Yamato DP-6200 User manual

DP-6200 Technical Manual
YCO-1112-04-06
YAMATO CORPORATION
1775 S. Murray Blvd.
Colorado Springs, CO 80916 USA
Tel (719) 591-1500 Fax (719) 591-1045
YAMATO TECH CORPORATION
#112-19425 Langley By-Pass
Surrey, B.C. V3S 6K1 Canada
Tel (604) 533-2338 Fax (604) 533-0827

YCO-1112-04-06
DP-6200 Technical Manual
I. Safety Instructions 1
II. Description 2
Name and Function of Parts
Specications
III. Calibration 5
IV. Changing the Scale Setup 6
Changing Parameters
User Mode
Test Mode
System Mode
Factory Mode
Standard Parameter Tables
Parameter Descriptions
V. Troubleshooting 17
VI. Wiring Diagram 19
VII. Parts List 20
Table of Contents

YCO-1112-04-06
DP-6200 Technical Manual
I Safety Instructions
1
Before using the scale, carefully read, understand, and follow the “Safety Instructions” described in this manual. Observe the
advice given in the “Directions for Use” section to ensure proper operation. Keep this operation manual handy for ref-
erence.
To prevent injury
1) DO NOT SHOCK LOAD the scale, Never step on or sit on the scale. Not only will the scale be damaged, but you may
also be injured.
2) DO NOT APPLY DIRECT WATER SPRAY to the scale if it is powered by an AC adaptor!! You could receive a shock. The
scale is not washdown unless the AC jack is plugged with a water-tight seal.
4) Do not operate the scale if there is smoke or a burnt smell coming from the scale. Remove the batteries or unplug the
AC adaptor immediately. After making sure that there is no danger, consult your dealer. Never try to repair the scale
by yourself.
5) This scale is not an explosion-proof model. Do not use the scale in an atmosphere containing ammable gases or
explosive fumes. A re or an explosion can result.
6) Place the item to be weighed in the center of the platform. Items placed on the edge of the platform may fall off and
cause injury.
7) When weighing a heavy, large or unbalanced item, make sure the item is stable on the platform, otherwise an accident
may occur.
8) When carrying or moving the scale, be sure to hold it by the bottom of the weighing platform with both hands. If you hold
it by the column, the column may become detached causing the scale to fall. This will damage the scale and could cause
injury.
9) Do not insert your nger into the gap or holes in the scale. You may be injured.
10) The DP-6200 uses a liquid crystal display. If the LCD breaks and the liquid leaks from the LCD, do not touch it with your
ngers. The liquid is toxic if ingested. Be especially careful around small children.
To prevent damage to the scale
1) Do not push the indicator or keys with sharp objects. They may break or puncture the switch membrane panel.
2) Use the specied power supply and choose a suitable environment. If you do not, the weight readings may be inaccurate
and the scale may be damaged.
3) The scale is a sensitive weighing instrument, avoid physical shocks. If you drop something on the scale, overload the
scale, step on the platform, or drop the scale, the scale may be damaged and lose accuracy.
4) If the scale becomes dirty, wipe it with a soft cloth. For stubborn stains, apply a little neutral detergent and then wipe the
scale with a dry cloth. Do not use thinner, benzene, hot water, or chemical agents, all of which can cause deformation, dis-
coloration, or deterioration of the scale.
5) Never remove the case. The ne adjustment section may be damaged and you may be injured by sharp edges on the
internal parts.
6) Do not place the scale upside down.
7) When the low battery indicator appears, replace all six of the batteries. When installing the batteries, install them accord-
ing to the polarity markings in the case (+, -). If the scale will not be used for a long period, remove the batteries.
8) Never connect or disconnect the RS-232 cable while the DP-6200 and a connected computer are on. The DP-6200 and/or
the computer may be damaged. RS-232 ports are not hot-swappable.
To keep the scale working efciently
1) Place the scale on a at stable surface that will support the scale and the load.
2) Do not place the scale in an area exposed to direct sunlight or to wind currents from an air conditioner, otherwise, the mea-
surements will not be accurate.
3) Do not place the scale near machines that create vibrations or electromagnetic disturbance, such as microwave ovens,
portable phones, or large motors. This will affect the accuracy.
4) The operating temperature range is from –5°C to +35°C (23°F to 95°F). Do not subject the scale to sudden temperature
changes; allow the scale to adjust to the new temperature before use.
5) If the scale is sealed, do not break the seal. If you break the seal, the scale will not be considered legal for trade. In this
case, contact your dealer.
6) Do not disassemble or modify the scale, you will void the warranty. Modied scales will not be legal for trade.

