Trotec BE40 User manual

TRT-BA-BE40-TC2019-51-003-EN
BE40
EN
OPERATING MANUAL
CLAMP METER

2 EN
clamp meter BE40
Table of contents
Notes regarding the operating manual.................................2
Safety .....................................................................................2
Information about the device................................................4
Transport and storage...........................................................6
Operation ...............................................................................7
Maintenance and repair ......................................................12
Errors and faults..................................................................12
Disposal ...............................................................................12
Notes regarding the operating manual
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of
persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average
risk level which, if not avoided, can result in serious
injury or death.
Caution
This signal word indicates a hazard with a low risk
level which, if not avoided, can result in minor or
moderate injury.
Note
This signal word indicates important information
(e.g. material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry
out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the
operating manual must be observed.
You can download the current version of the operating manual
and the EU declaration of conformity via the following link:
BE40
https://hub.trotec.com/?id=43773
Safety
Read this manual carefully before starting or using the
device. Always store the manual in the immediate vicinity
of the device or its site of use!
Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may
result in electric shock, fire and/ or serious injury.
Save all warnings and instructions for future
reference.
• The device is supplied with a warning sign. Prior to initial
start-up, make sure to paste the corresponding warning
sign in your local language over the one present at the rear
of the device as described in chapter Operation.
• Do not use the device in potentially explosive rooms.
• Do not use the device in aggressive atmosphere.
• Protect the device from permanent direct sunlight.
• Do not remove any safety signs, stickers or labels from the
device. Keep all safety signs, stickers and labels in legible
condition.
• Do not open the device.
• Observe the storage and operating conditions as given in
the Technical data chapter.

EN 3
clamp meter BE40
Intended use
This device is intended for measurements in the measuring
range and category (CATIII, 600V) specified in the technical
data.
To use the device for its intended use, only use accessories and
spare parts which have been approved by Trotec.
Improper use
Do not use the device in potentially explosive atmospheres,
when wet or very humid.
Unauthorized modifications of the device are forbidden.
Personnel qualifications
People who use this device must:
• master the 5safety rules (1De-energise, 2Secure against
restart, 3Verify de-energised state (bipolar), 4Earth and
short-circuit, 5Cover neighbouring live parts).
• use the voltage detector following safe working
procedures.
• be aware of the dangers that occur when working with
electric devices in damp areas.
• take measures to protect themselves from direct contact
with live parts.
• have read and understood the operating manual, especially
the Safety chapter.
Residual risks
Warning of electrical voltage
Electric shock due to insufficient insulation! Check the
device for damages and proper functioning before each
use.
If you detect damages, do not use the device any
longer.
Do not use the device when wither the device or your
hands are damp or wet!
Do not use the device when the battery compartment
or the housing is open.
Warning of electrical voltage
Electric shock due to contact with live parts! Do not
touch any live parts. Secure neighbouring live parts by
covering them or by switching them off.
Warning of electrical voltage
When performing non-contact measurements of the
current, disconnect the measuring cables from the
device beforehand.
Warning of electrical voltage
Work on the electrical components must only be
carried out by an authorised specialist company!
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may
use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the
hands of children.
Warning
Dangers can occur at the device when it is used by
untrained people in an unprofessional or improper way!
Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To avoid damages to the device, make sure that the
correct measuring range is selected before carrying out
a measurement.
If you are unsure, select the largest measuring range.
Unplug the measuring cable before changing the
measuring range.
Note
Test the device for proper functioning at a known
voltage source before each measurement.
Note
To prevent damages to the device, do not expose it to
extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the
device.

