11. Unbolt the four-point motor mounting bracket from the
blower housing.
12. Disconnect the motor wiring harness at the quick-con-
nect plug.
13. Pull the motor, motor mounting bracket, removable inlet
collar and blower wheel out of the blower housing.
14. Be sure to mark on the motor shaft the location of the
blower wheel so that the new motor shaft can be lined
up correctly with the required blower wheel position to
fit properly inside the blower housing.
Note: After the motor is installed, but before replacing the
plenum access panel, turn the motor on to ensure that rotation is
in the clockwise direction (when viewed from the motor side of
the blower).
9.2 Replacing the Blower Motor (Model 1286 Series
6-foot)
1. Turn the unit off and disconnect it from the power
source.
2. Remove all dress panels.
3. Remove the access panel.
4. Unfasten the two latches connecting the blower to the
air plenum.
5. Remove the two bolts toward the front of the cabinet
that connect the blower housing brackets to the top of
the cabinet.
6. Loosen - do not remove- the four remaining bolts that
connect the blower housing brackets to the top of the
cabinet.
7. Slide the blower housing toward the rear of the cabinet
as far as it will go.
8. Disconnect the motor wiring harness at the quick-con-
nect plug.
9. Unbolt the removable inlet collar from the blower hous-
ing.
10. Pull the motor, motor mounting bracket, removable inlet
collar, and blower wheel out of the blower housing.
11. Be sure to mark on the motor shaft the location of the
blower wheel so that the new motor shaft can be lined
up correctly with the required blower wheel position to
fit properly inside the blower housing.
9.3 Replacing the Filters, All Models
Dispose of the old filters per established laborato-
ry practices. If necessary, consult a bio-safety
officer or other appropriate person.
1. Turn the unit off and disconnect it from the power
source.
2. Close the window.
3. Remove the screw on the bottom left side, which secures
the control panel, and swing the panel open.
4. Remove the eight screws securing the dress panel.
Remove the panel and set it aside.
5. Remove the four hex nuts securing the lower cross brace
and set the brace to the side.
6. Remove all hex nuts securing the pressure plate.
7. Pry the pressure plate loose from the gasket and set it to
the side.
8. Pull the Velcro boot loose from the exhaust filter
plenum.
9. Release the two latches securing the supply plenum to
the blower housing.
10. Disconnect the vinyl tubing that connects the Mag gauge
to the plenum.
11. Remove the hex nuts (4 on the four-foot models, 6 on
the six-foot models), springs, washers and hold-down
brackets securing the plenum.
12. Remove the front filter hold-down studs (2 on the four-
foot models, 3 on the six-foot models) in front of the
plenum.
13. Slide the plenum from the cabinet.
14. Remove the supply filter and clean the filter flange.
15. Loosen - do not remove- the four bolts, springs and
washers that secure the exhaust filter.
16. Slide the exhaust filter out, clean the filter flange and
install the new filter, ensuring that the gasket is on the
top.
17. Tighten the hex nuts to secure the new exhaust filter in
position.
18. Install the new supply filter with the gasket side down.
19. Reinstall the supply plenum and assemble the compo-
nents in reverse order.
The cabinet must be recertified after filter replacement.