VitalGo Total-Lift Bed VGTLB-HS 425 User manual

Elevating Patient Care
Total-Lift Bed™
ASSEMBLY, OPERATING AND MAINTENANCE MANUAL
DEALER: This manual MUST be given
to the user of the product.
USER: BEFORE using this product, read
this manual and save for future reference.
© 2008 Vitalgo Systems Ltd. All rights reserved. Vitalgo, the Vitalgo logo, Total-Lift Bed™, Joy Bath, Joy Spa, Total-Lift Armchair and Total-Lift Toilet Seat are trademarks of Vitalgo Systems Ltd.
Other names are the property of their respective owners.
For more information regarding
VitalGo products, parts, and services,
please visit www.vitalgosys.com
Bed Model:
VGTLB-HS 425

WARNING
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL.
IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS, CONTACT A
HEALTHCARE PROFESSIONAL, DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS
EQUIPMENT - OTHERWISE INJURY OR DAMAGE MAY RESULT.
THE INITIAL SET UP OF THIS BED MUST BE PERFORMED BY A QUALIFIED TECHNICIAN.
PROCEDURES OTHER THAN THOSE DESCRIBED IN THIS MANUAL MUST BE PERFORMED BY A QUALIFIED
TECHNICIAN.
FOR DEALERS ONLY - SET-UP AND ASSEMBLY INSTRUCTIONS ARE INCLUDED IN THIS MANUAL. THESE
PROCEDURES MUST BE PERFORMED BY A QUALIFIED TECHNICIAN ONLY.
PLEASE NOTE: Updated versions of this manual are available on www.vitalgosys.com.
SPECIAL NOTES
Signal words are used in this manual and apply to hazards or unsafe practices which could result in
personal injury or property damage. Refer to the text below for definitions of the signal words.
SIGNAL WORD MEANINGS:
DANGER - Danger indicates an imminently hazardous situation which, if not avoided, will result in death or
serious injury.
WARNING - Warning indicates a potentially hazardous situation which, if not avoided, could result in death or
serious injury.
CAUTION - Caution indicates a potentially hazardous situation which, if not avoided, may result in property
damage.
NOTICE
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE.
Vitalgo products are specifically designed and manufactured for use in conjunction with Vitalgo
accessories. Accessories designed by other manufacturers have not been tested by Vitalgo and are not
recommended for use with Vitalgo products.
2
WARNING / CAUTION SUMMARY
• KEEP HANDS AND FEET CLEAR OF ALL MOVING PARTS.
• DO NOT ALLOW SMALL CHILDREN ON OR NEAR BED DURING OPERATION.
• DO NOT ALLOW THIS DEVICE TO BE OPERATED BY SMALL CHILDREN.
• WHEN OPERATING THE HI-LO, KNEE, TILTING, LEG REST OR BACK FUNCTION OF THE BED, ALWAYS
ENSURE THAT THE INDIVIDUAL CONFINED TO THE BED IS POSITIONED PROPERLY WITHIN THE
CONFINES OF THE BED. DO NOT LET ANY EXTREMITIES PROTRUDE OVER THE SIDE OR BETWEEN THE
BED RAILS WHEN PERFORMING ANY FUNCTIONS.
• DO NOT USE UNAUTHORIZED SIDE RAILS.
• WARNING/CAUTION LABELS APPLIED TO THE BED OUTLINE HAZARDS OR UNSAFE PRACTICES THAT