4 EN
clamp meter BE40
Information about the device
Device description
Using the clamp meterBE40 you can easily measure the
amperage of alternating currents, ACand DCvoltages as well as
resistances.
Furthermore, you can check the continuity of circuits, fuses and
contacts and test the forward voltage of diodes without the need
for contact.
The current measurement is effected without contact via the
electromagnetic field, which is why the electric circuit doesn't
have to be interrupted for this method. Therefore, running
systems, which cannot be switched off separately, can also be
checked.
Owing to the galvanic isolation, the measuring signal is also
potential-free towards the variable to be measured.
Device depiction
1
2
3
4
5
6
7
8
9
10
11 11
13
14
12
12
No. Designation
1 Clamp
2 Rotary switch
3MAXbutton (maximum value display)
4HOLD button (freezing the display)
5 Display
6 Input connection for measuring cable (red)
7 COM connection for measuring cable (black)
8 Diode test/ continuity test button
9 Display illumination on/off button (hold for3s)
10 Lever for opening the clamp
11 Test probes
12 Protection against contact
13 Red measuring cable
14 Black measuring cable
Rotary switch
22
21
20
19
18
17
16
15
No. Designation
15 Measuring alternating current of upto 1000A
16 Measuring alternating current of upto 200A
17 Measuring alternating current of upto 20A
18 Diode test/ continuity test
19 Measuring resistance
20 Measuring DCvoltage
21 Measuring ACvoltage
22 OFF (for switching the device off)

EN 5
clamp meter BE40
Display
34 35
33
24
29
28
27
26
25
32
30
31
36 23
No. Purpose of the indication
23 Battery loading status
24 Unit for amperage (A)
25 Unit for voltage (mV/ V)
26 Unit for resistance (kΩ/ MΩ)
27 Resistance measurement
28 Measurement value display
29 Minus sign (positive and negative poles confused)
30 ACmeasurement
31 DCmeasurement
32 Automatic measuring range selection
33 Diode test
34 Continuity test
35 MAXindication (indication of the maximum value)
36 HOLDindication (display of the measured value is frozen)
Technical data
Parameter Value
Model BE40
Size of the clamp opening approx. 30mm
Diode test test current of 0.3mA
open-circuit voltage: 1.5VDC
Continuity test threshold <120Ω;
test current <1mA
Measuring range exceeded OL is displayed
Measuring rate 2 per second
Input impedance 7.8MΩ (VDC and VAC)
Frequency range alternating
current
50/60Hz (AAC)
Frequency range ACvoltage 50/60Hz (VAC)
Operating temperature -10°C to 50°C
(-14°F to 122°F)
Storage temperature -30°C to 60°C
(-22°F to 140°F)
Relative humidity up to 85%
Overvoltage protection categorylll, 600V
Battery 2x 1.5V batteries of typeAAA
Automatic switch-off after approx. 15minutes
Dimensions
(lengthx widthx height)
229x 48x 80mm
Weight approx. 303g

6 EN
clamp meter BE40
Measuring ranges
Measuring
level
Measuring range Resolution Accuracy
(% of the
reading)
Alternating current
20A 0A to 20.00A 0.01A ±(2.5%+
10digits)
200A 0A to 200.0A 0.1A ±(2.5%+
10digits)
1000A 0A to 1000A 1A ±(3.0%+
8digits)
DC voltage
200mV 0mV to 200.0mV 0.1mV ±(0.8%+
5digits)
2V 0mV to 2000mV 0.001V ±(1.2%+
3digits)
20V 0V to 20.00V 0.01V ±(1.2%+
3digits)
200V 0V to 200.0V 0.1V ±(1.2%+
3digits)
600V 0V to 600.0V 1V ±(1.5%+
3digits)
AC voltage
200mV 0mV to 200.0mV 0.1mV ±(1.5%+
30digits)
2V 0mV to 2000mV 0.001V ±(1.5%+
3digits)
20V 0V to 20.00V 0.01V ±(1.5%+
3digits)
200V 0V to 200.0V 0.1V ±(1.5%+
3digits)
600V 0V to 600.0V 1V ±(2.0%+
4digits)
Resistance
200Ω 0Ω to 200.0Ω 0.1Ω ±(1.0%+
4digits)
2kΩ 0Ω to 2000Ω 1Ω ±(1.5%+
2digits)
20kΩ 0kΩ to 20.00kΩ 10Ω ±(1.5%+
2digits)
200kΩ 0kΩ to 200.0kΩ 100Ω ±(1.5%+
2digits)
2MΩ 0MΩ to 2.000MΩ 1kΩ ±(2.0%+
3digits)
20MΩ 0MΩ to 20.00MΩ 10kΩ ±(3.0%+
5digits)
Scope of delivery
• 1x Device
• 1x Red measuring cable
• 1x Black measuring cable
• 2x 1.5V battery AAA
• 1x Bag/ holster
• 1x Manual
Transport and storage
Note
If you store or transport the device improperly, the
device may be damaged.
Note the information regarding transport and storage of
the device.
Transport
For transporting the device, use the bag included in the scope of
delivery in order to protect the device from external influences.
Storage
When the device is not being used, observe the following
storage conditions:
• dry and protected from frost and heat
• protected from dust and direct sunlight
• with a cover to protect it from invasive dust if necessary
• The storage temperature is the same as the range given in
the Technical data chapter.
• Remove the batteries from the device.

