
Safety Measures S7 / OPMI VISU 160
Version 6.0
Page 16 G-30-1528-en
• In order to prevent phototoxic damages to the patient's eye during sur-
gery, use a retina protection filter (blue barrier filter) and/or a retina pro-
tection device that are integrated into the light source.
• Adjust the intensity of the irradiation and the resulting irradiation time by
selecting the appropriate illumination setting in order to prevent photo-
toxic injuries to the patient's eye. You will find the values recommended
by ZEISS in the table "Maximum radiation exposure times" on Page 21.
Any deviation from these values is only permissible when medically justi-
fied.
• In order to prevent phototoxic damages, never look directly into the light
source, e.g. on the microscope or the light guide.
• If the light guide is not connected at both ends, there may be a risk of fire
or burning when the illumination system is activated.
• Never leave a device unattended with the light source still switched on.
• If you have mounted the RESIGHT fundus viewing system on the underside
of the microscope and steeply tilt the microscope, the RESIGHT fundus
viewing system may unintentionally move in and injure the patient.
Remove the RESIGHT fundus viewing system before steeply tilting the mi-
croscope.
• Defective or unidentified accessories may lead to increased leakage cur-
rent on the system and injure the patient. Never connect any defective or
unidentified accessories and never touch the power outlet or video inter-
faces while in contact with the patient.
• If you need to change the lamp shortly after it has failed, the lamp will still
be very hot. Use heat-resistant protective gloves when replacing the lamp
to prevent burns.
• In case of a faulty cable or a malfunction in the motor's electronic system,
the microscope can inadvertently move down on the patient. Turn off the
system at the power switch.
• Malfunctions in the motor's electronic system may cause failure of the
main functions (X-Y movement, focus, zoom, light control) and impair
other functions. Proceed in manual mode.
• In the event that the zoom function can no longer be controlled via the
foot control panel or the control panel, adjust the zoom factor manually
at the microscope.
• In the event that the focusing function can no longer be controlled via the
foot control panel or the control panel, use the suspension arm to focus
the surgical field. This may lead to patient injury or the damaging of at-
tached equipment. Ensure to provide sufficient clearance between equip-
ment and patient (min. 40 mm) when focusing manually. Ensure that any
installed accessories do not come into contact with other objects.