Turner Designs 10-AU Use and care manual

10-AU Calibration for Chlorophyll
3 Main Steps
1. Setting the basic operating level
2. Blanking the instrument
3. Calibration with a liquid primary standard that has a
known concentration
This procedure should be performed the 1st time you use the
instrument, or any time you switch to a different application,
change filters or lamps.
1. Turn on the 10AU and allow the instrument to warm up
for at least 30 minutes.
#1 Setting the Basic Operating Level
Chlorophyll (ug/l) Rhodamine WT (ppb) Range % FS (+/-5)
240 - 180 100 - 75 High 80%
180 - 160 75 - 65 High 70%
160 - 130 65 - 55 High 60%
130 - 110 55 - 45 High 50%
110 - 80 45 - 35 High 40%
80 - 60 35 - 25 High 30%
60 - 40 25 - 15 High 20%
24 - 18 10 - 7.5 Medium 80%
18 - 16 7.5 - 6.5 Medium 70%
16 - 13 6.5 - 5.5 Medium 60%
13 - 11 5.5 - 4.5 Medium 50%
11 - 8.0 4.5 - 3.5 Medium 40%
8.0 - 6.0 3.5 - 2.5 Medium 30%
6.0 - 4.0 2.5 - 1.5 Medium 20%
2.4 - 1.8 1.0 - 0.75 Low 80%
1.8 - 1.6 0.75 - 0.65 Low 70%
1.6 - 1.3 0.65 - 0.55 Low 60%
1.3 - 1.1 0.55 - 0.45 Low 50%
1.1 - 0.8 0.45 - 0.35 Low 40%
0.8 - 0.6 0.35 - 0.25 Low 30%
0.6 - 0.4 0.25 - 0.15 Low 20%
2. Table 1 is an example of how set the sensitivity of the 10-AU.
Users should determine the appropriate range and %FS
for their calibration standard.
Table 1 is an example
of how to set the
scale for a given
linear range of the
10-AU.
For this example
we are using a
20µg/L standard.
3. Prepare your standards and acetone blank in
test tubes so you are ready to set the sensitivity and
calibrate the instrument.
Be sure to clear past calibrations on screen # 2.6
4. Access screen 2.43 and set range control to manual
and range to the appropriate level for your standard.
For our example the range should be set to MED on
Screen 2.42.
Change to MAN
4. Access Screen 3.2
5. Loosen the sensitivity locking screw using allen wrench.
6. Insert and cover your 20µg/L standard.
7. Using the sensitivity knob slowly turn the knob to adjust
the %FS to the value appropriate for your standard’s
concentration. A deviation ±5% is acceptable.
Sensitivity Knob
8. When the %FS reaches the
designated value, tighten
the sensitivity locking screw.
For our example this is ≈80% of medium
The Basic Operating Level is now set, and you can remove the
standard for the 10-AU.
Setting the Basic Operating Level (aka Sensitivity)
of your 10-AU is now complete.
You can access Screen 2.43 again and return range
control to AUTO and begin calibration.
#2 Blanking the Instrument
Now that the sensitivity is set, you should blank the
instrument.
1. Access screen 2.1
2. Insert and cover your blank (acetone).
Press 1 to Run blank.
*You can also choose to have
the 10-AU subtract the blank
signal from the sample
readings on screen 2.12.
Wait for the readings to stabilize and press “0”.
Wait for 15 seconds for the instrument to blank.
If Blanking is successful you will receive a
“Blank Finished” message
#3 Calibration with Your Standard
Now that the blank is set, you should calibrate with
your standard.
1. Access screen 2.0
2. Press <2> to access screen 2.2 and input the actual
concentration of your standard. Return to screen 2.0
3. Insert and cover your standard.
4. Press 3 to run standard. Wait for the reading to stabilize.
Press the <*> key and the calibration point is set.
Wait 15 seconds and you should receive a “FINISHED”
message.
You can test/check your calibration by running your standards
and other samples or standards.
I ran the 50µg/L and got:
I ran the blank and got:
I ran a 10µg/L standard and got:

Other manuals for 10-AU

2

Other Turner Designs Measuring Instrument manuals

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs TD-107 5.0 User manual

Turner Designs

Turner Designs TD-107 5.0 User manual

Turner Designs C-FLUOR User manual

Turner Designs

Turner Designs C-FLUOR User manual

Turner Designs AquaFlash User manual

Turner Designs

Turner Designs AquaFlash User manual

Turner Designs AquaFlash User manual

Turner Designs

Turner Designs AquaFlash User manual

Turner Designs 10-AU Guide

Turner Designs

Turner Designs 10-AU Guide

Turner Designs Trilogy User manual

Turner Designs

Turner Designs Trilogy User manual

Turner Designs FluoroSense User manual

Turner Designs

Turner Designs FluoroSense User manual

Turner Designs Little Dipper 2 User manual

Turner Designs

Turner Designs Little Dipper 2 User manual

Turner Designs Enviro-T User manual

Turner Designs

Turner Designs Enviro-T User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs TD-4100XD E09 Installation and operating instructions

Turner Designs

Turner Designs TD-4100XD E09 Installation and operating instructions

Turner Designs Little Dipper User manual

Turner Designs

Turner Designs Little Dipper User manual

Turner Designs Opti-Check User manual

Turner Designs

Turner Designs Opti-Check User manual

Turner Designs 8000 Manual

Turner Designs

Turner Designs 8000 Manual

Turner Designs TD-5100 ECA User manual

Turner Designs

Turner Designs TD-5100 ECA User manual

Turner Designs C-FLUOR User manual

Turner Designs

Turner Designs C-FLUOR User manual

Turner Designs C3 User manual

Turner Designs

Turner Designs C3 User manual

Turner Designs 10-AU User manual

Turner Designs

Turner Designs 10-AU User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs Opti-Trace User manual

Turner Designs

Turner Designs Opti-Trace User manual

Turner Designs TD-20/20 User manual

Turner Designs

Turner Designs TD-20/20 User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs C3 User manual

Turner Designs

Turner Designs C3 User manual

Popular Measuring Instrument manuals by other brands

VWR avantor pH 1100 L operating manual

VWR

VWR avantor pH 1100 L operating manual

Diablo AVI-CR user manual

Diablo

Diablo AVI-CR user manual

tibelec EM3082A instructions

tibelec

tibelec EM3082A instructions

Hanna Instruments HI 83214 instruction manual

Hanna Instruments

Hanna Instruments HI 83214 instruction manual

JRI LoRa TEMP user guide

JRI

JRI LoRa TEMP user guide

cobas h 232 Quick reference guide

cobas

cobas h 232 Quick reference guide

Ditel DINOS DT203/NE Quick installation guide

Ditel

Ditel DINOS DT203/NE Quick installation guide

Raycus RFL-C30000M-CE user guide

Raycus

Raycus RFL-C30000M-CE user guide

Autometers Systems Q-180 quick start guide

Autometers Systems

Autometers Systems Q-180 quick start guide

KEPCO 520K Series instructions

KEPCO

KEPCO 520K Series instructions

Hanna Instruments HI 755 instruction manual

Hanna Instruments

Hanna Instruments HI 755 instruction manual

HKC APL-3 Series Installation, operation & maintenance manual

HKC

HKC APL-3 Series Installation, operation & maintenance manual

KPS TP5000HY user manual

KPS

KPS TP5000HY user manual

Felix Instruments F-900 Operation manual

Felix Instruments

Felix Instruments F-900 Operation manual

BC Biomedical SPO-2000 user manual

BC Biomedical

BC Biomedical SPO-2000 user manual

ATIM Cloud Wireless TH-I user guide

ATIM

ATIM Cloud Wireless TH-I user guide

Lutron Electronics AM-4220 Operation manual

Lutron Electronics

Lutron Electronics AM-4220 Operation manual

SPM VibChecker user guide

SPM

SPM VibChecker user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.