YCO-1112-04-06
DP-6200 Technical Manual
II Description
2
FIG. 1 FIG. 2
FIG. 3
Name and Function of Parts

YCO-1112-04-06
DP-6200 Technical Manual
���
�
FIG. 4
1) Low Battery Indicator Time to replace the batteries.
Hold Indicator The scale is in Hold mode (not legal-for-trade).
Test Mode Indicator The scale is in Test Mode.
Stable Indicator Indicates the weight indication is stabilized.
2) LCD Display Shows all scale indications and weight readings.
3) Unit Indication KG, LB, OZ and LB-OZ (not legal-for-trade).
4) Over Indicator The current weight is over the user’s maximum setpoint.
Accept Indicator The current weight is within the user’s setpoints.
Under Indicator The current weight is below the user’s minimum setpoint.
5) Mode Indicator The scale is in setup mode.
Net Indicator The scale is indicating net weight.
Gross Indicator The scale is indicating gross weight.
Center of Zero Indicator The scale is at gross zero.
6) Net/Gross Key Toggles between gross and net weight readings.
Decrease Key Decreases setpoints.
Unit Selection Key Press both keys simultaneously to toggle between kilogram and
either pound, ounce or pound-ounce units.
7) On/Zero Key Turns the power on, and zeroes the display.
Next Key Moves to the next setpoint.
8) Off Key Turns the power off.
9) Tare Key Tares off the current weight.
Increase Key Increases setpoints.
Unit Selection Key Press both keys simultaneously to toggle between kilogram and
either pound, ounce or pound-ounce units.
II Description
3
Name and Function of Parts

YCO-1112-04-06
DP-6200 Technical Manual
Specications
Weighing system: Strain-gauge load cell
Platform: 13.8” x 19.7” (350 x 500 mm), stainless steel
Weight display: Single face display
Type: 7 segment LCD
Character size, etc.: 0.5” (12 mm) (W) x 0.9” (23 mm) (H), 5 digits
Optional equipment: Batteries, AC adaptor, RS-232C communications (the scale is not
washdown with the RS-232C option installed.)
Power supply: 9V DC - six “D” size batteries or AC adaptor (the scale is not
washdown unless the AC jack is plugged with a water-tight seal).
Consumption: 0.12 W (max.)
Battery life: 600 hours of continuous use with alkaline batteries. Using the RS-
232C option will substantially reduce battery life.
Operating temperature: 23°F to 95°F (-5°C to 35°C)
Operating humidity: 25% to 85% RH
Waterproof classication: Complies with NEMA 4 (washdown) when the AC jack is plugged
with a water-tight seal. Is not washdown with the RS-232C option
installed.
Capacities and divisions:
* Combined units, such as pound-ounce, are not legal-for-trade.
The DP-6200 is available in four capacities. It displays units of kilograms and one of the following
three: decimal pounds, continuous ounces, or pounds and ounces. It is important to note that various
parameter settings depend upon the mode (decimal pound, continuous ounce, or pound and ounce)
option selected when the scale was purchased. See section IV - Changing the Scale Setup for the
standard settings for each mode.
The DP-6200 is not washdown unless the AC adaptor jack is sealed, or if the RS-232 option is installed.
Remove the batteries when operating the scale with the AC adaptor. Use only the AC adaptor supplied
or an adaptor of the same output characteristics.
II Description
4
Kilogram Pound Ounce Pound-Ounce*
30 kg x 0.01 kg 60 lb x 0.02 960 oz x 0.5 oz 60 lb x 0.5 oz
60 kg x 0.02 kg 150 lb x 0.05 lb 2400 oz x 1 oz 150 lb x 1 oz
150 kg x 0.05 kg 300 lb x 0.1 lb 4800 oz x 2 oz 300 lb x 2 oz

