Quick Start Manual for Assistina 3x3
/
3x2
This Quick Start Manual is NOT a substitute for the instructions for use Assistina!
If you have any questions or queries please always pull the instructions for use for help.
1. Insert / take out cartridge.
The cartridges must be at room temperature
when they are inserted.
The cartridge spike is sharp.
Be careful not to injure yourself on the
projecting spike.
a)Insert the cartridge.
> Slide in the cartridge.
> You should hear the
cartridge engage.
b)Take out the cartridge.
> Press the button for
releasing the cartridge.
> Pull out the cartridge.
4. Carry out initial filling after cartridge change.
a)Close the cover. Press
the Start button
for approx. 5 sec..
The initial filling
process will take
approx. 2 minutes.
If the Assistina unit has been out of use for an extended
period of time (approx. 14 days), repeat the initial filling
process.
The Start LED light
up green. The
device is ready for
use.
a)Firmly tighten the
union nut anti-
clockwise by hand
(leak tightness).
5. General notice
W&H Dentalwerk Bürmoos GmbH t + 43 6274 6236-0
Ignaz-Glaser-Straße 53, Postfach 1 f+ 43 6274 6236-55
5111 Bürmoos, Austria [email protected] wh.com
3. Insertion / removal of the external cleaning ring
(Assistina 3x3)
a)Insertion of the
external cleaning ring.
> Push the connecting
hoses parallel onto the
connection bushes.
> Clamp the connection
hoses into the hose
guides.
b)Removal of the
external cleaning ring.
> Take out the
connecting hoses from
the hose guides.
2. Open the cover.
a)Open the cover.
Overview error messages
> Standby mode:
Press the Start button.
> Device not connected to the
mains supply:
Connect the power supply.
> Electronics defective:
Send to an authorized W&H
service partner.
> Rotate the care attachment
manually.
> Cartridge
W&H Service Oil F1, MD-200
or
W&H Activefluid, MC- 1100
empty or not in the unit:
a) Replace / insert the
cartridge.
b) Press the Start button
1x.
c) Carry out the initial filling.
> Electronics defective:
Send to an authorized W&H
service partner.
> Air supply fails.
a) Re-establish the
main air supply.
b) Press the Start button 1x.
c) Carry out the initial filling
> Cover opened during the
preparation process:
a) Close the cover.
b) Press the Start button 2x.
> Waste drawer full:
Empty the waste drawer
Press the Start button 1x.
> Waste drawer open or not in
the unit:
Close or insert the waste
drawer.
> Waste drawer opened
during the preparation
process:
a) Close the waste drawer.
b) Press the Start button 2x.
> Preparation process not
correctly completed:
Press the Start button 2x.
> Start button pressed during
the preparation process:
Press the Start button 2x.
Sensor dirty or covered:
> Wipe-disinfect the sensor.
> Check position of hose
external cleaning ring and
clip in place if necessary.
> Press the »Start« button 2x,
the preparation process will
then start.
50741 AEN Rev. 003 / 10.02.2016
Subject to alterations.
W&H Dentalwerk Bürmoos GmbH t + 43 6274 6236-0
Ignaz-Glaser-Straße 53, Postfach 1 f+ 43 6274 6236-55
5111 Bürmoos, Austria [email protected] wh.com
Quick Start Manual for Assistina 3x3
/
3x2

This manual suits for next models

1

Other W&H Medical Equipment manuals

W&H PROXEO HP-44M User manual

W&H

W&H PROXEO HP-44M User manual

W&H Implantmed SI-1010 User manual

W&H

W&H Implantmed SI-1010 User manual

W&H piezomed PLUS SA-435 M User manual

W&H

W&H piezomed PLUS SA-435 M User manual

W&H LINA MB17 User manual

W&H

W&H LINA MB17 User manual

W&H Implantmed User manual

W&H

W&H Implantmed User manual

W&H Assistina One User manual

W&H

W&H Assistina One User manual

W&H Implantmed Plus SI-1010 User manual

W&H

W&H Implantmed Plus SI-1010 User manual

W&H Implantmed Plus SI-1010 User manual

W&H

W&H Implantmed Plus SI-1010 User manual

W&H piezomed SA-320 User manual

W&H

W&H piezomed SA-320 User manual

W&H implantmed SI-923 User manual

W&H

W&H implantmed SI-923 User manual

W&H MN-111 User manual

W&H

W&H MN-111 User manual

W&H piezomed classic User manual

W&H

W&H piezomed classic User manual

Popular Medical Equipment manuals by other brands

DirectHome MEDICAL Transcend quick guide

DirectHome MEDICAL

DirectHome MEDICAL Transcend quick guide

Medical Quant Handy Cure S' user manual

Medical Quant

Medical Quant Handy Cure S' user manual

BIOFLEX Arthritis manual

BIOFLEX

BIOFLEX Arthritis manual

Tunstall FlamencoIP+ manual

Tunstall

Tunstall FlamencoIP+ manual

GDS 10+ OXYMED quick start guide

GDS

GDS 10+ OXYMED quick start guide

Interacoustics Titan operating manual

Interacoustics

Interacoustics Titan operating manual

BMC Polymate YH-1000C user manual

BMC

BMC Polymate YH-1000C user manual

OMNI ARC HIP SYSTEM SURGICAL TECHNIQUE

OMNI ARC

OMNI ARC HIP SYSTEM SURGICAL TECHNIQUE

Nonin Avant 4000 Operator's manual

Nonin

Nonin Avant 4000 Operator's manual

BMC E-20C-H-O user manual

BMC

BMC E-20C-H-O user manual

PLUSOPTIX PowerRef 3 instruction manual

PLUSOPTIX

PLUSOPTIX PowerRef 3 instruction manual

MIR Spirotel Service guide

MIR

MIR Spirotel Service guide

brewer BrewerElement manual

brewer

brewer BrewerElement manual

LEMI AYURVEDA V Original instructions

LEMI

LEMI AYURVEDA V Original instructions

Erbe ESM 2 Service manual

Erbe

Erbe ESM 2 Service manual

Somnetics Transcend user manual

Somnetics

Somnetics Transcend user manual

TECHCON SYSTEMS TS6500 user guide

TECHCON SYSTEMS

TECHCON SYSTEMS TS6500 user guide

Bionics PalmCare plus Operation manual

Bionics

Bionics PalmCare plus Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.