YAMAN Bloom EYE User manual

User’s Manual
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
When Using Outside of Japan. . . . . . . . . . . . . . . 8
Names and Functions of Components . . . . . . 9
Before Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
How to Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
After Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Additional Information . . . . . . . . . . . . . . . . . . . . . . . 19
● Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
● Frequently Asked Questions. . . . . . . . . . . . . . 21
● Product specifications . . . . . . . . . . . . . . . . . . . . 22
After-sales Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Table of
Contents
To be retained
Home Beauty Device Bloom EYE SEYE 1

1
* ‘Property damage’ refers to damage affecting a dwelling, household belongings, or domestic animals and pets.
DANGER Improper use of this device poses a severe and
likely risk of death or serious injury.
WARNING Improper use of this device may pose a risk of death
or serious injury.
ATTENTION
Improper use of this device may result in injury or
property damage*.
This symbol shows you what you must do.
This symbol shows you what you must be careful with.
Thank you for purchasing the YA-MAN product.
• Please read this User’s Manual carefully for information on how to use the device properly and safely.
• After reading the contents hereof, be sure to keep this User’s Manual in an accessible location.
• Make sure to include this User’s Manual if you are transferring ownership of the device to another person.
● Classifications of Precautions ● Symbols (the following are examples of the symbols)
• Read these Safety Precautions carefully before using the device in order to use it correctly.
• Use the device sensibly in accordance with the prescribed steps.
• Immediately discontinue use and arrange to have the device repaired if any abnormality occurs.
• Do not use the device in a defective state, as doing so could cause physical harm to the user.
• The precautions listed in the User’s Manual are very important. For your safety, please follow them carefully.
• Precautions are classified as DANGER, WARNING, or ATTENTION in order to clearly specify the level of danger and possible injury or damage that may occur
if the device is misused.
Introduction
Safety Precautions
This symbol shows you what you must not do.

2
Use of the device could cause electronic medical devices to malfunction.
◆ Do not use in conjunction with any medical electronic equipment, including:
pacemakers or other implanted electronic medical devices; artificial heart-lung machines or other electronic life-support medical devices, and electrocardiographs or
other worn electronic medical devices.
● The following individuals should not use the device.
DANGER (regarding use and storage)
Failure to comply may cause burn injuries.
● Do not use continuously on the same area. Be sure to move the device across the skin while it is in use. Do not use on areas that cannot be reached
by the entire electrode head surface.
Failure to comply may cause an accident or other problems.
• If you are under a doctor’s care • If you are a child or adolescent • If you have an acute disease • If you have tuberculosis • If you have a fever
• If you are suffering from hypertension • If you have an infectious disease • If you have a malignant tumor • If you have a blood disease • If you have a visceral disease
• If you have a skin disease • If you have an allergic constitution • If you have atopic dermatitis • If you have skin problems • If you have sensitive skin
• If you have unhealthy skin blemishes caused by physical irritation • If you are experiencing skin irritation or redness due to sunburn • If you are severely fatigued
• If you are physically unwell • If you are under the influence of alcohol • If you are pregnant, menstruating, or lactating • If you take medication that induces drowsiness
• If you have capillary dilatation due to long-term use of steroid hormonal drugs or a liver disorder; • If you have excessive myopia (-6D or higher)
• If you have had an eye or face operation • If you have diseases of the eyes or face or you are under treatment;
Failure to comply may cause an accident or other problems.
• Cuts • Sites where plastic surgery was performed • Sites where cosmetic surgery was performed
• Anywhere inside the mouth, around the pubic region, or any mucous membrane • Acne • Sites where skin suffers from purulent inflammation
• Sites with itchiness or a hot sensation • Bulging moles • Viral warts • The Adam’s apple • Above implants • Eyeballs • Eyelids, eye sockets
* If you are uncertain as to whether it is safe to use the device, consult a physician prior to use.
WARNING (Use and storage)
● Do not use under the following conditions:
● Do not use the device on the areas indicated below or outside of specified areas.
Prohibition
Prohibition

3
WARNING (Use and storage)
● Do not use the device in the bathroom or near a bathtub,
shower, sink, or other vessels filled with water.
Failure to comply may cause an electric shock, short circuit, fire,
or failure of the device.
● Do not splash water on the device (main body, etc.) or
immerse it in water. Do not use the device in a high-humidity
location or where it may be splashed with water
Failure to comply may lead to a fire, electric shock, or malfunction.
Exposure
to water
is prohibited
● Do not overuse the device. For usage time and frequency,
follow the instructions in the User’s Manual.
Failure to comply may cause physical injury.
● Do not use the device continuously for an overly long period
(for example, through commercial use).
Failure to comply may cause an accident, a failure of the device, or an
injury.
● Do not use the device for a long time over sites where the
skin is thin or sites above bones.
Failure to comply may cause skin problems or physical ailments.
● Do not use the device near a flammable substance (such as
alcohol, thinner, benzene, nail polish remover, or spray).
Doing so may cause the device to explode or catch fire.
Prohibition
● Use by unattended children or individuals with physical
disabilities should be avoided. Minors should use the
device only under the supervision of a guardian who has
given their consent.
Failure to comply may cause an accident or an injury.
● Do not use the device on individuals who are unable to
express their will or who have reduced skin sensitivity.
Failure to comply may cause an accident or an injury.
● Keep the device and generic AC adapter away from
children and do not allow them to use either. Do not allow
children to play with the device.
Failure to comply may cause an accident or an injury.
User
restrictions
● Do not place the device in an unstable location. Do not
drop or bump the device (either the main body or
attachments).
Failure to comply may cause a failure of the device. An injury may also be
caused if the device is dropped on a foot or other exposed body part.
● Do not place heavy items on the device, drop the device
(especially during use), or otherwise subject the device to a
strong impact.
Failure to comply may cause an injury or a failure of the device.
Dropping of
the device
is prohibited
● Do not use the device for any purpose other than those
stated in the User’s Manual.
Failure to comply may cause a failure of the device or an injury. Non-intended uses
are prohibited.
● Do not disassemble, attempt to repair, or modify this
device in any way.
Failure to comply may cause a fire, electric shock or other injuries.
Contact your local distributor for advice on repairing the device.
Disassembly
is prohibited

4
WARNING (Use and storage)
● Do not leave the device in hot or humid places,
such as near flames, in bright sunlight, or in a bathroom. Do not
bring the device close to a
heater.
Failure to comply may lead to the generation of heat, a fire, or explosion.
● Do not place the device near fire or flames, and do not dispose of
the device in a fire. Do not use, charge, or store in a
high-temperature area or if the device is hot.
The device features a built-in rechargeable battery.
Failure to comply may lead to the generation of heat, a fire, or explosion.
● Be sure to check the plug configuration and the voltage of the
power outlet before use.
Failure to comply may cause the generation of heat, a fire, or a malfunction
of the device.
Combustion
warning
● If any problem occurs during use, including the presence of
an abnormal odor or excess heat, stop using the device.
Failure to comply may cause the device to become damaged or
catch fire.
Use is prohibited
when an abnormal
odor or excess heat
is detected
● Do not put the device into a microwave or pressure vessel.
Failure to comply may lead to the generation of heat, a fire, or explosion.
Explosion
● If you are currently undergoing dental treatment
or have teeth fillings, choose a lower intensity level and apply the
device on a different area.
You may feel strong irritation or discomfort by EMS on your cheeks or chin.
If you feel strong discomfort, consult a physician.
● If you sense an earthquake or see lightning, turn the power off or
disconnect the AC adapter while the device is charging.
Failure to comply may lead to a fire or electric shock.
● Do not use or store the device in the following locations:
Failure to comply may lead to a fire, electric shock, malfunction, failure of
the device, or other accidents.
• A high-humidity location such as a bathroom • Dusty areas • A location
where there is exposure to direct sunlight • A location where water or
other liquids may splash on the device • A tilted location, or a location
where there is vibration or shocks • An outdoor location • A location
where there are strong radio waves or magnetism
• Locations that are accessible to children • A location where there is
exposure to oil smoke or steam • A location that may become hot, such as
near a heating device (for example, a heater) • A location with drastic
temperature changes (for example, near an air conditioner or heater)
● If the device is dropped or damaged, turn off the power
immediately and request a repair/inspection.
Failure to comply may lead to a fire or electric shock.
● Do not leave the device unattended while it is turned on.
Doing so may cause the device to catch fire or may lead to an accident.
Mandatory
Compliance
● If water or foreign matter gets into the device, immediately stop
using it and request an inspection.
Failure to comply may lead to a fire or electric shock. Mandatory
Compliance