YCO-1112-04-06
DP-6200 Technical Manual
The DP-6200 can be setup to be calibrated with pound weights or with kilogram weights. Check
the setting of system parameter 78 (see section IV - Changing the Scale Setup). 000 sets the scale
to be calibrated with kilogram weights, 001 sets the 60 lb and 300 lb scales to be calibrated with
pound weights, and 002 sets the 150 lb scale to be calibrated with pound weights.
1) Open the indicator housing by pulling the release tab forward and
swinging the front panel of the housing forward. Short the two test
pins to the right of the CPU on the main board. The Test Mode
ennunciator and internal counts will display.
2) Press and hold the
key and press the
key. If the
scale indicates two point calibration mode, change it to three point
calibration mode by pressing the
key.
If it becomes necessary to abort the calibration procedure before
completing it, press the
key to return to the test mode.
3) Remove all items from the scale platform, wait for the stable indi-
cator to display and press the
�
key.
4) Place 1/2 the full capacity of the scale on the platform, wait for the
stable indicator to display and press the
�
key.
5) Place the full capacity of the scale on the platform, wait for the
stable indicator to display and press the
�
key.
6) The scale will return to Test Mode and indicate full scale counts of
30000 for all kg, 27215 for 60 lb and 300 lb, and 34025 for 150 lb
calibration. The full scale counts should be within +/- 3 counts of
the appropriate target. Press the
key to exit Test Mode.
If the scale indicates error 103, then reconrm that system pa-
rameter 78 is set correctly for the weights being used and the
capacity of the scale. Verify that the correct weights, 1/2 capac-
ity and full capacity, are being used. Make sure that nothing is
restricting the motion of the platform, pressing against the platform
or wedged between the upper and lower platform. Remove any
sources of vibration or electromagnetic radiation (i.e. - cell phones,
power cables, motors, compressors, etc.) Repeat the calibration
procedure. If the error repeats, see section V - Troubleshooting.
III Calibration
5
Test Mode
Two Point Mode
Three Point Mode
Zero Calibration
1/2 Capacity Calibration
Full Capacity Calibration
Full Scale Counts
Error 103

YCO-1112-04-06
DP-6200 Technical Manual
IV Changing the Scale Setup
6
Changing Parameters
The scale setup is controlled by three groups of parameters that determine how the scale operates.
These three groups are User Parameters, System Parameters and Factory Parameters.
User Parameters (00 - 09) can be changed in both User Mode and System Mode. They are easily
accessed, without breaking the seal or opening the indicator housing, and control functions such as
the auto-off, RS-232 communications and whether the scale starts in metric or standard units. The
User Parameters are included in System Mode for the convenience of the technician.
System Parameters (30 - 40) can be changed in System Mode. They control functions that the
technician may need to change under unusual circumstances. Since these functions can alter the
accuracy of the scale, the seal has to be broken to access System Mode through Test Mode.
Factory Parameters (51 - 99) can be change in Factory Mode. They control functions that directly
relate to the accuracy, capacity and divisions of the scale and are rarely changed outside of the
factory. Factory Parameters are usually only changed if a different standard mode (pound, ounce or
pound-ounce) is needed or if a different capacity load cell is installed. Since these functions can al-
ter the accuracy of the scale, the seal has to be broken to access Factory Mode through Test Mode.
Each parameter has two components, the parameter keyword and
the parameter value. The rst two digits are the keyword and de-
termine which parameter is being changed. The last three digits
are the value and determine what the selected parameter is being
changed to.
While each mode is accessed differently, all modes use the same keys to select and change the
parameter settings.
To select the desired keyword, use the keys as follows:
- increase the keyword by one
+
- decrease the keyword by one
To change a parameter value, use the keys as follows:
�
- increase the value by one
�
+
- increase the value by ten
- decrease the value by one
+
�
- decrease the value by ten
To save the last value changed it is necessary to advance to the next keyword before exiting the
setup mode. Press and hold the
key and press the
key to exit System Mode or Fac-
tory Mode and return to Test Mode. Press the
key to exit any setup or test mode.
�
Keyword Value

