5.4.4. Technique
-Daily maintenance using a surface DD product, in one application.
-Maintenance when patient is discharged or periodic maintenance using the Biocleaning procedure,
involving 3 operations:
oClean using a cloth soaked with a surface detergent or Detergent-Disinfectant (DD) solution.
All harsh or scouring products such as detergents, powders, solvents and bleach are
prohibited.
oRinse with a cloth rinsed in fresh water
oDisinfect using a cloth soaked with a surface disinfectant solution.
-Specific maintenance by service providers after removing the rail from the institution:
oDispose of the packaging after decontaminating the inside using a Detergent-Disinfectant
spray solution.
oBiocleaning operation, or
oSteam cleaning (accessory with a microfibre mop) of the various flat surfaces and bed slats.
Regularly change the mop heads to avoid excess water. Clean hard-to-access areas (wheels,
opened moving parts, angles, etc.) with the steam nozzle. Use the steam nozzle with
a microfibre cloth for the tubes. Do not aim the nozzle directly on electrical components
or actuators.
oDry moving parts with compressed air
oNB: Disinfect actuators, electrical components and remote controls with microfibre wipes
soaked with disinfectant.
Do not rinse or dry.
Check for normal function and the various operations.
Repair if necessary
NB:
-Take the measures recommended by the hospital’s internal hygiene department in case of additional
precautionary measures (Contact, Droplets, Air precautions)
-Using a greater than 5000 ppm bleach solution (0.5% active chlorine must be justified by a
microbiological risk and applied for the required time (Risk that certain materials - particularly colour -
will age over time).
-The concentration of alcohol-based surface disinfectant solutions must be less than 30%.
Comment: use of terminal disinfection is compatible with the hospital bed and its accessories.
-Isolate the fixed rail in a disinfection room with a particle filtration system and an outlet for washing floors
and walls after disinfection.
-Use a disinfectant product with bactericide, fungicide and virucide properties, either by uniformly
spraying the surfaces or applying with a single-use cloth, or by spraying a disinfectant aerosol from a
distance of 30 cm.
NB:
Adhere to the precautions indicated on the disinfectant products.
Leave to dry and keep the disinfected material separate from other, non-disinfected material with a film
including a label indicating the disinfection date.
5.5. Shelf life
The shelf life of accessories under normal usage and maintenance conditions is 5 years.