Checkout and Preventive Maintenance SL 800
Service Manual Confidential and Proprietary SM-30-4072
Order No. 2330-705 2 - 2 Version 02, 2019-10-24
The printout may be obsolete. Check the SIS for the most current and approved version of this document.
2.1 System Check-Out Procedure
The system check described below can be used in order to verify the proper function of the SL 800
after completion of any corrective maintenance activity.
Step Check Procedure Expected Check Result
Check all housings and covers
All covers as well as labels are installed properly,
clean and free of damage.
2 Check all visible screws All visible screws are fastened.
3 Check all cable connections All cable connections are connected.
4 Check all cable isolations All cable isolations are free of damage.
5 Check all joints and removable parts All joints are properly fixed.
6 Check the front objective The front objective is clean and free of any
The mirror is clean and free of any scratches or
damage
The manual microscope changer can be turned
easily by 360° without any friction
The motorized microscope changer can be turned
easily to all magnification steps without any
friction, by using the plus and minus buttons at the
The eyepieces are clean and free of any scratches
or damage.
10 Check the fixation of any additional
accessory, e.g. the tonometer.
The accessories are properly fixed and functional.
Check the full height adjustment of
the instrument base.
The instrument height adjustment can be properly
moved, by the joystick, to both end positions (up
and down) without any friction or obstacles.
12 Check the full range of motion of
the x-y instrument base.
The instrument base can be properly moved, by the
joystick, to all lateral end positions without any
Check the function of all brake
knobs.
All brake knobs can be used without any friction.
Check the illumination unit.
All filters and slit positions can be set without any
friction or obstacles. The slit image edges should be
parallel and can be fully closed. The LED
illumination can be set via the illumination setting
knob within the full range.
Check the external fixation light
(if applicable).
The external fixation light is fully functioning.
Check the chin rest adjustment
motion.
The chin rest adjustment is fully functional, to both
end positions (up and down) without any friction or