3 Main Steps
1. Setting the basic operating level
2. Blanking the instrument
3. Calibration with a liquid primary standard that has a
known concentration
This procedure should be performed the 1st time you use the
instrument, or any time you switch to a different application,
change filters or lamps.
1. Turn on the 10AU and allow the instrument to warm up
for at least 30 minutes.
#1 Setting the Basic Operating Level
Chlorophyll (ug/l) Rhodamine WT (ppb) Range % FS (+/-5)
240 - 180 100 - 75 High 80%
180 - 160 75 - 65 High 70%
160 - 130 65 - 55 High 60%
130 - 110 55 - 45 High 50%
110 - 80 45 - 35 High 40%
80 - 60 35 - 25 High 30%
60 - 40 25 - 15 High 20%
24 - 18 10 - 7.5 Medium 80%
18 - 16 7.5 - 6.5 Medium 70%
16 - 13 6.5 - 5.5 Medium 60%
13 - 11 5.5 - 4.5 Medium 50%
11 - 8.0 4.5 - 3.5 Medium 40%
8.0 - 6.0 3.5 - 2.5 Medium 30%
6.0 - 4.0 2.5 - 1.5 Medium 20%
2.4 - 1.8 1.0 - 0.75 Low 80%
1.8 - 1.6 0.75 - 0.65 Low 70%
1.6 - 1.3 0.65 - 0.55 Low 60%
1.3 - 1.1 0.55 - 0.45 Low 50%
1.1 - 0.8 0.45 - 0.35 Low 40%
0.8 - 0.6 0.35 - 0.25 Low 30%
0.6 - 0.4 0.25 - 0.15 Low 20%
2. Table 1 is an example of how set the sensitivity of the 10-AU.
Users should determine the appropriate range and %FS
for their calibration standard.
Table 1 is an example
of how to set the
scale for a given
linear range of the
10-AU.
For this example
we are using a
20µg/L standard.
3. Prepare your standards and acetone blank in
test tubes so you are ready to set the sensitivity and
calibrate the instrument.
Be sure to clear past calibrations on screen # 2.6
4. Access screen 2.43 and set range control to manual
and range to the appropriate level for your standard.
For our example the range should be set to MED on
Screen 2.42.
Change to MAN
4. Access Screen 3.2
5. Loosen the sensitivity locking screw using allen wrench.
6. Insert and cover your 20µg/L standard.
7. Using the sensitivity knob slowly turn the knob to adjust
the %FS to the value appropriate for your standard’s
concentration. A deviation ±5% is acceptable.
Sensitivity Knob
8. When the %FS reaches the
designated value, tighten
the sensitivity locking screw.
For our example this is ≈80% of medium
The Basic Operating Level is now set, and you can remove the
standard for the 10-AU.
Setting the Basic Operating Level (aka Sensitivity)
of your 10-AU is now complete.
You can access Screen 2.43 again and return range
control to AUTO and begin calibration.
#2 Blanking the Instrument
Now that the sensitivity is set, you should blank the
instrument.
1. Access screen 2.1
2. Insert and cover your blank (acetone).
Press 1 to Run blank.
*You can also choose to have
the 10-AU subtract the blank
signal from the sample
readings on screen 2.12.
Wait for the readings to stabilize and press “0”.
Wait for 15 seconds for the instrument to blank.
If Blanking is successful you will receive a
“Blank Finished” message
#3 Calibration with Your Standard
Now that the blank is set, you should calibrate with
your standard.
1. Access screen 2.0
2. Press <2> to access screen 2.2 and input the actual
concentration of your standard. Return to screen 2.0
3. Insert and cover your standard.
4. Press 3 to run standard. Wait for the reading to stabilize.
Press the <*> key and the calibration point is set.
Wait 15 seconds and you should receive a “FINISHED”
message.
You can test/check your calibration by running your standards
and other samples or standards.
I ran the 50µg/L and got:
I ran the blank and got:
I ran a 10µg/L standard and got:

Other manuals for 10-AU

2

Other Turner Designs Measuring Instrument manuals

Turner Designs AquaFluor User manual

Turner Designs

Turner Designs AquaFluor User manual

Turner Designs TD-20/20 User manual

Turner Designs

Turner Designs TD-20/20 User manual

Turner Designs C3 User manual

Turner Designs

Turner Designs C3 User manual

Turner Designs Enviro-T User manual

Turner Designs

Turner Designs Enviro-T User manual

Turner Designs AquaFlash User manual

Turner Designs

Turner Designs AquaFlash User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs 8000 Manual

Turner Designs

Turner Designs 8000 Manual

Turner Designs 10-AU User manual

Turner Designs

Turner Designs 10-AU User manual

Turner Designs C3 User manual

Turner Designs

Turner Designs C3 User manual

Turner Designs TD-107 5.0 User manual

Turner Designs

Turner Designs TD-107 5.0 User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs Little Dipper 2 User manual

Turner Designs

Turner Designs Little Dipper 2 User manual

Turner Designs C-FLUOR User manual

Turner Designs

Turner Designs C-FLUOR User manual

Turner Designs Little Dipper User manual

Turner Designs

Turner Designs Little Dipper User manual

Turner Designs C-FLUOR User manual

Turner Designs

Turner Designs C-FLUOR User manual

Turner Designs TD-5100 ECA User manual

Turner Designs

Turner Designs TD-5100 ECA User manual

Turner Designs Opti-Check User manual

Turner Designs

Turner Designs Opti-Check User manual

Turner Designs Ballast-Check 2 User manual

Turner Designs

Turner Designs Ballast-Check 2 User manual

Turner Designs Opti-Trace User manual

Turner Designs

Turner Designs Opti-Trace User manual

Turner Designs FluoroSense User manual

Turner Designs

Turner Designs FluoroSense User manual

Turner Designs AquaFlash User manual

Turner Designs

Turner Designs AquaFlash User manual

Turner Designs TD-4100XD E09 Installation and operating instructions

Turner Designs

Turner Designs TD-4100XD E09 Installation and operating instructions

Turner Designs 10-AU Guide

Turner Designs

Turner Designs 10-AU Guide

Popular Measuring Instrument manuals by other brands

SPX FLOW APV DELTA SW4 instruction manual

SPX FLOW

SPX FLOW APV DELTA SW4 instruction manual

Digitron 2000 series operating instructions

Digitron

Digitron 2000 series operating instructions

Unimax LM 63A user manual

Unimax

Unimax LM 63A user manual

Lake Shore F71 quick start guide

Lake Shore

Lake Shore F71 quick start guide

I-care TA01i User and maintenance manual

I-care

I-care TA01i User and maintenance manual

ProAir TPK 21 Functional description and operating instructions

ProAir

ProAir TPK 21 Functional description and operating instructions

Shineway Tech VLS Series user manual

Shineway Tech

Shineway Tech VLS Series user manual

globalwater FP111 user manual

globalwater

globalwater FP111 user manual

Promax PRODIG-5 manual

Promax

Promax PRODIG-5 manual

Hillrom Welch Allyn TM286 Instructions for use

Hillrom

Hillrom Welch Allyn TM286 Instructions for use

Ink Bird IBS-TH2 PLUS manual

Ink Bird

Ink Bird IBS-TH2 PLUS manual

AutomationDirect ProSense DPM2 Series user manual

AutomationDirect

AutomationDirect ProSense DPM2 Series user manual

MPFiltri ACMU user guide

MPFiltri

MPFiltri ACMU user guide

Powerline XL2G instruction manual

Powerline

Powerline XL2G instruction manual

Pro's Kit MT-7600 user manual

Pro's Kit

Pro's Kit MT-7600 user manual

LineEye LE-8500X-RT instruction manual

LineEye

LineEye LE-8500X-RT instruction manual

Almaco FLEXICOMPT AUTONOME+ operating guide

Almaco

Almaco FLEXICOMPT AUTONOME+ operating guide

Micro Direct MicroRPM operating manual

Micro Direct

Micro Direct MicroRPM operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.