- 13 -
. . . . CALIBRATION
Zero and span - true zero procedure
Differential pressure,gauge and level.
- Switch on the power supply.
- With no pressure applied to the transmitters, the value read
onthedigitalmilliammetershouldbe4 mA; ifit isnot 4mA turn
the zero screw for at least 1 second. After this operation the
reading should move to 4 mA: if no change occurs repeat the
operation.
-ApplytotheH(high)connectionapressureequaltotheupper
range value (URV) and allow time for the pressure to stabilize.
-Turnthe spanscrew forat least1 second: afterthis operation
the reading on digital milliammeter should be 20 mA and the
calibration procedure is complete. If no change occurs either
the calibration procedure was not correctly performed or the
span exceeds the limit; correct and repeat the operation.
Absolute pressure
- Switch on the power supply.
- Connect a vacuum source to the process connection and
draw the maximum possible vacuum obtainable. The value
readon thedigital milliammetershould be4mA ;if itis notturn
the zero screw for at least 1 second. After this operation the
reading should move to 4 mA : if no change occurs repeat the
operation.
- If the value of the calibration span (URV) is less than the
atmosphericpressuregentlyopentheventvalvesoincreasing
the pressure to the Upper Range Value. If the calibration span
(URV) is greater than the atmospheric pressure then connect
the pressure connection to a pressure source and generate a
pressurecorrespondingtotheURV.Allowtimeforthepressure
to stabilize.
-Turnthe spanscrew forat least1 second: afterthis operation
the reading on digital milliammeter should be 20 mA and the
calibration procedure is complete. If no change occurs the
calibration procedure was not correctly performed or the span
exceedsthelimit;applythecorrection andrepeattheoperation.
Zero suppression procedure
Differential pressure,gauge and level.
Two different methods (a) or (b) can be used :
a) After completion of the zero and span procedure above,
apply to the H ( high ) connection a pressure equal to the
pressuretobesuppressed.Allowtimeforpressurestabilization
and then turn the zero screw for at least 1 second. After this
operation the digital milliammeter reading should be 4mA and
the Upper Range Value automatically moved to a value equal
to the sum of the pressure to be suppressed and the previous
calibrated span.
b)Usethezeroandspanprocedureabovebut apply pressures
equal to the Lower Range Value (LRV) and then to Upper
RangeValue(URV),andturning,foratleast1second,thezero
and span screws respectively.
Absolute pressure
Usethe zero andspan procedureas previouslydescribed, but
apply to the process connection absolute pressures equal to
the Lower Range Value (LRV) and then to the Upper Range
Value(URV), turning, for atleast1second,the zeroand span
screws respectively.
Zero elevation procedure
Differential pressure and level
Two different methods (a) or (b) can be used :
a) After completion of the zero and span procedure above
apply to the L ( low ) connection a pressure equal to the
pressure to be elevated. Allow time for pressure stabilization
and then turn the zero screw for at least 1 second. After this
operationthe digitalmilliammeter readingshould be 4mAand
the Upper Range Value (URV) is automatically moved to a
value equal to the sum of the pressure to be elevated and the
previous calibrated span.
b)Usethezeroandspanprocedureabovebutapplypressures
equal to the Lower Range Value (LRV) and then equal to the
Upper Range Value (URV) and turning, for at least 1 second,
the zero and span screws respectively. The LRV pressure will
beappliedtotheLconnectionwhereastheURVwillbeapplied
totheLortotheHconnectiondependinguponthewhether the
range is all negative or crosses zero.
Gauge pressure
Apply to the process connection, pressures equal to the LRV
and then equal to the upper range value (URV) and
correspondingly turn the zero and span screws respectively.
Note - To prevent unauthorized calibration
operation refit the write protection link in position
P (Write Protect) (Fig. 9).
Note - If during the calibration procedure the
readings on the digital milliammeter are outside its
inherent accuracy, output trimming of the transmitter
mayberequested.Thisoperationcanonlybeperformed
using the Hand Held Terminal Communicator or the
PersonalComputerConfigurator. If thisequipmentis
not available the transmitter should be returned to a
Service Center for recalibration.
In some cases, expecially for tank level measurement, the
calibrationcanalsobeobtainedautomaticallybytheindication
of the actual output percentage, without any calculation for
LRV and URV. The operation is called Output % Reranging
and can be performed using a HART configuration tool on a
600T EN transmitter, in the Analog + HART version (see
Output % Reranging in the ADDENDUM FOR FLANGE-
MOUNTED TRANSMITTER).