3
COULD RESULT IN PERSONAL INJURY AND/OR PROPERTY DAMAGE.
• POWER AND WIRED HAND SET CORD MUST BE ROUTED AND SECURED PROPERLY TO ENSURE THAT
THE CORD DOES NOT BECOME ENTANGLED AND EVENTUALLY SEVERED DURING USE. MAKE SURE
ELECTRIC CORDS DO NOT GET TANGLED AROUND THE BED, SIDE RAILS OR LEGS DURING NORMAL
OPERATION OF THE BED.
• WHEN USING NASAL OR MASKED TYPE ADMINISTERING EQUIPMENT, OXYGEN OR AIR TUBING MUST
BE REMOVED BEFORE AND DURING THE NORMAL OPERATION OF BED.
THE USAGE OF SUCH EQUIPMENT IN CONJUNCTION WITH THE BED FUNCTIONS COULD RESULT IN
PERSONAL INJURY
AND/ OR PROPERTY DAMAGE
• KEEP ALL MOVING PARTS FREE OF OBSTRUCTIONS (I.E. BLANKETS/SHEETS, HEATING BLANKETS/
PADS, TUBING, WIRING, AND OTHER TYPES OF PRODUCTS)
Meaning of the Safety Symbols
In these instructions the following safety symbols are used:
Warning about injuries to persons
Dangerous voltage. Life Threatening.
General danger. Injury or life hazard.
Warning about property damages
Possibility of damage to motor, material or environment.
OTHER SYMBOLS
Useful tip. For easier operation or better understanding of the unit.
Company name and address
Manufacturing date
IPX4 Protected against splashing water
Consult instruction manual
The item meets all the essential requirements of the relevant European Directive(s).
This Symbol applies to European Union only as is defined as “Separate collection for
electrical and electronic equipment waste per Directive 2002/96/EC in the European Union”
Type B device
Reference to the hazards in the instruction manual
Max Patient Weight
Safe Working Load - indicates maximum bed weight handling capacity
The safety symbols used does not replace the text of the safety notes.
Therefore read the safety instructions and follow them exactly!
Danger!
Warning!
Caution!
Important!
!
!

Table Of Contents
WARNING ............................................................................................................................................................ 2
SPECIAL NOTES.................................................................................................................................................... 2
NOTICE................................................................................................................................................................. 2
WARNING /CAUTION SUMMERY ....................................................................................................................... 2
MEANING OF THE SAFETY SYMBOLS................................................................................................................. 3
PREFACE............................................................................................................................................................... 7
ABOUT THE VITALGO TOTAL-LIFT BED™ .......................................................................................................... 8
UNPARALLELED SAFETY................................................................................................................... 8
GENERAL GUIDELINES & SAFETY INSTRUCTIONS .............................................................................................. 9
OPERATING INFORMATION .............................................................................................................. 9
ENTRAPMENT WARNING ................................................................................................................. 9
REPLACEMENT PARTS/ACCESSORIES GUIDELINES .......................................................................... 9
ELECTRICAL GUIDELINES ................................................................................................................ 10
REPAIR OR SERVICE INFORMATION ............................................................................................... 10
RADIO FREQUENCY INTERFERENCE............................................................................................... 11
WEIGHT LIMITATIONS..................................................................................................................... 11
BEFORE PUTTING THE BED INTO OPERATION FOR THE FIRST TIME: ........................................... 11
WARNINGS FOR USERS AND CAREGIVERS .................................................................................... 11
DEFINITIONS OF INVOLVED PERSONS............................................................................................ 12
SAFETY INSTRUCTIONS FOR TECHNICIANS ................................................................................... 12
SAFETY INSTRUCTIONS FOR USERS AND CAREGIVERS................................................................. 13
VGTLB-HS BED DESCRIPTION............................................................................................................................ 14
PRODUCT DESCRIPTION.................................................................................................................................... 15
NORMAL OPERATING PROCEDURE ............................................................................................... 15
SPECIAL FEATURES.......................................................................................................................... 15
CLASSIFICATION .............................................................................................................................. 15
TECHNICAL SPECIFICATIONS ............................................................................................................................. 16
STRUCTURAL DESIGN ..................................................................................................................... 17
DESCRIPTION OF MATERIALS ......................................................................................................... 19
LABEL LOCATIONS .......................................................................................................................... 19
OPERATION........................................................................................................................................................ 20
4

