pH calibration pH/Oxi 340i
8
Note
If a printer isconnected to theinterface,a calibrationprotocol is automatically
printed after a valid calibration.
AutoCal TEC / AutoCal DIN procedure:
The following example shows the AutoCal TEC calibration. In the
AutoCal DIN calibration, Cd... instead of Ct... appears on the display. Other-
wise, both procedures are identical.
lConnect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status display pH
(pH measurement) or U(measurement of the ORP voltage) appears.
lPress the <CAL> key repeatedly until the display Ct1 and the function
display AutoCal TEC appears. The sensor symbol displays the evalua-
tion of the last calibration (or no sensor symbol in the delivery state or af-
ter the measurement parameter has been reset).
lImmerse the pH electrode into the first buffer solution.
lWhen measuring without a temperature sensor:
Enter the temperature of the first buffer solution with <▲> <▼>while
pressing the <RUN/ENTER> key.
lPress <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) or the buffer nominal value appears on the display. If
the measured value is stable, Ct2 appears.
Note
At this point, the AutoCal TEC calibration can be terminated
with <M>. This corresponds to a single-point calibration.
To do this, the instrument uses the Nernst slope (-59.2 mV/pH
at 25 °C) and determines the asymmetry of the electrode.
lImmerse the two-point calibration pH electrode in the secondbuffer so-
lution to continue.
lWhen measuring without a temperature sensor:
Enter the temperature of the second buffer solution with <▲> <▼>while
pressing the <RUN/ENTER> key.
lPress <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) or the buffer nominal value appears on the display.
lWhen the measured value is stable, the instrument displays the value of
the slope and the calibration evaluation.
lPress <RUN/ENTER>. The instrument displays the value of the asym-
metry.
lSwitch to the measuring mode with <M>.
8
4
2
t1
°
C
TP
pH
C
AutoCalTEC
8
4
2
t2
°
C
TP
pH
C
AutoCalTEC
94
pH
5
mV/pH
2
pH
mV