ADE M501660 User manual

Operating Manual
International Range
Electronic Wheelchairs
M501660
M501660-210512-Rev003-UM-e

1 M501660-210512-Rev003-UM-e
Contents
Contents 1
Introduction 2
Check carton contents 2
Safety 2
Assembly and set up 3
Alignment of the scale 3
Power supply 3
Replacing the rechargeable battery 3
Display 4
Front View and side view of the display 4
Display Description 4
ey Description 5
Weighing 5
Correct weighing 5
How to change the unit 5
Tare function 6
Weighing with preset tare values 6
HOLD function 6
BMI (Body Mass Index) 6
Evaluation of the Body Mass Index 7
Connecting additional devices 7
Print out a record 7
Error Messages 7
Cleaning and Maintenance 8
Cleaning 8
Storage and Transport 8
Warranty – Liability 8
Technical data 9
CE-marking and Declaration of Conformity 10
Declaration of CE Conformity 10

M501660-210512-Rev003-UM-e 2
Introduction
This symbol means: "Consult operating instructions" To achieve accurate results, read
the instructions for use carefully before use and follow all instructions provided therein
Keep the instructions for use for future reference
Your ADE electronic wheelchair-scale is a quality, sturdy and durable product manufactured in
accordance with German national regulations
ADE instruction manuals have been prepared to provide the user with all the necessary knowledge
about the safe use of ADE scales and their related accessory equipment
For further questions about how to use the products or about this or other service manuals, feel free to
contact your local ADE representative or visit our web page at www ade-hamburg de
In order to achieve the best possible results kindly read and follow the operation manual in all respects
Check carton contents
- Weighing platform
- Display plus connecting cable
- Wall bracket already assembled to the display
- Mains adapter
- Operating Manual
Safety
The ADE-display unit corresponds to the latest state of the art technology and complies with standard
regulations and directives Kindly follow the safety instructions in this operating manual, the country-
specific installation standards as well as the applicable safety conditions and accident prevention
regulations in all aspects
Operate the scale only with the original ADE mains adapter
Always connect the mains adapter to the scale BEFORE connecting it to the electrical outlet
Disconnect from the electrical outlet BEFORE disconnecting the adapter from the scale
In case the mains adapter became wet or show visible damages the scale can no longer be
operated by using this mains adapter In this case please contact your local dealer or service
partner
The voltage and power supply as mentioned on the type plate must match the mains voltage
and power type at the place of use
When using the scale with a mains unit ensure that the supply cable is routed in such a way as
to exclude any type of potential trip hazard
Remove the mains plug before cleaning the unit
Do not operate the unit in hazardous or unstable conditions
Set up the unit as far away as possible from equipment that generates electromagnetic or
other interference Portable and mobile HF-communication systems may influence electronic
medical devices
All actions and operations described in this operation manual may only be carried out by
professional personnel authorized by ADE or official appointed dealers

3 M501660-210512-Rev003-UM-e
Assembly and set up
The ADE wheelchair-scale is already been delivered in a fully assembled condition Remove the
packaging and place the scale on a firm and plane surface Please place your scale free of any
vibration and at a dry place
Alignment of the scale
1) Place the wheelchair on a plain, dry and level surface before weighing
3) The air bubble in the spirit level should be right in the centre of the circle
4) Please ensure that the scale has no other contact to the floor than by the four feet underneath the
platform
Power supply
Power is supplied either by the rechargeable battery in the display unit or by the mains power via the
supplied power unit The rechargeable battery pack is consisting of 4xAA NIMH rechargeable
batteries
Use the AC adapter to load the rechargeable batteries The imprinted voltage value at the mains
adapter must match the local mains voltage When the battery is being charged for the first time a
sufficient charging period of approx 24 hours is necessary to achieve a full battery load The wall
socket must be located near to the unit and be easily accessible Please use only the mains adapter
supplied by ADE The mains adapter is charging the battery pack at the same time
Replacing the rechargeable battery
The battery compartment is located at the back of the display
1) Open the battery compartment
2) Remove the battery cables carefully from the housing, connect the wires and insert the battery into
the housing
3) Close the battery compartment again
Only use original rechargeable batteries and charger units from ADE The warranty does not cover
any damage caused by using other types of batteries or charger units

M501660-210512-Rev003-UM-e 4
Display
Front View and side view of the display
Display Description
Name
Description
Status indicator for the current battery charge level When this symbol
appears is the scale in battery operation mode or will be charged
The scale is within the zero level
Standstill status
indicator (14) :
The scale is at a standstill, meaning the weight is stable and no
movement is on the platform This is a mandatory precondition for
functions such as resetting, printing and taring)
In case the LCD is not displayed is the scale not in the standstill mode
“Hold” Displayed when the hold function is activated
“Ib” Displayed when the weighing unit is displayed in pounds
“kg” Displayed when the weight is displayed in kilograms
RS232 interface
Power Input
Load Cell connector