WIRED HAND SET ........................................................................................................................... 20
CONTROL FUNCTIONS....................................................................................................................................... 21
MANUAL FUNCTIONS........................................................................................................................................ 21
ADJUSTMENT OF THE BACK REST.................................................................................................................... 21
RAISING & LOWERING LEG-REST ................................................................................................... 21
RAISING & LOWERING THE BED HORIZONTALLY.......................................................................... 22
FOR LOWEST SLEEPING POSITION ................................................................................................. 22
WHEELS POSITION .......................................................................................................................... 22
AUTOMATIC FUNCTIONS ................................................................................................................................. 22
SEATING WITH LEGS DOWN (WITH MEMORY FUNCTION) .......................................................... 22
SEATING WITH LEGS UP (WITH MEMORY FUNCTION) ................................................................. 22
SLEEPING POSITION ........................................................................................................................ 23
MEMORY FUNCTION ...................................................................................................................... 23
NURSE FUNCTIONS............................................................................................................................................ 23
SIDE RAIL CONTROL PANEL OPERATION.......................................................................................................... 24
INITIALIZATION OF THE BED............................................................................................................................. 24
STRAPPING SYSTEMS(OPTIONAL)..................................................................................................................... 24
NON AMBULATORY PATIENT STRAPPING SYSTEM ........................................................................................ 25
RETRACTABLE HEAD BOARD............................................................................................................................ 25
MATTRESS ATTACHMENT ................................................................................................................................. 26
TROUBLESHOOTING.......................................................................................................................................... 27
MAINTENANCE.................................................................................................................................................. 27
MAINTENANCE CHECKLIST ........................................................................................................... 28
ELECTRICAL INSPECTION AND MAINTENANCE............................................................................. 28
RECHARGEABLE BATTERIES MAINTENANCE................................................................................. 28
CLEANING AND DISINFECTING ...................................................................................................... 29
TOTAL-LIFT BED™ ACCESSORIES (OPTIONAL) ................................................................................................. 30
SIDE RAILS ....................................................................................................................................... 30
HOYER LIFT LEGS EXTENSION ........................................................................................................ 30
IV POLE ............................................................................................................................................ 30
TRAPEZE .......................................................................................................................................... 30
BEDSIDE TABLE ............................................................................................................................... 30
TOTAL LIFT BED – COMBINATION WEIGHING SYSTEM ................................................................................... 31
5

6
OPERATION ..................................................................................................................................... 31
TLB – BED SCALE PATIENT WEIGHT SYSTEM.................................................................................................... 33
OPERATION ..................................................................................................................................... 33
FOOT-LIFTER SCALE (PRESSURE SENSOR)– (OPTIONAL) .................................................................................. 34
OPERATION ..................................................................................................................................... 34
USE OF THE FOOT LIFTER-SCALE.................................................................................................... 34
LOW AIR LOSS MATTRESS SYSTEM – (OPTIONAL)........................................................................................... 35
EXPLANATION OF SYMBOLS USED ON THIS DEVICE .................................................................... 35
MAIN FEATURES OF THE VG-TLB-LALM4....................................................................................... 36
SUPPORT SURFACE (MATTRESS / OVERLAY)................................................................................. 36
OPERATING INSTRUCTIONS............................................................................................................ 37
INCLINOMETER – ANGLE SENSOR.................................................................................................. 39
LIMITED WARRANTY......................................................................................................................................... 40
ABOUT VITALGO SYSTEMS LTD. ....................................................................................................................... 41

7
Preface
Dear customer,
Thank you for the confidence you have shown in VitalGo Systems products.
The Total-Lift Bed™ has been factory checked for electrical safety. All functions have been checked and ar oper-
ating perfectly.
Please take your time to read this manual carefully before operating the bed. This manual applies normal use of
the bed and assembly and maintenance. It is also a practical reference guide. Keep the instructions in a handy
place. These instructions provide information for the operator and user regarding the convenient handling and
safe operation of the bed.
We hope that the VitalGo Total-Lift Bed™ alleviates any problems that you or your caregiver experience in get-
ting in and out of bed.
VitalGo Systems LTD.