5
Mandatory
Compliance
● Be sure to remove makeup before using the device on your
face.
If you use the device over makeup, the foundation may cause dark
staining of the skin.
●
Be sure to comply with the following items.
Failure to comply may cause an accident, injury, or a failure of the
device.
• Use the device appropriately based on your skin condition.
• Do not use the device after sunbathing, or in combination with
peelings, skin packs, or other treatments.
• Be sure to turn the device off after use.
● During treatment, carefully check that your skin is not
showing any abnormalities.
Discontinue use immediately if you start to feel pain or
fatigue or if you see any skin reaction during use.
Stop using the device immediately if the treatment does not agree with
your skin. If symptoms persist the following day and beyond, consult a
physician.
ATTENTION
(Use and storage)
● Before use, check to make sure the electrode head is not
cracked, deformed, or broken. Do not use the device if it is
damaged or malfunctioning.
Failure to comply may cause your skin to suffer an injury or other problems,
or lead to the worsening of existing conditions.
● Do not use continuously on the same area for more than five
minutes.
Failure to comply may cause skin problems.
● Do not press the device strongly or use it intensively on a single
spot.
Pressing the device stronger than necessary or using it intensively on a single
spot may damage the skin.
● Do not allow the device to come into contact with another
beauty device or electronic appliance while the power is on.
Do not use with other electronic products (electric blankets/bed
sheets, heated tables, etc.) at the same time.
Failure to comply may lead to an accident due to malfunction.
● Do not wipe the device with solvents such as alcohol, thinner,
benzene, nail polish remover, kerosene, chemicals, or polishing
powder.
Failure to comply may cause damage to the surface of the device.
● Avoid allowing the electrode head to come into contact with
any other metal surfaces.
Failure to comply may cause a failure of the device or an injury.
● Do not use the device on pets.
Failure to comply may cause injuries.
● Do not use the device above clothes.
Prohibition
● Do not use the device outdoors.
Failure to comply may cause a failure of the device.
Outdoor use
is prohibited

6
Mandatory
Compliance
● When using on the same area again, be sure to leave an
interval of 24 or more hours between treatments.
Excessive use may damage the skin. The above recommendations are
guidelines only; please observe your skin condition and use the device
accordingly.
●
After each use, remove the dirt, cosmetics, and other
particles that have adhered to the device according to
‘After Treatment.’
Failure to comply may cause a failure of the device or skin problems.
● When cleaning the device, be sure to turn off the power,
disconnect the USB cable from the device, and follow the
instructions in the User’s Manual.
Failure to comply may lead to an accident or malfunctions.
ATTENTION
(Use and storage)
● Use only Level 1 upon first use and until you are accustomed
to use.
Failure to comply may cause an accident or injury.
● When you have not used the device for a long period of
time, check to see that it is functioning normally before
re-starting use.
Failure to comply may lead to an accident or malfunctions.
● Remove all metal accessories (watches, necklaces, rings,
earrings, etc.) when using the device.
Failure to comply may cause an accident or an injury. Accessories may
also become entangled in the device.
● Reapply skin-care products or water during treatment to
ensure that there is always some liquid between the
electrode head and the skin.
If there is an insufficient amount of skin-care products or water, the
stimulation will weaken, causing skin problems.
● If you feel an unpleasant tingling on your skin, reduce the
intensity level or move to another area. Alternatively,
moisturize your skin with the included skin-care products
before use.
Failure to comply may cause skin damage.
Mandatory
Compliance