5 M501660-210512-Rev003-UM-e
ey Description
Name
Description
HOLD
Activates the Hold-Function
PRINT
Output data to serial communication port RS 232
BMI
Enter, after the user steps on the scale, in order to proceed with the BMI mode
UNIT
Switch button between primary kgs and secondary units Lbs
TARE
PRESET
Activates the tare function and the Preset Tare Function
ZERO
ON/OFF
Switch the scale on/off and zero function
During key operation use this key to switch between the units
During key operation use this key to switch between the units
During key operation use this key to switch between the numbers downwards
During key operation use this key to switch between the numbers upwards
Enter button to confirm the value
Weighing
Correct weighing
- Press the start key ON/OFF to enter the normal operating mode of the scale with no load on
the platform An automatic function test will be performed as well as the zero setting of the
scale When `0 0´appears you are in the kg weighing mode Additionally the display shows the
symbol ` ´ to confirm stability of the zero setting
- The scale is now ready to be used
- Place the patient on the scale and ensure that he is not moving
- After the weight is being displayed and the symbol ` ´ appears the utmost accuracy has
been achieved
- To turn off the scale press ON/OFF for a period of minimum 3 seconds Using the scale on
mains it may remain permanently switched on or can be shut off manually by pressing key
How to change the unit
The scale offers the possibility to select between the weighing mode kg and lb To change into another
mode press the UNIT key Please check weighing mode after switching on to get proper result

M501660-210512-Rev003-UM-e 6
Tare function
The tare function allows ignoring a fixed weight (e g wheelchair) that will be deducted after each
weighing procedure Proceed as follows:
- Press the start key with no load on the scale
- Place the additional weight of the object(s), which needs to be ignored, on the scale and store
the result by pressing the TARE key
- The display is being reset to zero The scale is now in NET weighing mode which is also
indicated on the display
- Remove now the objects from the platform The ignored weight is being displayed with a
minus
- Place the Patient and the additional weight (e g in the wheelchair) on the scale
- The scale determines the weight of the patient without the additional weight You may now
use the scale for as many weighing procedures and the stored value will always be deducted
as long as the scale is not turned off
- Enter the tare button again to get of the tare mode
- The tare value is likewise erased by the scale until pressing the tare button or being switched
off
Weighing with preset tare values
To input a pre set tare value press down the tare button more than 3 seconds NET will appear at the
screen Follow the same steps as before and place a load on the platform The last stored tare value
will appear at the display and ignored by the scale
HOLD function
The Hold function is being used if you like to hold the results at the display after the weight/load has
been removed from the scale Use this button to take care of the patient first and then note the weight
- Press HOLD key while scale is under load
- “HOLD” is being displayed at the screen
- The weight remains saved in the display after unloading the scale
- For deactivation of the Hold function press again the HOLD key
BMI (Body Mass Index)
The Body Mass Index is the ratio between height and weight squared The BMI is accepted world wide
– also by WHO (World Health Organization) - helping to evaluate the state of nutrition and with that
value also the state of health of a person The result is a tolerance value
- Start the scale by entering the ON/OFF Button The scale is automatically being set to zero
- Step on the scale or helping the patient to get into the wheelchair and on the platform
- Wait until the weight is being displayed in a stable condition and ´ appears on the screen
- Press the BMI key shortly
- The scale starts automatically with the last input Use the arrow buttons to enter the height of
the patient
Confirm the height with the enter button
.
- The BMI is now being displayed at the indicator
- To correct the given height press the BMI button once again and the display returns to the
height mode
- To get back into normal weighing mode and leave the BMI modus press the tare button

7 M501660-210512-Rev003-UM-e
Evaluation of the Body Mass Index
Compare the determined value which corresponds to those being used by the WHO
BMI below 18,5
The patient is under the regular standard and does not got enough weight An increase of
weight is recommended Please consult a professional
BMI between 18,5 and 24,9
The patient got a normal weight If he feels well there is no need for a change
BMI between 25 and 30
The patient is slightly overweighed to pre-obese A reduction of weight is indicated or, is
necessary if the state of health is already affected (e g diabetes, high end blood pressure,
heart diseases)
BMI over 30
Reduction of weight is imperative To visit a doctor is highly recommended
Connecting additional devices
Connection of printer, PC or remote-display will be done with the RS-232 interface For connection of
additional devices kindly ask your local representative or authorized ADE dealer for further instructions
Print out a record
The scales must be at standstill The standstill status symbol must appear at the screen Press the
ENTER key to send the data to the serial interface
Note: Printing is not possible if the net weight is 0 The print function only produces one print-out A reset
is required to print out another copy
Error Messages
Error Codes Description Solution
Load cell wires to indicator are
incorrectly connected; or shorted, or
opened; or ADC/ load cell are being
damaged
Make sure wires are in good order
and condition and correctly
connected Replace load cell or
ADC chip Service is required
Weight reading exceeds the zero limit
after switching the scale on
Make sure scale platform is empty
Weight reading is below the zero limit
after switching the scale on
Install platform on scale
Otherwise the zero calibration/
adjustment have to be done by
authorized personnel
Weight reading exceeds load limit Reduce load on scale until weight
value can be displayed
Weight reading below load limit Service by authorized personnel
required
Configuration or Calibration parameters
are not correctly set
Contact your local dealer / supplier
for calibration purpose