About The Vitalgo Total-Lift Bed™
The Total-Lift Bed™ sets new standards in patient and care givers treatment and comfort.
The Total Lift Bed™ is a Bed and so much more giving many advantages over conventional beds.
The revolutionary design allows the bed to raise the user to a fully standing position and all positions in be-
tween including chair positions.
Among all its clinical obvious benefits, the Total Lift Bed™ is the best solution for safe patient handling and safe
lifting.
The Total Lift bed™ is ideal for –
• Safe patient Handling and Safe Lifting
• Hospitals
• Nursing Homes
• Assisted Living Facilities.
• Long Terms Acute Care centers.
• Rehabilitation centers.
• Sports rehabilitation centers.
• Homecare – Aging in Place.
• Fall Prevention
With just a press of a button you can bring the patient to any position desired.
The Patented (pending) technology of the total Lift Bed™, moves the Foot Lifter ™ towards the feet of the pa-
tient / user, and only when making contact will start tilting him, thus preventing any danger of sliding in bed.
While the bed tilts, when in proper angle the Foot Lifter ™ will gradually move down so when the bed is in fully
standing position the Foot Lifter will be parallel to floor allowing to simply walk out of bed, with or without
assistance.
The patient can also be just tilted, for treatment, in any desired angle, for as long as needed.
The Total Lift Bed™ has all possible bed and chair positions.
The Total Lift Bed™ can come with powered and non powered mattress in any grade.
Weight Bearing control – The Total Lift Bed is the only real true “Weight Bearing control device” With the Foot
Lifter scale (optional), there is a full weight Bearing control knowing the exact amount of pressure imposed on
the legs/ body of the user, which can be increased or decreased by changing the beds angle.
The Total Lift Bed™ has all safety features and is FDA and CE registered conforming with the EN/ IEC 60601-2-38
and the Assessment Guidance to Reduce Entrapment.
UNPARALLELED SAFETY
Safety is the top priority of all Vitalgo products. The Total-Lift Bed™ incorporates numerous safety features and
mechanisms and meets or exceeds the highest standards for medical electric beds.
The Total Lift Bed is FDA and CE Registered. The Total Lift Bed has been tested and conformed with EN\IEC
60601-1:1996, EN\IEC 60-601-2-38:1996 and UL 60601-1:2003.
The integrated safety features include:
• Exclusive self-adjusting Foot-Lifter™ – Automatically moves up to rest under the user’s feet prior to the bed
tilting into an upright position, preventing the user from sliding down the bed. Embedded micro-switches
ensure the user is properly positioned and that when the bed is in an upright position, the user’s feet are at
ground level.
• Safety Rails (optional) – Can be added to the upper and lower part of the bed, on one or both sides.
Vitalgo has adopted all FDA clinical guidance to reduce Entrapment.
• Electronic braking – Hidden casters are electronically lowered to allow mobility and raised to secure stability;
they are automatically raised before the bed can be tilted or the sitting/reclining positioning adjusted.
8

General Guidelines & Safety Instructions
OPERATING INFORMATION
DO NOT use near explosive gases.
Keep the product a minimum of 12” away from any direct heat source.
Close supervision is necessary when this product is used by or near children or people with disabilities.
Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact a qualified
technician for further instruction.
After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely.
DO NOT let any individual underneath the bed or in between the raised bed frame components at anytime.
When bed is not to be used for an extended period, unplug electric bed from the wall outlet.
When operating the bed body weight should be evenly distributed over the surface of the bed. DO NOT lay, sit
or lean in such a way that your entire body weight is placed only on raised head or foot sections of the bed. This
includes when assisting the user in repositioning or transferring in or out of bed.
This Warning DOES NOT apply to the Foot Lifter™.
When operating/moving beds, always ensure that the individual utilizing the bed is positioned properly within
the confines of the bed. DO NOT let any extremities protrude over the side or between the bed rails when per-
forming these functions.
If the unit is not working properly, call a qualified technician to examine the unit and repair it.
Keep all moving parts, including the main frame, mattress deck (head and foot sections) and all drive shafts,
free of obstruction (i.e. Blankets/sheets, heating blankets/pads, tubing, wiring, etc. and other types of products
using electric cords which may get tangled around the bed, side rails or legs) during operation of the bed.
ENTRAPMENT WARNING
Proper patient assessment and monitoring, and proper maintenance and use of equipment is required to re-
duce the risk of entrapment. Variations in bed rail dimensions, and mattress thickness, size or density could
increase the risk of entrapment.
Visit the FDA web site at http://www.fda.gov to learn about the risks of entrapment.
After any adjustments, repair or service and before use, make sure all attaching hardware is tightened
securely. Assist rails with dimensions different from the original equipment supplied or specified by the
bed manufacturer may not be interchangeable and may result in entrapment or other injury. Mattress
MUST fit bed frame and assist rails snugly to reduce the risk of entrapment.
REPLACEMENT PARTS/ACCESSORIES GUIDELINES
VitalGo products are specifically designed and manufactured for use in conjunction with VitalGo accessories.
Accessories designed by other manufacturers have not been tested by VitalGo and are not recommended for use
with VitalGo products.
Use ONLY VitalGo’s Total-Lift Bed™ supplied mattress and connect it to the bed frame according to the instruc-
9