7
● Do not use the device if the USB cable is damaged, or if the
wall socket is loose.
Failure to comply may cause an electric shock, short circuit,
fire, or failure of the device.
Contact your local distributor for advice on repairing the
device.
● Unplug the USB cable when not in use.
Do not insert or disconnect the plug with wet hands.
Failure to comply may cause an electric shock, short
circuit, or fire.
● In the unlikely event that an abnormality occurs,
immediately unplug the USB cable.
Do not use the device if smoke, abnormal noises, or an odor is detected,
or if it is otherwise operating in an abnormal state.
● Use the USB cable included with this product.
Plug the AC adapter directly into a wall power outlet and
do not overload the electric circuit.
Failure to comply may cause an electric shock, short circuit, fire, or
failure of the device.
WARNING (The power supply)
● Do not damage, modify, bundle, or otherwise apply
undue force to the USB cable.
Failure to comply may cause the cable to break, resulting
in a fire or electric shock.
● Do not wrap the USB cable around the device when
storing the device.
Placing a load on the USB cable may cause a
disconnection, resulting in electric shock, or a short
circuit.
● When disconnecting the USB cable from the device or
generic AC adapter, be sure to hold the cable at the front
as you pull (do not touch the metal part).
Failure to comply may cause an electric shock, short circuit, fire, or
failure of the device.
Removing
the plug
Disconnecting
the AC adapter
Overloading of
an electric circuit
is prohibited.
Use of a
damaged
cable or plug
is prohibited
Careless or
reckless use
is prohibited
● Do not soak into, spill over or wash USB cable with water.
Failure to comply may cause an electric shock, short circuit, fire, or failure
of the device. Exposure
to water
is prohibited

8
WARNING (The power supply)
Mandatory
Compliance
● Before charging the device, check to ensure the charging terminal and USB cable are not wet.
Failure to comply may cause an electric shock, short circuit, or fire.
● Do not pull the USB cable when moving or lifting the device.
Failure to comply may cause an electric shock, short circuit, fire, or failure of the device.
● Be sure to use only with a 5 V USB power supply.
Make sure that the USB cable is fully inserted into the USB port.
Failure to comply may cause a fire, electric shock or other accident.
Failure to comply may cause a fire, electric shock, or a failure of the device.
● Be sure to use only with a 5 V USB power supply.
Failure to comply may cause the generation of heat, a fire, or a malfunction of the device.
● Be sure to check the plug configuration and the voltage of the power outlet before use.
Mandatory
Compliance
WARNING
• The plug configuration and voltage may vary depending on the region and the type of accommodation. Always be sure to check for compatibility before using the device.
• For information about plug adapters, consult a travel agency or a consumer electronics retailer.
Combustion
warning
When Using Outside of Japan

9
Names and Functions of Components
Level lamp
Main Body
Electrode head
Power button ( )
Charging terminal (USB Type-C™)
● Front view ● Rear
Level button
Charging Lamp ( )

10
● USB cable (USB Type-C™) ● User’s Manual
● Flawless Serum Moisture 20G
Flawless Serum Smooth 20G
Accessories
Safety functions
● Current sensor
Measures the output current value and controls the output.
● Temperature sensor
Output will be reduced automatically for safety reasons when the temperature on the skin or the electrode head rises.
If the temperature becomes too high, the power will be turned off for safety. (When this happens, the electrode head flashes and beeps five times)
If this happens, let the device rest for about 30 minutes.
● Motion sensor
Motion sensor detects movement of the device and pauses output for safety reasons when the electrode head has been used on the same area continuously.
When movement of the device slows, the motion sensor may reduce the level of warmth.
Auto Power Off / Standby
The treatment duration on the entire face is about five minutes. After approximately 2.5 minutes, a long beep will sound and the device will enter Standby mode.
Standby lasts for approximately 5 minutes and the Level lamp flashes (all outputs pause). Press the Power button ( ) to start the second half, which lasts approximately 2.5 minutes.
The device will automatically switch off after the second half, which lasts approximately 2.5 minutes (two short beeps).
The device will automatically switch off (Auto-Off function) after being left in Standby mode for approximately five minutes without any operations (two short beeps).
Functions
The design and specifications of the device and attached accessories are subject to change for improvement purposes without notice.
• USB Type-C™ and USB-C™ are trademarks of the USB Implementers Forum.
Charging Guide Display
When you turn the power on, the color of the Charging Lamp ( ) (either white or red) will indicate the remaining battery level. If the lamp lights red, recharge the battery
before using the device again. The remaining battery level display will differ according to the condition of the battery and the usage environment.

11
Before Use
Charge the device
ATTENTION
• If there is any moisture on or around the main body or charging terminal, be sure to wipe it dry. Charging the device while it is moist could cause damage to the battery.
• Fully charge the battery before using the device for the first time after purchasing.
• When using the device for the first time or when the device has not been used for an extended period of time, it is possible that you will be unable to use the device
for very long even with a full charge displayed. In such a case, recharge the device before using it.
• Use only the included USB cable.
Using a household power outlet
USB cable
Charging terminal
Household power outlet
AC adapter*
* Not included with the device.
• While charging, the Charge lamp ( ) will blink in white.
• When charging is complete, the Charge lamp ( ) will light white for approximately five seconds and then turn off. Remove the USB cable.
• At an indoor temperature of approximately 25°C, it will take about three hours to reach a full charge, at which point the device can operate for about 40 minutes.
Actual charging time and operating time will differ depending upon the remaining battery level and the charging environment.
* If charging does not begin after approximately 10 seconds, check whether the USB cable connection is loose.
Charging Lamp ( )
Charging Guide Display
Charging Lamp ( ) Display
The battery is sufficiently charged.
Low battery.
Charge the battery.
White lighting
Red lighting
Red flashing
You can check the remaining battery level
through the Charging lamp ( ).

12
Pre-testing on Skin
When using the device for the first time, always test for compatibility.
1Wash the inner side of
your arm with soap and
water, then wipe dry
with a towel.
2Apply the skin-care product (about
one pearl-sized drop) evenly to the
test site on your arm, turn the
power on, and then use the device
for about 2.5 minutes while moving
it across the skin (page 13).
Be sure to test at Level 1.
3Rub the skin-care products
remaining on your skin into
the skin and check for any
abnormalities.
Continue to use the product only if you
experience no reddening or burning
sensation on your skin within 24 hours.
ATTENTION
● Be sure to perform a pre-test in order to prevent skin problems from occurring.
● If a rash, reddening, itchiness, stinging sensation, or other abnormality occurs at the pre-testing site,
immediately stop the pre-test.
● Check carefully to make sure there are no skin abnormalities before using the device. Stop using the device
immediately if the treatment does not agree with your skin.
How to hold the device and use it on your skin.
Hold the main body as shown below,
moving it slowly so that the entire electrode
head remains in contact with your skin.
• Be sure to move the device with the entire
electrode head in contact with your skin.

13
Basic method of use
How to Use
Be sure to remove makeup and wash your face before using the device.
1Turn on the device
The Charging lamp ( ) and Level 1 lamp will be lit, and the
radio frequency (RF), EMS, and micro-current will be output.
Press and hold for
about one second.
Electrode head
2Select the level
Each time the Level button is pressed, levels can be adjusted.
You can check the treatment level from the level lamp.
Level 1
Warms up gradually
Gentle EMS
Level 2 Level 3
Warms up quickly
Strong EMS
Level button
Level 1 lamp
Charging Lamp ( )
Power button ( )
The device remembers the level being used when last treatment ended,
and returns to the same settings the next time it is switched on.
Personal memory function
Use only Level 1 upon first use and until you are accustomed to use.
ATTENTION

14
4Care for the skin after the treatment
3
Amount of
skin care product
Location to apply
skin care product
ATTENTION
See ‘How to move the device’ (pages 15-16)
for information on how to move the device correctly.
Areas you should not treat: Upper eyelids and the Adam’s apple
* The device can be used around the eyes
(lower eyelids, outer corner of the eye)
Apply the included skin-care product evenly on the site to be
treated and then move the device so that the entire surface
of the electrode head comes in contact with your skin
Treat one eye at a time (pages 15-16).
Apply one pearl-sized drop of the skin-care product around the eye for each eye.
After treatment is complete, rub the remaining
skin-care product into the skin and then use your
usual basic skin-care products to condition the skin.
• If the device does not move smoothly, apply more of the included skin-care product.
• Use in a position in which the entire surface of the electrode head comes in contact
with your skin.
• Be sure to move the device on your skin while using it.
• If it feels too hot, move it a little more quickly or lower the level.
• Try to use five times a week (once daily, up to five minutes).