YCO-1112-04-06
DP-6200 Technical Manual
7
IV Changing the Scale Setup
User Mode
With the scale on, press the
and
�
keys to enter the user setup mode. Press the
key to exit the user setup mode.
Test Mode
Open the indicator housing by pulling the release tab forward and
swinging the front panel of the housing forward. Short the two test
pins to the right of the CPU on the main board. The Test Mode
ennunciator and internal counts will display. Pressing the
�
key will cause the scale to cycle through the following displays: in-
ternal counts, initial counts, direct raw counts, average raw counts,
battery check/A to D conversion value (should be around 220),
ROM version and a display segment check. Press the
key
to exit Test Mode.
System Mode
Enter Test Mode and then press the
key and the
�
key.
The display will show system parameter 30. Use the
key to
cycle through the system parameters (30 - 40) and then the user
parameters (01 - 09). See the previous page for instructions on
changing the parameters. Press the
key and the
�
key
to return to Test Mode or press the
key to exit setup.
Factory Mode
Enter Test Mode, press the
key and the
key to enter
Calibration Mode, and then press the
�
key and the
key to enter Factory Mode. The display will show factory parame-
ter 99. Use the
key to cycle through the factory parameters
(50 - 99). See the previous page for instructions on changing the
parameters. Press the
key and the
�
key to return to
Test Mode or press the
key to exit setup.
�
�
Test Mode
System Mode
Factory Mode

YCO-1112-04-06
DP-6200 Technical Manual
IV Changing the Scale Setup
Standard Parameter Settings - User Parameters
Standard Parameter Settings - System Parameters
8
Key User Parameter All
Word Cap.
01 Auto-off timer 015
02 Display hold timer 000
03 Communication 001
04 Communication data 000
05 Communication speed 000
06 Character length 000
07 Parity 000
08 Stop bit length 001
09 Default mode 001
Key System Parameter All
Word Cap.
30 System ID 018
31 Gravity compensation 010
32 Linearity compensation 000
33 Steady-state sampling count 004
34 Steady-state count 012
35 Polarity steady-state count 008
36 Steady-state collapse count 020
37 Steady-state average count 004
38 Zero tracking timing 004
39 Neglectable change count 001
40 Simple test mode 003

YCO-1112-04-06
DP-6200 Technical Manual
9
Key Factory Parameter 30 kg 60 kg 150 kg
Word 60 lb 150 lb 300 lb
51 Scale Mode 004 004 004
52 Multi-increments 000 000 000
53 Weighing capacity base value 003 006 015
54 Weighing capacity index value 003 003 003
55 Increment for lighter range 000 001 002
56 Location of decimal point 002 002 002
57 Type of decimal point 000 000 000
58 Weighing unit 002 002 002
59 Display of weighing unit 001 001 001
60 Weighing capacity base value (lb) 006 015 003
61 Weighing capacity index value (lb) 003 003 003
62 Location of decimal point 002 002 001
63 Pound increment 002 005 001
64 Pound conversion coefcient 003 002 003
65 Zero point range 019 019 019
66 Zero point range on plus side 012 012 012
67 Over scale 003 003 003
68 Display hold function 000 000 000
69 Hold release against increased wght. 000 000 000
70 Stable mark 000 000 000
71 Tare function 002 002 002
72 Zero reset under tare operation 001 001 001
73 Power off with ON key 000 000 000
74 Special function for stability 000 000 000
75 Over - under function 001 001 001
76 Tare cancel by Zero key toggle 000 000 000
77 -bat- display toggle 000 000 000
78 Conversion coefcient for lb cal. 001 002 001
79 Center-of-zero ennunciator position 001 001 001
80 Internal resolution 010 010 010
81 1 timer (ms) 060 060 060
82 3 timer (ms) 045 045 045
91 Mechanical zero 1 auto auto auto
92 Mechanical zero 2 auto auto auto
93 Span coefcient 1 (small) auto auto auto
94 Span coefcient 2 (small) auto auto auto
95 Span coefcient 1 (large) auto auto auto
96 Span coefcient 2 (large) auto auto auto
97 Regional # and gravity for span adj. auto auto auto
99 Factory setting - Do not change 000 000 000
IV Changing the Scale Setup
Standard Parameter Settings - Factory Parameters, Decimal Pounds