tions in this manual (see page 32)
Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails. These bed rails are
used for the purpose of preventing an individual from inadvertently rolling out of bed. The bed rails are NOT
intended nor may be used for restraint purposes. If an individual is capable of injuring himself/herself, a physi-
cian or a healthcare professional should be consulted for alternative means of safe restraint.
DO NOT use the side rails as push handles for moving the bed.
Once patient assessment concludes that the patient’s condition increases the chance of entrapment, the bed
MUST be in the flat position when left unattended.
Trapeze units must not be used in conjunction with the bed’s operation.
Trapeze units are to be used only for assisting the patient in repositioning or transferring into or out of bed and
must be moved away from the bed as soon as the patient finished using them.
Replacement mattresses and bed side rails with dimensions different than the original equipment supplied or
specified by the bed frame manufacturer are not interchangeable.
Variations in bed side rail design, width and thickness or firmness of the mattress could cause/contribute to
entrapment. Use only authorized VitalGo replacement parts and/or accessories otherwise the warranty is void.
VitalGo will not be responsible for any damage or injury that may result.
ELECTRICAL GUIDELINES
DANGER
Connect the bed directly to an electrical outlet. Do not use extension cords and/or multiple outlet extension
cords.
Use of three prong adapters can result in improper grounding and present a shock hazard to the user.
NEVER operate if the unit has a damaged cord or plug. If it is not working properly, call a qualified technician
for examination and repair.
Keep all electrical cords away from heated or hot surfaces.
Ensure all cables and cords are routed such that they will not become entangled or pinched. Otherwise damage
or injury may result.
DO NOT unplug power cord from junction box.
The pendant and power cords must be routed and secured properly to ensure that the cords DO NOT become
entangled, pinched and/or severed during operation of the electric bed.
Refer servicing to qualified personnel only. Grounding reliability depends upon a properly grounded wall outlet.
REPAIR OR SERVICE INFORMATION
DO NOT open assemblies such as the motors, pendant, junction boxes or gear boxes. No user serviceable parts
are inside. Only qualified technicians are permitted to repair these parts. If unqualified individuals perform any
work on these beds, the warranty is void.
Unplug the power cord from its power source before performing any maintenance on the bed.
10

DO NOT unplug the power cord from the junction box. Damage to cord will result.
RADIO FREQUENCY INTERFERENCE
Electronic equipment may be influenced by Radio Frequency Interference (RFI). Caution should be exercised
with regard to the use of portable communications equipment in the area around such equipment. If RFI causes
erratic behavior, unplug the electric bed IMMEDIATELY. Leave unplugged while transmission is in progress.
WEIGHT LIMITATIONS
The total weight limit of the VitalGo Total-Lift Bed™ is 425 pounds (195kg) patient weight.
DO NOT permit more than one person on/in the bed at any time.
When the bed is an operation body weight should be evenly distributed over the surface of the Total-Lift Bed™.
DO NOT lay, sit or lean in such a way that your entire body weight is placed only on raised head or foot sections
of the bed.
This includes when repositioning or transferring in or out of bed.
This Warning DOES NOT apply to the Foot Lifter™.
BEFORE PUTTING THE BED INTO OPERATION FOR THE FIRST TIME:
Read these instructions carefully in order to avoid damage or incorrect operation. Before using the Total-Lift
Bed™, the user shall ensure that it is in Proper working order and free of defects, and be aware of the instruc-
tion manual. This applies also to the accessories.
The VitalGo Total-Lift Bed™ complies with all requirements of the guidelines for medical products. It is classi-
fied as a Medical Equipment Electrical Operated Hospital Bed in accordance with standard IEC60601(US) and
EN60601-2-38(EU).
The VitalGo Total-Lift Bed™ was checked by an independent examining institute. As with all technical, electrical
devices improper handling can lead to damage and/or injury.
Observe your obligations as operator in accordance with medical devices Operators guidelines for Medical Prod-
ucts in order to ensure a dependable and safe operation of this medical device without endangering patients,
users and third parties.
This manual contains safety guidelines, which must be observed. All persons who work with the VitalGo Total
Lift Bed™ must familiarize themselves with these instructions and follow the safety guidelines.
WARNINGS FOR USERS AND CAREGIVERS
Position the power cable in such a way that during normal operation of the bed the cable will not be stretched,
driven over or interfere with moving parts of the bed.
Before any relocation of the bed it is imperative that the power cable is pulled from the wall socket and that
the cable cannot fall or be dragged over the floor.
Check the power cord regularly for damage (abrasions, exposed wires, kinks, pressure marks, etc.) In particular:
• After every larger mechanical strain (e.g.: Rolling over the power cord with the bed or with an equipment
cart.
• After a strong pulling or bending load caused by the bed rolling away with the power cord still attached to
the wall outlet.
11