EN 7
clamp meter BE40
Operation
Inserting the batteries
Insert the batteries contained in the scope of delivery (2x 1.5V
AAA) before first use.
Note
Make sure that the surface of the device is dry and the
device is switched off.
Please proceed as follows to insert the batteries into the device:
1. Loosen the screw at the rear of the device and open the
cover of the battery compartment.
2. Insert the batteries into the battery compartment with
correct polarity.
3. Close the battery compartment and fasten the cover in
place.
Info
Please note that moving from a cold area to a warm
area can lead to condensation forming on the device's
circuit board. This physical and unavoidable effect can
falsify the measurement. In this case, the display
shows either no measured values or they are incorrect.
Wait a few minutes until the device has become
adjusted to the changed conditions before carrying out
a measurement.
Attaching the warning sign
Prior to initial start-up, check whether the warning sign at the
rear of the device is in your local language, if not, paste the
proper one over it. A warning sign in your native language is
supplied along with the device. Please proceed as follows to
attach the warning sign to the rear of the device:
1. Remove the label in your local language from the supplied
film.
2. Affix the label in the intended position at the rear of the
device.
Performing a voltage measurement
Warning of electrical voltage
Risk of electric shock and risk of injury!
• Make sure to observe the device's nominal voltage
specified in the technical data during the measurement.
• Check the measuring lines for damaged insulation before
each measurement.
• Test the device for proper functioning at a known voltage
source before each measurement.
• When holding onto the measuring cables, make sure not to
reach behind the protection against contact(12):

8 EN
clamp meter BE40
Measuring DCvoltage
1. Connect the black measuring cable(14) to the
COMconnection(7) and the red measuring cable(13) to
the Input connection(6).
2. Turn the rotary switch(2) to the measuring DCvoltage
position(20).
ðThe DCmeasurement indication(31) appears on the
display.
3. Tap the desired measuring points of the electric circuit
with the test probes(11) at the measuring cables with
correct polarity.
ðThe measured value will be displayed in the
measurement value display(28).
ðThe voltage unit(25) is displayed.
Info
If the minus sign(29) appears in front of the measured
value, you have confused the measuring object's plus
pole with the minus pole. Check the connection and
connect the measuring cables with correct polarity.
Example:
20
Measuring ACvoltage
1. Connect the black measuring cable(14) to the
COMconnection(7) and the red measuring cable(13) to
the Input connection(6).
2. Turn the rotary switch(2) to the measuring ACvoltage
position(21).
ðThe ACmeasurement indication(30) appears on the
display.
3. Tap the desired measuring points of the electric circuit
with the test probes(11) at the measuring cables.
ðThe measured value will be displayed in the
measurement value display(28).
ðThe voltage unit(25) is displayed.
Example:
21

