6. MAINTENANCE
1. Be sure there is no resin or material sticking to the surface of the light guide.
2. If materials are sticking on the light guide immediately clean it with so cloth. Keep light guide clean for better
light intensity and curing performance.
3. e light guide should be sterilized aer each use. It can be safely autoclaved at 121 °C. For no more then 15
minutes. Let cool before handling.
4. e other accessories such as curing light hadpiece and eye protector should be scrubbed with pure water and
sheathed in accordance with standard disinfection procedures.
5. Keep curing light handpiece in charging station while not in use.
6. Dispose Lithium battery according to the federal and state laws.
7. e User should wear colored protection glasses to avoid eyes injury due to LED light.
7. TROUBLESHOOTING
• Light intensity seems low. Check the light guide and ensure it’s clean and free from cracks.
• e mode on my handpiece changes on its own.
Check the function button and ensure it is not
stuck in.
• My light will not change modes.
Check the function button and ensure it is not
stuck.
• My light guide does not feel secure in the hand piece.
Remove the light guide and ensure that
there is an o‐ring installed in the end of the hand piece.
• None of the mode or status lights illuminate.
Check all your connections to the main power
adapter. If you are able to measure voltage into the main power adapter but no DC voltage out
then you may need to replace the main power adapter.
• I am able to measure DC voltage out of the main power adapter but my light still does not work.
Send the unit in for factory service.