Insome cases the lowest range (range 1=highest sensitivity)is
not sufficient to adjust the full switch point.This would make
another filling procedure necessary.
For this reason we recommend setting and noting the empty
switching point in all three meas.ranges.Set the meas.range
selection switch (2)to the next higher range and repeat the setting.
Also note the values for the next ranges.
7Reset meas.range selection switch (2)to the next lower range in
which the control lamp lights green [lights red].
8Fill the vessel until the probe is completely covered.
9Turn the potentiometer (1)very slowly clockwise until the control
lamp lights green [lights red].
10 Note the position of the potentiometer (1). We recommend
documenting the value of the empty switch point and the full switch
point as well as the range.
11 If the control lamp does not light green [light red], then you have to
set the meas.range switch (2)to the next higher stage and repeat
the setting with the potentiometer until the control lamp lights
green [lights red].
12 Set the potentiometer (1)to the average value of the two noted
values.
The measuring system is now ready for operation.
Empty adjustment Full adjustment
range 1
range 2
range 3
Tab.1:Note the position of the potentiometer.
Note:
Ifyou do not find the full switch point in one of the ranges,we
recommend setting the meas.range selection switch (2)to the lowest
range in which you have found the empty switch point.Set the
potentiometer (1)to the average value between empty switch point
and 10.
Vertically mounted probes
1Set mode switch (3)to mode max.
2Set meas.range selection switch (2)to range 1.
3Fill the vessel up to the requested level.
4Turn potentiometer (1)to position 10.
When the control lamp (6)lights red:set the measuring range
selection switch (2)to the next higher measuring range.
When the control lamp (6)lights green:continue with the next item.
5Turn the potentiometer (1)very slowly anticlockwise until the
control lamp (6)lights red.
Mode max.
(max.detection)
20 VEGACAP 63 • - Relay (DPDT)
6Set up
30008-EN-090604