• After relocating/moving and before plugging in the power cable.
• During prolonged operating by the user.
Do not use the area under the Total-Lift Bed™ as a “parking spot” for any utensils.
Lower the bed to sleeping positions (see Page 24) and lift side rails if present when leaving the patient unat-
tended. This reduces the risk of injury to the patient getting in and out of the bed.
Keep the wired hand set safe from accidentally falling during non-use (Hang it on the its holder). Take care that
the cable is not damaged by the moving parts of the bed.
To protect the patient and especially children from unintentionally adjusting the Wired hand set, place it be-
yond their reach.
Place the wired hand set out of the patient’s reach when:
• The patient is not able to control the bed safely or he is unable to get out of dangerous positions without
help.
• When bed has functioning problem.
• Unsupervised children are in the room with the Total-Lift Bed™.
Initialization/ assembly functions can only be preformed by a properly trained personal When the Total-Lift
Bed™ is located in institutions (e.g.: Hospitals, nursing homes etc.).
When in home use the Total-Lift Bed™ Initialization/ assembly functions should only be used by a qualified tech-
nician.
Definitions of involved persons
In this manual the following persons are involved:
TECHNICIAN
Technicians (e.g.: Medical Supply Houses, Dealers and representatives, health insurance companies) are natural
persons or legal entities, who utilize the Total-Lift Bed™ or authorize its use.
The technician is responsible for instructing the user in the proper use of the unit.
CARE GIVER
Caregivers are persons who as a result of training and experience are authorized to operate the Total-Lift Bed™.
The care giver can recognize and avoid possible dangers, and judge the clinical condition of the patient.
USER
In this manual a the user is an infirm or injured or disabled person who uses the Total-Lift Bed™ (or others who
might find it useful).
SPECIALIST STAFF
Specialist staff are employees of the suppliers who are authorized as a result of their education and training to
deliver, assemble, dismantle and transport the Total-Lift Bed™. Furthermore, they are familiar with the regula-
tions for cleaning and disinfecting the unit.
Safety instructions for Technicians
Before first use, instruct every user in the safe operation of the Total-Lift Bed™ in accordance with these safety
instructions, which must be supplied along with the unit.
Call attention to the dangers of improper use of the unit, especially in regards to the electrical drives and side
rails.
12

When the Total-Lift Bed™ is located in institutions (e.g.: Hospitals, nursing homes etc.) The unit should only be
operated by properly trained personnel.
When in home use the Total-Lift Bed™ should only be used by users who read and understood thoroughly the
operation manual.
Make sure that representatives are also familiar with the operation of the VitalGo Total-Lift Bed™.
During long term use it is recommended that the Total-Lift Bed™ be inspected for function and visible damage
at regular intervals (Recommendation: yearly) (see Maintenance Checklist - Page 35).
Connect the bed directly to an electrical outlet. Do not use extension cords and/or multiple outlet extension \
cords.
When attaching other equipment (e.g. compressors for positioning systems) make sure that they are mounted
and can function in a secure and safe manner.
Do not put multiple outlet electrical extensions under the bed. Leaking liquids can cause a fire hazard.
Pay special attention to: Securely locating of all wiring, cables, tubes, etc.
For further information please contact the manufacturer of the accessories or VitalGo Systems directly.
Make sure that your personnel follow the safety instructions.
Safety instructions for users and caregivers
Let the technician instruct you in the safe use of the bed.
Before each use make sure that the bed is in proper and faultless condition.
Take care that no obstacles, such as furniture or slanted ceilings interfere with the adjustment functions (See
clearing page 19-20).
Important!
Patients under 5.5’ (170cm) tall, should make sure their feet touch the Foot Lifter™ before entering tilting posi-
tion.
Pay attention that when using additional electrical components, such as patient lifts, reading lights or compres-
sors for positioning systems that their electrical cords do not get entangled or damaged by the moving parts of
the bed.
Connect the bed directly to an electrical outlet. Do not use extension cords and/or multiple outlet extension
cords.
Make sure that when attaching other appliances (e.g. Compressors for positioning systems) that a safe mount-
ing and function is guaranteed.
Remove moveable items from the bed before starting the tilting operation.
Make sure you can see the hand control functions.
Do not put multiple electrical outlets under the nursing bed. Leaking liquids can be a fire hazard.
Pay special attention to securely locating all wiring, cables, tubes, etc.
For further information please contact the dealer or person from which you purchasedhte bed or VitalGo Sys-
tems directly.
13