YCO-1112-04-06
DP-6200 Technical Manual
Key Factory Parameter 30 kg 60 kg 150 kg
Word 960 oz 2400 oz 4800 oz
51 Scale Mode 007 007 007
52 Multi-increments 000 000 000
53 Weighing capacity base value 003 006 015
54 Weighing capacity index value 003 003 003
55 Increment for lighter range 000 001 002
56 Location of decimal point 002 002 002
57 Type of decimal point 000 000 000
58 Weighing unit 002 002 002
59 Display of weighing unit 001 001 001
60 Weighing capacity base value (lb) 006 015 003
61 Weighing capacity index value (lb) 002 001 002
62 Location of decimal point 001 000 000
63 Pound increment 005 001 002
64 Pound conversion coefcient 001 001 002
65 Zero point range 019 019 019
66 Zero point range on plus side 012 012 012
67 Over scale 003 003 003
68 Display hold function 000 000 000
69 Hold release against increased wght. 000 000 000
70 Stable mark 000 000 000
71 Tare function 002 002 002
72 Zero reset under tare operation 001 001 001
73 Power off with ON key 000 000 000
74 Special function for stability 000 000 000
75 Over - under function 001 001 001
76 Tare cancel by Zero key toggle 000 000 000
77 -bat- display toggle 000 000 000
78 Conversion coefcient for lb cal. 001 002 001
79 Center-of-zero ennunciator position 001 001 001
80 Internal resolution 010 010 010
81 1 timer (ms) 060 060 060
82 3 timer (ms) 045 045 045
91 Mechanical zero 1 auto auto auto
92 Mechanical zero 2 auto auto auto
93 Span coefcient 1 (small) auto auto auto
94 Span coefcient 2 (small) auto auto auto
95 Span coefcient 1 (large) auto auto auto
96 Span coefcient 2 (large) auto auto auto
97 Regional # and gravity for span adj. auto auto auto
99 Factory setting - Do not change 000 000 000
10
IV Changing the Scale Setup
Standard Parameter Settings - Factory Parameters, Continuous Ounce

YCO-1112-04-06
DP-6200 Technical Manual
11
Key Factory Parameter 30 kg 60 kg 150 kg
Word 60 lb 150 lb 300 lb
51 Scale Mode 005 005 005
52 Multi-increments 000 000 000
53 Weighing capacity base value 003 006 015
54 Weighing capacity index value 003 003 003
55 Increment for lighter range 000 001 002
56 Location of decimal point 002 002 002
57 Type of decimal point 000 000 000
58 Weighing unit 002 002 002
59 Display of weighing unit 001 001 001
60 Weighing capacity base value (lb) 006 015 003
61 Weighing capacity index value (lb) 002 001 002
62 Location of decimal point 001 002 002
63 Pound increment 005 001 002
64 Pound conversion coefcient 001 001 002
65 Zero point range 019 019 019
66 Zero point range on plus side 012 012 012
67 Over scale 003 003 003
68 Display hold function 000 000 000
69 Hold release against increased wght. 000 000 000
70 Stable mark 000 000 000
71 Tare function 002 002 002
72 Zero reset under tare operation 001 001 001
73 Power off with ON key 000 000 000
74 Special function for stability 000 000 000
75 Over - under function 001 001 001
76 Tare cancel by Zero key toggle 000 000 000
77 -bat- display toggle 000 000 000
78 Conversion coefcient for lb cal. 001 002 001
79 Center-of-zero ennunciator position 001 001 001
80 Internal resolution 010 010 010
81 1 timer (ms) 060 060 060
82 3 timer (ms) 045 045 045
91 Mechanical zero 1 auto auto auto
92 Mechanical zero 2 auto auto auto
93 Span coefcient 1 (small) auto auto auto
94 Span coefcient 2 (small) auto auto auto
95 Span coefcient 1 (large) auto auto auto
96 Span coefcient 2 (large) auto auto auto
97 Regional # and gravity for span adj. auto auto auto
99 Factory setting - Do not change 000 000 000
IV Changing the Scale Setup
Standard Parameter Settings - Factory Parameters, Pound and Ounce