EN 9
clamp meter BE40
Measuring resistance
Warning of electrical voltage
Risk of electric shock and risk of injury when
measuring the resistance!
Make sure that the power of the electric circuit is
switched off and all capacitors are completely
discharged.
• When holding onto the measuring cables, make sure not to
reach behind the protection against contact(12).
Please proceed as follows to measure the resistance:
1. Connect the black measuring cable(14) to the
COMconnection(7) and the red measuring cable(13) to
the Input connection(6).
2. Turn the rotary switch(2) to the measuring resistance
position(19).
ðThe resistance measurement indication(27) appears on
the display.
3. Tap the desired measuring points of the electric circuit
with the test probes(11) at the measuring cables.
ðThe measured value will be displayed in the
measurement value display(28).
ðThe resistance unit(26) is displayed (k=kilo=
1000ohms, M=mega=1millionohms).
19
Continuity testing
Warning of electrical voltage
Risk of electric shock and risk of injury when
checking the continuity!
Make sure that the power of the electric circuit is
switched off and all capacitors are completely
discharged.
• When holding onto the measuring cables, make sure not to
reach behind the protection against contact(12).
Please proceed as follows to check the continuity of the circuit
to be tested:
1. Connect the black measuring cable(14) to the
COMconnection(7) and the red measuring cable(13) to
the Input connection(6).
2. Turn the rotary switch(2) to the diode test/ continuity test
position(18).
3. Press the diode test/ continuity test button(8) until the
continuity test indication(34) appears on the display.
ðThe resistance measurement indication(27) appears on
the display.
ðOL is indicated on the measurement value display(28).
4. Tap the desired measuring points of the electric circuit
with the measuring cables' test probes(11).
ðIf the measured resistance value amounts to less than
120Ω, the device will emit an acoustic signal and the
measurement value display(28) will indicate the
resistance of the electric circuit.
ðIf the measured resistance value amounts to more than
120Ω, the device will not emit an acoustic signal and
the measurement value display(28) will continue to
indicateOL.
18
34 8
27

10 EN
clamp meter BE40
Diode testing
Warning of electrical voltage
Risk of electric shock and risk of injury when
testing diodes!
Make sure that the electric circuit to be tested is
switched off and de-energized.
• When holding onto the measuring cables, make sure not to
reach behind the protection against contact(12).
Please proceed as follows to test the forward voltage of a diode:
üThe electrical circuit is deactivated.
üAll capacitors are completely discharged.
1. Connect the black measuring cable(14) to the
COMconnection(7) and the red measuring cable(13) to
the Input connection(6).
2. Turn the rotary switch(2) to the diode test/ continuity test
position(18).
3. Press the diode test/ continuity test button(8) until the
diode test indication(33) appears on the display.
4. Connect the test probe(11) of the red measuring
cable(13) to the diode's anode(+).
5. Connect the test probe(11) of the black measuring
cable(14) to the diode's cathode(-).
ðThe diode's approximate forward voltage will be
displayed in the measurement value display(27).
ðThe voltage unit(25) is displayed.
ðIf OL is indicated on the display, the test probes are not
positioned correctly.
18
33
Measuring alternating current(AC)
Warning of electrical voltage
Risk of electric shock and risk of injury!
• When performing non-contact measurements, disconnect
the measuring cables(13, 14) from the device.
• For current measurements align the conductor to be tested
in the centre of the clamp.
• When holding onto the clamp meter, make sure not to
touch the clamp(1).
Info
Only measure one conductor at a time to ensure an
unambiguous measurement result.