VGTLB-HS Bed Description
14

Take the bed out of operation if damage or a malfunction is suspected: Immediately unplug the elec-
trical plug from the wall outlet; Call a qulified technician. do not use the bed until a technician has
innspected the bed and approved it for use.
Product description
NORMAL OPERATING PROCEDURE
The Total Lift Bed, Hereafter called bed, is a, one of its kind, lifting Bed which can take the user to a standing
position, seating position with legs down and up and many other positions as described in this manual.
The Total Lift Bed has a state of the art programmable system, which allows many functions and possibilities, for
the safety, comfort and independence of the user and Care giver.
The Total Lift Bed is the perfect bed to assist the recommendation of OSHA for minimal lift and caregiver injury
prevention.
The Bed can be used by one Care Giver and by the user himself, if capable.
The Total Lift Bed is the perfect solution for Aging in Place allowing the users to age with comfort dignity and
independence.
Total Lift Bed safe working load is 616lb (280 kg) and max patient capacity is 425lb (193 kg).
For users under 5’ (150cm), please use Tilting Function only up to 45 degrees, as the foot support may not
support user, if not placed low enough on the support surface.
The Bed can be used with or without a strapping system (see page 30 ) according to the user’s medical
condition.
The Total Lift Bed may only be operated under the operating instructions laid out in this instruction manual.
Any other use is deemed to be inappropriate.
SPECIAL FEATURES
HEIGHT ADJUSTMENT
Electrical height adjustment of the reclining surface from approx. 18”(46cm). to 29.5”(75cm).
Electrical movement of the Foot Lifter™ is from approx. 8”(20cm). into the bed (over the mattress) to about 4”
(10cm) outside the bed.
LYING/SITTING POSITION
Electrical adjustment of the backrest from 0° to about 65°.
Electrical adjustment of the thigh rest from 0° to about 7.5° (legs down)
and from 0° to about 25° (legs up).
Electrical adjustment of the Leg rest from 180° to about 50°.
IN STANDING POSITION
The bed joints work in synchronization bringing the bed to around 82°.
The bed is driven with four castors which become functional only when bringing the bed to “WHEELS” position
accessible through the hand set.
Reclining surface of the Total-Lift Bed™ is 77.5”x35”(197 x 89cm), divided into four areas;
Outside dimensions are 83.5”x39”(212 x 99cm).
Side rails on both sides can be lowered.
CLASSIFICATION
EU - The unit is classified as Class I with B type of applied part, continuously operated, movable and without
signal input output parts. Water resistance comply with IPX4 the device is not intended for use in presence of
flammable mixtures.
FDA - Device Class II, adjustable electric hospital bed, AC-powered regulation number 880.5100
The tested product satisfies the requirements of EN/IEC 60601-1:1990; UL 60601-1:2003, EN/IEC 60601-2-38:1996
15

16
Technical Specifications
Bed Dimensions:
• Length:
Overall Length (in regular position).............................83.5” (212 cm)
Overall Length – Fully extended (Foot lifter out)....87.5” (222cm)
• Width:
Overall Width – With Side -Rails................................40.5” (103 cm)
Overall Width – Without Side- Rails...........................39” (99cm)
• Height:
Low Position – deck to floor........................................18” (45.7cm)
High Position – deck to floor.........................................29.5” (75cm)
• Clearance:
From frame to floor.........................................................1.25” (3.18cm)
In Standing Position:
From top of Bed to floor...........................................80” (203.20cm)
Length including Foot Lifter..........................................87” (221cm)
• Mattress Dimensions– (Foam):
Length..........................................................................77.5” (197cm)
Width..............................................................................35” (89cm)
Height.............................................................................6” (15.20cm)
* Air Loss Mattress (Optional)
• Casters & Breaking System:
Diameter.........................................................................5” (12.70cm)
Smart BreakTM – Casters are lifted from floor when not in “Wheels” position
Indevidual mechanical Break on 3 casters
Directional Lock – On one caster for easy maneuvering of bed.
• Foot Lifter:
Moves over Mattress 7.90” (20cm) towards Head Rest.
Moves out of Bed – 3.95” (10cm).
Moves up in Chair Position until contact with feet or
Max. of 7.90” (20cm).
• Positions:
Sleeping
Chair Position (Legs down with foot lifter support) with reclining.
Leg Rest - Up & Down.
Back Rest Up.
Reverse Trend.
Tilting positions up to Standing Position.
• Control:
Control by Smart pre-programmed Hand Set.
Tilting function is locked for Nurse / Care giver use only.
Automatic functions with memory.
Hand Set works with constant press only.
Additional control, for basic functions, on side-rail,for patient and care giver.
• Strapping System:
One strap in Waist area.
Up to three straps in lower extremities area