YCO-1112-04-06
DP-6200 Technical Manual
IV Changing the Scale Setup
12
Key Std Function Description
Word Value
01 015 Auto-off timer 000 Disables auto-off.
001~240 Number of minutes unused to activate auto-off.
02 000 Display hold timer 000 No display hold for changes less than four divisions.
001~030 Display hold duration, in seconds, for changes less
than four divisions.
031 No display hold.
03 001 Communication 000 No serial interface.
001 Specied commands acceptable.
002 Sends zero detection.
003 Sends motion detection.
004 Sends continuously.
04 000 Communication data 000 Net weight (1 batch)
001 Net weight, tare weight, gross weight (1 batch)
002 Net weight, tare weight, gross weight (3 batches)
003 Display content (1 batch)
004 Serial printer format (1 batch)
05 000 Communication speed 000 9600 bps
001 600 bps
002 1200 bps
003 2400 bps
004 4800 bps
005 9600 bps
006 19200 bps
007 38400 bps
06 000 Character length 000 8 bits
001 7 bits
07 002 Parity 000 Non
001 Odd
002 Even
08 001 Stop bit length 000 1 bit
001 2 bits
09 001 Default mode 000 kg
001 lb/oz
Parameter Descriptions - User Parameters

YCO-1112-04-06
DP-6200 Technical Manual
13
Key Std Function Description
Word Value
30 018 System ID 018 Do not change.
31 010 Gravity compensation 000 No compensation.
001~016 Compensation for a specied area (Japan only.)
017~150 Acceleration due to gravity formula, (mm/s2 )-9700.
32 000 Linearity compensation 000 No compensation.
001~127 Positive compensation.
128~255 Negative compensation.
33 004 Steady-state sampling count 000~255
34 012 Steady-state count 000~255
35 008 Polarity steady-state count 000~255
36 020 Steady-state collapse count 000~255
37 004 Steady-state average count 000~255
38 010 Zero tracking timing 000 Disable zero tracking
001~255 Zero tracking at specied counts
39 001 Neglectable change count 000~010 Number of counts
40 000 Simple test mode 000 User parameter valid
001 User parameter invalid
002 User parameter and simple test mode valid
003 User parameter and simple test mode invalid
Parameter Descriptions - System Parameters
IV Changing the Scale Setup

YCO-1112-04-06
DP-6200 Technical Manual
14
IV Changing the Scale Setup
Parameter Descriptions - Factory Parameters
Key Function Description
Word
51 Scale mode 000 Fixed single increment.
001 Multi-increments.
002 Do not use.
003 Do not use.
004 Change-over / gram mode - decimal pound mode.
005 Change-over / gram mode - lb/oz mode.
006 lb/oz mode.
007 Change-over / gram mode - continuous ounce mode.
52 Multi-increments, 000 Fixed single increment.
complex increment mode 001 Fixed accuracy, 3 increments.
002 Fixed accuracy, 2 increments.
003 Increment change at 50% full scale, 2 increments.
004 Increment change at 80% full scale, 2 increments.
005 Increment change at 64% full scale, 2 increments.
006 Increment change at 40% full scale, 2 increments.
53 Weighing capacity base value 000~099
54 Weighing capacity index value 001~004
55 Increment for lighter range 000 1
001 2
002 5
003 10
004 20
005 50
006 100
007 200
56 Location of decimal point 000 0 (No decimal point.)
001 0.0
002 0.00
003 0.000
004 0.0000
57 Type of decimal point 000 Period (.)
001 Comma (,)
58 Weighing unit 000 No unit.
001 g
002 kg
003 lb.
004 oz.
59 Weighing unit display 000 No display.
001 Specied unit displayed.
60 Weighing capacity base value 000 Not used in pound mode.
(pound mode) 001~099
61 Weighing capacity index value 000~004
(pound mode)

YCO-1112-04-06
DP-6200 Technical Manual
15
IV Changing the Scale Setup
Parameter Descriptions - Factory Parameters (continued)
Key Function Description
Word
62 Location of decimal point <Decimal pound and decimal ounce modes>
(lb , oz or lb/oz mode) 000 0 (No decimal point)
001 0.0
002 0.00
003 0.000
004 0.0000
<Pound/ounce mode>
000 0 : 0.00
001 0 : 0.0
002 0 : 0
003~004 Not used.
63 Pound increment 000~008
64 Pound conversion coefcient 000 Conversion coefcient = 1.00000
001 Conversion coefcient = 0.705479
002 Conversion coefcient = 0.881849
003 Conversion coefcient = 1.10231
65 Zero point range 000~100
66 Zero point range on plus side 000~100
67 Over scale 000~010
68 Display hold function 000 No hold
001 Hold for weight more than net + 10 divisions.
002 Hold for weight more than net + 10 divisions, hold key invalid.
003 Hold for weight more than net + 20 divisions.
004 Hold for weight more than net + 20 divisions, #10 invalid.
005 Hold for weight more than net + 20 divisions, hold key invalid,
#10 invalid.
69 Hold release against 000 Hold against any increased weight.
increased weight 001~005 Hold release against a specied increase.
70 Stable mark 000 Display the stable mark.
001 Stable mark not displayed.
71 Tare function 000 No tare function.
001 One-time tare.
002 Consecutive tare.
72 Zero reset under tare operation 000 Zero reset enabled.
001 Zero reset disabled.
73 Power off with ON key 000 ON key disabled for power off.
001 Power off after pressing ON key for two seconds.
74 Special function for stability 000 No function.
001 Enforcement.
002 EU mode.
75 Over - under feature 000 Disabled.
001 Enabled.
76 Tare cancel by zero key toggle 000 Disabled.
001 Enabled.
77 -bat- display toggle 000 Enabled.
001 Disabled.

YCO-1112-04-06
DP-6200 Technical Manual
16
Key Function Description
Word
78 Conversion coefcient for 000 Kilogram calibration for all capacities.
pound calibration 001 Pound calibration for 60 lb and 300 lb capacity scales.
002 Pound calibration for 150 lb capacity scales.
79 Center-of-zero ennunciator 000 Alternate.
position toggle 001 NTEP.
80 Internal resolution 000 Do not use.
001~255
81 1 timer (ms) 000 Do not use.
001~255
82 3 timer (ms) 000 Do not use.
001~255
91 Mechanical zero 1 000~255 Automatically set at span adjustment.
92 Mechanical zero 2 000~255 Automatically set at span adjustment.
93 Span coefcient 1 (small) 000~255 Automatically set at span adjustment.
94 Span coefcient 2 (small) 000~255 Automatically set at span adjustment.
95 Span coefcient 1 (large) 000~255 Automatically set at span adjustment.
96 Span Coefcient 2 (large) 000~255 Automatically set at span adjustment.
97 Regional number and gravity 000~155 Automatically set at span adjustment.
for span adjustment
99 Factory setting 000 Always displayed. * Do Not Change *
IV Changing the Scale Setup
Parameter Descriptions - Factory Parameters (continued)

YCO-1112-04-06
DP-6200 Technical Manual
17
The scale can check various functions automatically. When a problem is detected, the following mes-
sages will be displayed. Solutions are listed in order of decreasing probability and increasing com-
plexity.
The batteries are running low. When the batteries are running
low, this indicator will come on. Replace all of the batteries with
new ones.
The batteries are low. When the batteries are too low for accu-
rate operation, this indication will be displayed and the scale may
automatically power off. The batteries must be replaced to use
the scale.
Weight out of range high. When the weighed value is 6 or more
divisions heavier than the scale’s capacity this message will be
displayed. Remove any excess weight from the scale platform.
Calibrate the scale. Replace the logic board and/or load cell.
Weight out of range low. When the weighed value is a few divi-
sions below calibrated zero, this message will be displayed. It is
usually displayed when the platform is removed or has something
wedged under it. Place the platform on the platform support or re-
move any objects wedged under the platform. Calibrate the scale.
Replace the logic board and/or load cell.
Initial error on startup, high. When the initial weight is heavier
than the permissible zeroing range and the
key is pressed,
this message will be displayed. Remove whatever is on the plat-
form and press the
key again. Calibrate the scale. Replace
the logic board and/or load cell.
Initial error on startup, low. When the initial weight is lighter
than the permissible zeroing range and the
key is pressed,
this message will be displayed. It is usually displayed when the
platform is removed or has something wedged under it. Place the
platform on the platform support or remove any objects wedged
under the platform, and press the
key again. Calibrate the
scale. Replace the logic board and/or load cell.
V Troubleshooting

YCO-1112-04-06
DP-6200 Technical Manual
18
Weighing sensor error. When the weight detecting circuitry reg-
isters an error, this message will be displayed. Turn the power off
and then on again. Calibrate the scale. Replace the load cell.
Electronic circuitry error. When the operating circuitry registers
an error, this message may be displayed. Turn the power off and
then on again. Replace the logic board.
Calibration points out of range error. Occurs after the scale is
calibrated, but the value of the calibration points are outside the
range of the software. Conrm that system parameter 78 is set
correctly for the weights being used and the capacity of the scale.
Verify that the correct weights, 1/2 capacity and full capacity, are
being used. Make sure that nothing is restricting the motion of the
platform, pressing against the platform or wedged between the
upper and lower platform. Remove any sources of vibration or
electromagnetic radiation (i.e. - cell phones, power cables, motors,
compressors, etc.) Repeat the calibration procedure. Replace the
logic board and/or load cell.
Electromagnetic disturbance error. This indication is occa-
sionally displayed due to strong static electricity and/or electro-
magnetic disturbance. Move the scale away from any sources
of electromagnetic radiation (motors, generators, power cables,
transmitters, etc.) Turn the power off and then on again. If an
AC adapter is being used, try using a different power circuit or AC
adapter. Replace the logic board.
V Troubleshooting
Other manuals for DP-6200
2
Table of contents
Other Yamato Scale manuals

Yamato
Yamato CPC-2000 User manual

Yamato
Yamato AW-WPS User manual

Yamato
Yamato DP-6900 User manual

Yamato
Yamato SPC-2002 User manual

Yamato
Yamato DSY-1100 User manual

Yamato
Yamato SPC Series User manual

Yamato
Yamato PPC-200W User manual

Yamato
Yamato AW-WPS User manual

Yamato
Yamato BC-10 User manual

Yamato
Yamato AW-300P User manual

Yamato
Yamato AW-WSM-5KG User manual

Yamato
Yamato PPC-200 User manual

Yamato
Yamato DP-6100GP User manual

Yamato
Yamato Clock MINI DF301 User manual

Yamato
Yamato PPC-300 User manual

Yamato
Yamato SPC - Solar User manual

Yamato
Yamato Acuu-Weigh SPC-2002 User manual

Yamato
Yamato DP-5602 User manual

Yamato
Yamato DP-6700 User manual

Yamato
Yamato DSR-400 User manual