also
displayed.
Print
Lights up during data transmission to the report
printer
Open configuration menu
Fuel type / fuel number
Depending on the set fuel, one of the icons (solid
fuel, oil, gas) and the associated fuel number lights
up.
, , , Battery capacity.
Indication of the remaining capacity of the
rechargeable battery by fill level of the battery icon:
• 3 segments 75-100%
• 2 segments 50-75%
• 1 segment 25-50%
• no segment <25%
3 Reading display line 1
See Area versions, page 17.
4 Reading display line 2
See Area versions, page 17.
5 Function key assignment:
Icon Possible assignment
Left function key:
Start print process
Select measurement type

4 Product description
14
Icon Possible assignment
OK
Start
Stop
Middle function key:
Confirm input
Start measurement
Stop measurement
Set
Right function key:
Open configuration menu
Switch to the next parameter:
Pos: 35 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Sc hnittstellen @ 12\mod_133 7587126580_79.doc x @ 119307 @ 2223 @ 1
4.1.4. Connections
1 Charging socket for mains unit (Micro USB)
2 Cable to the flue gas probe
3 Gas outlet
Pos: 36 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Ger äteanschlüsse @ 12\mod_13 37587072729_79.doc x @ 119273 @ 22233 @ 1
4.1.5. Condensate outlet/interface
1 Infrared interface
2 Condensate outlet

4 Product description
15
Pos: 37 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Ger äterückseite @ 12\mod_1 337588347371_79.doc x@ 119341 @ 22133 @ 1
4.1.6. Rear view
1 Attachment for carrying strap
2 Condensate trap
3 Magnetic holder
4 Gas outlet
5 Magnetic holder
CAUTION
Damage to other equipment caused by strong magnets!
>Keep a safe distance from products which could be damaged
by magnets (e.g. monitors, computers, pacemakers, credit
cards).

4 Product description
16
Pos: 38 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Abga ssonde @ 12\mod_13324177 85833_79.docx @ 114062 @ 2232443445555 @ 1
4.2. Flue gas probe
1 Thermocouple
2 Probe shaft
3 Probe handle
4 Connecting cable
5 Removable filter chamber with window, particle filter, and
sealing plug for differential pressure measurement
Pos: 39 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Gebi etsversionen @ 12\mod_ 1332425442045_79. docx @ 114268 @ 2212255 @ 1

4 Product description
17
4.3. Area versions
The instrument can be configured for four area versions. Country-
specific calculation formulas, associated measurement parameters
and fuels are activated with this setting. Similarly, the setting
influences the date and time format. The area version is set in the
instrument configuration menu / commissioning, see Instrument
configuration menu / commissioning, page 21.
Area
version
(ArEA )
Countries
(recom-
mendation)
Parameters Fuels
1 USA, HU,
IN, KR
Line 1:
O2, T, CO ,
P1, P2, uCO
Line 2:
CO, Eff, T,
EXA, CO2
nAt GAS – Natural gas
ProP GAS – Propane
FUEL OIL2 – Fuel oil 2
Bioh 5 – Biomass 5%
Uood 20 – Wood 20%
2 GB, RU, IT,
CZ, DK, AU,
JP, CN
Line 1:
rat, O2, T,
P1, CO ,
P2, uCO
Line 2:
CO2, CO,
EXA,
Eff.net,
Eff.gross,
T, O2
nAt GAS – Natural gas
LPG GAS – LPG
ProP GAS – Propane
But GAS – Butane
LI OIL – EL fuel oil
CEro SEnE – Kerosene
HE OIL – Heavy fuel oil
Uood PELL - Wood
pellets
3 NL, SE, TR,
RO
Line1:
O2, P1, T,
GI, uCO, P2
Line 2:
CO, CO2,
T, Eff, qA,
λ, CO
nAt Hb – Natural gas Hb
nAt Ho – Natural gas Ho
ProP Hb – Propane Hb
ProP Ho – Propane Ho
but Ho – Butane Ho
LPG Ho – LPG Ho
LI OIL – EL fuel oil
Uood PELL – Wood
pellets

4 Product description
18
Area
version
(ArEA )
Countries
(recom-
mendation)
Parameters Fuels
4 DE, AT, CH,
FR, ES, BE,
PL, PT, AR,
BR
Line1:
T, CO, uCO,
T, O2, P1,
P2, CO
Line 2:
O2, qA,
CO2, Eff, λ,
T, T
nAt GAS - Natural gas
ProP GAS - Propane
but GAS - Butane
CoO GAS - Coke oven
gas
Toun GAS - Town gas
LI OIL - EL fuel oil
HE OIL - Heavy fuel oil
Uood 15 - Wood 15%
Pos: 40 /TD/Produk tbeschreibung/ Übersicht/testo 3xx/ testo 310/testo 310 Mess wertanzeige @ 12\mod_1 340612549682_79.doc x@ 135 653 @ 22122222353533 5535533 @ 1
4.4. Reading display
Display Measurement parameter
T Ambient temperature
T Flue gas temperature
CO Carbon monoxide
O2 Oxygen
CO Ambient carbon monoxide
qA Flue gas loss without due consideration of the
calorific value range
Eff.net Net efficiency (without due consideration of
the calorific value range)
Eff.gross Gross efficiency (with due consideration of
the calorific value range)
Eff. Efficiency
λAir ratio
P2 Differential pressure
CO2 Carbon dioxide
P1 Flue draught

4 Product description
19
Display Measurement parameter
uCO Carbon monoxide undiluted
rat Ratio
EXA Air surplus
Pos: 41 /TD/Übersc hriften/6. Produkt verwenden @ 0\mod_ 1173774928554_79. docx @ 328 @ 22223353555533 @ 1

5 Using the product
20
5Using the product
Pos: 42 /TD/Produk t verwenden/testo 310/ testo 310 Netzteil_A kku @ 12\mod_133241915102 9_79.docx @ 114234 @ 23323543 54553535555553534455 5544 @ 1
5.1. Mains unit / rechargeable battery
The rechargeable battery is permanently installed and can
only be changed by a Testo service centre.
The measuring instrument is supplied with a partially
charged rechargeable battery.
> Charge the rechargeable battery fully before using the
measuring instrument.
5.1.1. Charging the rechargeable battery
The rechargeable battery can only be charged at an ambient
temperature of ±0 to +35°C. If the battery has been discharged
completely, the charging time at room temperature is approx.
5-6 hrs.
Charging in the measuring instrument
1. Connect the plug of the mains unit to the mains unit socket on
the measuring instrument.
2. Connect the mains plug of the mains unit to a mains socket.
- The charging process will start. The charge status will be shown
on the display. The charging process will stop automatically
when the battery is fully charged.
Battery care
> Do not fully exhaust the rechargeable battery.
> Only store instrument with battery charged and at low
temperatures, but not below 0°C (best storage conditions with a
charge level of 50-75% = 2 segments, at an ambient
temperature of 10-20°C, recharge completely before use).
> The rechargeable battery life depends on the storage, operating
and ambient conditions. The available useful life of the
rechargeable battery reduces more and more with frequent use.
If the useful life is significantly shortened, the rechargeable
battery should be replaced.