17
77.5” (197cm)
87” (221cm)
35.4” (89cm)
80” (203.20cm)
18” (45.7cm)
40.5” (103cm)
39”
(99cm)
3.95” (10cm)
7.90” (20cm)
83.5” (212cm)
29” (75cm)
1.25” (3.18cm) 5” (12.5cm)

18
• Maximum user weight: 425 lb.(193kg)
• Maximum Safe loading capacity – 616 lb (280kg)
• Casters/ brakes: Electrically controlled and individual mechanical break on each caster.
• Electronic Wired hand set: Integrated 12 functions + caregiver function
• Electricity requirements: Voltage - 120V~ AC 230V~ AC
Current - 5A Max 3A Max
Frequency - 60HZ 50HZ
Power Consumption - Max. 90W
• Environmental conditions : Ambient Temperature in Operation : +10oC (50oF) to +40oC (104oF)
Ambient Temperature in Transport : +5oC (41oF) to +50oC (122oF)
Relative Humidity Range: 30% to 75%
Atmospheric Pressure Range of 700 hPa to 1060 hPa.
• Battery backup: Optional
Reclining surface frame
The reclining surface frame has four sections: a movable backrest, a moveable thigh rest, a movable leg rest
and a movable Foot Lifter™. All parts can be adjusted by electric motors. The horizontal height of the reclin-
ing surface can be adjusted and inclined. All adjustment functions are controlled by a Wired hand set and basic
controls are available through the hand switch, located on the handles (when applicable).
The mattress base of the Total-Lift Bed™ consists of washable HPL;
The Total-Lift Bed™ has side rails on both sides which can be raised as a barrier or removed when not needed.
This safeguards the patient from accidentally falling out of the bed.
The rails are designed according to FDA and CE recommendation to avoid entrapment (Hospital Bed System
Dimensional and Assessment Guidance to Reduce Entrapment).
Electric adjustment system
The electrical adjustment system of this bed is error protected, flame-retardant (V0) and consists of An external
transformer unit which consists of a power cable, the transformer, and a low voltage connection cable.
The transformer creates a 24 volt low voltage, which is safe for patient and user. The transformer supplies all
drive motors with the 24 volt safety low voltage. The connecting socket on the bed frame is moisture proof.
22” (56cm)
!
!

Label Locations
19
SN XXXX
DESCRIPTION OF MATERIALS
The bed is mostly constructed out of steel sections, their surfaces are mostly powder coated and some
are painted. Some steel parts have black oxyde coating and chrome-nickel coating. There is also use of
aluminum.
The Washable HPL are made of HPL (High Pressure Laminate).
Foot-Lifter™ is made of ABS plastic.
Side Rails and Headboard are ABS.
All surfaces are harmless to skin contact.

59
31
246 8
OPERATION
CAUTION
Allow a slight pause between adjustments and avoid pressing multiple buttons at the same time unless indicat-
ed. If wired hand set buttons are depressed too rapidly or wrong button combinations are pressed at the same
time, the desired feature may not activate.
Simply release the wired hand set button, permit a slight pause and then activate the next operation.
Before placing the bed into use, test it by operating it through all phases of its operation.
If any problems arise during the test, recheck all electrical connections and mechanical hook ups.
WARNING
DO NOT place wired hand set under or between objects. This may unintentionally press the buttons and may
cause injury or damage.
Side rail control
6
8
10
12
CPR
1
3
5
7
9
11
2
4
6
8
10
12
Wired hand set
20
Table of contents
Other VitalGo Medical Equipment manuals
Popular Medical Equipment manuals by other brands

fiagon
fiagon Navigation Unit Cube 4D Instructions for use

Steris
Steris 4085 Operator's manual

Otto Bock
Otto Bock 28U22 Series Instructions for use

Seca
Seca CT3000/P80 How-to guide

NURSE ROSIE
NURSE ROSIE Rosebud VC troubleshooting guide

Medical Developments International
Medical Developments International Space Chamber Plus Usage instructions