
T129–TECHNICAL BULLETIN NanoDrop Lite
Page 2
T129 – Rev 1/2012
Thermo
Scientific
NanoDrop
Products
Wilmington, Delaware USA Technical support:nanodrop.tech[email protected] 302-479-7707 Toll free (US & Canada) 877-724-7690 www.thermoscientific.com/nanodrop
A: No. The Protein A280 application of the NanoDrop Lite is designed for measuring purified proteins that absorb at
280nm. Calculations are based upon Beer’s Law, using a protein specific extinction coefficient. Colorimetric assays
require standard curve generation and absorb light at wavelengths other than 280 nm. If you are currently using a
colorimetric assay to measure proteins, it is recommended that you use one of the preprogrammed colorimetric
methods available on the NanoDrop 2000/2000c.
Q: Is simply wiping the pedestal surface adequate to prevent sample carryover?
A: Yes. The highly polished quartz stainless steel surfaces of the sample retention system are resistant to sample
adherence. Wiping with dry lint-free lab wipes remove samples very effectively. However, if a sample is left to dry on
the pedestal, more extensive cleaning is required. Refer to the technical document, Cleaning and Reconditioning for
the NanoDrop Lite, for additional information.
Q. How do I clean the pedestals?
A: An application of water at the end of a measurement session is generally all that is necessary to keep the measure
surfaces clean and conditioned. If additional cleaning is required, refer to the technical document, Cleaning and
Reconditioning for the NanoDrop Lite, for additional information. Do not use detergents or isopropanol as cleaning
agents as their use may result in the pedestals becoming unconditioned.
Q: How do I keep my sample from spreading on the pedestal?
A: There are many reagents and organic solvents that could compromise the pedestal surface properties, causing
samples to flatten out rather than bead up. Use the NanoDrop PR-1 reconditioning compound as a rapid means of
reconditioning the pedestals when the pedestal surface properties have been compromised and samples spread out on
the pedestal. PR-1 kits are available through Thermo Fisher Scientific or your local distributor.
Q: What is an appropriate blanking solution?
A: The blanking solution should always be the solvent or buffer used to dissolve the sample, at the same pH and ionic
strength.
Q: Why do I have negative absorbance values?
A: A blank measurement was made either using a solution with more absorbance than the sample buffer or on a dirty
pedestal. Clean the pedestal and make a new blank measurement with a fresh aliquot of the appropriate buffer.
Q: Where is the spectrum from my measurement?
A: The NanoDrop Lite does not collect spectral data. It measures absorbance at three different wavelengths: a reference
wavelength, 260 nm and 280 nm.
Q. What information is included in the sample output data?
A. The sample output data includes sample number (auto-generated 1 through 500), time/date of measurement, analyte
being measured, absorbance (A260 or A280), and concentration. In addition, the 260/280 purity ratio is supplied for
nucleic acid measurements.