EN 11
clamp meter BE40
Please proceed as follows to measure the alternating
current(AC):
1. Depending on the desired measuring range turn the rotary
switch(2) to position(15) for a current of up to 1000A, to
position(16) for a current of up to 200A or to position(17)
for a current of up to 20A.
ðThe ACmeasurement indication(30) lights up on the
display.
2. Squeeze the lever(10) to open the clamp and insert the
conductor to be measured centrally in the clamp.
3. Release the lever again to close the clamp of the clamp
meter.
ðThe measured value will be displayed in the
measurement value display(28).
ðThe unit for amperage(24) is displayed.
= AC MAX. 20 A
= AC MAX. 200 A
= AC MAX. 1000 A
Freezing the displayed measured value
The hold function allows you to freeze the currently measured
result in the measurement value display(28). To do so, please
proceed as follows:
1. Carry out a measurement.
2. Press the HOLD button(4).
ðAn acoustic signal is emitted.
ðThe current measured value is frozen in the
measurement value display(28).
ðThe HOLD indication(36) appears on the display.
3. Press the HOLD button(4) again to deactivate the
HOLDfunction.
436
Displaying the maximum value
üYou have carried out several measurements.
1. Press the MAXbutton(3).
ðThe MAXindication(35) appears on the display.
ðThe highest of the measured values is displayed in the
measurement value display(28).
2. Press the MAXbutton(3) again to deactivate the maximum
value display function.
335
Switching the display illumination on/off
1. Press and hold the display illumination on/off button(9) for
3seconds to switch on the display illumination.
2. With activated display illumination, press and hold the
display illumination on/off button(9) for 3seconds to
switch the display illumination back off.
Switch-off
The device switches off automatically after 15minutes of
non-use.
1. Turn the rotary switch(2) to the switch positionOFF(22).
ðThe device switches off.

12 EN
clamp meter BE40
Maintenance and repair
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make
sure that no moisture enters the housing. Do not use any
sprays, solvents, alcohol-based cleaning agents or abrasive
cleaners, but only clean water to moisten the cloth.
Repair
Do not modify the device or install any spare parts. For repairs
or device testing, contact the manufacturer.
Errors and faults
The device has been checked for proper functioning several
times during production. If malfunctions occur nonetheless,
check the device according to the following list.
Display segments are only faintly visible or flicker:
• The battery voltage is too low. Exchange the batteries
immediately.
The device displays implausible measured values:
• The battery voltage is too low. Exchange the batteries
immediately.
• The opening of the current clamp is dirty. Clean the device
as described in the Cleaning chapter.
Disposal
The icon with the crossed-out waste bin on waste
electrical or electronic equipment stipulates that this equipment
must not be disposed of with the household waste at the end of
its life. You will find collection points for free return of waste
electrical and electronic equipment in your vicinity. The
addresses can be obtained from your municipality or local
administration. For further return options provided by us please
refer to our website www.trotec24.com.
The separate collection of waste electrical and electronic
equipment aims to enable the re-use, recycling and other forms
of recovery of waste equipment as well as to prevent negative
effects for the environment and human health caused by the
disposal of hazardous substances potentially contained in the
equipment.
In the European Union, batteries and accumulators must not be
treated as domestic waste, but must be disposed of
professionally in accordance with Directive 2006/66/EC of the
European Parliament and of the Council of 6September 2006
on batteries and accumulators. Please dispose of batteries and
accumulators according to the relevant legal requirements.

Trotec GmbH
Grebbener Str. 7
D-52525 Heinsberg
+49 2452 962-400
+49 2452 962-200
www.trotec.com
Other manuals for BE40
3
Table of contents
Other Trotec Measuring Instrument manuals

Trotec
Trotec T210 User manual

Trotec
Trotec BQ16 User manual

Trotec
Trotec BM22 User manual

Trotec
Trotec BM22WP User manual

Trotec
Trotec ProMarker OEM User manual

Trotec
Trotec BW10 User manual

Trotec
Trotec BO21 User manual

Trotec
Trotec HG 125 User manual

Trotec
Trotec BC 20 User manual

Trotec
Trotec BB20 User manual
Popular Measuring Instrument manuals by other brands

Roche
Roche cobas e 411 Interface manual

Arthur Grillo
Arthur Grillo DA2000 Series Installation and operation manual

DAKOTA ULTRASONICS
DAKOTA ULTRASONICS MX-5 DL Operation manual

SWR ENGINEERING
SWR ENGINEERING SOLIDFLOW operating instructions

Armada
Armada Pro95 instruction manual

KYOWA
KYOWA DBV-120A-8 instruction manual