15
Repeat the movements ① to ② within the specified time for a total of five sets as shown below.
How to move the device
STEP1:
Gently move from under the eyes to the temples,
then move slowly in a circle.
②①
STEP2:
Gently move in circles between the eyebrows, then slowly
pull upward along the bone above the eyes.
* If you want to focus on the under-eye area,
use only step 1 for 2.5 minutes.
Left side
Approximately
2.5 minutes
After about
2.5 minutes
One long beep
● After approximately 2.5 minutes, a long
beep will sound and the device will enter
Standby mode. To restart, press the power
button ( ).
● Recommended speed for moving the device: 1 to 2 cm/second
● Recommended duration of use: approximately 2.5 minutes per eye for a total of
5 minutes (entire area around both eyes)
● If you want to stop the treatment during use, press the Power button ( )
for at least one second to turn off the power.

16
Right side Automatic
Power-off
Approximately
5 minutes
Beeping sound
ATTENTION
●
Do not use continuously on the same area for more than five minutes.
Failure to comply may cause skin problems.
The device will
automatically switch
off approximately five
minutes of treatment
after it has been
turned on.
①
STEP1:
Gently move from under the eyes to the temples,
then move slowly in a circle.
STEP2:
Gently move in circles between the eyebrows, then
slowly pull upward along the bone above the eyes.
* If you want to focus on the under-eye area,
use only step 1 for 2.5 minutes.
Approximately
2.5 minutes

17
How To Clean
Clean the device after each use for hygienic reasons.
1Wipe off any dirt using a
well-wrung soft cloth or similar
material.
• For stubborn stains, wipe gently with a cloth
soaked in a diluted neutral detergent.
• Remove any skin-care products remaining
between the electrode ring of the electrode head
with a cotton bud or cotton pad.
WARNING ● When cleaning the device, be sure to turn it
off and disconnect the USB cable from the
main body.
Failure to comply may cause an electric shock, short circuit,
or fire.
● Do not wash the device in water.
This device is not waterproof. Doing so
may cause a failure of the device.
ATTENTION
After Treatment
● Do not wipe the device with solvents such as alcohol,
thinner, benzene, nail polish remover, kerosene,
chemicals, or polishing powder.
● Do not leave the device with the included skin-care product on it.
Benzene
Thinner
Cleaning agent

18
Storage
● Store in a clean, well-ventilated place out of the reach of children.
● Remove the USB cable from the main body when not in use for a long period of time.
● Do not leave the device in hot or humid places, such as near flames, in bright sunlight, or in a bathroom.
Maintenance and Inspections
● If you have not used the device for an extended period, inspect it carefully and check that it works properly and safely before using it again.
● If the device is malfunctioning, go to ‘Troubleshooting’ (page 19) and attempt to identify the problem before requesting a repair. If the problem persists, contact your local distributor,
making sure that you have your purchase details at the ready when you call.
● If the device malfunctions, appears damaged, or needs service, please contact your local distributor.

19
Troubleshooting If you perceive any abnormality while using the device, verify the following points before arranging to have it repaired.
Problem Possible Causes and Solutions
The device will automatically switch off approximately five minutes after the treatment started, or after five minutes of standby.
The battery level is low. Charge the battery (page 11).
The battery has expired. Contact your local distributor.
The device has run out of power. Charge the battery (page 11).
The device does not turn on.
The device shuts off.
The device only operates
for a few minutes
even after charging.
Insert the included USB cable into the main body (page 11). Check to make sure the cable is not loose.
Connect the generic AC adapter to a power outlet (page 11).
The device does not start charging.
Additional Information
Lower the level. Be sure to use Level 1 on the first use.
The device has been pressed against the same area of skin for too long. Be sure to keep the roller moving over the skin during operation.
The radio frequency (RF) feels
too hot Move the device a little more quickly.
Press and hold the Power button ( ) for about one second.
Pressing the Power button and Level button down simultaneously will reset the device and turn the power off.
Be careful not to press these buttons while using the device.
This manual suits for next models
1
Table of contents
Other YAMAN Personal Care Product manuals

YAMAN
YAMAN Photo PLUS Shiny User manual

YAMAN
YAMAN HRF-15 User manual

YAMAN
YAMAN Photo PLUS Prestige SS M21 User manual

YAMAN
YAMAN Bloom 6 User manual

YAMAN
YAMAN Medi Lift PLUS User manual

YAMAN
YAMAN Photo PLUS Shiny NEO User manual

YAMAN
YAMAN M30 User manual

YAMAN
YAMAN Waterpik WF-03 User manual

YAMAN
YAMAN MIYABI YJFG0 User manual

YAMAN
YAMAN GR-16 User manual