M501660-210512-Rev003-UM-e 8
Capacity parameters are not correct Contact your local dealer / supplier
for a new set-up
Calibration error The incoming signal of
loaded weight is too small, too big,
unstable or un-linear
Contact your local dealer / supplier
for a new set-up
Not turn on.
Power cord is not properly connected
Power outlet not supplying electricity
Batterie is discharged Other failure
Check power cord connections
Make sure power cord is plugged
into the power outlet Check power
source Replace batteries
Cannot zero the
display or will
not zero when
turned on.
Load on scale exceeds allowable limits
Load on scale is not stable
Load cell damage
Remove load from the scale
Wait for load to become stable
Service required
No Battery
symbol or
display of Lo.bAt
Batteries are discharged Charge batteries
Cleaning and Maintenance
Cleaning
Before cleaning disconnect the scale from the socket For cleaning just use a damp cloth or common
disinfection fluids Do not use any aggressive cleaning fluids, abrasive or acid cleaners or white spirits
During he process of cleaning take care that no water or cleaning fluids are penetrating the scale
Storage and Transport
Please keep all original parts and pieces for an eventual sending back of the scale
in order to avoid any possible damage during transport, which automatically leads to non warranty
Disconnect all cables before transport and avoid unnecessary damages
Warranty – Liability
ADE will repair or replace this product within a period of 2 years from date of delivery applying to
defects occurred due to poor material or workmanship (presentation of invoice is necessary) All
moveable parts – batteries, cables, mains units, rechargeable batteries etc – are excluded Warranty
does not include natural wear, damages caused by incorrect or negligent handling, overuse or
unsuitable or improper storage The cost of transport will be borne by the customer should the
equipment be located anywhere other then the customer’s premises In the event of transport
damage, claims under warranty can be accepted subject to using the complete original packaging and
the equipment secured the way it was packed and secured originally All the packaging should
therefore be retained
Any warranty claim will be rejected if the equipment is opened by persons not authorized expressively
by ADE to do so
Customers abroad are kindly requested to contact their local dealer for any warranty claim, customer
service and spare parts

9 M501660-210512-Rev003-UM-e
Technical data
Max Capacity: 300 kg / 660 lbs
Graduation: 100 g / 0,2 lb
Power supply: Mains and rechargeable batteries
Work current: ≤15mA (with backlight off, no optional Bluetooth Module and no load
cells)
Mains adapter: AC Adapter: 6-9V
DC,
≥500mA,
Central negative: 4xAA NIMH rechargeable batteries pack
Display: 6-digit,7-segment , 1"(25mm) LCD with 16 annunciators and blue
backlight
Working temperature: -10°C to 40°C
Storage temperature: -20°C to 70°C
Humidity: 10 to 90% RH without condensation
Voltage: 5Vdc
Max Current: 60mA (can power 4-350 ohm bridge, take care the limitation when USB
is power supply)
Signal connection: 4 or 6 lead with sense leads
Max Sensitivity: -3mV/V to +3mV/V
Serial port: Full-duplex RS232

M501660-210512-Rev003-UM-e 10
CE-marking and Declaration of Conformity
Declaration of CE Conformity
Manufacturer
:
ADE Germany GmbH
Neuer Hoeltigbaum 15
22143 Hamburg / Germany
℡ +49 40 432 776-0
+49 40 432 776-10
info@ade-germany de
www ade-germany de
Declaration of conformity by the manufacturer
These scales have been manufactured in accordance with the harmonised European standards
They conform to the regulations of the EC-directives listed below:
EMC Directive 2014/30/EU
Low Voltage Directive 2014/35/EU
RoHS Directive 2011/65/EU
in the currently valid versions
This declaration loses its validity if modifications are made to the scale without our approval
Hamburg, May 2021
ADE Germany GmbH
Neuer Hoeltigbaum 15
22143 Hamburg / Germany

11 M501660-210512-Rev003-UM-e
authorised dealer: